Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_1110_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
02.09.2026
Размер:
20 Мб
Скачать
8 Benign Anorectal Diseases
Fig. I.9. Diagram of echographic equip-
ment with amplitude mode (A-mode) display: transmitter (a), receiver (b), oscilloscope (c), trigger signal (d), pulse electrical signal (e), piezoelectric trans­ducer (f), pulse mechanical signal (g), medium 1 (h), medium 2 (i), medium 3 (l)
distance from the probe along a single observa­tion direction (A-mode, or amplitude mode), or the return signals may be displayed as luminous dots on the screen (B-mode, or brightness mode), the brightness of which is amplitude dependent. Grey scale means that the brightness is assigned to one of a fixed series of shades of grey, usually 256 in number. The illuminate dots are displayed on the screen in the same sequence of time and geometry according to which the ultrasonic signal encounters a differing acoustic impedance along its path while the ultrasonic beam moves in the scanning plane. This system permits two-dimen-
sional imaging of a section (tomography) in the scanning plane. If a sequence of scans is per­formed at a sufficiently high frequency, a continu­ous,real-time display is supplied of sections of the organ even though the organ is in motion. Information on the absorption of different tissues and the depth of the organ to be studied allow a choice of probes for various applications. The highest frequencies give better resolution and bet­ter display of detail, but the tradeoff for this is a short focal zone and limited penetration. With a mechanically rotated single crystal, this is fixed by the design of the transducer. Focusing describes the optimum range of resolution for the transduc-
er to work at. For the B-K Medical 10-MHz trans­ducer, the axial resolution is <0.05 mm and the lateral resolution 0.5–1 mm, with a focal range of 5–45 mm [1–2].
Artefacts
Reverberation is an artefact due to a gross mis­match of acoustic impedance at an interface, usu­ally an air/tissue interface, causing a very strong reflection which is only partially absorbed by the transducer and which is reflected back to the interface where it is again reflected back to the transducer. The beam therefore reverberates back and forth from the interface and transducer, cre­ating a series of reflections separated by a con­stant distance, which indicates the distance of the reflection from the transducer (Fig. I.10). Reverberation echoes are common in anal
endosonography when there is loss of acoustic contact within the canal (Fig. I.11).
Interpretative Principles
The prime determinant of the strength of ultra­sonic echoes is the impedance mismatch between adjacent tissue components. The larger the mis­match, the stronger the echo.The echogenicity of any structure may be characterized by the level of echoes within it (hyper- or hypoechoic).The echo pattern may be homogeneous or inhomoge­neous. Water has the lowest reflectivity and appears black.Fat varies in reflectivity but is usu­ally moderately reflective. Collagen is more reflective and whiter whereas muscle fibers are
Section I • Fundamental Principles of Ultrasound Imaging 9
Fig. I.10. Reverberation echoes develop
when there is a very strong reflection returning to the transducer. Part will be reflected from the transducer back into the tissue. This, in turn, will be reflected back to the transducer, and so on. This creates a series of equally spaced reflec­tions. The gap indicates the distance between the transducer (T) and the initial reflection
a b
Fig. I.11. The image has been taken with the probe free standing in air. The gas impedance mismatch at the cone/air interface
creates major reflection echo patterns with an equally spaced series of concentric rings (a).The probe has been inserted beyond the anal canal into the rectal ampulla.Acoustic contact has been lost from 9–11 o’clock, producing a reverberation echo (b)
poorly reflective and blacker. Solid tumors are usually poorly reflective and inhomogeneous. While uniform regions of fiber or fat are echo­poor, admixtures between them and watery tis-
References
1. Bartram CI, FrudingerA (1997) Basic principles of ultra­sonography. In: Bartram CI, Frudinger A. Handbook of anal endosonography. Wrightson Biomedical, Peters­field, UK
sues give stronger echoes. For a given impedance mismatch, a region that contains a large number of scatterers is more echogenic than one where they are spread out.
2. Santoro GA, Di Falco G (2004) Basic principles of ultrasonography. In: Santoro GA, Di Falco G. Atlas of endoanal and endorectal ultrasonography. Springer Italy, Milan
SECTION II
New Technical Developments
in Endoanal and Endorectal
Ultrasonography
New Technical Developments in Endoanal and
Endorectal Ultrasonography
In order to obtain meaningful ultrasonic images, the operator must have an overall understanding and therefore correct use of the controls available on the ultrasound device. In fact,unsuitable regu­lation of the equipment produces poor images and can lead to false positive or negative diagno­sis.
Many types of ultrasound probes have been developed to evaluate the rectal wall and anal sphincter. Most of these have been made to exam­ine the prostate gland and are not suitable for evaluating the wall of the anorectum and the immediately adjacent tissues. The types of
endorectal probes include mechanical sector probes with a single transducer,which may have a limited field of view of 120–210° or may incorpo­rate radial probes with a full 360° field of view and linear and curved array probes, which have a lim­ited 120–210° field of view. Some of these probes are biplanar and can be changed from the axial plane to the longitudinal (sagittal, coronal) plane with the press of a button.All of these probes have a frequency range from 5 to 10 MHz. The focal range of a 7.0-MHz transducer is 2–5 cm whereas a 5.5-MHz transducer has a focal range of 1–4 cm. The longer focal length provides an advantage in examining perirectal structures. The 10.0-MHz transducer is preferred in the anal canal as it gives improved resolution of the sphincter mechanism
over the 7.0-MHz transducer.
We currently use an Hawk 2102 EXL B-K Medical scanner (B-K MedicalA/S, Mileparken 34, DK-2730 Herlev, Denmark) (Fig. II.1) with a 1850 radial array probe,which gives a 360° axial view of the rectal wall (Fig. II.2) [1–3]. The radial probe has a 24-cm metal shaft with a rotating transduc-
G.A. Santoro, B. Fortling
Fig. II.1. B-K Medical scanner: Hawk 2102 EXL
14 Benign Anorectal Diseases
Fig. II.2. B-K Medical rotating endoprobe type 1850
a
er at its tip. This 8539 transducer has a frequency range from 5 to 10 MHz with a focal length of 2–5 cm and a 90° scanning plane and is rotated at 4–6 cycles per second to get a radial scan of the rectum and surrounding structures (Fig. II.3) [4–5]. The B-K unit must be assembled carefully (Fig. II.4). The rectal tube (UA 0878) fits over the central axle that holds the transducer and slot into the handle with a screw collar to tighten the tube onto a rubber compression ring. The rotating transducer is pushed into the end of the axle and is covered by a latex balloon that is held in place
b
Fig. II.3. B-K Medical transducer type 8539 (a, b)
by two metal rings, one of which screws onto the metal shaft (Fig. II.5).A syringe filled with 50 ml of tap water is manipulated to remove air bubbles then attached to a spigot at the base of the probe.
Fig. II.4. B-K Medical anorectal probe type 1850 (dismantled), with rectal tube, 10 MHz transducer, plastic cone,and latex bal-
loon
Section II • New Technical Developments in Endoanal and Endorectal Ultrasonography 15
Fig. II.5. Transducer, metal rings, and latex balloon for
endorectal ultrasound
Fig. II.6. B-K Medical anorectal probe type 1850 assembled
with latex balloon
Water is gently injected into the balloon; the water is then withdrawn again with the transducer tip in the dependent position.In this way,all of the air is removed from the water-filled balloon. This pro­cedure may have to be repeated to remove all of the air (Fig. II.6). Latex balloon filled with degassed water allows acoustic coupling between the transducer and the rectal wall (Fig. II.7). The rectum can be of varying diameters and therefore the volume of water in the balloon may have to be adjusted intermittently.
Endoluminal ultrasound is usually performed
with the patient in the left lateral decubitus posi-
tion. Before the probe is inserted into the rectum, a digital rectal examination may be performed to identify the size, fixation, morphology, and loca­tion of the tumor, if it is low enough. If there is a stenotic annular lesion, the finger can check to determine whether it will allow easy passage of the probe [6–7]. The entire shaft of the probe is coated with a thin layer of warm gel using a paper towel.The probe tip is gently inserted through the anal canal and then angled posteriorly and advanced cephalad to as high a level as possible, with the bony sacrum used as a curved landmark. The patient should be instructed before the exam-
ination that no pain should be experienced. If
Fig. II.7. Diagrammatic representation of
an endorectal ultrasonography
16 Benign Anorectal Diseases
Fig. II.8. B-K Medical transducer type
6005
pain should occur, the study should be halted until the cause of the pain is elucidated.Under no circumstances should force be used to advance the probe. The examiner should never try to push the tip through a narrow stenotic lesion.However, in most instances, passage can be achieved although the volume of the fluid in the balloon will have to be substantially reduced in order to withdraw the probe through the stenotic portion. In some instances, a plastic cone filled with water will facilitate the imaging of a stenotic lesion, or it may be necessary to use a smaller probe, 7 mm in diameter (7 MHz,type 6005, focal range: 0.5-3 cm)
(Fig. II.8). Once the tip is advanced to as high a level as possible, usually 10–14 cm from the anal verge, the balloon can be inflated with 50 ml of water. The amount of water may have to be increased to provide complete acoustic coupling with the rectal wall. The examiner should never distend the balloon with more than 80 ml of degassed water, as it may rupture. If this occurs, the probe must be removed from the rectum and cleaned, a new balloon installed, and the whole procedure started over.If air or stool gets between the balloon and rectal wall, it will prevent visual­ization of the wall. To avoid this, we administer an enema 2 h before the examination, but despite this, problems can arise, and it may be necessary to remove the probe and suction out the rectum
with reintroduction of the probe in order to opti­mize the image.
With the probe at the highest level possible and with good visualization of the rectal wall, images are obtained at 1-cm intervals as the probe is withdrawn. The tip of the ultrasound probe should be maintained in the center of the rectal lumen to gain optimal imaging of the rec-
tal wall and perirectal structures. Some adjust­ments may have to be made in the gain of the
ultrasound unit to provide optimal imaging.
Occasionally, it is possible to perfectly depict all
five layers of the rectum circumferentially, but usually, only a portion of the rectal wall at a time will be optimally imaged, and minor adjustments will have to be made in the location of the probe relative to the rectal wall at various locations to optimally image all five layers. The exact level of the transducer tip can be read off the metal shaft of the ultrasound probe. More closely spaced images (0.5 cm) are obtained in the area of any abnormality.The balloon may have to be deflated and reinflated to maintain good acoustic contact with the rectal wall as the probe is withdrawn down the rectum. Once the entire rectum down to the anal sphincter has been evaluated, the bal-
loon is fully deflated, and the probe is removed from the rectum.
In most instances,the use of a large-bore proc­toscope serves several purposes (A.4522,Sapimed, Alessandria, Italy) (Fig. II.9).Firstly,it allows visu­al examination of rectal lesions with exact deter­mination of their location, both with respect to circumferential involvement of the rectal wall and the distance from the anal verge. Secondly, it allows suctioning of any residual stool or enema fluid that might interfere with the acoustic path­ways of the ultrasound waves, which may distort
the image. Most importantly, however, it allows easy passage of the probe to insure complete imaging of the rectum. This is of extreme impor­tance in the preoperative staging of rectal cancer. Once the 20-cm scored mark on the shaft of the probe is at the proximal end of the proctoscope, the proctoscope is then pulled back on the probe
Section II • New Technical Developments in Endoanal and Endorectal Ultrasonography 17
Fig. II.9. Rectosigmoidoscope for endorec-
tal ultrasonography (A.4522, Sapimed)
as far as possible,thus exposing the transducer for 2 cm beyond the end of the proctoscope. The bal­loon is then instilled with 30–60 cc of water, the volume of fluid usually needed to gain optimal imaging.
If the anus is being evaluated, the same probe is used with a transducer that has a shorter focal zone of 1–4 cm. A water-filled, hard-plastic cone (WA 0543) made of sonolucent polymethyl pen­tene (TPX) plastic and 1.7 cm in outer diameter is used in place of the latex balloon (Fig. II.10) [8].A hole at the top allows air to escape as the assembly is filled with degassed water. A gel-containing condom is then placed over the probe, and a thin layer of water-soluble lubricant is placed on the
exterior of the condom. Any air interface will cause a major interference pattern. The probe is now ready for insertion. The outer walls of this cone are parallel so that the probe may be moved within the anal canal without causing any anatomical distortion (Fig. II.11). This is very important when assessing the canal at different levels. There does not appear to be any benefit from asking the patient to “squeeze” their pelvic floor to assess sphincter contraction.At the origin of the canal, the“U”-shaped sling of the puborec­talis is the main landmark and should be used for final adjustment [9].
When the patient is in the left lateral position and the spigot for introducing water into the probe is pointing toward the ceiling, by conven-
tion, the anterior aspect of the anus will be supe­rior (12 o’clock) on the screen,right lateral will be left (9 o’clock) on the screen, left lateral will be right (3 o’clock) on the screen, and posterior will be inferior (6 o’clock) on the screen (just as in the image on axial CT scan) (Fig. II.12).
Future Directions
When we display a normal two-dimensional (2-D) ultrasound cross-sectional view, there are many elements of the image that will not be correctly recognized as components of a three-dimensional (3-D) structure – or at least not perceived in their true spatial relationships. With ultrasound imag­ing, we are usually looking at a 3-D structure that contains a solid volume of echoes and that there­fore does not readily translate onto a 2-D projec­tion. In routine clinical scanning, the operator forms a mental representation of the 3-D anatom­ic or pathological structure while viewing a large
series of 2-D slices interactively. In this case, the operator is using manual sense information about the physical location of the individual slices in building up 3-D subjective impressions. Three­dimensional and, indeed, four-dimensional (4-D) ultrasound has been promoted by different ultra­sound companies for several years. The acquisi­tion of a 3-D data volume and the underlying techniques are, however, different from applica­tion to application.Some areas of the human body
require extremely high-resolution 3-D volumes of data for adequate and precise diagnostic evalua­tion (anal sphincter,rectal wall,fistula,urethral sphincter complex, etc.); other areas require less when it comes to sampling the 3-D data volume (baby face, kidney, pancreas). For endoanal and endorectal ultrasound in general, the problems will be in assessing anatomy and in characterizing tissue structure and type.
Three-dimensional reconstruction of 2-D images is possible by connecting the ultrasound apparatus to a computer equipped with special software (BK 3Di) (Fig. II.13) [10].A normal dig-
18 Benign Anorectal Diseases
a b
Fig. II.10. Transducer and plastic cone for endoanal ultrasonography (a, b)
ital ultrasound image is displayed using a display matrix of around 700¥700 pixel elements, with each pixel assigned a value between 0 and 255
(256 levels of grey). The result seen on the ultra­sound monitor is a 2-D image with no depth information. Adding the third dimension to the pixels means that the pixels are transformed to voxels (voxel is a pixel, X- and Y-plane only, in the ultrasound image with an added Z-plane), each of which also will have an assigned value between 0 and 255. The depth of the voxel is crit-
ical to the resolution of the 3-D image, and this depth is directly related to the spacing between two adjacent images. Ideally, the voxel should form an exact cube, but because generally sam­pling with slightly lesser resolution in the Z­plane (due to acquisition speed considerations) compared with the 700¥700 matrix, the resolu­tion in Z-plane is marginally lower. High-resolu­tion data volumes may consist of typical voxel sizes around 0.15¥0.15¥0.2 mm (or 0.25 mm). This means that an acquisition based upon, for
Fig. II.11. Diagrammatic representation
of an endoanal ultrasonography
Section II • New Technical Developments in Endoanal and Endorectal Ultrasonography 19
a
Fig. II.12. Lesion localization with patient in left lateral posi-
tion when the spigot for introducing water into the balloon is pointing toward the ceiling (a,b)
example, sampling of transaxial images over a distance of 60 mm in the human body will result in a data volume block consisting of between 240 and 300 transaxial images. This is far more than any 4-D technique can handle using computer power as we know it today. High-resolution 3-D technique and consequently much slower acqui­sition speed does not function in 4-D mode. Four to five transaxial images are, as mentioned, sam­pled per 1-mm distance (in the Z-plane). Acquisition of a high number of parallel transax­ial images can be performed using a special col­orectal pullback mover (UA0552) with the B-K
Medical ultrasound probe type 1850 (Fig. II.14). The colorectal pullback mover is a computer­controlled, motor-driven device that can be operated at different levels of resolution. For the endoanal application, the usual setting is
0.2–0.3 mm between adjacent transaxial images. Scanning the anal canal with these settings over a pullback distance of 35 mm will typically yield 175 parallel images (Fig. II.15). This approach provides a neat way of uniformly scanning a solid volume, the acquisition taking typically 60 s. However, it is subject to the important lim­itation of being rather bulky. An alternative approach is to use a manual technique. In the case of manual acquisition, the samples will not be uniform, and they may be anisotropic in the sense that a voxel gives different values. Extreme
cases will result in voxels either having no sam­ples, that is, having data gaps, or having multiple samples. Gaps cause visually distracting arte­facts in the resulting displays. The new B-K Medical 2050 anorectal transducer solves most
b
Fig. II.13. Hardware (BK 3Di) for acquisition, reconstruction,
and visualization of three-dimensional ultrasound images