Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3752_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
31.08.2026
Размер:
24 Мб
Скачать
151
Reconstruction of CT Data in 64-Detector CT Systems
With 64-detector CT systems, each revolution of the gantry enables imaging of 3.2–4.0 cm of the heart (64 detectors times the width of each detector). To cover the entire heart, the required length of most cardiac scans is 9–10 cm; this necessitates combining data from multiple table positions to yield a fi nal volume of data. The greatest challenge in the reconstruction of this volumetric cardiac CT data is the maintenance of temporal uniformity: that is, one should be able to provide cardiac images at each table position from the same part of each cardiac cycle. A breakdown of temporal uniformity can lead to characteristic “stairstep” artifacts.
The most commonly used reconstruction algorithm in spiral cardiac CT is the half-scan reconstruction method , which involves the use of scan data from a single gantry rotation to generate an axial CT image. Half-scan reconstruction excludes the fan beam width (approximately 30°), so that approximately 210° of rotation are necessary to generate a single axial image. Using the half-scan method,
ab
cd
Fig. 8.7 Sharp and smooth convolution kernels are employed to improve visualization of stents and calcifi ed vessels while reducing blooming artifacts. ( a and c ) Depict long axis and short axis oblique thin-slice projections of an LAD stent using a standard smooth (B26f)
kernel. ( b and d ) are similar projections using a sharp (B46f) convolu- tion kernel. Blooming artifacts are reduced, and the edge of the stent is more clearly delineated at the expense of increased noise in the remain­der of the image
Pitch
Pitch 1
Pitch 2
itch 0.5
Pitch:
(in Multislice CT)
Table feed per rotation
Total width of collimated beam
(n detectors ¥ collimation)
Fig. 8.8 Pitch refers to the table feed per gantry rotation during a spiral acquisition divided by the width of the collimated beam. A pitch of exactly 1 ( central diagram ) implies that there are no data gaps and there is no overlap of data. When pitch is greater than 1 ( top diagram ), there are gaps in data acquisition, whereas when pitch is less than 1 ( bottom diagram ), there is data overlap. Retrospectively gated cardiac CT exam­inations are performed with a pitch of approximately 0.2
8 Post-processing and Reconstruction Techniques for the Coronary Arteries
https://t.me/medicina_free
152
the temporal resolution is approximately 60 % of the rotational speed of the scanner (due to the fan beam width exclusion). For modern single source 64-slice CT scanners which have gantry rotation times ranging from 330 to 375 ms, the temporal resolution using the half-scan recon­struction method is approximately 200–225 ms. In patients with heart rates of 60 beats per minute or less, the mid-diastolic diastasis period of minimal coronary motion is long enough to allow effective reconstructions in mid diastole in the majority of patients.
In patients with heart rates faster than 80 beats per min­ute, the duration of the diastasis period decreases consider­ably and may be only 100–200 ms. This makes it nearly impossible to reconstruct motion-free images using the half­scan reconstruction method. In these cases, multi-segment reconstruction may be utilized to improve the effective temporal resolution of the CT scanner. Multisegment recon­struction relies on additional data overlap with slower table movement and decreased pitch during CT acquisition [ 11 , 12 ]. This overlap results in the same table position being available for imaging at multiple heart beats from multiple detectors (Fig. 8.9 ). By combining views at a single table
position from different subsequent gantry rotations, one simulated half-scan rotation is generated. This results in improved image quality with fewer motion artifacts. By combining images from 3 cardiac cycles, the effective temporal resolution can be improved to as much as 65 ms. Multi-segment reconstruction relies heavily on a consistent R-R interval on the consecutive beats used to generate the fi nal axial image. Irregularities from atrial fi brillation or sinus arrhythmia during breath holding may cause misregis­tration artifacts.
Reconstruction of CT Data in 320-Detector CT Systems
With 320-detector CT systems, one axial rotation can cover up to 16 cm of tissue, although typically all 320 detectors cannot be employed simultaneously to image the heart. To accomplish this, the use of a wide X-ray beam and a wide cone angle is required. Figures 8.10 and 8.11 demonstrate the concept of the fan angle and the cone angle. The cone angle determines the coverage of the X-ray beam in the lon­gitudinal, or z-axis while the fan angle determines the cover­age in the x/y plane. The use of a wide cone angle gives the 320-detector scanner the ability to cover up to 16 cm of tissue
60°
60°
60°
67 ms 67 ms
67 ms
67 ms
180°
Continuous
spiral scan & feed
Image data
Delay
Z – Position
slower
pitch
Volume
gaps
Recon
Recon
Recon
Fig. 8.9 ( a and b ) Multisegment reconstruction of a CT acquisition involves the acquisition of data at a single table position over several cardiac cycles. The volumetric data are combined to yield a fi nal summed volume. The major requirement for multisegment reconstruc­tion is data overlap, which results in a slower pitch and higher radiation dose during the CT acquisition. Using this technique, effective temporal resolution can be improved to 67 ms for a scanner with a half-scan acquisition time of 200 ms
Fig. 8.10 The fan angle represents the spread of the X-ray beam in the x/y plane and is a factor in determining the diameter of the fi eld of view (Reproduced with permission from Toshiba America Medical Systems)
S.V. Gurudevan
https://t.me/medicina_free
153
in one axial rotation [ 13 ]. As a result, it is not necessary to move the patient through a spiral scan line. Instead, the patient can remain stationary and the entire heart to be imaged in a single gantry revolution. This has the potential to substantially improve image clarity by eliminating stairstep and misregistration artifacts that are sometimes seen with 64-detector systems, but will have no impact on temporal resolution, based upon the gantry rotation speed (as described above).
While use of a wide cone angle offers the capability of whole organ coverage, it also presents a challenge to image reconstruction algorithms employed in 64-detector systems. Traditional reconstruction algorithms cannot successfully incorporate the highly tilted X-ray planes created by the cone angle trajectory. As a result, one can get cone-beam artifacts including shading, ghosting, and diminished resolution as shown in Fig. 8.12a . With more advanced reconstruction algorithms specifi cally designed for 320-detector systems that account for the cone angle, these artifacts can be elimi­nated, shown in Fig. 8.12b [ 14 ].
Post Scan Related Post-processing Parameters
After the scan has been completed and datasets have been generated, additional post-processing techniques on a car­diac CT workstation are essential to accurately interpreting cardiac morphology as well as coronary artery anatomy and disease burden. The presence of an isotropic data set in which the spatial resolution is identical across all planes of examination facilitates manipulation of the data on a workstation.
Fig. 8.11 The cone angle quantifi es the spread of the X-ray beam in the z-direction, along the length of the patient (Reproduced with per­mission from Toshiba America Medical Systems)
ab
Fig. 8.12 Panel A ( top ) illustrates cone beam artifact. The image demonstrates shading, ghosting, and diminished resolution. This is a result of a reconstruction algorithm that does not properly take into account the wide cone angle. Panel B ( bottom ) shows an identical
image reconstructed with a novel reconstruction algorithm that takes into account the wide cone angle, thereby eliminating the artifact (Reproduced with permission from Toshiba America Medical Systems)
8 Post-processing and Reconstruction Techniques for the Coronary Arteries
https://t.me/medicina_free
154
The raw axial images are the most reliable for diagnosis, as they refl ect the source data in the order the images were acquired. To assist the reader in processing large volumes of data and illustrating key fi ndings, additional rendering techniques have been developed [ 15 , 16 ]. These include the multiplanar reformatting, maximum intensity projection, the volume averaging and volume rendering techniques, and the curved multiplanar projection. All involve rendering data contained within a 3 dimensional slab (the thickness of which the user can change) of data as a single 2-dimensional projection.
Multiplanar Reformatting (MPR)
Multiplanar reformatting involves the selection of an arbitrary image plane in a cardiac CT volume. This tech­nique requires an isotropic volumetric data set with equal spatial resolution in the X, Y, and Z axes. The plane may not conform to the conventional axial, coronal, and sagittal imaging planes and can be modifi ed by the user, as shown in Fig. 8.13 . MPR is the mainstay for analysis of cardiac CT data sets and is the most reliable method for the reader to arrive at the correct diagnosis. It can be performed using a single slice or with differing numbers of stacked slices. When more than one slice is selected, a rendering option must be selected to display a composite image of the multi­ple slices.
Maximum Intensity Projection
The maximum intensity projection (MIP) involves the pro­jection of data in a 3-dimensional slab so that only the voxels of highest HU are displayed on a 2-dimensional image (Fig. 8.14 ). Initially developed by Rubin and colleagues [ 17 , 18 ]. for use in peripheral CT angiography, the MIP is now used for nearly all CT angiography applications and is the mainstay of coronary artery interpretation. The MIP is ideal for the display of coronary artery images from a contrast CT examination as the maximum intensity in the coronary arter­ies is usually the intraluminal contrast. The coronary arteries
are surrounded by low-attenuation epicardial fat, resulting in an angiogram- like image with excellent edge defi nition (Fig. 8.14 ). Due the selection of the highest intensity voxels within a slab, the MIP tends to overestimate stenosis severity in calcifi ed vessels and stented segments (bright objects like calcium and metal get further enhanced). Overreliance on MIPs can also lead the reader to overlook subtle fi ndings in the coronary arteries, including motion and misalignment artifacts. In general, readers should always reconfi rm fi nd­ings on MIP with the source axial data.
Volume Averaging and Volume Rendering
Volume averaging (VA) involves the projection of data in a 3-dimensional slab so that the intensity of all the voxels in the slab is averaged on a fi nal 2-dimensional image (Fig. 8.15 ). While edge defi nition is poorer than with MIP, VA enables the reader to “see through” a dense object in a slab and can be useful in interpreting stenoses in calcifi ed vessels. When specifi c colors are assigned to specifi c ranges of HU in a 3-dimensional volumetric slab, this is termed vol­ume rendering (VR); this is available on virtually every car­diac workstation. The relative position and 3-dimensional relationship of the coronary arteries, cardiac veins, and car­diac chambers is possible using 3-dimensional VR (Fig. 8.16 ). While especially helpful in evaluating coronary
Objects in volume Image plane
Selectsd plane
Fig. 8.13 Multiplanar Reformat Projection (MPR)
Objects in volume
Image plane
Selected “Slab” of images
Fig. 8.14 Maximum Intensity Projection (MIP)
Objects in volume Image plane
Selected “Slab” of images
Fig. 8.15 Volume averaging
S.V. Gurudevan
https://t.me/medicina_free
155
anomalies and bypass grafts, the reader should never attempt to interpret stenoses purely on the basis of a 3D VR images: as calcium and intraluminal contrast have attenuation ranges that are near one another, signifi cant coronary artery stenoses in calcifi ed vessels can be misinterpreted.
Curved Multiplanar Projection
The curved multiplanar projection (CMP) is a centerline method for analysis of vessels whereby a virtual plane is cre­ated using the center of the column of contrast visualized on a series of consecutive axial slices (Fig. 8.17 ). A virtual plane in which the vessel is “stretched out” can then be dis­played. CMP is useful for confi rming and illustrating the appearance of stenoses identifi ed using standard axial multi­planar projections. It is particularly useful for tortuous ves­sels and those vessels that cannot be easily tracked in a single plane or thin slab using MIP. The CMP is only as reliable as the accuracy of the centerline, and should not be used as a fi rst-line assessment of coronary stenosis severity. Artifacts from inaccurate centerlines can lead to incorrect assessment of stenosis severity.
References
1. Lu B, Mao SS, Zhuang N, et al. Coronary artery motion during the cardiac cycle and optimal ECG triggering for coronary artery imag­ing. Invest Radiol. 2001;36(5):250–6.
2. Mao S, Lu B, Oudiz RJ, Bakhsheshi H, Liu SC, Budoff MJ. Coronary artery motion in electron beam tomography. J Comput Assist Tomogr. 2000;24(2):253–8.
3. Kim WY, Stuber M, Kissinger KV, Andersen NT, Manning WJ, Botnar RM. Impact of bulk cardiac motion on right coronary MR angiography and vessel wall imaging. J Magn Reson Imaging. 2001;14(4):383–90.
4. Isma’eel H, Hamirani YS, Mehrinfar R, et al. Optimal phase for coronary interpretations and correlation of ejection fraction using late-diastole and end-diastole imaging in cardiac computed tomog­raphy angiography: implications for prospective triggering. Int J Cardiovasc Imaging. 2009;25(7):739–49.
5. Gurudevan SV, Narula J. Prospective electrocardiogram-gating: a new direction for CT coronary angiography? Nat Clin Pract Cardiovasc Med. 2008;5(7):366–7.
6. Flohr TG, Schaller S, Stierstorfer K, Bruder H, Ohnesorge BM, Schoepf UJ. Multi-detector row CT systems and image­reconstruction techniques. Radiology. 2005;235(3):756–73.
7. Cademartiri F, Mollet NR, Runza G, et al. Improving diagnostic accuracy of MDCT coronary angiography in patients with mild heart rhythm irregularities using ECG editing. AJR Am J Roentgenol. 2006;186(3):634–8.
8. Ehara M, Kawai M, Surmely JF, et al. Diagnostic accuracy of coro­nary in-stent restenosis using 64-slice computed tomography: com­parison with invasive coronary angiography. J Am Coll Cardiol. 2007;49(9):951–9.
9. Nieman K, Cademartiri F, Raaijmakers R, Pattynama P, de Feyter P. Noninvasive angiographic evaluation of coronary stents with multi-slice spiral computed tomography. Herz. 2003;28(2): 136–42.
10. Maintz D, Seifarth H, Flohr T, et al. Improved coronary artery stent visualization and in-stent stenosis detection using 16-slice com­puted-tomography and dedicated image reconstruction technique. Invest Radiol. 2003;38(12):790–5.
11. Kachelriess M, Ulzheimer S, Kalender WA. ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart. Med Phys. 2000;27(8):1881–902.
12. Ohnesorge B, Flohr T, Becker C, et al. Cardiac imaging by means of electrocardiographically gated multisection spiral CT: initial experience. Radiology. 2000;217(2):564–71.
13. Tuy HK. An inverse formula for cone-beam reconstruction. SIAM J Appl Math. 1983;43:546–52.
14. Feldkamp LA, Davis LC, Kress JW. Practical cone beam algorithm. J Opt Soc Am. 1984;1:612–9.
15. Addis KA, Hopper KD, Iyriboz TA, et al. CT angiography: in vitro comparison of fi ve reconstruction methods. AJR Am J Roentgenol. 2001;177(5):1171–6.
Fig. 8.16 3D volume rendering
Objects in volume Image plane
Fig. 8.17 Curved Planar Reformat (CPR)
8 Post-processing and Reconstruction Techniques for the Coronary Arteries
https://t.me/medicina_free
156
16. Pavone P, Luccichenti G, Cademartiri F. From maximum intensity projection to volume rendering. Semin Ultrasound CT MR. 2001;22(5):413–9.
17. Rubin GD, Dake MD, Napel S, et al. Spiral CT of renal artery ste­nosis: comparison of three-dimensional rendering techniques. Radiology. 1994;190(1):181–9.
18. Prokop M, Shin HO, Schanz A, Schaefer-Prokop CM. Use of maxi­mum intensity projections in CT angiography: a basic review. Radiographics. 1997;17(2):433–51.
S.V. Gurudevan
https://t.me/medicina_free
157
© Springer International Publishing 2016 M.J. Budoff, J.S. Shinbane (eds.), Cardiac CT Imaging: Diagnosis of Cardiovascular Disease, DOI 10.1007/978-3-319-28219-0_9
Coronary CT Angiography: Native Vessels
Stephan Achenbach
Abstract
Coronary CT Angiography has become increasingly stable and robust and accuracy to iden­tify and rule out high-grade stenosis of the coronary arteries is high. The main clinical application is to rule out signifi cant coronary artery disease in patients with a relatively low pre-test likelihood of disease. For this application, several guidelines in the US and Europe endorse the use of cardiac CT. Other applications of coronary CTA include the support of coronary interventions, especially in the context of chronic total coronary occlusion, the identifi cation of coronary anomalies, and, to some extent, the assessment of non-obstructive coronary atherosclerosis. However, the use of coronary CTA for screening purposes is cur­rently not supported by offi cial recommendations.
Keywords
Coronary CT Angiography • Coronary CTA • Coronary Artery Disease • Computed Tomography • Stenosis • Atherosclerosis • Plaque
Introduction
Visualization of the coronary arteries has been the major focus of cardiac CT in the past years. Non-invasive “coronary CT angiography” has tremendous clinical potential for detect­ing or ruling out coronary artery stenoses in selected patients (see Figs. 9.1 and 9.2 ). As a consequence of continuous and substantial progress regarding image quality and robustness of the investigation, it is incorporated in several recent offi cial guidelines and recommendations. In addition, imaging of coronary atherosclerotic plaque may play a potential role in risk stratifi cation. However, spatial resolution and temporal resolution of CT imaging, even with the latest scanner gen­erations, are not equal to those of invasive coronary angiogra­phy. Interpreters of coronary CT angiography data sets must
therefore be aware that artefacts can occur and may lead to false-positive and, less frequently, to false- negative results. Diagnostic accuracy is impaired when image quality is reduced and image quality, in turn, is infl uenced by many fac­tors such as the patient’s heart rate, body weight, ability to cooperate, and extent of coronary calcifi cation. Therefore, the clinical utility of coronary CT angiography signifi cantly depends on the specifi c clinical situation and patient under investigation. The specifi c advantages and disadvantages of coronary CT angiography must be carefully considered before using this method in the workup of a patient with known or suspected coronary artery disease.
Imaging Protocol
Since the small dimensions and the rapid motion of the coro­nary vessels pose tremendous challenges for non-invasive imaging, high-end CT equipment and adequate imaging pro­tocols must be used. Currently, 64-slice CT is considered the minimum requirement for coronary artery imaging, and newer
S. Achenbach , MD Department of Cardiology , University of Erlangen , Ulmenweg 18 , Erlangen 91054 , Germany e-mail: Stephan.Achenbach@uk-erlangen.de
9
https://t.me/medicina_free
158
ab
cd
e
Fig. 9.1 Normal anatomy of the coronary arteries in transaxial images. ( a ) Level of the left main origin from the aortic root. The bifurcation of the left main into the left anterior descending ( large arrow ) and left circumfl ex coronary artery ( small arrow ) can be seen. The arrowheads point at a coro- nary vein. ( b ) A few mm further distal, the left anterior descending coro- nary artery ( large arrow ) has given rise to a diagonal branch ( arrowhead ). ( c ) Level of the right coronary ostium. A short section of the right coronary
artery is visible ( double arrows ). Large arrow : Mid left anterior descending coronary artery, small arrow : left circumfl ex coronary artery. ( d ) Mid- ventricular level. The left anterior descending coronary artery ( large arrow ), left circumfl ex coronary artery ( small arrow ), and right coronary artery can be seen ( double arrows ). ( e ) Distal segment of the right coronary artery ( double arrow ), which ends in the posterior descending artery ( small arrow ). The arrowhead points at a right ventricular branch
S. Achenbach
https://t.me/medicina_free
159
technology [ 1 ], such as Dual Source CT and scanners that allow simultaneous acquisition of 256 or 320 cross- sections, provide for more robust image quality and further improved image quality.
A basic prerequisite for CT imaging of the coronary arteries is the patient’s ability to understand and follow breathhold commands. Even slight respiratory motion
during data acquisition will cause substantial artefact and may render the coronary arteries (or parts of them) unevalu­able. Therefore, patients should be able to reliably hold their breath for approximately 10 s. Otherwise, coronary CT angiography should not be performed. Heart rate should be regular and preferably low (optimally below 60/min, even though this is not as strictly required for Dual Source CT)
ab
c
d
Fig. 9.2 Same patient as in Fig. 9.1 . Various forms of post-processing have been used to visualize longer segments of the coronary arteries. ( a ) Curved multiplanar reconstruction (curved MPR) of the right coronary artery. ( b ) Maximum Intensity projection (MIP) of the right coronary
artery ( arrows ) in a double-oblique plane. ( c , d ) 3-dimensional, surface- weighted Volume Rendering Technique (VRT) reconstructions in two different angulations
9 Coronary CT Angiography: Native Vessels
https://t.me/medicina_free
160
[ 1 , 2 ]. It is usually recommended that patients receive pre- medication with short acting beta blockers to lower the heart rate. Beta blockers can be administered orally approxi­mately 1 hr prior to scanning, or intravenously immediately before the scan. Sometimes, a combination of both is neces­sary. Nitrates should be given to all patients who have no contraindications in order to achieve coronary dilatation, which substantially improves image quality [ 1 ].
Intravenous contrast enhancement is necessary for coro­nary CT angiography and typically, 50–100 ml of iodine­based, high concentration contrast agent are injected. High fl ow rates are recommended, injection should be between 4 and 7 ml/s. Synchronization of contrast injection and data acquisition can be achieved either through a “bolus tracking” method or by using a separate “test bolus” acquisition to measure the contrast transit time.
Subsequent data acquisition can follow various princi­ples. The obtained data need to be synchronized with the heart beat, and this can either be achieved through retrospec­tive ECG gating or prospective ECG triggering [ 1 ].
Retrospectively gated scans are acquired in spiral mode and usually provide for robust and high image quality, fl exi­bility to choose the cardiac phase during which images are reconstructed, as well as the ability to reconstruct “func­tional” data sets throughout the cardiac cycle in order to analyse left ventricular function and regional wall motion. In order to limit radiation exposure, the output of the x-ray tube can be “modulated” during the acquisition, with lower out­put in systole, and higher output in diastole, when the most relevant image reconstructions are usually performed.
Prospectively triggered scans are associated with substan- tially lower radiation exposure. Images are acquired in axial mode without table movement and the patient table is advanced by one detector width following the acquisition, with subsequent images acquired in the next or second to next cardiac cycle. Less fl exibility to reconstruct data at dif­ferent time instants in the cardiac cycle as well as greater susceptibility to artefacts caused by arrhythmia are trade-offs for the advantage of lower dose. Especially in young patients – in whom radiation dose may be of major concern – prospectively triggered scans should be strongly considered [ 25 ]. Heart rate must be low so that artefact-free images can be guaranteed at the time instant of radiation exposure. The lowest radiation dose is achieved when the x-ray exposure window in each cardiac cycle is only as long as one-half rotation of the gantry requires (just long enough to recon­struct one set of transaxial slices), but “padding” (x-ray exposure over a longer time period in each cardiac cycle) may be used to provide some fl exibility regarding the time instant of image reconstruction [ 6 , 7 ].
A combination of spiral acquisition and prospective trig­gering is the so-called “Flash Mode” ( prospectively ECG triggered high pitch spiral acquisition ). It is only available
with a limited number of scanners that have either two detec­tors or a very wide detector. Images are acquired during con­tinuous, very fast motion of the table, and the volume of the heart is typically covered within 150–250 ms. In the cranio­caudal direction, the data for each subsequent image are acquired with a very slight temporal offset as compared to the previous image (approximately 0.5 ms), so that the con­secutive images represent ever so slightly different time instants within the cardiac cycle. Since the transition is smooth and image acquisition is typically performed in dias­tole with very little cardiac motion, this offset is not notice­able in the data set and in reconstructed images. This mode of data acquisition provides high image quality at very low doses, but requires stable heart rates below 60 beats/min to avoid artefact [ 811 ].
Radiation Exposure
Unless specifi c measures are taken to limit radiation dose, the exposure during coronary CTA can be high. A landmark study performed several years ago demonstrated that in individual centers, the average estimated effective radiation exposure was as high as 30 mSv (while in the same study, sites at the lower end of the spectrum performed coronary CTA with an average exposure of only 4–5 mSv) [ 12 ]. Since then, substantial progress has been achieved regarding radi­ation exposure (see Table 9.1 ). ECG-based tube current modulation in spiral acquisition and prospectively ECG­triggered image acquisition avoid x-ray exposure during the entire cardiac cycle and limit x-ray tube output to those phases of the heart beat which are likely to be used for image reconstruction. This limits the fl exibility of recon­structing images during different parts of the R-R interval, which would be desirable to assess ventricular function (a question, however, that is rarely relevant in coronary CTA), and which is also advantageous when motion artefacts are present, to identify a phase with no or little artefact. Hence, low heart rates, which make it extremely likely that artefact­free images are obtained in diastole, facilitate the use of these techniques and in this way, reducing the heart rate by premedication contributes to lower radiation exposure [ 2 ]. While the tube voltage for coronary CTA used to uniformly be 120 kV, it has been observed that depending on patient size, it is possible to reduce tube voltage to 100 or 80 kV, or in very selected cases even 70 kV [ 1317 ]. The increase in noise is tolerable depending on patient size, and to some extent is offset by higher iodine contrast. A “rule of thumb” is that tube voltage can be reduced to 100 kV for all patients with a body weight below 100 kg. Finally, there is a linear relationship between tube current and image noise, so that again, especially in patients with low body weight there is potential to reduce exposure.
S. Achenbach
https://t.me/medicina_free