N. Kato et al. Chapter 20 Endografts for Chronic Dissections
https://t.me/med1917
stents can easily dilate it. Therefore, endografts can al­most fully expand and become round at both the top and the bottom ends after the treatment even if the true lumen is markedly compressed by the false lumen and is paper-thin (Fig. 20.6). Thus, stents could be simply cylindrical with a round shape at both ends. In con­trast, endografts do not always become round in the case of chronic dissection, especially at the distal end since the intimal flap is fibrotic and rigid (Fig. 20.7). Thus, stents have to be able to accommodate to the shape of the deformed true lumen in the case of
Fig. 20.5. So-called first-generation endograft which is fabri-
cated with Z-stents and polytetrafluoroethylene
chronic dissection. Z-stents have sufficient compressibil­ity to fulfill this demand.
An anchoring device such as a hook is mandatory for the treatment of true aneurysms of the thoracic or the abdominal aorta to avoid migration. However, it
which are fabricated with Z-stents and polyester or ex­panded polytetrafluoroethylene (ePTFE), still work well in the treatment of aortic dissection as well as in that of thoracic aortic aneurysms [4, 16] (Fig. 20.5).
The design of the stent configuration is important when chronic dissection is treated. In the case of acute dissection the intimal flap is stretchable enough that
might not be preferable in the case of dissection since it could make a new intimal tear when there is an inti­mal flap at the level where the hooks are supposed to be located. In addition, such an anchoring device may not be necessary because the landing zone is tapered distally in most dissection cases and such tapering of the landing zone could prevent migration of devices.
203
ab
Fig. 20.6. The bottom end of the endograft, which was narrow and deformed before treatment (a), fully expands and is round
1 year after endografting (
ab
Fig. 20.7. The bottom end of the endograft seems almost the same before (a) and after (b) endografting, i.e., it does not fully ex-
pand and is still deformed 1 year after endografting
b) in the case of acute dissection
204
https://t.me/med1917
IV. Dissection
The size of the endografts is always a key to success­ful treatment. Endografts the diameters of which are approximately 10±15% larger than that of the landing zone are used as in the case of thoracic aortic aneu­rysms. The diameter of a circle which has the same cir­cumference as the landing zone needs to be estimated since the landing zone is usually deformed in the case of aortic dissections. It can be calculated by dividing the circumference of the landing zone by p. More easily the mean of the shorter and the longer diameter of the landing zone can be substituted for it.
The other thing to be memorized is that the intimal flap does not always expand fully in the case of chronic dissection, especially when dissections older than 1 year are treated. Therefore, tapered endografts might be suit­able for chronic dissections instead of straight endo­grafts, which are usually sufficient for acute dissections, since the use of a straight one could lead to intimal in­jury because of too much hoop strength at the more de­formed and narrower true lumen at the bottom end.
20.6.2 Procedures
Since the procedure itself is basically the same as that used for the treatment of thoracic aortic aneurysms, the details of the procedures are abbreviated in this chapter. The only difference from true aneurysms is that there are two lumens in the case of aortic dissections. A cath­eter and a wire inserted from the femoral artery can easily go into the false lumen when there is a large re­entry tear. Thus, careful monitoring of the advancement of a catheter with occasional test injections at various levels to confirm the catheter is in the true lumen is mandatory. When it is difficult to enter the true lumen from the femoral artery, the true lumen can be obtained by advancing a catheter and a wire from the brachial artery to the femoral artery.
Completion angiography after deployment of the de­vice must be carefully appreciated in the case of aortic dissections. The false lumen is sometimes rapidly opaci­fied by contrast medium entering from natural fenestra­tions or reentry and it may resemble type I endoleak, although the false lumen is thrombosed soon or later even in such cases.
When type I endoleak is observed after deployment of an endograft, balloon dilatation can be applied. How­ever, meticulous inflation is required to avoid making new intimal injuries although the possibility should be small compared with that in the case of acute dissec­tion.
20.6.3 Postoperative Care
Postimplantation syndrome including high fever con­tinuing for about 1 week is almost always observed. It can be easily controlled with antifebrile. Transient in­crease of C-reactive protein and white blood cells is also observed as well. However, the levels also subside to­gether with fever [18, 28] and there is no need to treat them.
Paraplegia, which is one of the most catastrophic complications accompanying surgical graft replacement, reportedly develops in approximately 5% of patients un­dergoing endografting, which is comparable to or less than that of surgical intervention [4, 5]. It may develop immediately or late after endografting [14]. To avoid this challenge, cerebrospinal fluid drainage seems useful as well as in the case of graft replacement for the treat­ment of thoracoabdominal aortic aneurysms [3, 10, 35].
Strict control of blood pressure and close follow-up by CT even after endografting should be a must. Those patients who have suffered aortic dissection may have a higher chance of development of a new dissection at an­other segment of the aorta. In addition, once dissected the aorta may not be normal even if the false lumen is completely thrombosed by endografting. Fortunately, there is no patient who underwent endografting for the treatment of chronic aortic dissection and had a new dissection thereafter in our hospital; type A aortic dis­section developed in one patient who underwent endo­grafting for the treatment of acute type B dissection. Type A aortic dissection with a new intimal tear in the ascending aorta developed in another patient who un­derwent endografting for the treatment of acute type A dissection with the intimal tear in the descending tho­racic aorta.
Follow-up CT is obtained 1 week after the treatment, 1 month, 3 months, 6 months, 1 year, and yearly there­after in our hospital. It is not rare that only the true lu­men is densely opacified in the early phase of contrast­enhanced CT since the flow into the false lumen of the descending thoracic aorta coming up from fenestrations or reentry is relatively slow. Therefore, delayed scan is mandatory to recognize the exact extent of thrombosis in the false lumen. Whenever any problem, including type I endoleak and new intimal injury, is observed, ad­ditional procedures have to be planned immediately.
The false lumen of the descending thoracic aorta sometimes remains patent because blood flow coming from fenestration or reentry is sometimes directed to such aortic branches as bronchial arteries or intercos­tals arteries which are supplied totally from the false lu­men. However, the false lumen finally thromboses dur­ing follow-up in most cases unlike type II endoleak, which is often observed in cases of abdominal aortic aneurysms.
N. Kato et al. Chapter 20 Endografts for Chronic Dissections
https://t.me/med1917
20.7 Results of Endografting
Nienaber et al. [29] should be credited as being first to objectively prove the efficacy of endografting for the treatment of chronic aortic dissection. They treated 24 patients with chronic aortic dissection: 12 patients with endografting and 12 patients with surgical graft re­placement. Endografting was associated with no mortal­ity or morbidity, whereas conventional surgery was as­sociated with four deaths and five serious adverse events in their series. Although the number of the pa­tients and the follow-up term in their series are ex­tremely limited, their results were so impressive that those who had been hesitating to repair chronic dissec­tion with endografts were encouraged.
From 1997 to 2004, 29 patients with chronic aortic dissection were treated with endografting in our hospi­tal [19, 21]. ªChronicº is referred to as those dissections present for 1 month or more. The characteristics of the patients are listed in the Table 20.2. Indications for en­dografting are shown in Fig. 20.8. There was no death within 30 days (Table 20.3). Intraoperative complica­tions were observed in two patients. Minor intimal in­jury at the external iliac artery caused by inserting the delivery system was repaired with stenting in one pa­tient. Another patient with arterio-venous fistula for he­modialysis in the left forearm complained of left arm pain immediately after anesthesia wore off. Since the origin of the left subclavian artery was closed with an endograft, ischemia was deemed the cause of pain. The symptom disappeared after creating a bypass between both axillary arteries.
There was no major complication including cerebral infarction, spinal ischemia, and type I endoleak. The false lumen of the descending thoracic aorta was par­tially patent in two patients. Type II endoleak to which the bronchial artery contributed was observed in one patient. In the other patient blood entering from the natural fenestration came up the false lumen of the des­cending thoracic aorta and flowed into the intercostal artery. The false lumen of the descending thoracic aorta was completely thrombosed in the other 27 patients. Early postoperative complication was observed in one patient who suffered wound infection at the skin inci­sion site.
During a mean follow-up term of 40 Ô 26 months (range 1±93 months) there was no late death (Table
20.4). In addition, no device-related complications were observed. ULP developed at the descending thoracic aor­ta in two patients 2 and 3 years after endografting, re­spectively. Hemoptysis developed in one patient 3 months after endografting, although its cause was not elucidated. Complications which were not related to aortic dissection or devices were observed in two patients, lung cancer in one patient, and abdominal aortic aneurysm independent of aortic dissection in other patient.
Fig. 20.8. Indications of endografting. 1 aortic diameter greater
than 50 mm on CT, 2 aortic diameter greater than 40 mm on CT in a dissection present for 1 month or less, 3 enlargement of an ulcerlike projection during follow-up, 4 persistent back pain despite medical therapy
Table 20.2. Patient characteristics
Age (years) 60 Ô13 Male sex 24 (83%) Maximal diameter of the descending thoracic
aorta (cm) Dissection-related complications 0 (0%) Dissection-unrelated complications 24 (83%)
Hypertension 18 (62%) History of ascending aortic replacement 4 (14%) Abdominal aortic aneurysm/thoracic aortic
aneurysm Coronary arterial disease 3 (10%) Renal failure 2 (7%) Marfan syndrome 2 (7%) Coagulopathy 1 (3%)
Table 20.3. Early mortality and morbidity
Mortality 0 (0%) Complication 5 (17%)
Aorta-related 2 (7%)
Intimal injury 0 (0%) Type I endoleak 0 (0%) Type II endoleak 2 (7%) Aneurysmal degeneration 0 (0%)
Aorta-unrelated 3 (10%)
Stroke 0 (0%) Paraplegia 0 (0%) Left arm ischemia 1 (3%) Dissection of iliac artery 1 (3%) Wound infection 1 (3%)
47 Ô9
3 (10%)
Remodeling of the aorta occurred during follow-up. The aorta and the false lumen shrank, while the true lu­men expanded. However, it is not as conspicuous as in case of acute dissection, in which the false lumen often disappears and the aorta recovers its original diameter.
205
206
https://t.me/med1917
IV. Dissection
Table 20.4. Late mortality and morbidity
Mortality 0 (0%) Complication 5 (17%)
Aorta-related 4 (7%)
Intimal injury 0 (0%) Type I endoleak 0 (0%) Type II endoleak 2 (7%) Aneurysmal degeneration 2 (7%)
Aorta-unrelated 1 (3%)
Hemoptysis 1 (3%)
20.8 Summary
Although further follow-up is mandatory, endografting for the treatment of chronic type B aortic dissection is seemingly a superior alternative to surgical graft re­placement in terms of short and midterm results, even if first-generation endografts are used. To expand indi­cation more sophisticated devices such as branched or fenestrated endografts should be necessary.
References
1. Acute aortic dissection. Lancet 1998; 2:827±828.
2. Blum U, Voshage G, Lammer J, Beyersdorf F, Tollner D, Kretschmer G, Spillner G, Polterauer P. Endoluminal stent-grafts for infrarenal abdominal aortic aneurysms. N Engl J Med 1997; 336:13±20.
3. Coselli J, LeMaire S, Schmittling Z, Kæksoy C. Cerebro­spinal fluid drainage in thoracoabdominal aortic surgery. Semin Vasc Surg 2000; 13:308±314.
4. Dake MD, Miller DC, Mitchell RS, Semba CP, Moore KA, Sakai T. The ªfirst generationº of endovascular stent-grafts for patients with aneurysms of the descending thoracic aorta. J Thorac Cardiovasc Surg 1998; 116:689±704.
5. Dake MD, Kato N, Mitchell RS, Semba CP, Razavi MK, Shimono T, Hirano T, Takeda K, Yada I, Miller DC. Endo­vascular stent-graft placement for the treatment of acute aortic dissection. N Engl J Med 1999; 340:1546±1552.
6. DeBakey ME, McCollum CH, Crawford ES, Morris GC, Howell J, Noon GP, Lawrie G. Dissection and dissecting aneurysms of the aorta: Twenty-year follow-up of five hundred twenty-seven patients treated surgically. Surgery 1982; 92:1118±1134.
7. DeSanctis RW, Doroghazi RM, Austen WG, Buckley ML. Aortic dissection. N Engl J Med 1987; 317:1060±1067.
8. Elefteriades JA, Hartleroad J, Gusberg RJ, Salazar AM, Black HR, Kopf GS, Baldwin JC, Hammond GL. Long­term experience with descending aortic dissection: The complication-specific approach. Ann Thorac Surg 1992; 53:11±21.
9. Fann JI, Miller DC. Aortic dissection. Ann Vasc Surg 1995; 9:311±323.
10. Fleck T, Hutschala D, Weissl M, Wolner E, Grabenwæger M. Cerebrospinal fluid drainage as a useful treatment op­tion to relieve paraplegia after stent-graft implantation for acute aortic dissection type B. J Thorac Cardiovasc Surg 2002; 123:1003±1005.
11. Glower DD, Fann JI, Speier RH, Morrison L, White WD, Smith LR, Rankin JS, Miller DC, Wolfe WG. Comparison
of medical and surgical therapy for uncomplicated des­cending aortic dissection. Circulation 1990; 82(Suppl IV): IV39±46.
12. Glower DD, Speier RH, White WD, Smith LR, Rankin JS, Wolfe WG. Management and long-term outcome of aortic dissection. Ann Surg 1991; 214:31±41.
13. Hagan PG, Nienaber CA, Isselbacher EM, et al. The inter­national registry of acute aortic dissection (IRAD). JAMA 2000; 283:897±903.
14. Huynh D, Miller C, Safi H. Delayed onset of neurologic deficit: significance and management. Semin Vasc Surg 2000; 13:340±344.
15. Inoue K, Hosokawa H, Iwase T, Sato M, Yoshida Y, Ueno K, Tsubokawa A, Tanaka T, Tamaki S, Suzuki T. Aortic arch reconstruction by transluminally placed endovascu­lar branched stent graft. Circulation 1999; 100(Suppl II): II316±321.
16. Ishida M, Kato N, Hirano T, Shimono T, Takeda K. Endo­vascular stent-graft treatment for thoracic aortic aneu­rysms: short- to midterm results. J Vasc Interv Radiol 2004; 15:361±367.
17. Juvonen T, Ergin MA, Galla JD, Lansman SL, McCullough JN, Nguyen K, Bodian CA, Ehrlich MP, Spielvogel D, Klein JJ, Griepp RB. Risk factors for rupture of chronic type B dissections. J Thorac Cardiovasc Surg 1999; 117:776±786.
18. Kato M, Bai H, Sato K, Kawamoto S, Kaneko M, Ueda T, Kishi D, Ohnishi K. Determining surgical indications for acute type B dissection based on enlargement of aortic diameter during the chronic phase. Circulation 1995; 92 (Suppl II):II107±112.
19. Kato N, Hirano T, Shimono T, Ishida M, Takano K, Nishide Y, Kawaguchi T, Yada I, Takeda K. Treatment of chronic aortic dissection by transluminal endovascular stent-graft placement: preliminary results. J Vasc Intervent Radiol 2001; 12:835±840.
20. Kato N, Hirano T, Kawaguchi T, Ishida M, Shimono T, Yada I, Takeda K. Aneurysmal degeneration of the aorta after stent-graft repair of acute aortic dissection. J Vasc Surg 2001; 34:513±518.
21. Kato N, Shimono T, Hirano T, Suzuki T, Ishida M, Saku­ma H, Yada I, Takeda K. Mid-term results of stent-graft repair of aortic dissection: comparison between acute and chronic dissection ± the complication-specific approach-. J Thorac Cardiovasc Surg 2002; 124:306±312.
22. Kato N, Hirano T, Ishida M, Cheng SH, Shimono T, Take­da K. Stent-graft placement combined with percutaneous fenestration for the treatment of aortic dissection with a risk of renal ischemia. Ann Thorac Surg 2004; 78:1072±5.
23. Kouchoukos NT, Dougenis D. Surgery of the thoracic aor­ta. N Engl J Med 1997; 336:1876±1888.
24. Miller DC. Surgical management of aortic dissections: in­dications, perioperative management, and long-term re­sults. In: Doroghazi RM, Slater EE, editors. Aortic dissec­tion. McGraw-Hill, New York, 1983. pp. 193±243.
25. Miller DC. The continuing dilemma concerning medical versus surgical management of patients with acute type B dissections. Sem Thorac Cardiovasc Surg 1993; 5:33±46.
26. Miller DC, Stinson EB, Oyer PE, Rossiter SJ, Reitz BA, Griepp RB, Shumway NE. Operative management of aortic dissections. Experience with 125 patients over a sixteen­year period. J Thorac Cardiovasc Surg 1979; 78:365±382.
27. Miller DC, Mitchell RS, Oyer PE, Stinson EB, Jamieson SW, Shumway NE. Independent determinants of operative mortality for patients with aortic dissections. Circulation 1984; 70(Suppl I):I153±164.
28. Neya K, Omoto R, Kyo S, Kimura S, Yokote Y, Takamoto S, Adachi H. Outcome of Stanford type B acute aortic dis­section. Circulation 1992; 86(Suppl II): II1±7.
29. Nienaber CA, Fattori R, Lund G, Dieckmann C, Wolf W, von Kodolitsch Y, Nicolas V, Pierangeli A. Nonsurgical re-
N. Kato et al. Chapter 20 Endografts for Chronic Dissections
https://t.me/med1917
207
construction of thoracic aortic dissection by stent-graft placement. N Engl J Med 1999; 340:1539±1545.
30. Slonim SM, Nyman URO, Semba CO, Miller DC, Mitchell RS, Dake MD. Aortic dissection: percutaneous manage­ment of ischemic complications with endovascular stents and fenestration. J Vasc Surg 1996; 23:241±253.
31. Slonim SM, Nyman URO, Semba CP, Miller DC, Mitchell RS, Dake MD. True lumen obliteration in complicated aor­tic dissection: endovascular treatment. Radiology 1996; 201:161±166.
32. Smith DC, Jang GC. Radiological diagnosis of aortic dis­section. In: Doroghazi RM, Slater EE, editors. Aortic dis­section. McGraw-Hill, New York, 1983. pp. 193±243.
33. Sueyoshi E, Matsuoka Y, Sakamoto I, Uetani M, Hayashi K, Narimatsu M. Fate of intramural hematoma of the aorta: CT evaluation. J Comput Assist Tomogr 1997; 21:931±938.
34. Svensson LG, Crawford ES, Hess KR, Coselli JS, Safi HJ. Dissection of the aorta and dissecting aortic aneurysms. Circulation 1990; 82(Suppl IV):IV24±38.
35. Tefera G, Acher C, Wynn M. Clamp and sew techniques in thoracoabdominal aortic surgery using naloxane and CSF drainage. Semin Vasc Surg 2000; 13:325±330.
36. Wheat MW. Current status of medical therapy of acute dissecting aneurysms of the aorta. World J Surg 1980; 4:563±569.
37. Williams DM, Lee DY, Hamilton BH, Marx MV, Narasim­ham DL, Kazanjian SN, Prince MR, Andrews JC, Cho KJ, Deeb GM. The dissected aorta: percutaneous treatment of ischemic complications ± principles and results. J Vasc In­terv Radiol 1997; 8:605±625.
Problems Encountered During
https://t.me/med1917
and After Stent-Graft Treatment of Aortic Dissection
Jong Yun Won, Do Yun Lee
Chapter
21
Contents
21.1 Introduction ......................
21.2 Materials and Methods ................ 209
21.2.1 Patients .................... 209
21.2.2 Imaging .................... 210
21.2.3 Stent-Graft ................... 210
21.2.4 Procedures .................. 210
21.2.5 Follow-Up ................... 210
21.3 Problems During the Procedure ...........211
21.3.1 Stent-Graft Migration ............ 211
21.3.2 Complications Related to the Size
oftheStent-Graft...............
21.3.3 Appearance of a Hidden Intimal Tear
intheDTA ..................
21.3.4 Type I Endoleak ............... 212
21.4 Problems During the Follow-Up ...........212
21.4.1 Postimplantation Syndrome ........ 212
21.4.2 Neurologic Complication .......... 214
21.4.3 Persistent Type I Endoleak ......... 214
21.4.4 Type II Endoleak ...............215
21.4.5 Progressive Abdominal Aortic Aneurysm . 215
21.4.6 New Intimal Tear ............... 217
21.4.7 Mechanical Failure of the Stent-Graft . . . 219
21.5 Conclusion ....................... 221
209
211
212
21.1 Introduction
Aortic dissection is one of the life-threatening diseases involving the aorta whether it is acute or chronic [7, 23]. The conventional treatment strategy depends on the extension of the dissecting process: surgical repair of the aortic replacement with prosthetic graft when it involves the ascending aorta (Stanford type A), and medical treatment when it involves the descending tho­racic aorta (DTA) (Stanford type B). However, more than 20% of patients with Stanford type B aortic dissec­tion need surgical intervention for the various indica­tions: an expanding false lumen of dissection, branch vessel and leg ischemia, recurrent pain, or impending rupture [8, 31]. Because of high surgical morbidity and
a mortality rate around 20±60% [9, 34], endovascular stent-graft placement has emerged as a new treatment modality as a substitute for open surgery in aortic dis­section with a descending aortic intimal tear and many authors have reported their encouraging results of more than 60% of complete thrombosis or resolution rate [1, 3, 5, 21, 37].
However, because of the very limited number of pa­tient populations and the duration of follow-up, many questions are still controversial. Especially, only a few reports have concerned the various problems which can be encountered in stent-graft treatment of the aortic dissection [15, 16]. In this article, we present our ex­periences and review other authors' reports about these problems encountered during the procedure of stent­graft placement and the follow-up period in patients with aortic dissections.
21.2 Materials and Methods
21.2.1 Patients
Between July 1994 and December 2003, 60 patients with aortic dissections who underwent stent-graft placement in the DTA were included in this study. The protocol was approved by the Institutional Review Board, and informed consent was obtained from all patients. The 39 men and 21 women ranged in age from 37 to 88 years (mean 59 years). The inclusion criteria for the stent-graft were type A with retrograde dissection and an intimal tear located in the DTA, type B dissection with dynamic obstruction of abdominal aortic branch vessels, persistent or recurrent pain, aortic rupture, and an aortic diameter greater than 40 mm in the acute phase of dissection or greater than 60 mm in the chronic phase [22]. Five of nine patients with type A dissections had retrograde dissection with a primary entry tear located distal to the left subclavian artery (SCA). Therefore, a stent-graft was available for the treatment of this rare type of dissection. The other four
210
https://t.me/med1917
IV. Dissection
patients, including two with Marfan syndrome, with type A dissection had previously undergone surgical re­placement of the ascending thoracic aorta and the aor­tic arch, leaving a large reentry tear in the DTA. Ten patients presented acute (less than 2 weeks) symptoms; 50 patients had chronic (2 weeks or more) symptoms [24]. The location of the primary entry tears were at the proximal DTA in 38 patients and at the mid to dis­tal DTA in 22 patients. All aortic dissections simulta­neously involved the abdominal aorta below the level of the diaphragm, with contiguous extension. The distal extents of the dissections were variable from the supra­renal abdominal aorta to both iliac arteries. All patients showed patent false lumen flows from the proximal pri­mary entry tear to the distal several reentries in the ab­dominal aorta or the iliac artery.
21.2.2 Imaging
Before the procedures, all patients underwent contrast­enhanced spiral computed tomography (CT) scans and digital subtraction aortograms in order to measure the dimension of the aorta and to evaluate entry and reen­try tears. Conventional spiral CT scans of the aorta were evaluated using the following parameters: collimation of 7 mm with a table feed of 7 mm/s, 2-mm reconstruc­tion, 120 kV, and 220 mA s. The patients received 120± 150 ml of contrast medium at a rate of 3.0 ml/s. Re­cently, we used a 16-channel multidetector row CT scan (Somatom Sensation 16, Siemens, Forchheim, Germany) with the following parameter: table feed of 15 mm, gan­try rotation time of 0.5 s, tube voltage of 120 kV, amper­age of 150 mA m, 2-mm reconstruction. Between 10 and 15 s after the end of the arterial phase acquisition, a second acquisition was performed. The aortography was performed with the use of a calibrated marker cath­eter (Cordis, Roden, The Netherlands). We injected 25± 30 ml of contrast medium at a rate of 15 ml/s and ob­tained six to ten images per second for the thoracic aor­ta and two to six images per second for the abdominal aorta. Angiographies of the femoral, external iliac, and common iliac arteries allowed accurate sizing of these arteries and evaluation of tortuosity for the adequate vascular access.
21.2.3 Stent-Graft
We used two different kinds of stent-graft. Until the end of 1999, we implanted a homemade Z-shaped stent covered with balloon-dilated poly(tetrafluoroethylene) (Impra, Tempe, AZ, USA) material in 20 patients. Since then, we have implanted our self-designed percutaneous separated stent-graft that can be deployed percuta­neously through a 12-French low-profile introducer. The
new device consists of two separate bodies: the main body with two anchoring stents at both ends connected by a Dacron graft and the second body of an inner niti­nol bare stent. This unique structure enables percuta­neous deployment and less resistance against blood flow during deployment, so we could implant the stent­graft without any blood pressure, lowering medication [14]. Each stent-graft was individually constructed and manufactured to be 10±15% larger than the diameter of the unaffected aorta proximal to the entry tear of dis­section. The mean diameter and the mean length of the stent-graft were 33 Ô 4 mm (24±40 mm) and 9.6 Ô 2.0 cm (6.0±15.0 cm).
21.2.4 Procedures
The method of Z-shaped stent-graft placement was al­most the same as those described in a previous report [20]. All procedures were performed in angiography units under local or epidural anesthesia and full hemo­dynamic monitoring while placing the patients in the supine position. Vascular access was obtained through the femoral artery following surgical cut-down. After inserting an exchange (260-cm) 0.035-in.-diameter Am­platz superstiff guidewire (Boston Scientific, Watertown, MA, USA), we advanced the 20- or 22-French stent­graft delivery sheath (Keller-Timmerman Sheath, Cook, Bloomington, IN, USA) over the guidewire. The patient was anticoagulated with intravenous heparin (5,000 IU) before the insertion of the stent-graft delivery sheath. After that, we deployed the stent-graft by rapidly with­drawing the introducer sheath. Aortography via a con­tralateral percutaneous femoral artery approach with a 5-French multisidehole catheter (Cook, Bloomington, IN, USA) was performed before and after the stent-graft deployment in order to confirm the exact location of the stent-graft, the absence of perigraft leakage, and the patency of adjacent branch vessels.
Deployment of the percutaneous separated stent­graft needs slightly different steps from conventional Z­shaped stent-graft. First, the main body of the stent­graft is deployed and subsequently the bare stent is de­ployed inside the main body. In addition, deployment of a percutaneous separated stent-graft was performed through a 12-French low-profile introducer under local anesthesia without surgical cut-down. Recently the puncture site has been closed using a Perclose device (Abbott Vascular Devices, Redwood City, CA, USA).
21.2.5 Follow-Up
Follow-up spiral CT scans were performed within 1 month of the procedures, then at 3±6-month intervals for 2 years, and annually thereafter. Each patient was
J. Y. Won, D.Y. Lee Chapter 21 Problems Encountered During and After Stent-Graft Treatment of Aortic Dissection
https://t.me/med1917
211
Table 21.1. List of problems encountered in stent-graft place-
ment for aortic dissection
Problems during the procedure
Stent-graft migration Stent-graft torsion Complication related to the size of the stent-graft Appearance of a hidden intimal tear in the descending thoracic aorta Type I endoleak
Problems during follow-up
Postimplantation syndrome Neurologic complication Persistent type I endoleak Type II endoleak Progressive abdominal aortic aneurysm New intimal tear Mechanical failure of the stent-graft
followed for 9±103 months and the mean follow-up was 31 months. Technical success was defined as the com­pletion of the stent-graft deployment at the target area without stent-graft failure. Clinical success was defined as complete exclusion of the primary entry tear from the circulation without a significant adverse cardio­vascular event by serial CT. We assessed all technical problems during the procedure and complications which occurred during the follow-up period (Table
21.1).
on the follow-up CT scan. The patient is under close observation for 16 months. Since changing the device to a separated stent-graft, the migration problem has no longer happened. Many investigators are known to de­sign their own devices which enable a better proximal fixation, using stent appendages such as hooks or barbs and a rapid deployment system using a pull-string to avoid this unfavorable complication [12, 35].
21.3.2 Complications Related to the Size of the Stent-Graft
The method of measuring the adequate size of the stent-graft in aortic disease seems to be an subject that needs general consensus, especially in aortic dissection. Investigators determine the diameter of a stent-graft in many different ways. We use the diameter of the proxi­mal unaffected aorta as a base line and manufacture the
21.3 Problems During the Procedure
21.3.1 Stent-Graft Migration
Migration of the stent-graft caused by the ªwind sockº effect of ventricular ejection results predominantly in technical failure, particularly when proximal fixation is at or near the aortic arch [2]. To avoid this serious technical problem, lowering the blood pressure with a vasodilator or beta-blocker drugs has been done [16, 25]. Despite this effort, stent-graft migration has re­ported to occur with an incidence of 2±20% by many authors [29].
In our experience, we observed two cases (3%) of stent-graft migration, which occurred in two patients in our early period when we used a single stent-graft sys­tem and nitroprusside to lower the blood pressure dur­ing deployment of the stent-graft. Primary entry tears of those two patients were located at the proximal DTA, 3 and 5 cm from the left SCA. We attempted to correct the location of the stent-graft with a balloon, but failed. We deployed an additional stent-graft and successfully excluded the primary entry tear in one patient. How­ever, in the other patient, an additional stent-graft of adequate size was not readily available. Later, the pa­tient refused further intervention and his false lumen remained patent without demonstrable interval changes
a
b
Fig. 21.1. Stent-graft folding. A 56-year-old man with De Bakey
type III chronic aortic dissection. mography (CT) scan shows the true lumen and the false lumen of the descending thoracic aorta. CT scan after the stent-graft treatment. Despite inward folding of the stent-graft strut (arrow), complete resolution of the tho­racic false lumen is achieved
a Preoperative computed to-
b Fifty-two-month follow-up
212
https://t.me/med1917
IV. Dissection
stent-graft 10±15% larger. Shimono et al. [33] use a stent-graft 10% larger than the average of the maximum and minimum diameters of the thoracic true lumen in acute dissection and 20% in chronic dissection. Choos­ing a stent-graft diameter that is too small can cause an endoleak or migration, whereas oversizing of the pros­thesis may lead to infolding of the graft. In the experi­mental report of Schurink et al. [32], oversizing and folding of the stent-graft were significantly related to the grade of the endoleak, and balloon expansion can not help to reduce the endoleak.
Among 60 patients of our study, two stent-grafts (3%) were infolded during deployment, which was caused by overestimation of the stent-graft size. On the follow-up CT scan, those stent-grafts were not expanded even after the aortic true lumens had been fully ex­panded. Fortunately, primary entry tears of these two patients were completely excluded and there was no thrombus formation around the deformed stent-graft on follow-up CT scans taken 46 and 55 months after the procedure (Fig. 21.1). We keep following these patients closely because of the possibility of blood flow distur­bance inside the aorta by the protruding stent strut.
21.3.3 Appearance of a Hidden Intimal Tear in the DTA
have further intervention because a suitably sized de­vice was not readily available. Despite persistent con­trast enhancement of the thoracic false lumen in these two patients, gradual decrease of the aortic diameter was seen on serial follow-up CT scans at 6 and 27 months each (Fig. 21.2).
In our institute, very thin sections of spiral CT scans of less than 2-mm slice thickness and a three-dimen­sional reconstruction image combined to aortograms with a high frame rate of more than 6 frame/s and oblique views perpendicular to the aortic intimal flap are obligatory for the preoperative evaluation for the aortic dissection. Recently, the multidetector CT scan became available and enables more accurate evaluation of aortic dissection by its greater spatial resolution, thinner slice thickness, and faster scan time [17]. De­spite careful evaluation of the preoperative images, we lost three cases of proximal and distal hidden intimal tears which were relatively small compared with the ad­jacent primary entry tear. Pamler et al. [26] suggested obtaining many preoperative aortograms with the cath­eter tip at different levels. They also propose covering every intimal tear in the thoracic aorta with a stent­graft to prevent distal reperfusion of the false lumen.
21.3.4 Type I Endoleak
Many authors mentioned the importance of verifying every intimal tear of the thoracic aorta on preoperative imaging. Especially in long-standing chronic dissec­tions, it is more important owing to the possibility of multiple fenestrations in the DTA and the abdominal aorta. In chronic dissections, along with fibrosis, con­traction of the myointima, and dilated outer layers, even a small intimal tear could disturb reattachment of the intima to the outer media and adventitia [17, 20]. Ac­cording to our experience, every thoracic false lumen of acute dissection was decreased in size only with the ex­clusion of the primary entry tear. However, in chronic dissection, the size of the thoracic false lumen did not decrease until every intimal tear including the primary entry tear was excluded. In addition, the time required for complete thrombosis of the thoracic aortic false lu­men was longer in the chronic dissections, as was the resolution time. Those differences might be related to the formation of the multiple thoracic reentry tears, the status of endothelialization in the false lumen, and the endotension caused by reentry tears.
In three patients (5%) with primary entry tear in the distal DTA of our series, initially undetected hidden in­timal tears appeared within 2 cm of the proximal (n=1) and the distal (n=2) margins of the stent-graft which had been successfully deployed upon the primary entry tear. All of their dissections were acute. One of them underwent additional stent-graft deployment upon the hidden intimal tear. The other two patients could not
Twelve endoleaks (20%) were detected immediately after the successful deployment of the stent-graft and all of them were type I endoleaks (attachment site endoleak) at the proximal margin of the stent-graft. Ten of the twelve patients with endoleaks had a primary entry tear in the aortic arch and the proximal DTA and one of them was located in the lesser curvature side.
21.4 Problems During the Follow-Up
21.4.1 Postimplantation Syndrome
Postimplantation syndrome can be manifested in var­ious symptoms and signs, such as fever, leukocytosis, elevated C-reactive protein level, pleural effusion, and decreased platelet count.
Fever, leukocytosis, and elevated C-reactive protein level, of which the incidence is known to be 20±60%, are suggested to be a nonspecific systemic inflammatory reaction rather than true infection. However, it has been reported to result in little prolonged complication by many authors [20, 36]. In our early experiences, we rou­tinely medicated prophylactic antibiotics; however, this did not help to lower the incidence of postimplantation syndrome. All the patients recovered with conservative treatments in 2±10 days. Many institutes manage these
J. Y. Won, D.Y. Lee Chapter 21 Problems Encountered During and After Stent-Graft Treatment of Aortic Dissection
https://t.me/med1917
213
a
c
Fig. 21.2. Emergence of hidden intimal tear. A 66-year-old man
with De Bakey type III acute aortic dissection. aortogram demonstrates a large entry tear located at the distal descending thoracic aorta near the diaphragm (arrow). No other entry tear is defined in the descending thoracic aorta. Immediately after the stent-graft deployment, the completion
b
d
a Preoperative
aortogram shows exclusion of the entry tear and nonvisualiza­tion of the false lumen. ment, emergence of a hidden entry tear (arrow) is depicted just above the stent-graft on CT (
b
right anterior oblique view (d). Contrast enhancement of the thoracic false lumen is still noted
c,d One week after the stent-graft treat-
c) and aortograms of 508
complications only with anti-inflammatory agents or even without any medication and recommend judicious use of antibiotics [12, 25, 30].
Pleural effusion and decreased platelet count are not well-known complications of stent-grafts in aortic dis­section. In our series, 34 patients (57%) with pleural ef-
fusion and 22 patients (37%) with decreased platelet count to more than 50% of the base line were observed during the follow-up. All patients with pleural effusion, which we suggest to originate from foreign-body irrita­tion to the adjacent pleura, showed complete resolution on 1-month (n=25) or 3-month (n=9) follow-up CT