CHAPTER 12 Cardiovascular hybrid operatingrooms 123
https://t.me/med1917
Fig. 12.2 Aortic stent graft visualized with 3D CT-like
imaging. Courtesy of Professor Fosse, Rikshospitalet, Oslo.
Copyright: Siemens AG HIM AX, Forchheim, Germany
and mortality decreases in comparison to conventional surgery [14,15].
Heart rhythm disturbances
The combination of the surgical epicardial
approach with the interventional endocardial
approach for the treatment of rhythm disturbances
in particular atrial fibrillation offers theoretically
advantages over conventional therapy. First reports
emphasize the potential benefits [16].
Endovascular aortic repair
Endovascular aortic repair (EVAR) for the
abdominal aorta in chronic aneurysms has become
a valid alternative to open repair with superior
survival [17]. EVAR is also increasingly used for the
thoracic aorta (Fig.12.2). In selected cases EVAR
in combination with open surgery is even used
forpathologies of the aortic arch and distal ascending aorta [18]. Endoleaks are common complications of EVAR and may be missed by angiographic
evaluation. Computed tomography (CT) like
imaging with the angiographic C-arm enables the
surgeon to diagnose this complication intraoperatively and correct it, if it is deemed necessary [19].
Pacemaker and implantable
cardioverter defibrillator implantation
Pacemakers and implantable cardioverter defibrillators (ICDs), particularly biventricular systems,
may be optimally implanted in a hybrid OR
environment. This is because the hybrid OR offers
the required superior angulation and imaging
capabilities in comparison to mobile C-arms and
the higher hygienic standards compared to cath
labs.
Interdisciplinary use
The need for hybrid operating theaters is not
restricted to cardiovascular disease. Vascular and
neuro specialists have equally developed hybrid
procedures necessitating angiography systems in
a surgical environment. Imaging needs, hygienic
requirements, and room set up – particularly
for neuro interventions – may be considerably
different. Other surgical disciplines may want to
introduce navigation systems, magnetic resonance
imaging, endoscopy, biplane angiography systems,
or a lateral position of anesthesia equipment.
However, the hybrid ORs are more commonly
shared between vascular cardiac surgeons,
interventionalists including cardiologists, interventional radiologists, electrophysiologists, neuroradiologists, and pediatric cardiologists. Their
specific needs have tobe carefully considered and
weighted when planning the hybrid theater.
Basics of the hybrid room
Location
The hybrid room is used by an interdisciplinary
team of surgeons, interventional cardiologists,
anesthesiologists, and others and it is good practice
to involve all stakeholders deeply in the planning
and keeping of such a facility. Ideally, the hybrid
OR is located next to interventional suites and
operating rooms in order to keep logistics
simple.However, if the ORs are separated from the
interventional cath labs it is recommended to
establish the hybrid OR next to the other operating
rooms [20], because all OR equipment and personnel (e.g. heart–lung machine and perfusionists)
are immediately ready and anesthesia and intensive
care are available.
Room size and preparation
Interventional rooms have excellent imaging
capabilities but frequently lack the prerequisites,
size, and equipment required for formal ORs.
Operating rooms meet those required standards,

124 PART II Structural heart disease
https://t.me/med1917
Fig. 12.3 Set-up of a hybrid OR at Rikshospitalet in Oslo,
Norway. The robotic multi-axis system offers high
exibility and advanced 3D imaging capabilities in the
operating theater. Anesthesia space is maximized in the
but usually lack high-level imaging capabilities.
Ahybrid suite should be larger than a standard OR
and the basic principle for planning is the larger
thebetter, because not only the imaging equipment
needs sufficient space. Staff calculations have
shown that in hybrid procedures 8–20 people are
needed in the team including anesthesiologists,
surgeons, nurses, technicians, perfusionists, experts
from device companies, and so forth [21]. Expert
opinions recommend for newly built ORs atleast
2
70 m
[22]. Additional space for a control room and
a technical room is mandatory, which along with
washing and prep rooms, total approximately 150 m
2
for the whole area. If a fixed C-arm system is considered, an OR size of 45 m
2
is the absolute lower limit.
Rebuilding in terms of lead shielding (2–3 mm) will
be needed. Depending on the system it may be
necessary to enforce the ceiling or the floor to hold
the weight of the stand (c. 650–1800 kg).
Planning
Planning of the hybrid room is truly an interdisciplinary task. Clinicians and technicians of all
involved disciplines should define their requirements and form a responsible planning team. The
concrete planning is refined in several steps by specialized architects and vendors of OR equipment
park position when the system is not needed. Courtesy of
Professor Fosse, Rikshospitalet, Oslo. Copyright: Siemens
AG HIM AX, Forchheim, Germany
and imaging systems in a close feedback loop with
the planning team. Virtual visualization of the
room, visits of established hybrid rooms, and
information exchange with experienced users
help tremendously during the planning process.
Inthe recent literature a number of case studies are
published for planning guidance [23–26].
Lights, monitors, and other devices
In general, all members of the team need access
to all important information. Therefore, multiple
moveable and flexible booms need to be installed
in the hybrid OR. If there are two booms, one
boom on each side of the table will best serve the
team. Collision of the ceiling-mounted displays
with operating lights or other ceiling-mounted
equipment should be avoided. Large displays are
now available capable of showing multiple video
inputs in various sizes, decreasing the need for
multiple screens. A dedicated ceiling plan with
all ceiling-mounted components including air
conditioning should be drawn up to ensure the
function and usability of all devices.
Conventional surgical lights may collide with
the imaging equipment, particularly with ceilingmounted systems. If a laminar air flow is present in
combination with a ceiling-mounted system, light

CHAPTER 12 Cardiovascular hybrid operatingrooms 125
https://t.me/med1917
pendants need very long arms, making them cumbersome to handle. An alternative may be new light
concepts with ceiling-integrated multiple theater
lights, as developed at the Interventional Centre at
Rikshospitalet in Oslo, which solve the problem
ofcollision with the imaging equipment. A remote
control offers the possibility of focusing light
where it is needed (www.lightor.com, accessed 2
November 2012) (Fig.12.3).
Hygiene
Hygienic requirements differ from country to
country and even among surgical disciplines, with
the highest standards in orthopedic surgery. In order
to guarantee highest flexibility in room usage, hospitals tend to equip all ORs according to the highest
standards and that includes a laminar air flow
ceiling. Some hospitals even require skirts around
the laminar air flow field and this set up may preclude ceiling-mounted systems. Ceiling-mounted
systems with running parts above the operating
field, which are difficult to clean and interfere
with the air flow by causing turbulences, are not
recommended from a hygienic standpoint [20].
heat storage capacity of more than 1million HU is
recommended by cardiology societies for cath labs
to avoid overheating and a dangerous shutdown
during complex procedures, which may occur in
mobile C-arms for example during complex EVAR
procedures. For these reasons expert consensus
recommends the use of fixed C-arms [20]. A semimobile system with a fixed generator (80 kW;
AXIOM Artis U, Siemens AG, Forchheim,
Germany) may accommodate high-power imaging
demands even in average-sized ORs too small to
house a fixed C-arm (<45 m
Image intensifier and flat
panel detector systems
2
).
Modern fixed C-arm systems are equipped with a
flat panel detector. Contrast resolution is far higher
compared with image intensifier systems, leading
to a higher image quality in detector systems.
Additionally in image intensifier systems the image
is slightly distorted at the edges compared with the
center. As a consequence image intensifier systems
are not capable of 3D imaging with soft tissue
contrast resolution.
Imaging equipment
Mobile and fixed systems
Mobile C-arms have been commonly used in
vascular and cardiac surgery and they are readily
available in every department, e.g. for pacemaker
implantation. Mobile C-arms may depict larger
stents or catheters well. However, their technical
specifications do not meet the recommendations
of the cardiology societies [27]. The power
(2–25 kW versus 80–100 kW recommendation)
and frame rate (up to 25 frames/s 50 Hz or 30
frames/s 60 Hz versus 30–60 frames/s 50 Hz
recommendation) are below the standards of the
American College of Cardiology (ACC). The
cardiology recommendations need to be met,
because cardiology or neuroradiology interventions are an integral part of the hybrid procedure.
Thin guide wires (0.2 mm) and stents must be
visualized even in obese patients and stenoses
of small vessels have to be quantified, which
requires adequate power. Mobile C-arms generally
have a heat storage capacity of up to 300,000 heat
units (HU) (except rare water-cooled systems). A
Ceiling- and floor-mounted systems
Expert consensus recommends floor-mounted
systems for hygienic reasons. In fact, some hospitals
do not allow running parts immediately above the
operative field, because dust may fall down and
cause infections. Despite these facts, a large number
of hospitals decide to have ceiling-mounted
systems as these cover the whole body with more
flexibility and – most importantly – without
moving the table, which is a sometimes difficult
and dangerous undertaking during surgery because
many lines and catheters have to be moved too.
Some ceiling-mounted systems are capable of 3D
imaging from a surgical position, perpendicular
toLthe patient from both the right and left table
side. Moving from a parking to a working position
during surgery, however, is easier with a flexible
floor-mounted system, because the C-arms just
turn in from the side without interference with the
anesthesiologist, whereas ceiling-mounted systems
can hardly move during hybrid procedures in the
park position at the head side without colliding
with the anesthesia equipment.

126 PART II Structural heart disease
https://t.me/med1917
Fig. 12.4 Example of a room plan with a robotic oor-
mounted imaging system. A room size of 70 m
recommended for hybrid ORs because of the large amount
2
or larger is
Monoplane and biplane systems
In an overcrowded environment like a hybrid suite,
biplane systems add to the complexity and interfere
with anesthesia except for neurosurgery, where
anesthesia is not at the head side. Monoplane
systems (Fig. 12.4) are therefore clearly recommended for rooms mainly used for vascular, cardiac,
and orthopedic procedures. There are exceptions; if
pediatric cardiologists, electrophysiologists, or neuroradiologists are important stakeholders in room
usage, a biplane system may also be considered.
Table considerations
The operating table should meet the expectations
of both surgeons and interventionalists. This is in
fact a special challenge, because the expectations
may be mutually exclusive. Surgeons expect a table
with a breakable tabletop. For imaging reasons the
table has to be radiolucent and should allow
of equipment and people needed. Copyright: Siemens
AGHIM AX, Forchheim, Germany
coverage of the patient over a wide range. Therefore
carbon-fiber tabletops are used that are not
breakable. Cardiovascular surgeons in general do
not have very sophisticated positioning needs and
are used to having fully motorized movements of
the table and the tabletop. For the positioning of
the patients, inflatable cushions are sometimes
used for positioning if a breakable table is not
available. Interventionalists require a floating
tabletop to allow fast and precise movements during angiography, and in some countries floating
tabletops are among the technical requirements for
performing coronary angiographies or are at least
highly recommended by expert consensus [28].
Floating tabletops are not available with conventional OR tables. The radiolucent area of the OR
table only meet the needs in pediatric cases –
complete coverage of an adult can not be achieved
with today’s systems.

CHAPTER 12 Cardiovascular hybrid operatingrooms 127
https://t.me/med1917
As a compromise, tables with a floating tabletop
and vertical and lateral tilt are recommended [21].
Special rails for mounting special surgical equipment such as retractors or camera holders should
be available on the table. Placing the operating
table in a diagonal position in the hybrid suite may
gain space. A crucial element when selecting the
imaging system and table is the possibility of
having access to the patient from all sides and being
able to tilt the table both head up and down and
sidewise. In order to perform 3D imaging with the
operating table, the C-arm has to be fully integrated
with the table because a fast and precise rotation
around the patient lying in the isocenter is
necessary. Breakable OR tables are currently not
fully integrated and therefore 3D CT-like imaging
on these tables is impossible.
Imaging methods and
technologies
Fluoroscopy and acquisition are the basic and most
important imaging modes and are offered by all
systems. Since fluoroscopy needs much less
radiation dose, brilliant fluoroscopy images are the
predominantly used images during the procedure.
However, modern angiography systems offer
advanced imaging and post-processing capabilities
including image fusion with any type of previously
acquired 3D volumes (e.g. CT, magnetic resonance
(MR), positron emission tomography (PET),
single-photon emission computed tomography
(SPECT) images), guidance, or 3D imaging.
Computer tomography-like 3D
imaging with the angiography system
Surgery very much depends on 3D visualization of
the anatomy and therefore 3D CT-like imaging
with the angiography system is an important
feature, because it enables the surgeon to navigate
in real time in 3D volumes. In principle, CT-like
imaging of the heart is performed by one or two
sweeps of up to 220° of the C-arm around the
patient. During the rotation several hundred images
are acquired and then reconstructed as a 3D
volume. If the acquisition is gated by an
electrocardiogram (ECG) 3D volumes over time
can be generated to depict the beating heart.
Radiation dose with new protocols is approximately
Fig. 12.5 Deployment of an Edwards Sapien valve with
3Dguidance. 3D images are acquired at the start of the
procedure with the patient lying on the OR table within
5seconds. The 3D volume of the aortic valve and ascending
aorta are automatically reconstructed and segmented in less
than 1 minute. The coronaries are automatically detected
and marked with coloured dots. A red ring is reconstructed
1 cm below the sinuses. On the workstation the 3D volume
can be rotated so that the ring becomes a line. Once this
isachieved the projection is perpendicular to the aortic
annulus. Through the push of a button the C-arm moves into
a position to give the surgeon exactly the projection shown
on the workstation. The volume is then overlaid over the
live two-dimensional uoroscopy images and supports precise
placement of the aortic valve under rapid pacing. Courtesy of
the Heart Center, Leipzig; prototype software under clinical
trial. Copyright: Siemens AG HIM AX, Forchheim, Germany.
one-third that of a conventional multislice CT. The
OR staff can move out of the OR completely during
a CT-like run, because it lasts in most cases 5 seconds. Reconstruction is performed within 1 minute.
Accurate information of the cardiac anatomy in the
OR supports planning of complex procedures like
redo operations, surgery for complex congenital
heart disease, and transcatheter valve replacement.
Segmentation of anatomic structures and overlay of
the 3D volumes over live fluoroscopy (3D roadmap) enables the surgeon to virtually navigate in
3D anatomy (Fig.12.5). First investigations demonstrate the value of this new technology in transcatheter aortic valve replacement (Video 12.1) [29],
particularly with anatomic adjustable valve prosthesis such as the Medtronic Engager valves
(Medtronic, Minneapolis, MN). In pulmonary
valve replacement the 3D valve anatomy in relation
to the coronary artery anatomy is of great importance inorder to avoid obstruction. In EVAR [19]

128 PART II Structural heart disease
https://t.me/med1917
Fig. 12.6 Functional imaging will be a future application
in a hybrid OR. Angiography systems will be capable of
acquiring blood pool and perfusion-related parameters
by DynaCT imaging. The image shown depicts the
endoleaks can be evaluated in the OR and corrected,
if deemed necessary.
Future trends in hybrid
operating rooms
The wider use of hybrid ORs started recently and the
first dedicated surgical applications and systems aiming to support specific workflows are currently being
developed. There are some specifics in surgery that
need to be addressed further. Extensive training for
surgeons to operate systems is difficult, because these
complex systems are not their foremost specialty.
Therefore specific software applications for surgical
procedures are necessary that are easy to operate with
little or no interaction with the surgeons. A first such
prototype software is demonstrated in Video12.1.
Surgeons are used to 3D anatomy and not so much to
two-dimensional images. Acquisition and reconstruction of 3D volumes with DynaCT and their live
display during fluoroscopy will become standard as
fluoroscopy is today (Video 12.2). The combination
of robotic endovascular navigation with 3D imaging
will allow completely new technologies such as
complete endovascular stenting of the aorta and in
vivo construction of fenestrated branched grafts
(Video12.3). Functional imaging will gain importance since the functional not the anatomic result of
the surgical intervention determines patient outcomes. First steps have been made with the imaging
of tissue blood volume (Fig.12.6).
parenchymal blood volume in a 70-year-old patient with
a non-vascularized tumour. Courtesy of Dr. Mawad,
Houston, TX. Copyright: Siemens AG HIM AX, Forchheim,
Germany
Conclusions
The hybrid OR facilitates a whole new spectrum of
cardiovascular therapies and will therefore become
an essential resource of every cardiovascular center.
The trend towards hybrid techniques is more a revolution than an evolution due to the rapid integration
into the surgical techniques. All areas of cardiac
therapies – procedures for ischemic, structural,
and rhythm heart disease – are deeply affected.
Fluoroscopy represents the basic imaging mode.
Furthermore, image fusion, 3D/4D imaging, soft
tissue visualization, modeling, and navigation allow
very advanced surgical applications. The hybrid
operating suite itself represents an extremely complex
working environment that demands careful planning
by all stakeholders. Bundling of clinical, technical,
and architectural expertise as well as a realistic view
of what is achievable is key for a successful hybrid OR
project. Due to wide variations in utilization, generic
recommendations are only of limited use for these
highly individual rooms and certainly cannot replace
the diligent work of the project team. However,
once the room is successfully established it really
transforms surgical techniques and paves the way for
revolutionary new minimally invasive therapies.
References
1 Fogarty, AM. Angioscopy. New developments in vascular
surgery. AORN J 1991;53(3):725–8.
2 Neumann FJ. EuroPCR (abstracts book) 2009.

CHAPTER 12 Cardiovascular hybrid operatingrooms 129
https://t.me/med1917
3 Sternbergh WC, 3rd, Tierney WA, Money SR.
Importance of access to fixed-imaging fluoroscopy:
practice implications for the vascular surgeon. J
Endovasc Ther 2004;11:404–10.
4 Greelish JP. Routine angiography after bypass. Paper
presented at the Trans Catheter Therapeutics Meeting,
Washington, October 12–15, 2008.
5 Bacha EA, Marshall AC, McElhinney DB, et al.
Expanding the hybrid concept in congenital heart
surgery. Semin Thorac Cardiovasc Surg Pediatr Cardiol
Surg Annu 2007;146–50.
6 Sivakumar K, Krishnan P, Pieris R, et al. Hybrid approach
to surgical correction of tetralogy of Fallot in all patients
with functioning Blalock Taussig shunts. Catheter
Cardiovasc Interv 2007;70:256–64.
7 Li SJ, Hu SS, Zhang H, et al. One-stop hybrid cardiac sur-
gery for neonates and young children with congenital
heart disease. Zhonghua Xin Xue Guan Bing Za Zhi
2009;37:986–9.
8 Bacha EA, Daves S, Hardin J, et al. Single-ventricle palli-
ation for high-risk neonates: the emergence of an
alternative hybrid stage I strategy. J Thorac Cardiovasc
Surg 2006;131:163–71.e2.
9 Gutgesell HP, Lim DS. Hybrid palliation in hypoplastic
left heart syndrome. Curr Opin Cardiol 2007;22:55–9.
10 Walther T, Chu MW, Mohr FW. Transcatheter aortic
valve implantation: time to expand? Curr Opin Cardiol
2008;23:111–16.
11 Vahanian A, Alfieri OR, Al-Attar N, et al. Transcatheter
valve implantation for patients with aortic stenosis: a position statement from the European Association of CardioThoracic Surgery (EACTS) and the European Society of
Cardiology (ESC), in collaboration with the European
Association of Percutaneous Cardiovascular Interventions
(EAPCI). Eur J Cardiothorac Surg 2008;34(1):1–8.
12 Mack MJ. Intraoperative coronary graft assessment. Curr
Opin Cardiol 2008;23:568–72.
13 Zhao DX, Leacche M, Balaguer JM, et al. Routine intraop-
erative completion angiography after coronary artery
bypass grafting and 1-stop hybrid revascularization results
from a fully integrated hybrid catheterization laboratory/
operating room. J Am Coll Cardiol 2009;53(3):232–41.
14 Bonatti J, Schachner T, Bonaros N, et al. Simultaneous
hybrid coronary revascularization using totally endoscopic left internal mammary artery bypass grafting and
placement of rapamycin eluting stents in the same interventional session. The COMBINATION pilot study.
Cardiology 2008;110:92–5.
15 Reicher B, Poston RS, Mehra MR, et al. Simultaneous
“hybrid” percutaneous coronary intervention and minimally invasive surgical bypass grafting: feasibility, safety,
and clinical outcomes. Am Heart J 2008;155:661–7.
16 Bisleri G. Combined surgical and interventional
electrphysiology approaches to atrial fibrillation: the hybrid
ablation procedure. Paper presented at the Trans Catheter
Therapeutics Meeting, Washington, October 12–15, 2008.
17 Schermerhorn ML, O’Malley AJ, Jhaveri A, et al.
Pomposelli F, Landon BE. Endovascular vs. open repair
of abdominal aortic aneurysms in the Medicare
population. N Engl J Med 2008;358:464–74.
18 Walsh SR, Tang TY, Sadat U, et al. Endovascular stenting
versus open surgery for thoracic aortic disease: systematic
review and meta-analysis of perioperative results. J Vasc
Surg 2008;47:1094–8.
19 Biasi L, Ali T, Ratnam LA, et al. Intra-operative DynaCT
improves technical success of endovascular repair of
abdominal aortic aneurysms. JVasc Surg 2008;Nov 26.
20 Bonatti J, Vassiliades T, Nifong W, et al. How to build
a cath-lab operating room. Heart Surg Forum
2007;10:E344–8.
21 ten Cate G, Fosse E, Hol PK, et al. Integrating surgery and
radiology in one suite: a multicenter study. J Vasc Surg
2004;40(3):494–9.
22 Benjamin ME. Building a modern endovascular suite.
Endovasc Today 2008;3:71–8.
23 Eagleton, MJ, Schaffer, JL. The vascular surgery operating
room. Endovasc Today 2007;8:25–30.
24 Hirsch R. The hybrid cardiac catheterization laboratory
for congenital heart disease: from conception to completion. Catheter Cardiovasc Interv 2008;71:418–28.
25 Peeters P, Verbist J, Deloose K, et al. The catheterization
lab of the future. Endovasc Today 2008;3:94–6.
26 Sikkink CJ, Reijnen MM, Zeebregts CJ. The creation of
the optimal dedicated endovascular suite. Eur J Vasc
Endovasc Surg 2008;35:198–204.
27 Bashore TM, Bates ER, Berger PB, et al., American
College of Cardiology, Task Force on Clinical Expert
Consensus Documents, American College of Cardiology/
Society for Cardiac Angiography and Interventions
Clinical Expert Consensus Document on cardiac catheterization laboratory standards. A report of the American
College of Cardiology Task Force on clinical expert consensus documents. J Am Coll Cardiol 2001;37:2170–214.
28 Hamm CW, Bösenberg H, Brennecke R, et al., German
Society of Cardiology-Heart and Cardiovascular Research.
Guidelines for equipping and managing heart catheter
rooms (1st revision). Issued by the governing body of the
German Society of Cardiology-Heart and Cardiovascular
Research. Revised by order of the Committee of Clinical
Cardiology [in German] Z Kardiol 2001;90:367–76.
29 Kapadia SR. Role of CT angiography to evaluate patient
eligibility and guide transcatheter aortic valve eligibility.
Paper presented at the Trans Catheter Therapeutics
Meeting, Washington, October 12–15, 2008.
Videoclips
This chapter contains the following videoclips:
Video 12.1 Aortic valve replacement with the
help of intraoperative DynaCT

130 PART II Structural heart disease
https://t.me/med1917
imaging and overlay of
reconstructed 3D volumes on live
uoroscopy. e video is described
comprehensively in the legend of
Fig.12.3. Courtesy of the Heart
Center, Leipzig; prototype soware
under clinical trial.
Video 12.2 e abdominal aortic aneurysm was
diagnosed with multislice CT. e
rendered 3D volume of the aorta was
registered with an intraoperatively
obtained DynaCT volume. Contours
ofthe 3D volume are then overlayed on
live uoroscopy and are automatically
adjusted by changes of projection,
tableposition, zoom factor, etc.
Courtesy of St. Olav’s Hospital,
Trondheim; prototype soware under
clinical trial.
Video 12.3 Intraoperative DynaCT of a fenes-
trated stent-gra for the repair of
acomplex abdominal aneurysm.
Courtesy of Dr. Roy Greenberg,
Cleveland Clinic Foundation,
Cleveland, OH.
They can be accessed at www.wiley.com/go/
valverepair.

13
https://t.me/med1917
CHAPTER 13
Access techniques for transcatheter
aortic valves
Jacques Kpodonu
University of California Irvine,
Cardiovascular Hybrid Interventions, Hoag Heart and Vascular Institute, Newport Beach, CA, USA
Introduction
Aortic stenosis is a common valvular degenerative
process in the adult population, with increased
prevalence in advanced age. It is estimated that
300,000 patients have severe aortic stenosis in
theUSA, and approximately 60,000 undergo aortic
valve replacement (AVR) every year. Within 3 years
of the onset of angina, syncope, or congestive
heart failure, 75% of patients die (2% per month
mortality) unless treated with aortic valve
replacement. The 3-year survival rates have been
reported to be 87% and 21% for operated and
non-operated patients, respectively [1]. Even in
apparently asymptomatic patients, when the valve
area is 0.8 cm
without AVR [2]. The mortality rate for AVR is
approximately 3–4% in the USA, but it increases
with higher baseline risk factors, reaching 20–50%
in the highest risk groups [3]. Patients in the
highest risk group are considered inoperable. High
surgical risk can be defined as a logistic EuroScore
calculated risk of mortality ≥15%, the Society of
Thoracic Surgeons’ score risk of mortality ≥10%,
or the presence of other comorbidities rendering
conventional aortic valve surgery difficult, such as
a porcelain aorta, previous cardiac surgery with
thepresence of patent grafts, or severe adhesions,
previous radiation therapy, liver cirrhosis, need to
2
or less, the mortality rate is high
avoid sternotomy due to patient immobilization,
or marked patient frailty. The introduction of a
transcatheter option for AVR [4–6] may provide a
potential alternative solution for patients considered high risk for aortic valve replacement.
Patient selection
The selection of patients for transcatheter AVR
requires a thorough assessment of the access vessel
sites [7]. Safe performance of transcatheter aortic
valve procedures and other complex endovascular
procedures for structural heart disease requires
zero tolerance for major access-related complications. Thorough preoperative planning, understanding the pathology of aorto-iliac occlusive
disease, advanced endovascular skills, and ability
to use alternate access sites (including being able
to perform an iliac conduit via a retroperitoneal
approach) are necessary to achieve excellent results.
Furthermore, deliverability of large sheath devices
through tortuous anatomy or old graft material
may be improved by more proximal access provided
by an iliac conduit [1]. An evaluation of the iliac
and common femoral arteries is important to make
certain that these vessels will allow the delivery of
sheaths ranging from 18 to 25 Fr. A minimum
diameter of 9 mm (3 Fr ≈ 1 mm) may be necessary
Endovascular and Hybrid Therapies for Structural Heart and Aortic Disease, First Edition.
Edited by Jacques Kpodonu and Raoul Bonan.
© 2013 John Wiley & Sons, Ltd. Published 2013 by John Wiley & Sons, Ltd.
131

132 PART II Structural heart disease
https://t.me/med1917
for safe access via the common femoral arteries.
A detailed quantitative angiogram and computed
tomography (CT) are performed and analyzed for
diameter, tortuosity, and calcification; occasionally
intravascular ultrasound of these vessels is performed for further clarification. Expect knowledge
of the devices, access techniques, and bail-out
maneuvers are necessary for a successful transcatheter aortic valve program.
Transcatheter aortic
valve devices
The two front runners seeking to bring transcatheter
aortic valves to the market in the USA include
Carpentier-Edwards Lifesciences (Irvine, CA; the
Sapien transcatheter aortic valve) and CoreValve
Inc. (Irvine, CA; the revalving device). A thorough
knowledge of both devices and their deployment
characteristics are essential for successful deployment of transcatheter aortic valves.
The CoreValve prosthesis consists of a trileaflet,
bioprosthetic, porcine, pericardial tissue valve,
which is mounted and sutured in a self-expanding
nitinol stent (Fig.13.1). The prosthetic frame/stent
is manufactured by a laser cutting tool and has an
overall length of 50 mm. The lower portion of the
prosthesis has high radial force to expand against
the calcified leaflets and to avoid recoil. The middle
portion carries the valve and is constrained and
narrower to avoid the coronary arteries, while the
upper portion is flared to center and fix the stent
firmly in the ascending aorta and to provide
longitudinal stability. The device is deployed via
a delivery system into which the valve is loaded
shortly before implantation. The CoreValve
transcatheter valve is manufactured in two sizes,
with the smaller prosthesis generally used for aortic
annulus sizes ≤23 mm; the larger valve fits annulus
sizes up to 27 mm in diameter. The sizes of the
CoreValve delivery systems have been gradually
reduced over time to allow easier deployment.
Generally, a femoral artery is chosen for device
placement due to the 18 Fr profile of the catheter,
however alternative access sites – including the
axillary artery with a cutdown procedure – are
possible in cases of inadequate femoral access.
The Sapien transcatheter aortic valve is composed
of a balloon-expandable, stainless steel frame with
an integrated trileaflet, bovine, pericardial valve
(Fig.13.2). The pericardial leaflet material is treated
with a similar process to the oneused for the surgical Carpentier-Edwards Perimount Magna pericardial valves (thermafix anticalcification treatment,
leaflet deflection testing for matched elasticity, and
proprietary tissue processing). The valve is available
in two sizes (23 and 26 mm) to achieve optimal matching with the aortic annulus dimensions. The valve
is crimped on a balloon just before implantation
Corevalve
Fig. 13.1 The CoreValve transcatheter aortic valve system
consists of pericardial leaets attached to a self-expanding
nitinol frame. In the deployed state, the ared distal end
assists in anchoring in the ascending aorta. The stent
covers the coronary ostia, but cell size is designed to allow
coronary artery perfusion and later coronary
catheterization.
Fig. 13.2 The Carpentier-Edwards Sapien aortic valve.