Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_5807_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
31.08.2026
Размер:
26 Мб
Скачать
Chapter 2: Ultrasonography: physics and principles
Normal
Incident ray
θ
1
Fig. 2.4. Reection, transmission, and refraction of sound waves.
Reflected ray
Interface
θ
2
Transmitted (refracted) ray

From sound to image

The creation of an image from sound is achieved in three steps – producing a sound wave, receiving echoes, and interpreting those echoes.
Producing a sound wave
The transducer (probe) is a device that converts electrical signals into ultrasound waves at the desired frequency and vice versa (Figure 2.5). These sound waves bounce obody tissues and make echoes. The part of the transducer that converts electrical impulse to ultrasound and vice versa is known as the crystal or the active element. Transducers are made from materials that exhibit the property of piezoelec­tricity (a Greek word, piezein: to squeeze or press). Piezoelectricity is the ability of some materials to change their dimensions when an electric eld is applied to them and conversely to develop electrical charges when they are deformed. Medical transducers are made from a synthetic ceramic material (lead zirconate titanate) that is red in a kiln and therefore can be model ed into almost any shape. To establish an electrical connection, thin layers of silver are evaporated onto the surface to form electrodes. This device will expand and contract when a voltage is applied to it but will also create a voltage when subjected to a small pressure such as a returning echo might exert. Obviously the voltages generated when receiving echoes are normally much smaller than those applied to create the ultrasound wave in the rst instance. The material on the face of the transducer is usually a rubbery coating, a form of impedance matching to enable the sound to be transmitted eciently into the body. Thus, transducers in simple terms are metallic electrodes attached to a piezoelectric substance that converts electrical energy into acoustic energy when it is switched onand acoustic energy to electrical energy when it is listening.
There are dierent types of transducers. All types have advantages and disadvantages and no single transducer can perform all functions. There are several classication of transducers. A linear transducer is a rectangular-shaped probe in which the elements are arranged in a line and involves a large number of parallel scan lines, whereas a sector (annular) transducer has the elements in the form of ring
shape and arranged concentrically. This results in a series of lines that all originate from a single location and travel out­ward in a pie-shaped wedge. Linear arrays are usually cheaper than sector scanners but have wider skin contact and there­fore make it diculttofocusonsomeorganssuchasthe heart. The linear arrays use a ring sequence of alternate groups of 3–4 elements. A curved array is similar to a linear array except that the image created is a sector type. A linear phased array applies voltage pulses to all elements as a group but with small time dierences (phasing). The time dierence is chang ed each time so that the sound pulses will be sent out in dierent directions. A phased array results in a sector image and has several advantages such as being smaller in size, the fact that the at face allows better gel coupling, and the fact that the electronics create high frame rates compared with annular arrays. The main disadvantage is the poor supercial visualization because true phased arrays come to a point in the midline of the probe as the beam is made up of the whole group of elements. This can be overcome by exploiting the principles of p hased and linear arrays to form a hybrid called vector phased array. Some transducers operate in a burst-excited mode that converts 1–2 cycles of alterna­ting voltage bursts into alternating pressure, resulting in a sound pulse. It then receives echoes and converts them into voltage bursts. Others use a shock-excited mode. These create ultrasound pulses and also receive echoes and convert them into voltage bursts. The most common mechanical trans­ducers are the oscillating probes and the rotating probes driven by a motor. These use a combination of single-element oscillation, multiple-element rotation, or a single element and set of acoustic mirrors to generate the sweeping beam for 2D mode. Mechanical probes are subject to wear but produce excellent images. On the other hand, electrical probes are not subject to wear but are generally more expensive.
Most transducers are only able to emit one frequency because the crystals have a certain inherent frequency. Accordingly, most ultrasonographers use multiple probes. Multifrequency probes multiple crystals with dierent frequencies and a specic fre­quency is selected by the user. They are convenient because they save time in not having to switch to dierent probes. Nevertheless, they do have slower frame rates and therefore are useful only for imaging static structures.
do exist, however. These probes have
Receiving the echoes
The return of the sound wave to the transducer results in the same process that it took to send the sound wave, but in reverse. The return sound wave vibrates the transducer. The transducer turns the vibrations into electrical pulses that travel to the ultrasonic scanner where they are processed and transformed into a digital image.
Forming the image
The ultrasound scanner must determine two things from each received echo. First, how l ong did the echo take to be
13
Section 1: Imaging techniques
Acoustic
Transmitting
crystal
Crossed-beam
sound path
barrier
Receiving
crystal
Case
N
Beam from group of elements compared with single
D
D = diameter of aperture
received from the time the sound was transmitted? From this the focal length for the phased array is deduced, ena­bling a sharp image of that echo a t that depth. Second, how strong was the echo? From these two an swers the scanner determines which pixel in the image to light up and with what intensity and at what hue if frequency is processed. Furthermore, for the ultrasound scan to operate in real time – i.e., any real movement in tissue is instantly asso- ciated with a corresponding movement in the displayed image – it has to avoid judder such as can be seen on early cinema movies and the object being imaged must not move excessively between successive views. This can be maintained by a suciently high frame rate,whichisthe rate at which the image is updated or refreshed. To avoid judder, the image must be updated at a rate of approxi­mately 25 times per second or higher. Scanners are also equipped with a facility often labeled frame f reeze whereby the same image is written onto the screen about 25 times a second.
The returning echoes from tissues show a steady decline in amplitude with increasing depth due to attenuation. This is generally considered to be a nuisance and attempts are made to correct for it. The amount of amplication or gain given to the incoming signals is made to increase simultaneously with the arrival of echoes from the greater depth. This is called the time gain compensation (TGC) control and is now tted to virtually all ultrasound scanners. Of course, the assumption that all echoes should be made equal is not really valid. The operator still needs to use the TGC control with care so as not to produce misleading images; for example, excessive TGC can turn a normally echo-poor area within a uid-lled cyst into one that seems to have small echoes, thereby resembling a tumor.
The layout of the TGC controls varies from one machine to another. One of the most popular settings is a set of slider knobs. Normally each knob in the slider set controls the gain for a specic depth. It is the task of the operator to set each level for each patient and often it is necessary to adjust the TGC during the examination when moving from one ana­tomic region to another. TGC has also one important clinical application, to avoid acoustic shadowing and its opposite,
Fig. 2.5. A schematic representation of an
ultrasound transducer and the dierence in the beam shape between a single-element transducer and an array transducer.
2
N = D
N equals near zone
λ = wavelength
/4λ
aring or enhancement. Shadowing occurs when the desired organ to be examined is placed behind another organ that absorbs too much of the energy of the transmitted and received pulses. For example, we may see a break in the posterior uterine wall where it lies posterior to the fetal head. On the other hand, the posterior wall of an ovarian cyst may appear to be very bright because the path traveled by the pulse and its corresponding echoes is largely through cyst uid, which absorbs very little of the beam energy. In addi­tion, this can also be used to dierentiate between some solid masses that are quite homogeneous and whose image can be devoid of internal echoes (i.e. hypoechoic). These can be con­fused with a cyst, which would also be expected to be hypo­echoic. Nonetheless, the solid mass is much more likely to be absorptive than the cyst and hence the two can normally be distinguished by the presence or absence of aring or shad­owing posteriorly.

Modes of ultrasonography

Four dierent modes of ultrasound are used in medical imaging: A-mode: This is the simplest type of ultrasound wherein a
single-element transducer scans a line through the body with the echoes plotted on screen as a function of depth. It is therefore a one-dimensional view. Therapeutic ultrasound aimed at a specic tumor is A-mode, to allow for pinpoint accurate focus of the destructive wave energy.
B-mode: To produce a more useful two-dimensional (2D)
scan, it is necessary to obtain a series of A-mode scans and assemble them in a convenient format. This is done either by moving the transducer using a suitable mechanical device or else by having more than one transducer. This second option is preferred in modern scanners and the transducer, which is hand-held by the operator, in fact contains a row (array) of many transducers (typically 100–200). In this way, a series of A-scans can be obtained in a closely packed regular format. The amplitude (height) of each echo is represented by the brightness of a spot at the position. This display mode, in which the x and y directions relate to real distance of organs and the use
14
Transducer
(frequency f )
Chapter 2: Ultrasonography: physics and principles
Fig. 2.6. Doppler frequency shift and the
equation for calculating it.
Doppler frequency (f
Beam
θ
Higher doppler frequency obtained if: – velocity is increased – beam is aligned more to flow direction – higher frequency is used
Flow velocity = V
- doppler shift
f
d
c - speed of sound in tissue
- transmited beam
f
t
V - velocity of the blood θ - angle of incidence between the ultrasound beam and the direction of the flow
of gray scale to represent echo strength, is known as the B-mode scan.
M-mode: M-mode is a motion scan wherein a rapid sequence
of B-mode images follow each other in sequence on the screen to enable physicians to see and measure a range of motion. This mode can be useful when imaging heart valves, because the movement of the valves will make distinct patterns.
Doppler mode: This mode produces an image of ow and is
essentially obtained from measurements of movement. In ultrasound scanners, echoes from stationary tissue are the same from pulse to pulse. However, echoes from moving objects exhibit slight dierences in the time for the signal to be returned to the receiver – the Doppler eect. In general, this refers to a change in the received fr equency compared with the frequency emitted whenever there is relative motion between a sound source and the listener. These dierences can be measured as a direct time dierence or, more usually, in terms of a frequency shift from which the Doppler eect is obtained. This shift falls in the audible range of sound and is often presented audibly using stereo speakers to produce a very distinctive, although synthetic, pulsing sound. The Doppler frequency shift (Figure 2.6) depends on several factors:
) = 2. ft. V. cos θ
d
c
ABCD
skin
beam
direction
flow
Fig. 2.7. Eect of the Doppler angle on the sonogram. A higher-frequency
Doppler signal is obtained if the beam is aligned more in the direction of ow. In the diagram, beam A is more aligned than B and produces higher Doppler signals. The beam/ow angle in C is almost 90° and there is a very poor Doppler signal. The ow at D is away from the beam and there is a negative signal.
vessel
BA
Sonogram
C
D
1. Blood velocity: As velocity increases, so does the Doppler
frequency.
2. Ultrasound frequency: Higher frequencies give increased
Doppler frequency. However, this is a compromise between
better sensitivity to ow or deeper penetration.
3. The angle of insonation: The Doppler frequency increases as
the beam becomes more aligned to the direction of ow, i.e.,
as the angle between the beam and the direction of ow
becomes smaller (Figure 2.7). Therefore, the maximum
Doppler shift will occur at angles of 0° (maximum positive
Doppler shift) and 180° (maximum negative Doppler shift)
and at an angle of 90° there will be no Doppler shift as the
cosine of 90° is 0.
Types of Doppler ultrasound ow modes
The Doppler frequency shift information can be displayed graphically in various ways.
*
Color Doppler (directional Doppler) uses a computer to convert the Doppler measurements into an array of colors. The transducer elements are switched rapidly between B-mode and color ow imaging to give an impression of a combined simultaneous image. Thus, color visualization is combined with a standard ultrasound picture of a blood vessel to show the speed and direction of blood ow (Figure 2.8). The assignment of color to frequency shifts is usually based on direction and magnitude: red for ow toward the ultrasound beam and blue for shifts away from it
15
Section 1: Imaging techniques
Fig. 2.8. Color Doppler (directional) showing the ow along the umbilical
vein and arteries.
(a)
(a)
(b)
Fig. 2.9. Color power Doppler, showing its sensitivity to low ow. (a) Color
power angiogram of the circle of Willis in the fetal head. (b) Color power angiography of a submucosal broid; note the small vessels inside the tumor.
and dierent color hues or lighter saturation for higher
frequency shifts. Color Doppler is very sensitive to low ow
(Figure 2.9) and has the ability to render the directional
information in dierent colors (color ow maps). However,
it gives limited ow information and poor temporal
resolution/ow dynamics as the frame rate can be low when
scanning deep. The color Doppler image is dependent on
the general Doppler factors , particularly the need for a good
beam/ow angle. In practice, the experienced operator
alters the scanning approach to obtain good insonation
angles so as to get unambiguous ow images.
(b)
Fig. 2.10. Setting the color gain to minimize the signals (artifacts) from
surrounding tissue: (a) color gain = 71; (b) decreasing the color gain to 35.
*
Other factors that control the appearance of the color ow image include:
1. Power and gain: Color ow uses higher-intensity power than B-mode. The values are set to obtain good signal for ow and to minimize the signals from surrounding tissue (Figure 2.10).
2. Frequency selection: High frequencies give better sensitivity to low ow and have better spatial resolution. Nevertheless, low frequencies have better penetration and are less susceptible to aliasing at high velocities.
3. Velocity scale/pulse repetition frequency: Low pulse repetition frequencies should be used to examine low velocities but aliasing may occur if high velocities are encountered (Figure 2.11).
4. Region of interest: Because more pulses are needed to look at ow rather than for the B-mode image, reducing the width and maximum depth of the color ow area under investigation will usually improve frame rate and may allow a higher color scan line density with improved spatial resolution.
16
Chapter 2: Ultrasonography: physics and principles
(a)
(b)
(a)
(b)
Fig. 2.11. Color ow imaging with eects of pulse repetition frequency or scale.
(a) The pulse repetition frequency or scale is set low (yellow arrow). The color image shows ambiguity within the umbilical artery and vein and there is extraneous noise. (b) When the scale is set appropriately for the ow velocities, the color image shows the arteries and vein clearly and unambiguously.
5. Focus: The focus should be at the level of the area of interest. This can make a signicant dierence to the appearance and accuracy of the image.
*
Power Doppler (energy, amplitude ow, nondirectional Doppler) is a technique that is more sensitive in detecting blood ow than is color Doppler. It is able to obtain images that are dicult or impossible to obtain using standard color Doppler. It also provides greater detail of blood ow, especially in vessels that are located inside organs. However, it provides nondirectional information in some modes and has very poor temporal resolution and is susceptible to noise.
*
Spectral (pulsed) Doppler, where instead of displaying the Doppler measurements visually they are displayed graphically (Figure 2.12). It is used to provide a measure of the changing velocity in the sample volume gate.If an accurate angle correction is made, then absolute velocities can be measured. Spectral Doppler has the advantage of detailed analysis of distribution of ow and good temporal
Fig. 2.12. (a) Spectral Doppler of the common carotid artery and (b) setting up
the sample volume in a sonogram of the descending aorta. With the angle correction the peak velocities can be measured, where b is the direction of the Doppler beam, g is the gate or sample volume, and a the angle of correction.
resolution and it can examine ow waveform and allows calculations of velocity and indices. Spectral Doppler images are aected by the same factors as color Doppler as well as the gate size; a large gate may include signals from adjacent vessels (Figure 2.13).
Since color ow imaging provides a limited amount of information over a large region, and spectral Doppler provides more detailed information about a small region, the two modes are complementary and, in practice, are used as such. Color ow imaging is used to identify vessels requiring examination, to identify the presence and direction of ow, to highlig ht gross circulation anomalies, and to provide beam/vessel angle cor­rection for velocity measurements. Pulsed-wave Doppler is used to provide analysis of the ow at specic sites in the vessel under investigation. When using color ow imaging with pulsed-wave Doppler, the color ow/B-mode image is frozen while the pulsed wave Doppler is activated. Recently, some manufacturers have produced concurrent color ow imaging
17
Section 1: Imaging techniques
and pulsed-wave Doppler, sometimes referred to as triplex scanning. When these modes are used simultaneously, the per-
formance of each is decreased. Because transducer elements are employed in three modes (B-mode, color ow, and pulsed-wave
Fig. 2.13. Inuence of gate size. The spectral Doppler gate insonates an artery
and vein and the sonogram shows ow from both of these vessels. The calculation of mean velocity (arrow) is meaningless since velocities from one vessel subtract from those of the other.
Doppler), the frame rate is decreased, the color ow box is reduced in size, and the available scale is reduced, leading to increased susceptibility to aliasing.
When pulses are transmitted at a given sampling frequency (the pulse repetition frequency or the scale), the maximum Doppler frequency that can be measured unambiguously is half the scale. Therefore, if the blood velocity and beam/ow angle being measured combine to give a Doppler frequency value greater than half of the scale, ambiguity arises in the Doppler signal. This is aliasing and its bad eect can be cor­rected by reducing color gain or increasing the scale (Figure 2.14). The pulse repetition frequency is itself con­strained by the range of the sample volume. The time interval between sampling pulses must be sucient for a pulse to make the return journey from the transducer to the reector and back. If a second pulse is sent before the rst is received, the receiver cannot distinguish between the reected signal from both pulses and ambiguity ensues. As the depth of investigation increases, the journey time of the pulse to and from the reector is increased, reducing the pulse repetition frequency for unam­biguous ranging. The result is that the maximum Doppler frequency measured decreases with depth. Therefore, low pulse repetition frequencies are employed to examine low velocities (e.g., venous ow) as the longer interval between
(a) (b)
(c) (d)
18
Fig. 2.14. An example of aliasing and its correction. (a) Abrupt termination of the systolic peak, with the truncated part of the peaks showing below the baseline. (b)
The same case after correction by increasing the pulse repetition frequency and adjusting the baseline (downward). (c) Aliasing of the color ow (yellow arrows) is corrected by reducing the color gain and increasing the pulse repetition frequency (d).
Chapter 2: Ultrasonography: physics and principles
(a)
(b)
Fig. 2.15. The eects of pulse repetition frequency or scale on aliasing. (a) The
pulse repetition frequency is set low (yellow arrow). The color image shows ambiguity within the umbilical artery and vein and there is extraneous noise. (b) The scale is set appropriately for the ow velocities and the color image shows the arteries and vein clearly and unambiguously.
pulses allows the scanner a better chance of identifying slow ow. Aliasing will occur if low scale is used and high velocities are encountered (Figure 2.15). Conversely, if a high scale is used, low velocities may not be identied.
Modes of Doppler waves
In the sonographic community, the terminology Dopplerhas been accepted to apply to both the continuous-wave and pulsed-wave systems despite the dierent mechanisms by which velocity is detected.
Continuous-wave Doppler (CW), as the name suggests, uses continuous transmission and reception of ultrasound by two separate elements within the transducer, e.g., for listening to fetal heart rate. Doppler signals are obtained from all vessels in the path of the beam until it becomes suciently attenuated due to depth. These machines are unable to determine the specic location or velocities and cannot be used to produce color ow images.
Pulsed-wave Doppler (PW) machines transmit pulses of ultrasound, and then switch to receive mode. As such, the
reected pulse that they receive is not subject to a frequency shift, as the insonation is not continuous. However, the phase change in subsequent measurements can be used to obtain the frequency shift.
Blood ow measurements
*
Calculation of velocity
Theoretically, once the beam/ow angle is know n, velocities can be calculated from the Doppler equation. Nonetheless, errors may still occur due to:
(a) Use of multiple elements in array transducers. (b) Nonuniform insonation of the vessel lumen. (c) Insonation of more than one vessel. (d) Use of lters removing low-velocity components. (e) Use of high angles (>60°) may give rise to error
because of the comparatively large changes in the cosine of the angle that occur with small changes of the angle.
(f) The velocity vector may not be in the direction of the
vessel axis.
It is good practice to try to repeat velocity measurements, using a dierent beam approach, to gain a feel for the variability of measurements in a particular application.
*
Calculation of absolute ow
Total ow measurement using color or duplex Doppler ultrasound is fraught with diculties, even under ideal conditions. Errors that may arise include:
(a) Those due to inaccurate measurement of vessel cross-
sectional area (b) Those originating in the derivation of velocity. These errors become particularly large when ow
calculations are made in small vessels; errors in measurement of diameter are magnied when the diameter is used to derive cross-sectional area.
*
Flow waveform analysis
This has the advantage that derived indices are independent of the beam/ow angle. Furthermore, changes in ow waveform shape have been used to investigate both proximal disease (e.g., peripheral arterial circulation in adults) and distal changes (fetal circulation and uterine arteries). Many dierent indices have been used to describe the shape of ow waveforms. All are designed to describe the waveform in a quantitative way. In general, they are a compromise between simplicity and the amount of information obtained.
The most commonly used indices available on most
commercial scanners are:
1. Resistance index (RI) (also called resistive index)
2. Systolic/diastolic ratio (S/D) ratio, sometimes called the
A/B ratio
3. Pulsatility index (PI)
19
Section 1: Imaging techniques
Fig. 2.16. Indices of measurement of the ow waveform shape.
These indices are all based on the maximum Doppler shift waveform and their calculation as described in Figure 2.16. Although PI takes slightly longer to calculate, it does give a broader range of values – for instance, in describing a range of waveform shapes when there is no end-diastolic ow. In addition to these indices, the ow waveform may be described or categorized by the presence or absence of a particular feature, e.g., absent end-diastolic ow in fetal compromise.

Safety issues

Despite its impressive safety record of ultrasound to date, the intensity (or acoustic output) level of ultrasound used to scan the fetus in utero has increased almost eightfold over the level that was allowed in the early 1990s. Therefore, the comfort obtained from the absence of any harm based on epidemiolog­ical evidence must be tempered by the fact that there are not enough studies appropriate and adequate for guiding current clinical practice.
On the basis of some concerns about the theoretical eects of ultrasound on the developing fetus, researchers have conducted epidemiological studies looking for associations between ultra­sound exposure and various traits, particularly brain develop­ment (dyslexia, non-right-handedness, and delayed speech development), reduced birth weight, and childhood cancers. Meta-analyses of randomized controlled trials of adverse eects show only that there is a just-signicant increased tendency to non-right-handedness in the ospring of women who have had scans; however, the complexity of the study makes the observa­tion difficult to interpret [1]. Nevertheless, continual vigilance is necessary particularly in areas of concern such as the use of pulsed Doppler in the rst trimester.
Ultrasound causes heating, referred to as thermal and non- thermal eects. The main areas of concern among nonthermal eects are cavitation and microstreaming, but eects due to movement of cells in liquids, electrical changes in cell mem­branes, and pressure changes also exist.
Thermal heating is a consequence of the absorption of the ultrasound wave by tissue. Absorption increases with increas­ing frequency, and the temperature rise caused by an ultra­sound beam depends on many factors such as beam intensity and output power, focusing, beam size and depth, tissue absorption coecient, tissue-specic heat and thermal conduc­tivity, time, and blood supply. The Consensus Rep ort on Potential Bioeects of Diagnostic Ultrasound in 2007 [2] stated that Due to the movement of the transducer and of the struc­tures being imaged during clinical examination, the acoustic eld remains xed over a given structure or volume of tissue for brief periods of time, typically measured in seconds or fractions of a second. Under these conditions, the probability of local tissue or organ heating is small and unlikely to be of clinical signicance.
Cavitation (bubble formation) is the growth, oscillation, and decay of small gas bubbles under the inuence of an ultrasound wave. These bubbles often grow to some limiting size and continue to vibrat e at the ultrasound frequency. The growth and collapse of these microbubbles focuses and trans­fers energy and produces extremely high localized pressures and temperatures that add further stress to cell boundaries. When bubbles expand and contract without growing to crit­ical size, the activity is termed stable cavitation. Unstable cavitation does not occur in the therapeutic range in normal tissues except in air- lled cavities, most notably adult lung and intestine. Luckily, the fetal lung and intestine do not contain obvious air bubbles. Cavitation is limited by low­intensity and pulsed Doppler because there will be enough time for bubbles to regain their initial size during the o period.
Microstreaming when ultrasound passes through liquid causes a sort of stirring action termed acoustic streaming. As the acoustic pressure of the ultrasound increases, the ow of liquid speeds up. Cavitation sets up eddy currents in the uid surrounding the vibrating bubbles and the eddy currents in turn exert a twisting and rotational motion on nearby cells. In the vicinity of vibrating gas bubbles, intracellular organelles are also subjected to rotational forces and stresses. This stirring action, in theory, could occur in uid-lled parts of a patients body, such as blood vessels, the bladder, or the amniotic sac. In experimental animals shearing can occur when streaming liquid comes near a solid object, and this can damage platelets and lead to abnormal blood clotting (thrombosis). It is not clear to what extent this eect occurs in humans exposed to diag­nostic ultrasound.
Accordingly, the conclusion should always be that the diag­nostic procedure valid medical indication, with the lowest possible ultrasonic exposure setting to gain the necessary diagnostic information. This requires self-regulation on the part of the manufacturer and in part of the operator to keep the time limit as short and informative as possible.
should be performed only when there is a
20

References

1. Miller DL. Safety assurance in obstetrical ultrasound. Semin Ultrasound CT MR. 2008; 29(2): 156–64.
2. Barnett SB, Duck F, Ziskin M. WFUMB symposium on safety of ultrasound in medicine: conclusions on recommendations on biological eects and safety of ultrasound contrast agents. Ultrasound Med Biol 2007; 33(2): 2334.

Suggested reading

Elvy M. Physics of medical ultrasound. http://www.qmseminars.co.nz/ PDF/ElvyPhysicsMedicalUltrasound.pdf (Accessed May 10, 2008).
Kremkau FW. (2005). Diagnostic Ultrasound Principles and Instruments, 7th edn. Philadelphia, WB Saunders, 2005.
Deane C. Doppler ultrasound: Principles and practice. http://www. centrus.com.br/DiplomaFMF/SeriesFMF/doppler/capitulos-html/ chapter_02.htm (Accessed April 20, 2008).
Evans T. Physics and instrumentation. In: Chudleigh T, Thilaganathan B, eds. Obstetric Ultrasound: How, Why and When. 3rd edn. Edinburgh, Churchill Livingstone, 2004; 1–15.
Chapter 2: Ultrasonography: physics and principles
21
Chapter

Hysterosalpingography

3
Shawky Z. A. Badawy, Stuart J. Singer and Amr Etman

Introduction

The evaluation of the pelvic organs and pathology related to them has always been dependent on proper pelvic examination, rectal examination, and external palpation of the abdomen and pelvic areas. That there were marked limitations to such meth­ods of evaluation was long realized by physicians. About a century ago various investigators developed technologies to visualize the pelvic organs. One of the earliest technologies was the introduction of air into the abdominal cavity using a needle, and there was discussion about what type of air medium should be introduced. Investigators started by using oxygen, but they real ized that this gas takes many hours to be absorbed, thus subjecting the patient to unnecessary pain and discomfort after the procedure. They replaced the oxygen medium by carbon dioxide and found that within 15–20 minutes carbon dioxide is very easily absorbed into the circulation; patients therefore will not have any lasting discomfort and may go home comfortably after the procedure. The practice of intro­ducing carbon dioxide into the abdominal cavity, producing pneumoperitoneum, developed in association with radiological science. After the production of pneumoperitoneum, radiogra­phy is used to visualize many organs in the abdominal cavity, including tumors and adhesions. Contraindications to the use of transabdominal pneumoperitoneum are, of course, the pres­ence of large masses or the suspicion of massive adhesions; the technique is also contraindicated in patients who are suspected of having heart prob lems. This technique was useful for limited evaluation of the pelvic cavity and for outlining the pelvic organs.
Isodor Clinton Rubin introduced the technology of trans­cervical carbon dioxide insuation for diagnosis of tubal path­ology [1]. He used an apparatus that allowed him to monitor the ow of carbon dioxide as well as the pressure during the procedure. Carbon dioxide was introduced through the cervix into the uterus using a cannula with a rubber end that tted onto the cervix and produced a seal with the external os. The carbon dioxide was then allowed to ow and Rubin noted that the pressure usually rose to 60–100 mmHg and then began to drop, indicating that carbon dioxide had easy access through the tubes into the peritoneal cavity and that at least one tube was
patent. If the pressure continued to rise, reaching almost 200 mmHg without any drop, the patient started to suer pain; this suggested that tubes were blocked and the procedure was then terminated. This was an elegant procedure to make a diagnosis of tubal factors in infertility. However, the limitations at that time involved whether one tube or both were patent, which could not be ascertained with this technique, depending on reading the intrauterine pressure with a special manometer. The procedure was supplemented by using a stethoscope to listen suprapubically to the sound of air passing through the tube if it was patent.
Limitations on the use of Rubins insuation test include acute or subacute pelvic infections and also the presence of cervical infection as diagnosed by purulent uid discharging from the cervix.
The principle of transuterine insuation appealed to other investigators, who introduced modication of the technique for producing pneumoperitoneum, which, associated with radio­graphy of the pelvis, proved more useful in outlining pelvic pathology. Essentially, the patient was placed in the knee–chest position or Simms position, the cannula was introduced into the cervix in that position, and transuterine insuation was performed to produce pneumoperitoneum. Certainly if the tubes are open then pneumoperitoneum will be sucient to show on the radiographs, thus outlining the uterus and ovaries and any pathology in the pelvis. Some uterine anomalies might even be diagnosed by this technology.
Clearly, transuterine insuation was an important technol­ogy that laid the foundation for evaluation of the uterus, tubes, and pelvic organs and was an advance that preceded the use of dyes to outline the uterine cavity and the fallopian tube.
Hysterosalpingography
Hysterosalpingography is a technique introduced by Rubin used to visualize the uterine cavity and fallopian tubes. Many investigators have attempted to bypass this method in the evaluation of the infertile couple, going on to laparoscopic and hysteroscopic procedures instead. However, hysterosalpin­gography has withstood the test of time as a noninvasive pro­cedure that is used without any anesthesia, and much of the
Ultrasonography in Reproductive Medicine and Infertility, ed. Botros R. M. B. Rizk. Published by Cambridge University Press. © Cambridge University Press 2010.