Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5423_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
15.09.2026
Размер:
11 Мб
Скачать
☆
      143
https://t.me/medicina_free
Since development of a serotype‐monospecific TAb assay may not be possible due to high homology between AAV serotypes, AAV serotype specificity may need to be further characterized when patients are retreated with a different AAV cap­sid. To understand whether a particular TAb response in an individual is serotype‐ specific, a confirmatory test can be conducted, whereby excess amount of unlabeled AAV vector for the serotype of interest is added to determine if the assay signal is reduced.
6.2.4.3 Precision
Assay precision is a parameter to describe the degree of variability in assay raw signal, based on results obtained in multiple intra‐ and inter‐run tests[14]. Like protein biopharmaceutical ADA assays, PC samples prepared at a few concentra­tions in NC matrix (or surrogate) are tested in replicates (typically two per sample) and evaluated on several occasions (typically 3) by two or more analysts. As for regular ADA assays, variability is expressed as coefficient of variation percentage (CV%) and is expected to be less than 20%.
6.2.4.4 Matrix Interference and Selectivity
Impact from sample matrix is evaluated in interference and selectivity assess­ments. Several (typically 10) individual matrix samples are analyzed unspiked and spiked with PC control reagent at the low PC (LPC) and high PC (HPC) con­centration. This enables an understanding of whether inter‐individual variations in matrix components interfere with detection and if the assay can reliably detect PC at low concentrations across multiple donors. The test needs to be performed using healthy and disease population samples to understand if any difference in potential interference can be observed. It is understood that for rare diseases obtaining indication‐specific matrices may be challenging and alternative solu­tions, for example, use of sample collected in clinical study, may be required. It is recommended that at least 80% of unspiked samples should score TAb negative and the same percentage of LPC/HPC‐spiked samples should score TAb positive. In addition, the potential impact of known interfering factors, such as lipids, hemoglobin and bilirubin, and other matrix components that are expected to be present in the sample (e.g. anti‐coagulation factors) has been discussed in regula­tory guidance and needs to be considered[6].
6.2.4.5 Assay Cut-Point
TAb assay cut‐point is arguably one of the most important parameters to be deter­mined during assay development or validation. Assay cut‐point is a threshold value that is used to determine whether a given sample is positive or negative for the presence of AAV TAb, an approach that is very similar to the one used for
6 Bioanalytical Methods to Detect Preexisting and Post-administration
https://t.me/medicina_free
144
protein biotherapeutic ADA assays [14, 29]. Cut‐point values are assay‐ and laboratory‐specific and are separately defined for screening, confirmatory, and titration protocols. Frequently, screening and titration protocols utilize the same cut‐point value. As for the ADA methods, an AAV TAb assay cut‐point is deter­mined based on a statistical analysis of assay signals generated for a collection of treatment‐naïve individual donor samples that are preferably negative for preex­isting AAV immunity. Generally, 50 or more individual samples are tested in 3 or more runs to obtain a sufficient collection of data. The data are analyzed for sta­tistical and biological outliers that are excluded before determining the 95th or 99th percentile of distribution for screening (as well as titration) and confirmatory tests, respectively[14, 29].
Due to preexisting AAV antibodies, it may be challenging to obtain TAb negative treatment‐naïve individual donor samples. It may also be difficult to separate true TAb positive vs true TAb negative populations. Conducting a confirmatory test may offer a solution, with an assumption that samples that generate a signal that is inhibited in the presence of unlabeled AAV material by 50% or more contain spe­cific AAV TAb. Results generated for these reactive samples may be removed from the data before performing cut‐point data analysis.
6.2.5  TAb Assay Data Interpretation
Information about AAV TAb status of an individual patient may facilitate a deci­sion about AAV treatment eligibility and help understand potential clinical events of interest, including impact on safety or efficacy of the GTx treatment. Application of TAb methods for detection of preexisting AAV immunity has gained popularity since these methods are easy to develop and implement, rela­tively straightforward for development into a companion diagnostic (CDx) or similar routine test, and relatively robust in their application. If TAb assay results are used as treatment eligibility criteria, a threshold TAb titer for enrolment needs to be pre‐defined. The specific treatment eligibility TAb titer threshold may be dependent on disease, route of administration, assay sensitivity, nature of the detected antibodies, and other parameters. It continues to be a subject of ongoing debate. Examples include using the lowest detectable AAV antibody titer, 1:1000 titer[30] or no detectable antibody titer[1, 31] as a threshold for eligibility. For some routes of administration (e.g. intrathecal, intravitreal, and subretinal), concerns related to the impact of AAV antibodies may be reduced compared to a systemic, intravenous route of administration. Therefore, the TAb titer thresholds that would be applied when determining patient eligibility would also be different.
      145
expression
https://t.me/medicina_free
6.3   Considerations forCell-based Transduction  Inhibition Assays
6.3.1  Principle and Methodology ofCell-based AAV TI Assays
AAV TI assays are typically performed as cell‐based reporter‐gene assays[3, 18, 32–34]. A permissible cell line, such as HEK293T/17, is seeded in complete media on white clear‐bottom 96 well plates and grown overnight. On the next day, plasma or serum test samples are mixed with a viral AAV reporter gene construct carrying, for example, a CMV promoter‐driven luciferase reporter gene (AAV‐ Luc), in serum‐free media and incubated for 30minutes at room temperature to allow binding of AAV inhibitors to the vector capsid (Figure6.2).
Following this incubation, samples are added to the cells in replicate wells at a certain multiplicity of infection (MOI), generally between 5000 and 400,000 vg per cell[18, 34]. After a 1‐hour incubation at 37 etoposide solution in complete media may be added to stimulate DNA repair mechanisms that enhance productive transduction. After 2days, the cell culture medium is removed and a luciferase substrate, such as Steady‐Glo, is added for 10minutes at ambient temperature. Luminescence is measured, for example, by using a 500‐ms integration time on a VICTOR‐™X microplate reader. The relative
°C, a final concentration of 20 μM
mix sample w/ AAV-Luc
w/o AAV
inhibitors
Luciferase expression Decreased luciferase
Figure6.2  Principle of cell-based AAV transduction inhibition (TI) assays.
with AAV inhibitors
6 Bioanalytical Methods to Detect Preexisting and Post-administration
https://t.me/medicina_free
146
luminescence of a test sample is then normalized to that of a NC sample, which consists of pooled normal human plasma or serum without any detectable AAV transduction inhibitors. The results can be further converted to percent transduc­tion values relative to NC. Alternatively, test sample results may be expressed as percent inhibition values by subtracting the percent transduction value of a test sample from 100%. Either way, one needs to establish a cut‐point below or above which a test sample would be considered positive. Positive test samples are usu­ally reported with a semiquantitative TI titer value as a relative measure of their capacity to inhibit AAV transduction. Titer values are determined by serially dilut­ing a positive test sample to determine the highest sample dilution still detectable in the TI assay. The reciprocal of the highest detectable sample dilution is equal to the numerical neutralizing titer value.
6.3.2  AAV TI Assay Development: Designing forClinical Relevance
One of the key parameters to optimize during TI assay development is MOI. MOI represents the ratio between the quantity of AAV vector and number of plated cells. Cell lines that require a higher MOI to achieve productive transduction can be thought of as more difficult to transduce. Difficult to transduce cell lines can pose challenges for developing sensitive AAV TI assays, since higher test sample titers will be required to neutralize the higher quantity of AAV vector used in the assay, resulting in limited detectability of lower titers. Cell lines commonly used for TI assays are HEK293, HeLa, and HuH7. Depending on AAV serotype, a popu­lar choice is HEK293 and its derivatives, such as HEK293T/17 and 2V6.11[3, 35].
To optimize MOI, AAV vector titrations are performed, similar to how drug‐ response curves are needed to develop cell‐based anti‐protein therapeutic NAb assays. These titrations are performed to establish the dynamic range, within which a meaningful decrease in AAV vector concentration generates a distinctly measurable change in assay signal. The goal is to ensure that– at the selected MOI– the assay signal, such as measured luminescence, is sufficiently robust yet not fully saturated. Otherwise, the assay would remain unresponsive to a mean­ingful decrease in effective AAV vector concentration mediated by neutralizing factors in a test sample. In conclusion, MOIs should be selected to fall well within the dynamic range of the assay.
Dynamic ranges can be broad, and this is where strategic optimization may be employed to maximize clinical relevance of a cell‐based TI assay. The lower the MOI for a given number of plated cells, the more sensitive the assay will be, since a lower quantity of AAV vector will need to be neutralized by a particular volume of test sample. The question is how sensitive does the assay need to be? To answer this with clinical efficacy in mind, it can be helpful to determine the volume of test sample (plasma or serum) that is incubated together with a particular quantity of AAV
      147
https://t.me/medicina_free
vector in the assay, based on the selected MOI and number of plated cells. The AAV vector‐to‐sample ratio can then be compared to the maximal AAV vector plasma concentration anticipated in study participants, based on intended clinical dose level. For example, an adult clinical trial participant who is administered an AAV vector dose level of 6 E12 vector genome per kilogram (vg/kg) will receive a total vector dose of 4.2 E14 a theoretical maximal plasma concentration of 1.4 E11 total plasma volume of 3000 may be incubated with 42microliters of undiluted plasma test sample, resulting in a vector‐to‐sample ratio of 3.36 E09 vg/0.042 mL=8 E10 vg/mL, which is about two times less than theoretical maximal plasma concentration of 1.4 E11
To be relevant for clinical efficacy evaluations, the vector‐to‐sample ratio in a TI assay should be equal to or lower than the maximal plasma concentration of the AAV GTx theoretically achieved in study participants, as illustrated in the exam­ple above. Otherwise, the neutralizing potential associated with a particular vol­ume of plasma or serum could completely neutralize clinically administered GTx quantities, while barely inhibiting transduction of the cell line. Therefore, cell lines requiring exceedingly high MOIs may not be suitable for use in TI assays. In conclusion, to develop a TI assay with clinically relevant sensitivity, it is useful to concurrently optimize cell density per well, MOI, and test sample volume, while comparing vector‐to‐sample ratio invitro to anticipated GTx plasma levels invivo. Similar considerations have been successfully applied to interpolate clinical rele­vance of conventional anti‐biologics NAb assays[36].
vg, assuming average body weight of 70 kg. This would lead to
vg/mL, assuming average
mL. In a cell‐based TI assay, 3.36 E09 vg AAV vector
vg/mL.
6.3.3 Key Assay Validation Parameters
This section has been written in consideration of available regulatory guidance for immunogenicity assays[6, 37] and industry white papers[38, 39].
6.3.3.1  Screening and Titer Cut-Points
Test samples that contain AAV inhibitors are expected to decrease transduction in a cell‐based TI assay compared to NC. To adjudicate a sample as positive or nega­tive for AAV inhibitors, a cut‐point is required to define a meaningful decrease in transduction. This analytical cut‐point is established in reference to cell‐based assay procedures; it does not directly translate to the anticipated impact on clini­cal efficacy. This will be discussed further below, refer to: Data Interpretation: Determining a Meaningful Titer Threshold for Clinical Efficacy.
Two main strategies exist to establish TI assay cut‐points: (1) using a threshold of 50% transduction at a predetermined sample dilution compared to a NC sample[35]; (2) using a statistical cut‐point derived from treatment‐naïve, AAV inhibitor‐negative individual donor samples [3, 40], as recommended by regulatory guidance [6].
6 Bioanalytical Methods to Detect Preexisting and Post-administration
https://t.me/medicina_free
148
A 50% transduction cut‐point is commonly used in antiviral and vaccine research to report neutralizing titers against target viruses, and thus this strategy is often­times applied for measuring AAV inhibitors. However, this approach could miss less potent but potentially impactful inhibitors that only partially decrease trans­duction to 60–70%.
Whether reduced detectability of less potent AAV inhibitors poses a significant hurdle for clinical data interpretation will depend on the assay setup, for example, on how much lower the vector‐to‐sample ratio is compared to the theoretical max­imal vector plasma concentration in patients: If the vector‐to‐sample ratio is 10 times lower than the anticipated vector plasma concentration, then using an ana­lytical cut‐point of 50% transduction could be justified, simply because the quan­tity of AAV vector neutralized invitro would translate to as little as 5% of the maximal clinical vector concentration anticipated in the tested sample volume. It will also be critical when implementing a 50% transduction cut‐point to account for assay variability: Variability needs to be empirically evaluated and may be caused by different reagent lots, composition of individual sample matrices, pas­sage number of cells, and operators.
Statistical cut‐points have been discussed elsewhere[6, 14, 39] and are com­monly used for detection of anti‐drug antibodies to therapeutic proteins. This bio­analytical strategy has also been successfully implemented for clinical AAV TI assays[3, 40]. Statistical cut‐points are generally considered to be more objective, even though they can be difficult to establish due to a varying degree of preexist­ing AAV antibodies in healthy humans and patient populations alike [3, 41]. Ahigh number of antibody‐positive samples may confound the statistical analysis of data generated during cut‐point establishment. To mitigate cut‐point inflation resulting from using samples with preexisting AAV antibodies, samples may be prescreened prior to the cut‐point experiments to exclude those associated with less than 50% transduction, indicative of preexisting AAV immunity [3]. To be more rigorous, prescreening could include a confirmatory assay (described below) and any sample with more twofold increase in transduction following immuno­globulin depletion would be considered AAV NAb‐positive and excluded from cut‐point experiments. These approaches may require an initial analysis of a larger number of commercial samples to find 50individual samples without pre­existing AAV immunity. Alternative methods to deal with antibody‐positive sam­ples and analyze data generated during cut‐point establishment are discussed in the literature[29, 42–45].
6.3.3.2  Limit ofDetection
To characterize the analytical sensitivity of a cell‐based AAV TI assay, a limit of detection (LOD) can be determined using a surrogate positive‐control AAV anti­body. Both polyclonal and monoclonal antibodies may serve as PC, with polyclonal
      149
https://t.me/medicina_free
antibodies better reflecting the nature of humoral immune responses in future study samples. The LOD is defined as the concentration of a positive‐control anti­body in neat (undiluted) sample matrix that is detectable at the cut‐point of the TI assay (Figure6.3). Serial 1:2 or 1:3 dilutions of a positive‐control antibody are typically performed and tested. Dilutions expected to fall above and below cut‐ point should be included to accurately interpolate the LOD. The LOD may be inter­polated using either a fitted curve, as shown in Figure6.3, or a linear curve interval between the two antibody concentrations whose transduction values flank the cut‐point.
Regulatory agencies recommend a detection limit of 100
ng/mL or less for the positive‐control antibody in conventional anti‐protein therapeutic TAb assays and acknowledge that cell‐based NAb assays may not always meet this mark [6]. Meeting this specification for AAV TI assay will depend on the type and affinity of the positive‐control antibody and hence does not necessarily reflect the level of AAV inhibitors in study samples. Furthermore, it is presently unknown what level of AAV inhibitors in study participants will have a considerable impact on GTx efficacy or safety. This level may also depend on GTx vector‐specific factors, such as AAV serotype and route of administration. In conclusion, a clinically relevant LOD target value cannot be specified for AAV TI assays at this time.
Therefore, suitable assay sensitivity may need to be ascertained through alterna­tive avenues, such as (1) consideration of the MOI and vector‐to‐sample ratio, as discussed above; (2) adequate resolution of AAV inhibitor titers in a large number of individual subjects with pre‐existing immunity, which should range from low to high titers; (3) empirical verification of predicted dose‐neutralization curves, as described in the last paragraph; (4) correlation with efficacy and/or safety signals in nonclinical studies.
100
90 80 70 60 50 40 30
% Transduction
20 10
0
050 100 150 200 250 300 350400
Figure6.3  LOD interpolation at the assay cutpoint set for 50% transduction.
Antibody concentration (ng/mL)
6 Bioanalytical Methods to Detect Preexisting and Post-administration
https://t.me/medicina_free
150
6.3.3.3 Precision
Precision of cell‐based TI assays is generally lower than that of ligand‐binding assays, resulting in higher coefficients of variance (CV). Intra‐replicate, intra‐ assay, and inter‐assay precision of TI assays can be assessed using low (LPC) and high (HPC) PC AAV antibody‐spiked samples. CVs can be evaluated for raw lumi­nescence signals or normalized transduction values and should generally remain below 30%. Inter‐assay CV is also expected to remain below 30%, even though this limit could be elevated to 50%, if such higher variance does not affect the repro­ducibility of qualitative results (negative/positive) for low‐positive samples.
Titer precision is also important to consider, which can be assessed as the range of titer values measured repeatedly for the same sample. Titer precision is primar­ily dependent on the dilution scheme applied to the test samples and will typically fall within ±1 of the mean log
(titer) derived from repeated measurements if a
2
1:2 dilution scheme is used. This corresponds to the last detectable sample dilu­tion shifting one step upwards or downwards.
6.3.3.4 Specificity
Specificity for AAV TI assay addresses: (1) detection of neutralizing AAV antibod­ies versus non‐antibody‐based neutralizing factors (antibody specificity); (2) no detection of AAV inhibitors that exclusively react with other AAV serotypes (sero­type specificity); (3) detection of AAV inhibitors to both reporter gene construct and GTx (GTx specificity). For (1), refer to section: Confirmatory Steps to Ensure Specific Detection of Neutralizing AAV Antibodies. Serotype specificity (2) can be assessed by testing samples spiked with monoclonal antibodies that exhibit exclu­sive reactivity against related AAV serotypes, which should remain undetectable. This assessment is, however, of limited value and merely confirms that the reporter gene construct is not contaminated with another AAV serotype. Clinical test samples typically contain polyclonal mixtures of AAV antibodies that broadly cross‐react with multiple AAV serotypes[41].
The underlying question of (3) is whether the reporter gene construct ade­quately represents the GTx. Even though reporter gene construct and GTx have the same AAV serotype, they may have been produced by different manufacturing processes and differ in the size of packaged DNA, both of which could impact capsid antigenicity. One way to confirm GTx specificity in TI assays is to spike excess GTx into a sample containing AAV inhibitors. The GTx will bind to AAV inhibitors, preventing them from neutralizing the reporter gene construct, thus leading to increased transduction.
6.3.3.5  Confirmatory Steps toEnsure Specific Detection ofNeutralizing  AAV Antibodies
While TI assays detect AAV inhibitors, these assays do not routinely distinguish between neutralizing antibodies (NAbs) and non‐antibody‐based neutralizing
      151
https://t.me/medicina_free
factors[3]. This could obfuscate GTx patient eligibility when investigating preex­isting AAV immunity, since some non‐antibody‐based neutralizing factors decrease transduction invitro but are not impactful invivo, as demonstrated for an AAV5 GTx in nonhuman primates[4].
The nature of non‐antibody‐based neutralizing factors in plasma is heterogene­ous and poorly understood[46–49]. To ensure specific detection of NAbs in AAV TI assays, a confirmatory procedure may be implemented that depletes antibodies using resins coupled with immunoglobulin‐binding proteins, such as protein A, G, and L[33, 40, 50]. Samples that screened positive for AAV inhibitors are retested following immunoglobulin depletion, which restores transduction if and only if NAbs were present. A mock depletion of the same sample is performed in parallel to account for non‐specific depletion by unconjugated resin. Transduction follow­ing protein AGL depletion can be normalized to that observed following mock depletion, establishing an AGL/mock ratio[40]. For regulated studies, a statistical confirmatory cutpoint can be derived for the AGL/mock ratio; sensitivity, preci­sion, and selectivity of the confirmatory assay could also be validated.
Alternative strategies to confirm the specificity of AAV TI assays include deple­tion of test samples using resins conjugated with the AAV vector of interest [20,51], or the addition of empty or generic AAV vectors of the same serotype to absorb and functionally deplete AAV inhibitors in test samples[52]. Even though these strategies confirm the interaction of AAV inhibitors with a particular AAV serotype of interest, they do not immediately provide for a molecular classifica­tion of the involved neutralizing factors as antibodies, since any AAV inhibitor would be physically or functionally depleted.
6.3.3.6 Selectivity/Matrix Interference
Selectivity assessments for AAV TI assays are performed as described for TAb assays. At least 80% of spiked donor samples should test positive, while at least 80% of unspiked donor samples should test negative. A related assessment is interference from individual matrix components. For AAV TI assays, it is not advisable to use endogenously lipemic, hemolytic, and icteric test samples for interference studies, since results may be confounded by preexisting AAV inhibi­tors, falsely indicating additional interference. It may be more conclusive to add interfering factors under investigation (lipids, hemoglobin, or bilirubin) exoge­nously to individual or pooled samples previously confirmed to be negative for AAV inhibitors.
6.3.3.7  Stability
Stability studies can be performed by spiking low and high concentrations of positive‐ control AAV antibodies into negative test samples. Stability of non‐antibody‐based neutralizing factors could alternatively be assessed by retesting a positive donor sample over time. Neutralizing AAV antibodies are the most relevant AAV
6 Bioanalytical Methods to Detect Preexisting and Post-administration
https://t.me/medicina_free
152
inhibitors, and antibodies are generally stable if stored at −70 °C or colder[53, 54]. Thus, there is limited value in performing long‐term stability studies at low tem­peratures. Short‐term and freeze/thaw stability assessments may be more applica­ble; for example, samples may be stored at higher temperatures, such as −20
°C, or ambient temperature, given logistical constraints, or be subject to repeated
4 freeze–thaws to allow for retesting.
6.3.4  Sample Testing Strategy and Monitoring Assay Performance
Samples from GTx treatment‐naïve subjects can be positive for AAV inhibitors but seroprevalence typically does not reach 100%[3, 41]. This could justify tiered test­ing for pre‐dose samples: initial screening at minimally required dilution (MRD), followed by titering of screen‐positive samples only. Though for rare diseases, where sample batches are small, screening and titer assays are typically performed simultaneously to enable timely eligibility decisions. Screen‐positive samples may also be tested in a confirmatory assay, if deemed suitable for the serotype of interest.
In contrast, samples collected following systemic GTx administration have high AAV TI titers due to the formation of neutralizing AAV antibodies, usually with an incidence rate of 100% and persisting for at least 15 years[55–58]. Therefore, in most clinical scenarios, post‐dose samples could be titered directly. While post‐ dose samples might also contain non‐antibody‐based neutralizing factors, their titers will be negligible compared to those from treatment‐emergent neutralizing AAV antibodies. Hence, confirmatory assays for post‐dose samples may not be necessary or be restricted to a subset of samples to demonstrate that neutralizing AAV antibodies have indeed formed as expected.
TI assay performance can be monitored long‐term by trending luminescence signals and/or transduction/inhibition values for LPC and HPC, as well as titer values for HPC. Appropriate trending ranges may be established based on results from at least 10 assay runs.
°C,
6.3.5  Data Interpretation: Preexisting TI Titer and Clinical Efficacy
Preexisting AAV immunity detected in TI assays does not always correlate with decreased GTx efficacy. Lack of biological impact has been linked to low TI titers[2, 59] as well as non‐antibody‐based neutralizing factors, which are typi­cally present at low TI titers[3, 4]. This raises the question of whether there is a clinically meaningful correlation between preexisting TI titers and the likelihood of clinical success, which could inform GTx patient eligibility.
For TI data interpretation, it can be useful to determine the theoretical neutral­izing capacity of plasma/serum in relation to a particular TI titer, based on assay