Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3823_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
15.09.2026
Размер:
15 Мб
Скачать
☆
Aortic Length Measurements
Calculating the aortic length to repair requires several measurements: the lowest renal artery to the aortic bifurcation, the bifurcation to the right hypogastric artery, and the bifurcation to the left hypogastric artery. These measurements can prove challenging, especially in the case of a tortuous aorta or iliac arteries; however, this can be aided with the use of centerline reconstruction software.
For most devices appropriate wall apposition requires at least 10 mm of distance between the aortic bifurcation and the internal iliac artery. The need for longer iliac limbs typically occurs in patients with more tortuous iliacs. Patients with splayed aortic bifurcations may benefit from “balleting” (crossing) the iliac limbs, which also requires longer limb lengths (Figure 5.2).
The distal seal zone along the iliac vessels also plays a critical role in graft sizing. As with proximal fixation, distal fixation typically requires 10–20% oversizing of the measured iliac diameter and is crucial to prevent type IB endoleaks, iliac aneurysmal degeneration, or iliac thrombosis. Iliac arteries less than 7 mm or greater than 25 mm may render endovascular treatment unsuitable. For patients with concurrent iliac occlusive disease, treatment of such lesions should occur prior to endograft placement, as these aortic endografts do not provide the radial force necessary for treatment of stenotic lesions.
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
Figure 5.2 In cases where aortic anatomy is not suitable
for on‐label device use, graft alteration or “physician‐ modified endografts” as pictured above can provide an endovascular solution in patients who are otherwise poor candidates for difficult open repairs.
For patients with concurrent aneurysmal iliac disease, the Iliac Branch Excluder™ Device (W.L. Gore, Flagstaff, AZ, USA) allows for canalization and preservation of the iliac branch vessels, with exclusion of associated aneurysm. These devices are especially important in patients with occluded contralateral hypogastric vessels.
Step 1. Vascular Access
Percutaneous
Ultrasound‐guided percutaneous access has become increasingly routine in both EVAR and thoracic endovascular aortic repair (TEVAR) over the past decade. Multiple peer‐reviewed studies have demonstrated that percutaneous access offers patients shorter procedure times and length of stay, as well as decreased postoperative pain and access site wound complications [15, 16]. Percutaneous access can be safely achieved with sheaths up to 24 Fr, which accommodate most commercially available devices. Safe closure of the percutaneous access can be achieved with any of the large‐access closure devices available in the market.
As in device sizing and selection, the use of percutaneous access is dependent on patient selection. Those patients with scarred or hostile groins (i.e. previous surgery, radiation, or cancer) tend to benefit from open exposure. Vascular anatomy also plays a key role in decision‐ making. Patients with a high femoral bifurcation or significant calcific or occlusive iliofemoral disease may not be amenable to percutaneous access. Likewise, patients with small iliofemoral vessels, typically women, usually have limited options for delivery systems. While sometimes overlooked, one ought to remember that vascular access is one of the most crucial portions of the
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
EVAR procedure, the one from which most complications arise. Due to the inherent risks, the ability to convert to open for access and surgical closure must be available.
Initial access is similar to that of a standard lower extremity angiogram and should be performed under ultrasound guidance to limit the risk of complications. The optimal access target is along the anterior surface of the common femoral artery between the inguinal ligament and femoral bifurcation. Arterial cannulation can be obtained using either a standard “single‐wall” 18‐
gauge needle or with a “micropuncture” 21‐gauge needle, with wire choice dependent on operator preference and anatomic profile. Access location relative to the femoral head and subsequent placement of guidewire into the iliac system is then confirmed under fluoroscopic imaging.
Open
With the advent and widespread adoption of percutaneous access devices, open access has become less common; however, still has its place in specific patient populations. As mentioned above, open exposure plays an important role to avoid potentially devasting access complications. When performed, a vertical or oblique incision can be utilized. While a vertical incision allows for additional exposure of the femoral vessels, oblique incisions tend to be favored by operators due to decreased rates of wound complications [17].
Iliac Disease and Conduits
Even after successful groin access, iliac occlusive disease may be a major hindrance to endograft deployment. Evaluation of the iliac vessels on preoperative CTA allows for planning sheath navigation and device approach through potentially diseased and tortuous vessels. Wire selection and access during this portion of the procedure is crucial. Any placement of dilator sheaths and larger devices should occur over very stiff wires. Wire selection will be discussed in detail later in
the chapter. Additionally, it is crucial to maintain wire placement across the iliac arteries throughout the duration of the case. In many instances, the damage in diseased iliacs occurs with vessel injury or disruption upon removal of a large bore sheath. In these devastating moments, a stiff wire can provide life‐saving access.
There are several approaches to iliac occlusive disease, dependent on the degree of disease and device used. Often, focal iliac lesions can be treated prior to sheath insertion with balloon angioplasty. For patients with more significant disease that requires stenting, there are two key concepts to remember. First, these aortic endografts do not provide the radial force to maintain long‐term patency in the iliacs, and thus should not be used in lieu of a stent. Second, if stenting is required, it should be done following deployment of the endograft, given the tendency for many bare‐metal stents to migrate with repeated manipulation.
There are several endovascular techniques that allow for access in the case of small or diseased iliac arteries. One option is to create an “internal endovascular conduit” by placing oversized, covered stents in the common and external iliac arteries and balloon‐expanding them to a sufficient diameter. Another is the use of a recently developed balloon‐expandable sheath (SoloPath, Onset Medical Corp, Irvine, CA, USA). The sheath, which contains an incorporated angioplasty balloon, is initially inserted with a size of 14 Fr then expanded to size 24 Fr, enough to accommodate most devices.
In select cases, the use of a surgically placed graft conduit can be performed to completely bypass diseased and calcified iliac and femoral vessels. The conduit should be performed through a right or left oblique retroperitoneal incision. Taking extreme care to stay retroperitoneal, it is possible to expose an adequate segment of common iliac artery to allow for graft anastomosis. An end‐to‐side anastomosis in the distal common iliac artery with 10 mm prosthetic conduit. The conduit is then tunneled under the inguinal ligament into a femoral counter incision to be used appropriately
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
for the remainder of the procedure. At the completion of the EVAR, the conduit may be either ligated, a small segment left for later access, or anastomosed to the common femoral artery.
Step 2. Imaging
Equipment
Successful access and appropriate device selection mean little without quality fluoroscopy equipment. Many institutions today have at least one “hybrid” room, which combines a contemporary fixed‐imaging unit with traditional operating room capabilities. The procedure may also be performed with the traditional “C‐arm” fluoroscopy unit. In both situations, the operator must be comfortably using the equipment efficiently and safely.
Neck Angulation
Many infrarenal aneurysmal necks have varying degrees of angulation, often in an anterior orientation. In order to appropriately correct for this and to remove any parallax in the imaging, a degree of adjustment in the fluoroscopy unit is typically required. This can be calculated using the preoperative CTA in a centerline software system and involves creating an angle perpendicular to the neck of the aneurysm.
Renal Arteries
In a similar fashion, the left renal artery typically originates slightly posterior and lateral on the aorta, requiring adjusting the gantry angle with some degree of left anterior oblique (LAO) to allow for adequate imaging. For patients with accessory renal arteries and normal renal function, these can typically be covered with the endograft without the need for embolization.
Step 3. Wires
With bilateral femoral access established, wire access across the diseased aorta must be achieved. This can initially be done with a soft‐tipped wire (e.g. Glidewire, Terumo, Sunrise, FL, USA) under fluoroscopic imaging. This wire should be exchanged over a soft catheter for a stiff guidewire (e.g. Lunderquist and Amplatz), which should be placed in the distal thoracic aorta. One must ensure that these wires are not withdrawn or advanced during the procedure. Repositioning of a stiff guidewire without fluoroscopic guidance can result inadvertent cannulation of an arch vessel or aortic valve, aortic plaque or thrombus disruption, or aortic dissection.
Once stiff wire access is established via one femoral sheath, a pigtail catheter with radiopaque markers can be advanced over a soft wire in the contralateral sheath. If there is question about possible iliac occlusive disease that requires preprocedure treatment, or if there is concern regarding aneurysmal anatomy (aortic length, renal artery location), angiography and intervention can be performed at this time.
Step 4. Delivery and Deployment
Main Body
Graft Orientation
Prior to insertion, the contralateral gate should be oriented under fluoroscopy. Typically, this involves maintaining orientation based on fluoroscopic graft markers to ensure the gate opens aligned with the contralateral common iliac to facilitate cannulation. The decision to cross the graft limbs is an exception to this orientation. This decision is sometimes guided on initial wire access with splayed bifurcations. If the wires appear to sit crossed low in the aneurysm, the operator should consider crossing the limbs in order to limit endotension, especially in shorter body devices such as the Gore Excluder. For longer body devices like the Cook Zenith, which have the contralateral gate closer to the bifurcation, such a maneuver is less optimal.
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
With orientation confirmed the device should be advanced slowly under fluoroscopy to confirm that positioning is maintained. Any rotation to the graft should be performed while the graft is being advanced to avoid building up unnecessary torque in the device. Care should also be taken to avoid rotating the graft within the proximal neck due to risk of embolic events from aortic thrombus. If there is a large thrombus burden, consider systemic heparinization prior to graft or wire insertion.
If insertion of the endograft device proves difficult, due to tortuous iliacs or distal aorta, there are several troubleshooting options available. The contralateral wire can be exchanged to a stiffer wire to take out any residual tension in the aneurysm or aortic bifurcation. A “buddy wire” of a second stiff wire can also be advanced up the ipsilateral system, to further straighten any tortuous segment.
Proximal Landing Zone
Once the endograft is sufficiently advanced, and both contralateral gate position and appropriate gantry angulation have been set, initial diagnostic imaging can be obtained. The use of a contrast injector is required for adequate opacification of visceral vessels. We prefer an initial angiogram with a high rate and small volume of contrast, typically 20 ml/s with a total volume of 20 ml of full‐strength contrast. This should ensure visualization of both renal arteries.
With the ostium of the renal arteries marked on screen and the fluoroscopy table locked, initial deployment of the graft can begin. Initial deployment should begin above the target position, as many grafts tend to shift or “jump” distally (Figure 5.3). This also allows for the graft to be pulled down slowly during deployment, to a level just below the renal artery. Depending on the device, advancing the graft if it is positioned too low can prove difficult. Some grafts, such as the Excluder, have a repositionable proximal stent, which can be reconstrained within the sheath by the operator to ensure accurate deployment. The goal should be to place
the graft within 0–2 mm of the caudal edge of the lowest renal. Maximal overlap will mitigate the risk of further aneurysmal degeneration, graft migration, or type I endoleak. It is important to know graft specifics, as different grafts will have fabric located at varying distances from the proximal stent.
Figure 5.3 Repeat contrast angiography prior to
complete deployment of the main body endograft can ensure placement of the covered stent just below the level of the lowest renal artery. The above picture demonstrates filling of the left renal artery with the proximal graft partially deployed.
After the proximal landing zone is established, the endograft is further deployed until the contralateral gate is deployed. At this point devices with suprarenal bare stents can now be deployed to set the proximal landing
Данная книга находится в списке для перевода на русский язык сайта https://meduniver.com/
zone in place before gate cannulation and limb deployment.
Contralateral Gate Cannulation
For modular devices, the contralateral gate must be cannulated prior to completion of the main body deployment. This is however not the case in unibody devices such as the AFX™ graft (Endologix, Irvine, CA, USA). Several steps may assist the operator during main body deployment: orientation of the gate slightly anterior facilitates an easier angle of retrograde cannulation from the ipsilateral iliac sheath, balleting or crossing the limbs, and placement of the ipsilateral main body limb up the more tortuous iliac side (Figure 5.4).
Figure 5.4 Cannulation of the contralateral gate can
prove challenging, as demonstrated by the still image above. Operators must be facile with multiple catheters of various orientations to help ensure success.