Sterilization ofEndoscopic Instruments
https://t.me/med1917
6 Steps inEndoscopic Instrument
Reprocessing [21]
The steps in endoscopic instrument reprocessing include the following:
1. Dismantling
2. Decontamination
3. Pre-cleaning
4. Cleaning and rinsing
5. Rinsing-ultrasonic cleaner
6. Drying-packaging for sterilization
7. Sterilization/disinfection
8. Instrument storage
6.1 Dismantling
Most laparoscopic hand instruments exist as articulations of different parts which should be separated as the rst step in processing. This exposes the grooves and crevices which may har­bour tissue debris and microorganisms. The ports (trocars and cannulae) and sheaths used in hyster­oscopy should all be dismantled for proper processing.
29
6.2 Decontamination
This process should start in the operating room; it reduces the load of microorganisms and renders the instruments safe for handling prior to further processing and transfer to the CSSD.The trained healthcare worker (nurse or certied surgical tech­nician) should wipe off any visible tissue, debris, body uid or bloodstain from the instrument with a sterile but damp sponge. The instruments are then immersed inside a receiver containing disin­fectant such as 0.5% chlorine or hypochlorite solu­tion. They are allowed to soak for 10min. There is no benet in soaking for longer as this predisposes them to corrosive damage (Fig.1).
6.3 Pre-Cleaning
This should be done in the CSSD or in the the­atre, away from the operating room or any other controlled area within the theatre, but not
Fig. 1 Decontamination: the instruments should be immersed in the disinfectant
where sterile instruments are kept. The pre­ferred pre- cleaning method involves the use of enzymatic products such as protease, lipase and amylase. These proteases break down blood and other protein stains on the instru­ments; they should, however, be thoroughly removed from the instruments by cleaning. The use of detergents and soap scrubs may not be effective, because of their limited capacity to lyse protein debris especially on the surfaces of instruments with areas that are hard to reach for cleaning (Fig.2).
6.4 Cleaning andRinsing
Cleaning should be done with soft brushes so that debris can be removed from the groves of instru­ments. Rinsing is best done under running water
30
https://t.me/med1917
J. O. Imaralu et al.
The solution used for ultrasonic cleaning should have a neutral pH (to prevent corrosive damage) and instruments should be immersed in it for 10–15min. Instruments made of similar material—metals such as aluminium—should be cleaned together separate from other instruments made from, for example, stainless steel and other materials. To allow opti­mum penetration and effect, the ultrasonic cleaner should not be overloaded and all instruments must be open. Rinsing must be done for all instruments that have undergone ultrasonic cleaning.
6.5 Drying
It is necessary to exclude water droplets from the instruments before they are packed for steriliza­tion. This can be achieved with the use of an oven in the CSSD or an air gun.
6.6 Sterilization or Disinfection
Fig. 2 Enzymatic products are preferred for
pre-cleaning
to clear all particulate matter and chemicals used for decontamination, pre-cleaning and cleaning.
High-pressure running water taps or cleaning guns, where available, can be used for rinsing. These guns give a directed jet of water to long instruments and instrument grooves. Stagnant water in containers should be avoided during manual cleaning as it may retain particles and solutions which may get on to other instruments. All instruments should be open during the clean­ing process, to expose ratchets and locks and indeed all parts of the instrument to the cleaning agent. If cleaning and rinsing are done together, re-rinsing or proper rinsing must be done again after a brush or other forms of cleaning.
Ultrasonic cleaning is a method that is rapidly increasing in popularity. The ease of use, time saved and better efciency make it a better method than manual cleaning. Ultrasonic cleaning is 16 times more efcient than hand cleaning.
It is widely recommended that rigid endoscopic instruments should be sterilized or at least high­level disinfected where sterilization is not feasi­ble. Users of endoscopic instruments, however, should take note of manufacturer information regarding what to sterilize or high-level disinfect, the method of sterilization, the type of disinfec­tant to be used, their dilution instructions and the number of times or length of time a constituted high-level disinfectant solution can be used.
Sterilization is superior to HLD, for infection control; however, because of the fragile nature of some instruments, sterilization methods may not be feasible. The cost and availability of the steriliza­tion methods are also important considerations.
6.6.1 Sterilization
There are three major sterilization methods:
1. Steam sterilization
2. Gas sterilization
3. Peracetic acid sterilization
Steam Sterilization
The use of the autoclave for sterilization is one
of the commonest methods employed. It is
Sterilization ofEndoscopic Instruments
https://t.me/med1917
31
cheap and non-toxic yet an effective method. Autoclaving at 121°C for 15min is sufcient for all reusable metal instruments. This method is effective for all rigid endoscopic instruments and some telescopes. Packaging of instruments for steam sterilization should be such that all items that can be damaged by heat conduction like silicone tubing, rubber insulations and coatings on cables and instru­ments are shielded by wrapping in two layers of linen to prevent contact with the metal con­tainer of the autoclave.
Some Notes on Flash and Vacuum
Sterilization
1. Flash Sterilization This was dened by Underwood and
Perkins as sterilization of unwrapped items at 132°C for 3min at 27–28 lbs of pressure in a gravity displacement sterilizer [22]. It was designed for use in the theatre for emergent and immediate use of a forgotten or unanticipated instrument or one that was accidentally dropped on the oor. It is a modication of conventional steam sterilization whereby the ashed item is placed in an open tray or is placed in a specially designed, covered, rigid container to allow for rapid penetration of steam. The term “immedi­ate-use sterilization” currently replaced ash sterilization. It is sterilization with little or no dry time available. It is safe and efcacious when done correctly and when only used to process instruments intended for immediate use. It should not be used for reasons of conve­nience or to save time.
2. Vacuum Sterilization A sterilizer congured to run a vacuum
cycle is equipped with a vacuum system. A typical vacuum cycle will begin with a series of alternating steam pressure injections and vac­uum draws (also called pulses) to dramatically remove the air from the chamber. This will allow steam to be sucked into areas where it would otherwise have difculty penetrating.
• Low-Temperature Steam and Formaldehyde
Sterilization This method is especially suited for heat­sensitive instruments like cables, insulations, tubes and instruments with blind-ending
lumina. It employs heat at temperature ranges of 55–80°C, for about 5–6h. It is important to note that the use of formaldehyde for steriliza­tion is fast diminishing in popularity; some reasons for this include the potential risk for toxicity and irritation when in contact with the skin, eyes, nose and respiratory tract. Its ef­cacy in the sterilization process is being que­ried in reports from some studies.
Gas Sterilization Gas sterilization is useful for hand instruments that are coated with insulation material, light cables and tubing for gas and suction irriga­tion. Disposable instruments can also be reprocessed using this method. Ethylene oxide (ETO) is the gas that is used for gas steriliza­tion. Cold ETO sterilization is done at a temperature of 85 °C and instruments are exposed to the gas for 4h 30min, and they are then aerated for 12h. Warm ETO sterilization is, however, done at a temperature of 145 °C with instrument exposure of 2h 30min and a shorter aeration time of 8h. Gas sterilization has the advantage of ensuring that porous materials and deeper instrument areas are reached without corrosive damage to them. Its demerit, however, is that the set-up is expen­sive and the required steps of exposure and aeration consume a lot of time.
Newer Methods of Sterilization This method uses the plasma chamber (STERRAD) developed by Johnson and Johnson. It uses hydrogen peroxide vapour and low- temperature gas plasma. The entire process with this method takes only about 25 min. The instruments should have under­gone the earlier decontamination-cleaning cycle and should have been properly pack­aged. A vacuum is created inside the equip­ment chamber and 59% aqueous H2O2 is vaporized into it, enveloping all the items until when the H2O2 vapour is well diffused. The chamber pressure is then reduced to allow the generation of low- temperature gas plasma. A radiofrequency amplier is used at this time to introduce radiofrequency energy into the chamber—thus causing the generation of reactive free radical species from the vapourized H2O2 which comes in contact with
32
https://t.me/med1917
J. O. Imaralu et al.
the instruments. The reactive energy species also react with each other to form water vapour, oxygen and other non-toxic by­products. The instruments are then dried by the equipment and are ready for use or stor­age. The advantages of this new method that make it cost-effective include the fact that it can be used in areas where there is rapid turn­over and a shorter processing time is required. There are no toxic residues and there is no requirement for aeration or water hook- up within the set-up. The equipment is not big and does not take space; it requires only a reli­able source of electricity supply.
6.6.2 Disinfection
High-level disinfection (HLD) is particularly useful for semi-critical instruments and fragile heat-sensitive instruments. Its effect on spores is, however, relative and depends on the duration of exposure of microorganisms to disinfectant, the concentration and temperature of the disinfec­tant, the amount and kind of microorganisms, organic material (blood, other uids, tissues) and other matter (for example, dirt) present on the instrument or other item and the access to the microorganisms (for example, grooves can shield microorganisms). It is therefore pertinent that instruments for HLD undergo the preliminary phases for thorough decontamination.
HLD destroys all pathogenic microorganisms and some bacterial spores (e.g. Clostridium dif- cile) during short exposure. During long expo­sure, it is sporicidal for a higher number of bacterial endospores and has a limited sporicidal capability for mycobacteria, viruses and fungi. Mycobacteria are used as indicators of effective­ness. The agents that are popularly used for HLD include 2% glutaraldehyde, 6% stabilized hydro­gen peroxide and peracetic acid (a mixture of acetic acid and hydrogen peroxide).
The ideal chemical high-level disinfectant should have the following characteristics:
1. Rapid turnover rate HLD (<10min).
2. Effective against a wide range of organisms
including blood-borne viruses and prion proteins.
3. There should not be any disinfectant residue after rinsing.
4. Excellent compatibility with endoscopes and other instrument materials.
5. Long shelf-life.
6. Non-irritant and non-toxic to users.
7. Environment-friendly, i.e. have no disposal problems.
Glutaraldehyde
This is the most recommended chemical dis­infectant agent, by most laparoscopic instru­ment manufacturers. Some of its advantages include the following: it is safe, is relatively inexpensive, has good microbicidal activity, is non-corrosive to optics, and is effective in the presence of protein matter. It is, however, irri­tating to the eyes, skin (allergic contact der­matitis), and respiratory mucosa especially at higher concentrations 0.3 parts per million (ppm). It has a pungent and irritating odour and can coagulate and x blood and tissues to surfaces. It thus requires thorough rinsing off from the instrument before use. Most com­mercially available glutaraldehyde solutions can be used (after preparation) for between 14 and 30 days. The manufacturer’s preference for the duration of usage varies and it is advis­able that this is followed. The effectiveness of glutaraldehyde as an HLD depends on its con­centration, the instrument contact time, the physical conguration of the instrument, tem­perature, pH and the amount of organic matter on the instruments. The OSHA recommends that the maximum allowable concentration of glutaraldehyde is 0.2 ppm. Instrument expo­sure can however be varied to suit the HLD needs, for instance, bre-optic light source cables and telescopes should be soaked for a minimum of 10min. The endocamera (charge­couple device [CCD] only) can also be soaked in 2% glutaraldehyde for 10 min, but care must be taken to avoid soaking the plug part of the cable. Soaking of optical instruments should be for not more than 20min. Metallic hand or accessory, trocars, reducers, ring/clip applicators and dye injectors should be soaked
Sterilization ofEndoscopic Instruments
https://t.me/med1917
in 2% glutaraldehyde for 60min (long expo­sure) to achieve better destruction of myco­bacterial spores. Glutaraldehyde is marketed with trade names as Cidex or MetriCide. Another form known as rapid Cidex or Rapicide exists which provides for shorter instrument exposure of 5min.
A particular non-tubercolous mycobacte­rium, M. massiliense BRA100 strain, has been found to be resistant to high-level disin­fection with glutaraldehyde, even at high concentrations (7% glutaraldehyde), thus suggesting that glutaraldehyde may not be effective for rapidly growing mycobacteria. Orthophthalaldehyde (OPA) and peracetic acid may substitute glutaraldehyde for HLD with better efcacy [23].
Orthophthalaldehyde (OPA) This is sold as Cidex OPA and it exists in the concentration of 0.55%. This agent can pro­vide effective high-level disinfection and can destroy all bacteria, fungi and mycobacterial spores with instrument exposures of as short as 5–12 min. Its minimum effective concen­tration is 0.3%, and after preparation, the solu­tion can still be effective for 14 days. It is non-irritant to the eyes and mucus membranes and non-corrosive to instruments and has a weak odour. Its disadvantages, however, are that it is expensive and stains protein grey, and hypersensitivity reaction has been reported in some patients with a history of bladder cancer [19] (Fig.3).
Notes on Hydrogen Peroxide and Peracetic
Acid for High-Level Disinfection
The characteristics of an ideal chemical steril­ant used as a high-level disinfectant should include broad antimicrobial spectrum, rapid activity and masterial compatibility. The ideal HLD should be odourless, be safe, have pro­longed shelf-life, and be generally cost­effective. Hydrogen peroxide and peracetic acid are both oxidizing agents that generate free radicals which can attack membrane lip-
33
Fig. 3 Orthophthalaldehyde (OPA)
ids, DNA and other essential cell components. Cidex PA is one brand of high-level disinfec­tant that contains 0.08% peracetic acid and
1.0% hydrogen peroxide [24, 25]. It inacti­vates all microorganisms within 20min with the exception of bacterial spores. The risk of corrosive damage to instruments is a major drawback to this disinfectant. Olympus does not endorse Cidex PA on any of their endo­scopes and will not assume any liability for chemical damage. The manufacturers of Cidex PA currently are reformulating this product (altering the buffer system and changing anti­corrosive inhibitors) to improve its material compatibility.
34
https://t.me/med1917
J. O. Imaralu et al.
6.6.3 Quality Control Measures
forSterilization andDisinfection
Bacteria indicator count: Cultures swabs should be taken from the telescope and other instrument tips and surfaces routinely.
Strip checks on the disinfectant: Manufacturer directions should be followed.
Temperature and pH monitoring: They are also essential, for optimal effect of the steril­ants and chemical disinfectants.
Keep a log for accountability: This includes a log reporting entry and exit of instruments at every stage of reprocessing with timing prop­erly recorded. A log of both instrument and sterilant usage frequency should also be kept.
7 Storage ofProcessed
Endoscopic Instruments
The proper storage of processed unused instru­ments is very important. It is recommended that high-level disinfection be done when instruments are needed for use, because of the short process­ing time involved. However, because sterilization is more expensive, time-consuming and tedious and has greater potential to eradicate all forms of microbial existence, storage of sterilized instru­ments becomes expedient. In addition, third­party sterilized instruments would need storage when they arrive at centres where they are required before use, or before transport to centres where they are required.
Effective storage of endoscopic instruments
to prevent contamination should take note of the following points: the storage shelves and cabinets (whether open or closed) should be dry or have very low humidity, the material used for wrapping the instruments should be such that would remain intact and dry, han­dling of sterile packs not meant for use should be limited, and the storage time should be kept as minimal as possible, to prevent moisture and contamination.
8 Notes onEmerging Disease
Pathogens Related toEndoscopic Surgery
As minimal access surgery becomes increasingly available, it becomes more accessible to patients with chronic infectious diseases. Examples of these infections include hepatitis C virus, HIV, Clostridium difcile, Cryptosporidium, antibiotic- resistant microbes (MDR-TB, VRE, MRSA), Helicobacter pylori, newer strains of E. coli, coronaviruses (SARS-CoV and MERs­CoV) and Creutzfeldt-Jakob disease (no brain, eye, or spinal cord contact).
Standard disinfection and sterilization proce­dures for patient and equipment care are adequate to sterilize or disinfect instruments or devices contaminated with blood and other body uids from persons infected with emerging pathogens. Where there is doubt, disposable or single-use endoscopic instruments should be used.
9 Conclusion
Endoscopic surgical equipment and instruments are peculiar for their complexity and high cost; the level of disinfection or sterilization is depen­dent on the intended use of the item. Critical items (devices that contact sterile tissue such as surgical hand instruments) require sterilization, while semi-critical items (devices that contact mucus membranes such as endoscopes) require a mini­mum of high-level disinfection, and non- critical items (objects that contact intact skin such as electrosurgical return plates, patches and cables) require low-level disinfection. Cleaning should always precede disinfection and sterilization.
Learning Points
• Endoscopic instruments are complex in design
and contain many articulations and crevices
where tissue debris and microorganisms can
get lodged.
• The focus of endoscopic instrument repro-
cessing includes the prevention of hospital-
Sterilization ofEndoscopic Instruments
https://t.me/med1917
35
acquired surgical site infections and prolongation of instrument lifespan.
• Most laparoscopic instruments can be classi­ed as semi-critical devices and high-level disinfection is considered appropriate.
• All the preliminary steps in endoscopic instru­ment processing should be followed to aid effective sterilization or disinfection.
References
1. Manatakis DK, Georgopoulos N.Reducing the cost of laparoscopy: reusable versus disposable laparoscopic instruments. Minim Invasive Surg. 2014;2014:408171.
https://doi.org/10.1155/2014/408171.
2. Lau H, Lee F, Patil NG, Yuen WK. Two hundred endoscopic extraperitoneal inguinal hernioplasties: cost containment by reusable instruments. Chin Med J. 2002;115(6):888–91.
3. Adler S, Scherrer M, Rückauer KD, Daschner FD. Comparison of economic and environmental impacts between disposable and reusable instruments used for laparoscopic cholecystectomy. Surg Endosc Other Interv Tech. 2005;19(2):268–72.
4. Slater M, Booth MI, Dehn TC.Cost-effective lapa­roscopic cholecystectomy. Ann R Coll Surg Engl. 2009;91(8):670–2.
5. Vasios N, Verroiotou M, Komninos G, Arapoglou S, Fragandreas G.Study of the cost of mainly reusable equipment for laparoscopic cholecystectomy. Hell J Surg. 2013;85(2):101–4.
6. Lopes CLB, Graziano KU, Pinto TJA.Evaluation of single–use reprocessed laparoscopic instrument steril­ization. Rev Lat Am Enfemagem. 2011;19(2):370–7.
7. Stockley JM, Allen RM, Thomlinson DF, Constantine CE. A district general hospital’s method of post­operative infection surveillance including post­discharge follow-up, developed over a ve-year period. J Hosp Infect. 2001;49:48–54.
8. Sasmal PK, Mishra TS, Rath S, Meher S, Mohapatra D. Port site infection in laparoscopic surgery: a review of its management. World J Clin Cases. 2015;3(10):864–71. https://doi.org/10.12998/wjcc.
v3.i10.864.
9. Centers for Disease Control and Prevention. The National Healthcare Safety Network (NHSN) man­ual: patient safety component. Atlanta, GA: Division of Healthcare Quality Promotion, National Center for Emerging and Zoonotic Infections Diseases; 2019.
http://www.cdc.gov/nhsn/acute-care-hospital/index. html.
10. Lilani SP, Jangale N, Chowdhary A, Daver GB. Surgical site infection in clean and clean-contami­nated cases. Indian J Med Microbiol. 2005;23:249–52.
11. Brill A, Ghosh K, Gunnarsson C, Rizzo J, Fullum T, Maxey C, Brossette S. The effects of laparoscopic
cholecystectomy, hysterectomy, and appendec­tomy on nosocomial infection risks. Surg Endosc. 2008;22:1112–8.
12. Richards C, Edwards J, Culver D, Emori TG, Tolson J, Gaynes R.Does using a laparoscopic approach to cholecystectomy decrease the risk of surgical site infection? Ann Surg. 2003;237:358–62.
13. Hamzaoglu I, Baca B, Böler DE, Polat E, Ozer Y.Is umbilical ora responsible for wound infection after laparoscopic surgery? Surg Laparosc Endosc Percutan Tech. 2004;14:263–7.
14. Mangram AJ, Horan TC, Pearson ML, Silver LC, Jarvis WR.Guideline for prevention of surgical site infection, 1999. Centers for Disease Control and Prevention (CDC) hospital infection control prac­tices advisory committee. Am J Infect Control. 1999;27:97–132. quiz 133-134; discussion 96
15. Falkinham JO. Epidemiology of infection by non tuberculous mycobacteria. Clin Microbiol Rev. 1996;9:177–215.
16. Muthusami JC, Vyas FL, Mukundan U, Jesudason MR, Govil S, Jesudason SR. Mycobacterium fortuitum: an iatrogenic cause of soft tissue infection in surgery. ANZ J Surg. 2004;74:662–6.
17. Vijayaraghavan R, Chandrashekhar R, Sujatha Y, Belagavi CS.Hospital outbreak of atypical mycobac­terial infection of port sites after laparoscopic surgery. J Hosp Infect. 2006;64:344–7.
18. Duarte RS, Lourenço MC, Fonseca Lde S, Leão SC, Amorim Ede L, Rocha IL, Coelho FS, Viana-Niero C, Gomes KM, da Silva MG, etal. Epidemic of post­surgical infections caused by Mycobacterium massil­iense. J Clin Microbiol. 2009;47:2149–55.
19. Colak T, Ersoz G, Akca T, Kani KA, Aydin S.Efcacy and safety of reuse of disposable laparoscopic instru­ments in laparoscopic cholecystectomy: a prospective randomized study. Surg Endosc. 2004;18:727–31.
20. Occupational Safety & Health Administration. US Department of Labor. Bloodborne patho­gens. 29CFR1910.1030.1991. http://www.osha.
gov/pls/oshaweb/owadisp.show_document?p_ table=STANDARDS&p_id=10051.
21. Bhandarkar DS, Katara AN.Sterilization and main­tenance of laparoscopic equipment. In: Kriplani A, Bhatia P, Prasad A, et al., editors. Comprehensive laparoscopic surgery; 2007. p.23–8.
22. Carlo A.The new era of ash sterilization. AORN J. 2007;88:58–68.
23. Lorena NS, Pitombo MB, Côrtes PB, Maya MC, Silva MG, Carvalho AC, et al. Mycobacterium mas­siliense BRA100 strain recovered from postsurgical infections: resistance to high concentrations of glutar­aldehyde and alternative solutions for high level dis­infection. Acta Cir Bras. 2010;25:455–9.
24. Rutala WA, Weber DJ. Disinfection of endoscopes: review of new chemical sterilants used for high­level disinfection. Infect Control Hosp Epidemiol. 1999;20(1):69–76.
25. Rutala WA, Weber DJ.Disinfection and sterilization: an overview. Am J Infect Control. 2013;41:52–5.
The Electrosurgical Unit: Basic
https://t.me/med1917
Principles forPractice
JosephIfeanyichukwuIkechebelu andChukwuemekaO.Ezeama
1 Historical Perspective
Human beings have traditionally been dependent on re for their very existence; unsurprisingly its use for medicinal purposes was initiated shortly after its discovery [1, 2]. Cautery is mentioned in some very early medicinal texts credited to Hippocrates [2, 3]. It was used to stop bleeding from injuries, amputations, etc.
From the late eighteenth century up until the late nineteenth century, the discovery of electric­ity and its application rose exponentially. Several renowned scientists including Galvani, Becquerel, Sèrè, D’Arsonval, Doyen and Nagelschmidt were pioneers in the eld of elec­tricity and its use in medicine.
J. I. Ikechebelu (*) Gynaecology Endoscopy and Assisted Reproduction Unit, Department of Obstetrics and Gynaecology, Nnamdi Azikiwe University, Awka, Nigeria e-mail: ji.ikechebelu@unizik.edu.ng
C. O. Ezeama Gynaecology Endoscopy Unit, Department of Obstetrics and Gynaecology, Nnamdi Azikiwe University Teaching Hospital, Nnewi, Anambra State, Nigeria e-mail: co.ezeama@unizik.edu.ng
The basic electrosurgical unit was produced by William Bovie. His invention was tested in 1926, while its use during surgery for a vascular myeloma was carried out by Dr. Harvey Cushing [4]. Since the rst ‘Bovie’ unit was used, many modications have been introduced to make this instrument a safe and important part of the sur­geon’s armamentarium.
It works by applying the principle that passing an alternating current through tissues generates certain effects which the surgeon can leverage to improve surgical precision and reduce bleeding during procedures. Intimate knowledge of the principles of electrosurgery is paramount for cur­rent surgical practice [5].
2 Biophysical Principles
The three main properties of electricity are cur­rent (I), voltage (V), and impedance (R).
Current is a measure of the movement of electrons within a circuit. It is measured in
amperes.
Voltage expresses the force required to push electrons through a tissue; this is measured in
volts.
Impedance/resistance is a measure of the dif­culty the current must overcome to pass through a given substance (i.e., biological tissues); it is measured in ohms.
© The Editor(s) (if applicable) and The Author(s), under exclusive license to Springer Nature Switzerland AG 2022 J. E. Okohue et al. (eds.), Gynaecological Endoscopic Surgery,
https://doi.org/10.1007/978-3-030-86768-3_4
37
38
https://t.me/med1917
J. I. Ikechebelu and C. O. Ezeama
In an electric circuit, these three properties are
expressed by Ohm’s law: I=V/R.
Power is the ability to do work per unit of time and is measured in watts. It is mathemati­cally expressed as the product of voltage and cur­rent W=V × I
Since I=V/R,
Then W (power)=V × V/R=V2/R
Or W=V × I=I2R
Heat/energy produced=W × T (time)=I2RT
3 Types ofCurrent
3.1 Direct Current
As seen in computer and car batteries. Here the current is unidirectional and does not change polarity.
3.2 Alternating Current
where the current changes direction (polarity) periodically as seen in our home electrical outlets.
Application of direct current during surgery would lead to an electrolytic effect causing gen­eration of sodium hydroxide and hydrogen chlo­ride. These would have a seriously deleterious effect on the cells leading to necrosis. The use of low-frequency alternating currents (under 10,000Hz), apart from causing similar problems, also predisposes to stimulation of nerves and muscles (Faraday effect) [6]. Due to these unwanted effects, the current used in electrosur­gery is high-frequency alternating current (over 300kHz).
From the above, it can be surmised that the human body, when inserted into an incomplete circuit, completes the circuit and conducts electricity. When voltage is applied across the human body, its electrons are forced to move. The tissues of the body resist the ow of these electrons and in doing so cause a local rise in temperature. The thermal energy occurring in these areas of the body causes local tissue dam­age and destruction.
The biophysical principle equations above show clearly that the voltage is directly propor­tional to the power produced and also the tissue destruction it produces. Also, the time the elec­trode is applied is directly proportional to the thermal effects generated. The surface area of the electrode used or the tissue in contact is inversely proportional to the heat produced at the site of application per unit time.
3.3 Thermal Eects onTissue
Electrical currents when passed through tissues particularly those with a high uid content cause movement of molecules and subsequent heat generation. The tissue effect of this heat varies at different temperatures.
4 The Electrosurgical Generator
(Electrosurgical Unit [ESU])
In the earlier section on the type of current, the alternating high-frequency current was recom­mended as most appropriate for electrosurgery. To generate this current, special equipment called electrosurgical generators are used. They ramp up the normal current frequency seen in house/ theatre outlets from 50–60 Hz to 500,000– 3,000,000Hz [7]. This very high frequency does not allow the faradic effect.
The tissue effect of this current depends on several factors, but the thermal damage on tissue is generally directly proportional to the volume of current passing through it. This thermal effect is a function of current density.
The size/cross-sectional area of the elec­trodes determines the density of the current at its tip. Smaller electrodes have a larger current density than larger-sized electrodes. Various tis­sue types have different electrical resistance, which affects the rate of heating. Adipose tissue and bones have high resistance and are poor conductors of electricity, whereas muscle and skin are good conductors of electricity and have low resistance.
The Electrosurgical Unit: Basic Principles forPractice
https://t.me/med1917
39
Fig. 1 ESU output modes
Cutting mode
Blend mode
Coagulation
mode
Fulguration
mode
The output from the generators can be modi­ed to produce specic tissue effects. The vari­ous outputs are determined by the shape of the waveform and the duration of the specic duty cycle (the term duty cycle is used to describe the percentage of time that the ESU is producing a waveform—if it is 80% of the time, it is referred to as an 80% duty cycle).
Output modes include cutting, coagulation, blend, and fulguration modes (see Fig.1).
4.1 Cutting Mode
Peak Voltage
Average Voltage
Watts [9]. The coagulation mode can be used for desiccation or fulguration depending on the active electrode positioning to the target tissue.
4.3 Desiccation
Using the coagulation mode, the surgeon can desiccate (loss of water from the tissue cells but no signicant protein damage). The end result is deeper necrosis and greater thermal spread. For desiccation, direct contact of the electrode with the target tissue is necessary.
This mode has a continuously high current and low-voltage waveform, causing a very rapid rise in temperature of up to 100°C.As earlier noted, this causes vaporization of cell uid and disrup­tion of their membranes, thus resulting in cutting with no coagulation. The duty cycle in this mode is 100%. A power setting between 40 and 80 Watts is ideal for this mode [8].
4.2 Coagulation Mode
This is a high-voltage mode with the waves dampened and produced in bursts. The duty cycle in this mode is typically 6%. The pulsed nature of the current delivery allows for cooling of the tis­sues in between pulses and minimizes the cutting effect while allowing for the production of coag­ulum due to dehydration and denaturing of the tissue proteins. This coagulum is very effective for haemostasis. The coagulation current is oper­ated with a power setting between 30 and 50
4.4 Fulguration
This is coagulation in a non-contact position. The electrode is held 2–4 mm away from the target tissue. An arc of current is formed which nds its way to the tissue in a spray-like effect, hence the name ‘spray coagulation’. This is due to the rapid rise in temperature and subsequent rapid cooling resulting in supercial coagulation, typically to a depth of about 0.5 mm. It covers a wider area than the surface area of the electrode alone. Its benet is seen in small arteriolar and capillary bleeding.
4.5 Blended Current (Cut/ Coagulation)
Blended currents are actually modulated low­voltage waveforms. It produces different tissue effects by modulating the duration of the duty