Добавил:
Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Optical methods of analysis. Educational aid

.pdf
Скачиваний:
0
Добавлен:
07.09.2026
Размер:
2 Мб
Скачать
21
out by measuring the absorbance of the analyzed (or standard) solution with different light filters.
2. Optical density. The absorbance of a solution is influenced by a num- ber of factors: the nature of the solvent, the pH of the solution, the tempera­ture, the concentration of the electrolyte, and the presence of foreign matter. It is necessary to choose the analytical conditions so that small changes in these factors do not significantly change the absorbance. One of the ways is to measure the optical density relative to the reference solution, which can be used as a blank sample: a solvent, a sample without coloring reagents, a solvent with reagents.
The main source of random errors in photometric analysis is the meas­urement of optical density, i.e. measurement of the intensity of the light flux after passing through the test solution, as well as through the reference solu­tion. It has been shown that the determination of concentration by photometry is carried out with an error of 1–2 %, if the concentration of the analyzed sub­stance is such that the value of the optical density is in the range of 0.1–1.0. This is due to the fact that the error distribution in photometry has the form shown in Fig. 3.3.
Fig. 3.3. Relative error of photometric determination
3. Choice of cuvettes. A set of cuvettes with different distances between the inner working faces makes it possible to choose a cuvette of such a work­ing length that the measurements are carried out on the section of the optical density scale that gives the smallest relative measurement errors. A cuvette can be considered correctly selected if the value of optical density for the most concentrated solution does not exceed 0.7–0.8. The working faces of the cuvette in the cuvette holder are set perpendicular to the light beam. It should be remembered that on one of the two working plane-parallel faces
of the cuvette there are numbers that indicate the thickness of the light-ab­sorbing layer of the solution in millimeters. When working, these edges must be clean, free of liquid drops and fingerprints. The solution is poured into the cuvette a little higher than the risk, which is applied on one of the faces of the cuvette.
4. The lower limit of detection of a substance by the photometric method (method sensitivity). The sensitivity of photometric methods is quite high. The lower detection limit of a substance is calculated by the ratio:
с
min
=A
min
/(εl).
If we take A
= 0.01, l = 5 cm, ε = 103–105, then с
min
= 10-5–10-7 mol/l.
min
22
4 . M E T H O D S F O R D E T E R M I N I N G
C O N C E N T R A T I O N I N O P T I C A L M E T H O D S
OF A N A L Y S I S
4 . 1 . C a l i b r a t i o n c u r v e m e t h o d
Usually, four methods are used to determine the concentration from the measured value of the optical density of the test solution: the calibration curve method, the comparison method, the standard addition method, and the differential photometry method.
In the first three methods, a solvent or a solution prepared with the ad­dition of all reagents, but without the analyte, is used as a reference solution.
In the calibration curve method, a number of standard solutions with a precisely known concentration are prepared in volumetric flasks, the ap­propriate reagents (reagents) necessary for the formation of a colored com­pound are added, the solution volumes are adjusted to the mark with a sol­vent, and the optical densities of the resulting solutions are measured. Ac­cording to the measurement data, a dependency graph is built in coordinates: A – concentration. It should have a linear character when the Bouguer–Lam­bert–Beer law is fulfilled (Fig. 4.1). Moreover, to build a graph, it is neces­sary to prepare 5–7 standard solutions with different concentrations and carry out at least 5 parallel measurements of the optical density of solutions of each concentration. This will allow, with a minimum error, to build a calibration graph using the method of mathematical statistics – the method of least squares (LSM).
Similarly to the preparation of standard solutions, a solution of the in­vestigated (analyzed) solution is prepared, its optical density is measured, and, using the constructed calibration graph, the content of the substance in the volume of the analyzed sample is determined. The desired concentration can be determined by calculation, based on the equation of the obtained straight line
А = bс,
where b = tgφ, and φ is the angle of inclination of the straight line.
23
24
Then
с = А/ tgφ = А・k,
where k = сtgφ and is called the sensitivity coefficient.
The formula expression of the calibration dependence allows computer processing of the measurement results. If the dependence is non-linear, i.e. the Bouguer–Lambert–Beer law is not fulfilled, to find the concentration from the optical density value, only the calibration graph is used.
Fig. 4.1. Calibration curve
It is advisable to use the method for a large number of measurements of the same substance in the absence of interfering components in the ana­lyzed solution.
4 . 2 . C o m p a r i s o n m e t h o d
In the case when a single analysis is carried out, a comparison method is used. In the comparison method, the obtained value A of the test solution is compared with the value of the standard solution only if the Bouguer– Lambert–Beer law is fulfilled. For a standard solution and a solution with an unknown concentration, the following is true:
А
х
= εlсх and Аst = εlсst.
25
Solving the equation for cx, we get:
с
х
= сst Ах/Аst.
The method gives the most accurate results when the ratio of the optical densities of the studied and standard solutions is close to unity, i.e. for com­parison, choose the concentration of the standard solution that is closest to the one being determined:
А
х/Аst
= 1.
Under this condition, the Bouguer–Lambert–Beer law is satisfied with a higher probability, i.e. there is a directly proportional dependence of Ast on cst and Ax on cx.
To achieve this, 3–4 standard solutions are prepared, their optical den­sities are measured and the most suitable Ast value is selected from them for calculating cx.
4 . 3 . S t a n d a r d a d d i t i o n m e t h o d
If there are components in the test solution that contribute to the value of optical density, the calibration curve method becomes unsuitable, since the conditions for measuring the optical density of standard solutions without impurities and the test solution with interfering components are different. To eliminate the error in determining the concentration in this case, the method of standard additions is used.
In the standard addition method, equal volumes of the test solution are placed in a series of volumetric flasks (usually 4). Then, they do not add to the first flask, but add various volumes of the standard solution to the rest, add the appropriate reagents (reagents), bring the volume of the solution to the mark with a solvent, and measure the optical densities of the resulting colored solutions. It is nessesary build a graph of the dependence of optical density on the amount of additive. This is a calibration graph obtained against the background of the analyzed solution with interfering components. The concentration of the test solution (the amount of substance in the sample) is determined according to the graph (Fig. 4.2), where c1, c2, c3 are the con­centrations of the added standard solution, mg in the volume of the flask; Ax,
26
A
x+1
, A
x+2
, A
x+3
are the optical densities of the test solution and the test solu­tion with additives. The segment 0 - Cx expresses the desired concentration (amount) of the analyte in the volume (sample) of the solution.
The concentration of the test solution can also be determined by com­paring the optical density of the test solution and the solution with the small­est additive. We get two equations based on the Bouguer–Lambert–Beer law:
А
х
= εl・сх ,
А
х + n
= ε l с
х + n
= ε・l・(сх + сn),
where cn is the concentration of the additive in the analyzed solution.
Further, these equations are solved with respect to cx.
Fig. 4.2. Determination of the concentration of the solution by the method
of standard additions
The advantage of the addition method is that this method eliminates the systematic error introduced into the determination of concentration by the influence of impurities, since it creates the same conditions for photom­etry of the test solution and solutions with a standard additive. In this regard, it is particularly suitable for the determination of traces of analytes in the presence of large amounts of foreign components.
4 . 4 . D i f f e r e n t i a l p h o t o m e t r y m e t h o d
The differential photometry method is mainly used to determine large concentrations when the optical density value exceeds unity. In this method,
27
a series of standard solutions are prepared and their optical density is meas­ured relative to a reference solution, which is the standard solution with the lowest concentration from a series of given ones. The amount of the an­alyte is determined either according to the calibration curve A = f (c) (Fig. 4.3), or by calculation using the comparison method:
cх = с
comp
+ FАх,
where F is the conversion factor; Aх and Ast are the optical densities of the test and standard solutions, measured relative to the reference solution; cst is the concentration of the substance in the standard solution; c
comp
is
the concentration of the substance in the reference solution.
The method of differential photometry makes it possible to expand the range of concentrations determined using photometry and to carry out measurements with high accuracy.
Fig. 4.3. Calibration curve in the method of differential photometry
In some cases, extraction-photometric analysis is used to eliminate the influence of interfering components and increase the sensitivity of the de­termination. In this method, the microcomponent to be determined is con­verted into a water-soluble colored compound, extracted with a suitable sol­vent, and the extract is photometrically measured. The content of the compo­nent is determined by the methods discussed above, and the standard solutions are carried out through all stages of the analysis. The concentration of the mi­crocomponent in the solvent phase and the separation of interfering compo­nents occur during extraction. This method is used, for example, in the analy­sis of anionic surfactants in water, in the determination of metals in water in the form of dithizonates in an organic solvent that is added to the analyzed solution.
28
5 . L A B O R A T O R Y W O R K
ON S P E C T R O P H O T O M E T R Y A N D P H O T O M E T R Y
5 . 1 . L a b o r a t o r y w o r k o n p h o t o m e t r y
The purpose of photometric determination: calculation of the amount of a component from the value of the optical density of the solution.
Relationship equation: A = f (C).
Applied methods – calibration curve methods, additive method, calcu­lation method.
5 . 1 . 1 . C h o i c e o f o p t i m a l c o n d i t i o n s f o r p h o t o m e t r i c
det e r m i n a t i o n s
As follows from the basic law of absorption, the optical density A of a so­lution of a component with concentration C will be determined by the values of
ε
λ
and l, which depend on the conditions of the study. Since monochromatic radiation is not used in photometry, the absorption value will be determined by the characteristic of the light flux that has passed through the filter.
The choice of a light filter is made in such a way that the transmitted light flux is absorbed as fully as possible by the component to be determined. This is achieved when the minimum in the absorption spectrum of the optical filter coincides with the maximum of the absorption band of the component. It is also possible to select a light filter according to the principle of complementary color to the color of the component. The optimal light filter (from those available in the device) can be selected by comparing the optical density of the analyzed solution obtained with different filters. Solutions are poured into cuvettes (30 mm), and in one – a reference solution, in the other – one of the reference solutions of the component under study. Next, determine the optical density of the solution with all the filters available in the device. Choose such a light filter, using which the optical density will be the highest.
29
Cuvette selection. Relative errors of photometric determinations will be minimal when the optical density of solutions is 0.4–0.5. Therefore, they try to choose a cuvette with such a working length that the optical density of the reference solution with the highest concentration of the substance does not exceed 1.0.
In the future, the construction of a calibration curve and the measurement of the concentration of the test solution are carried out using the selected cuvettes and light filters.
5 . 1 . 2 . P h o t o m e t r i c d e t e r m i n a t i o n o f c o p p e r
The absorption value of solutions containing the Cu2+ ion is insufficient to determine low concentrations of copper. The colored ammonia complex of copper absorbs much more intensively in the visible part of the spectrum. The determined copper ion is converted into this complex:
Cu2+ + 4NH3·H2O = [Cu(NH3)4]2+ + 4H2O.
The optical density of the copper ammonia solution will be proportional to the concentration of copper. Determination of copper can interfere with cations, which also form colored ammonia (Co2+, Ni2+ and others).
Equipment and reagents:
1. Photometer with a set of cuvettes, 50 ml volumetric flasks, 5 ml vol-
umetric cylinder.
2. Copper salt standard solution with a known concentration (Cst) of
about 10-3 g/ml.
3. Solution NH3·H2O (1:1).
Construction of a calibration graph:
1. Prepare 5 copper ammonia standard solutions using volumes Vst of
1.00; 3.00; 5.00; 8.00; 10.00 ml of standard copper salt solution. First, a so­lution with the highest concentration is prepared, with the help of which a light filter and a cuvette are selected. If there are not enough volumetric flasks, then the flask must be thoroughly washed before preparing the next solution.
A certain volume of a standard copper salt solution is introduced into a 50 ml volumetric flask using a burette, 5 ml of ammonia solution is added
30
and the volume is adjusted to the mark with distilled water to prepare a ref­erence solution. Stopper the volumetric flask and mix the resulting solution thoroughly.
2. Calculate the concentration of copper in the prepared standard solu-
tions (C
Cu
2+
).
When constructing a calibration graph, the concentration of the С
ref
so-
lution can be expressed as a titer T (g/ml) or mg/ml.
The calculation method uses molar concentrations of solutions. There­fore, it is necessary to find the concentrations of standard solutions in other units (mol/l). The recalculation is carried out according to the formula:
С
ref
(mol/l) = С
ref
(g/ml)


.
The calculated concentrations are entered in the table.
3. The reference solution is prepared as follows: 5 ml of ammonia so­lution is placed in a 50 ml volumetric flask and adjusted with distilled water to the mark. In this case, it is acceptable to use distilled water as a comparison solution.
4. Using the standard solution with the highest concentration of copper, choose a filter and a cuvette with the optimal working length according to the method described in subsection 5.1.1.
5. Under the chosen conditions, measure the optical density A of the reference solutions, starting from the minimum concentration. Each de­termination is repeated 3 times (A1, A2, A3) finding the average values of Ā for each solution.
6. Calculate the value of ε for each solution.
7. The values, obtained during measurements and the calculated values entered in the Table 5.1.
Table 5.1
Results of determinations
standard
V
orig
,
ml
С
Cu
2+
,
g/ml
С
Cu
2+
,
mol/l
А1
А2
А3
Ā
ε
General
terms
t = l = С =
Light filter
8. Build a calibration graph in the coordinates Ā – СCu, g/ml.