Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_2656_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
47 Мб
Скачать
156 PHYSICAL EXAMINATION
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
UNIT 2
FIGURE 6.8 Peripheral pulse sites
Temporal Carotid
Apical Brachial
Radial Femoral
Popliteal
Posterior tibial Dorsalis pedis
Site
Peripheral pulses can be palpated where the large arteries are close to the skin surface. There are nine common sites for assessment of pulse, as indicated in
Figure 6.8.
When routine vital signs are assessed, the pulse is generally measured at one of two sites: radial or apical.
Measuring the apical pulse is indicated for consumers with an irregular pulse or known cardiac or pulmonary disease. The assessment of apical pulse can be accomplished through palpation but is most commonly accomplished through auscultation.
Radial pulse
To palpate the radial pulse, follow these steps.
E
1. Place the pad of your rst, second and/or third nger on the site of the radial
pulse, along the radial bone on the thumb side of the inner wrist (see
Figure 6.9).
2. Press your ngers gently against the artery with enough pressure so that you
can feel the pulse. Pressing too hard will obliterate the pulse.
3. Count the pulse rate using the second hand of your watch. If the pulse is
regular, count for 30 seconds and multiply by 2 to obtain the pulse rate per minute. If the pulse is irregular, count for 60 seconds.
4. Identify the pulse rhythm as you palpate (regular or irregular).
5. Identify the pulse volume as you palpate (using scales from Table 6.2).
N A P
Refer to section on Rate.
Apical pulse
To assess the apical pulse, follow these steps (see Figure 6.10).
E
1. Place the diaphragm of the stethoscope on the apical pulse site.
2. Count the pulse rate for 30 seconds if regular, 60 seconds if irregular.
3. Identify the pulse rhythm.
4. Identify a pulse decit (apical pulse rate greater than the radial pulse rate)
by listening to the apical pulse and palpating the radial pulse simultaneously.
N A P
Refer to section on Rate.
FIGURE 6.9 Palpation of the radial pulse
FIGURE 6.10 Auscultation of the apical pulse
E
ExaminationNNormal ndingsAAbnormal ndingsPPathophysiology
EXAMINATION REQUIREMENTS FOR EVERY CONSUMER 157
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
Rate
Normal pulse rates vary with age. Table 6.3 depicts ranges for normal pulse rates
N
by age. The heart rate normally increases during periods of exertion. Athletes commonly have resting heart rates below 60 beats per minute because of the increased strength and efciency of the cardiac muscle.
TABLE 6.3 Pulse rate: Normal range according to age
CHAPTER 6
AGE RESTING PULSE RATE
AVERAGE
(BEATS PER MINUTE)
Newborn 100–170 140 1 year 80–160 120 3 years 80–120 110 6 years 70–115 100 10 years 70–110 90 14 years 60–110 85–90 Adult 60–100 72
A
Tachycardia refers to a pulse rate faster than 100 beats per minute in an adult.
Psychophysiological stressors such as trauma, blood volume losses, anaemia,
P
infection, fear, fever, pain, hyperthyroidism, shock and anxiety can increase pulse rate because of increased metabolic demands placed on the body.
Some tachycardia may not have clinical signicance; however, in consumers
P
with myocardial disease, tachycardia can be a sign of decreased cardiac output, congestive heart failure, myocardial ischaemia or dysrhythmias.
A
Bradycardia refers to slow pulse rates. Pulse rates that fall below 60 beats per
minute in adults are considered to be bradycardic. Medications such as cardiotonics (digoxin) and beta blockers decrease the
P
heart rate. Bradycardia usually occurs with excessive vagal stimulation or decreased
P
sympathetic tone. Conditions that may cause bradycardia are eye surgery, increased intracranial pressure, myocardial infarction, hypothyroidism and prolonged vomiting.
A
Asystole refers to the absence of a pulse. Palpate or auscultate for a pulse for
10–15 seconds to establish asystole. Cardiac arrest resulting from biological or clinical death results in asystole.
P
Pulseless electrical activity (electromechanical dissociation) caused by, for example,
P
hypovolaemia, pneumothorax, cardiac tamponade or acidosis results in the absence of a pulse despite the presence of electrical activity in the heart muscle.
A
A pulse decit occurs when the apical pulse rate is greater than the radial pulse rate. Dysrhythmias (such as atrial brillation, premature ventricular contractions,
P
second-degree heart block, third-degree heart block) and heart failure can cause pulse decits because some heart contractions are too weak to produce a pulse pressure to the peripheral site. Severe vascular disease can also cause pulse decits.
Rhythm
N
Normal pulse rhythm is regular, with equal intervals between each beat.
A
Dysrhythmias, or arrhythmias, refer to pulse rhythms that are not regular.
They may consist of irregular beats that are random, or irregular beats that present in a regular pattern.
Cardiac dysrhythmias that are atrial and ventricular in origin cause abnormal
P
rhythms, such as atrial utter and ventricular brillation.
E
ExaminationNNormal ndingsAAbnormal ndingsPPathophysiology
158 PHYSICAL EXAMINATION
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
UNIT 2
Volume
The pulse volume is normally the same with each beat. A normal pulse volume
N
can be felt with a moderate amount of pressure of the ngers and obliterated
with greater pressure.
A
Small, weak pulses are referred to as weak or thready pulses or as pulses easily
obliterated with light pressure.
Decreased cardiac stroke volume caused by heart failure, hypovolaemic shock or
P
cardiogenic shock can result in weak pulses.
A low pulse amplitude occurs in states of increased peripheral vascular
P
resistance, such as in aortic stenosis and constrictive pericarditis.
Weak pulses occur in conditions in which ventricular lling time is decreased,
P
such as in dysrhythmias.
A
Bounding pulses are full, forceful pulses that are difcult to obliterate with pressure.
Hyperkinetic states such as exercise, fever, anaemia, anxiety and
P
hyperthyroidism can cause bounding pulses.
Early stages of septic shock are characterised by bounding pulses because of
P
decreased peripheral vascular resistance.
Temperature
Scales, variables, routes, and measurement methods for assessing temperature are outlined.
Celsius
42
41
40
39
38
37
36
35
34
To convert:
(9/5 3 temperature in Celsius) + 32° = temperature in Fahrenheit
5/9 3 (temperature in Fahrenheit – 32°) = temperature in Celsius
FIGURE 6.11 Correlation between Celsius
and Fahrenheit scales
Fahrenheit
107.6
105.8
104.0
102.2
100.4
98.6
96.8
95.0
93.2
Temperature scales
In Australia and New Zealand, degree Celsius is the most commonly used measurement scale for
temperature, which reects the degree of core body heat.
Figure 6.11 summarises and correlates both Celsius and Fahrenheit scales and gives
the conversion formulas.
Variables affecting body temperature
Core body temperature is established by the temperature of blood perfusing the area of the hypothalamus (the body’s temperature control centre), which triggers the body’s physiological response to temperature. An ideal thermometer would accurately measure central brain stem temperature at the hypothalamus. Invasive procedures that provide temperatures of the arterial blood, oesophagus or bladder are reliable indicators of core temperature, but are impractical. More practical methods for measurement of body temperature are less reliable and can result in variations in body temperature readings. In addition, there are physiological variables that affect body temperature. These include:
> circadian rhythm patterns. Normal body temperature (as well as pulse and
blood pressure) uctuates with a consumer’s activity level and the time of day.
lowest in the early morning just before awakening from sleep, and highest in
the afternoon or early evening. A 0.5°C to 1.0°C uctuation in body temperature
throughout the day is considered within the normal range.
> hormones. In women, increased production of progesterone at the time of
ovulation raises the basal body temperature about 0.35°C.
> age. Infants and young children are affected by the environmental temperature
to a much greater extent than adults because their thermoregulation
mechanisms are not fully developed. The elderly are more sensitive to extremes
of environmental temperature due to a decrease in thermoregulatory controls.
> exercise. Body temperature rises due to increased metabolic activity. > stress. Stimulation of the sympathetic nervous system increases the production
of adrenaline, resulting in increased metabolic activity and higher
body temperature.
E
ExaminationNNormal ndingsAAbnormal ndingsPPathophysiology
EXAMINATION REQUIREMENTS FOR EVERY CONSUMER 159
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
> environmental extremes of hot or cold. > health status. Temperature deviations can occur in illnesses such as infections.
as well as in hypothalamus dysfunction.
Measurement routes
There are four basic routes by which temperature can be measured: oral, rectal, axillary and tympanic. Each method has advantages and disadvantages. The advantages and disadvantages of each route are summarised in
TABLE 6.4 Advantages and disadvantages of four routes for body temperature measurement
ROUTE NORMAL RANGE ADVANTAGES DISADVANTAGES
Oral
Average 37.0°C 35.5–37.5°C Convenient;
accessible
Rectal
Average 0.4°C higher than oral
Axillary
Average 0.6°C lower than oral
Tympanic
Calibrated to oral or rectal scales
36.6–38.0°C Considered most accurate
35.5–37.5°C Safe; non-invasive Time frame: Glass thermometer must be left in place for 5 minutes to obtain
35.8–38.0°C Convenient, fast, safe, non-invasive; does not require contact with any mucous membrane
Table 6.4.
Safety: Consumers need to be alert and cooperative, and cognitively capable of following instructions for safe use. Physical abilities: Consumers need to be able to breathe through the nose and be without oral pathology or recent oral surgery; route not applicable for comatose or confused consumers. Accuracy: Oxygen therapy by mask, as well as ingestion of hot or cold drinks immediately before oral temperature measurement, affects accuracy of the reading.
Safety: Contraindicated following rectal surgery. Risk of rectal perforation in children less than 2 years of age. Risk of stimulating Valsalva manoeuvre in cardiac consumers. Physical aspects: Invasive, uncomfortable, and possibly embarrassing.
accurate measurement. Glass and mercury thermometers are no longer used in healthcare facilities. Placement and position of thermometer tip affect reading.
Accuracy:
with other body temperature measurements. Technique affects reading. Tympanic membrane is thought to reect the core body temperature.
Research is inconclusive as to accuracy of readings and correlations
CHAPTER 6
Oral method
E
1. Place the thermometer (Figure 6.12) at the base of the tongue and to the right
or left of the frenulum, and instruct the individual to close the lips and to avoid biting the thermometer. Ensure that it has been at least 15 minutes since the individual consumed a hot or cold beverage or food.
2. Leave the thermometer in the mouth until the device has signalled that the
maximum body temperature has been reached.
3. Remove the thermometer from the person’s mouth.
Rectal method
E
1. Position individual with the buttocks exposed. Adults may be more
comfortable lying on the side (with knees slightly exed), facing away from you, or prone.
2. Put on nonsterile gloves.
3. Lubricate the tip of the thermometer with a water-soluble lubricant.
4. Ask the individual to take a deep breath; insert the thermometer into the
anus 1.5 to 2.5 cm, depending on the person’s age.
5. Do not force the insertion of the thermometer or insert into faeces.
E
ExaminationNNormal ndingsAAbnormal ndingsPPathophysiology
FIGURE 6.12 Taking a consumer’s oral
temperature with an electronic thermometer
160 PHYSICAL EXAMINATION
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
E
6. Leave the thermometer in the rectum until the device has signalled that the
maximum body temperature has been reached.
7. Remove the thermometer from the individual’s rectum.
UNIT 2
FIGURE 6.13 Taking a consumer’s
axillary temperature
Axillary method
E
1. Place the thermometer into the middle of the axilla (Figure 6.13) and fold the
individual’s arm across the chest to keep the thermometer in place.
2. Leave the thermometer in the axilla until the device has signalled that the
maximum body temperature has been reached.
3. Remove the thermometer from the individual’s axilla.
Electronic thermometer
E
1. Remove the electronic thermometer from the charging unit (Figure 6.14).
2. Attach a disposable cover to the probe.
3. Using one of the methods described (oral, rectal or axillary), measure the temperature.
4. Listen for the sound or look for the symbol that indicates maximum body
temperature has been reached.
5. Observe and record the reading.
6. Remove and discard the probe cover.
7. Return the electronic thermometer to the charging unit.
Plastic holder containing digital thermometer probes and rechargeable battery
Rectal probe
Box of disposable probe covers
Power source
FIGURE 6.15 A tympanic thermometer
Disposable probe cover
FIGURE 6.14 A battery-operated thermometer
Oral probe
Plastic cord goes around nurse’s neck
Tympanic thermometer
E
1. Attach the probe cover to the nose of the thermometer (Figure 6.15).
2. Gently place the probe of the thermometer over the entrance to the ear
canal. If the individual is under 3 years old, pull the pinna down, aiming the probe towards the opposite eye. If the individual is over 3 years old, grasp the pinna and pull gently up and back, aiming the probe towards the opposite ear (Figure 6.16). Make sure there is a tight seal.
3. Press the start button on the thermometer handle.
4. Wait for the beep, remove the probe from the ear, and read the temperature.
5. Discard the probe cover.
6. Return the tympanic thermometer to the charger unit.
Normal body temperatures are described in Table 6.4.
N
A
Hyperthermia, pyrexia or fever are conditions in which body temperatures
exceed 38.5°C. Clinical signs of hyperthermia include increased respiratory rate and pulse, shivering, pallor and thirst.
E
ExaminationNNormal ndingsAAbnormal ndingsPPathophysiology
EXAMINATION REQUIREMENTS FOR EVERY CONSUMER 161
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
There can be many causes of hyperthermia (including infection), which results
P
from an increased basal metabolic rate.
A
Hypothermia occurs when the body temperature is below 34°C. Clinical
signs of hypothermia include decreased body temperature and initial shivering that ceases as drowsiness and coma ensue. output, lack of muscle coordination, and disorientation also occur as hypothermia progresses.
Hypothermia can be caused by prolonged exposure to cold, such as immersion
P
in cold water or administration of large volumes of unwarmed blood products. Hypothermia can be induced to decrease the tissues’ need for oxygen, such as
P
during cardiac surgery.
Hypotension, decreased urinary
URGENT FINDING
Hypothermia – importance of assessment
Hypothermia can cause bradycardia and cardiac arrest. The focus of care for those with hypothermia is on comprehensive and multiple methods of rewarming, with close cardiac monitoring. As the person’s temperature increases, arrythmias can develop. It is important to identify the consumer’s full history and assessment to ensure correct treatment. Resus­citation efforts do not stop until the person is at a normal body temperature and all other interventions have been exhausted.
FIGURE 6.16 Taking a consumer’s
temperature with a tympanic thermometer
CHAPTER 6
Blood pressure
Blood pressure measures (in millimetres of mercury [mmHg]) the force exerted
by the ow of blood pumped into the large arteries. Arterial blood pressure is determined by blood ow and the resistance to blood ow as indicated in the following formula:
MAP = CO × TPR mean arterial pressure (MAP) = cardiac output (CO) × total peripheral resistance (TPR)
Changes in blood pressure can be used to monitor changes in cardiac output. Ineffective pumping, decreased circulating volume, as well as changes in the characteristics of the blood vessels, can affect blood pressure. There is a diurnal variation in blood pressure, as characterised by a high point in the early evening and a low point during the early deep stage of sleep.
Korotkoff sounds
Korotkoff sounds are generated when the ow of blood through the artery is
altered by inating the blood pressure cuff that is wrapped around the extremity. Korotkoff sounds may be heard by listening over a pulse site that is distal to the blood pressure cuff. As the air is released from the bladder of the cuff, the pressure on the artery changes from that which completely occludes blood ow to that which allows free ow. As the pressure against the artery wall decreases, ve distinct sounds occur.
Phase I: The rst audible sound heard as the cuff pressure is released. Sounds like
>
clear tapping and correlates to systolic pressure (the force needed to pump the
blood out of the heart).
Phase II: Sounds like swishing or a murmur. Created as the blood ows through
>
blood vessels narrowed by the ination of the blood pressure cuff.
Phase III: Sounds like clear, intense tapping. Created as blood ows through the
>
artery, but cuff pressure is still great enough to occlude ow during diastole.
E
ExaminationNNormal ndingsAAbnormal ndingsPPathophysiology
162 PHYSICAL EXAMINATION
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
UNIT 2
> Phase IV: Sounds are mufed and are heard when cuff pressure is low enough to
allow some blood ow during diastole. The change from the tap of Phase III to the mufed sound of Phase IV is referred to as the rst diastolic reading.
Phase V: No sounds are heard. Occurs when cuff pressure is released enough to
>
allow normal blood ow. This is referred to as the second diastolic reading.
Measuring blood pressure
Systolic pressure represents the pressure exerted on the arterial wall during systole, when the ventricles are contracting. Diastolic pressure represents the pressure in the arteries when the ventricles are relaxed and lling. Blood pressure is recorded as a fraction, with the top number representing the systole and the bottom number(s) representing the rst diastolic sound is written over the second. For example, 120/90/80 indicates that 120mmHg is the systolic pressure, 90mmHg is the rst diastolic sound, and 80mmHg is the second diastolic sound. the diastolic and systolic blood pressures.
diastole. If the rst and second diastolic sounds are recorded, the
Pulse pressure is the difference between
Measurement sites
There are several potential sites for blood pressure measurement. The preferred site is the brachial pulse site, where the brachial artery runs across the antecubital fossa. The posterior thigh, where the popliteal artery runs behind the knee joint, can also be used. A site should not be used if there is pain or injury around or near the site; for instance, a postmastectomy consumer should have blood pressure assessed on the unaffected side. Surgical incisions; intravenous, central venous or arterial lines; or areas with poor perfusion should be avoided for blood pressure measurement. Consumers with arteriovenous (AV) stulas or AV shunts should not have blood pressure measured in those extremities.
FIGURE 6.17 Blood pressure cuffs come in
various sizes.
FIGURE 6.18 An aneroid manometer dial
Equipment
Blood pressure is measured indirectly with a Doppler or a stethoscope and a
sphygmomanometer, which consists of the blood pressure cuff, connecting tubes
and air pump, and manometer. Blood pressure cuffs come in several sizes ( The size of the cuff bladder should be 80% of the circumference of the limb being assessed (JNC, 2003). The cuff should completely encircle the limb.
A manometer is attached to the cuff via a second tube. The
manometer
the air pressure within the cuff (
is a calibrated dial with a needle that points to numbers representing
Figure 6.18).
aneroid
Figure 6.17).
HEALTH EDUCATION
Home blood pressure monitoring devices
Many consumers take their blood pressure at home for numerous reasons, for example to monitor antihypertensive medication treatment. Periodically, consumers should be instructed to bring their home devices to their healthcare provider appointment. The consumer should be observed taking their blood pressure to assess their technique. Reinforcing correct technique can positively inuence an individual’s compliance with their medical treatment.
It is also appropriate to compare the blood pressure reading on the home device with that obtained in the healthcare provider’s ofce. The measurements should correlate within 5mmHg of each other. This assists to determine if the readings from home are accurate and reect the consumer’s actual blood pressure.
EXAMINATION REQUIREMENTS FOR EVERY CONSUMER 163
A. When taking a blood pressure measurement, the nurse can manually elevate the arm to the level of the patient’s heart.
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
Blood pressure assessment
A Doppler ultrasonic stethoscope can also be used to obtain blood pressure. It is especially useful when blood pressure sounds are difcult to hear, such as with infants or very obese consumers.
E
1. Ensure that the consumer has not had any caffeine or tobacco products in
the past 30 minutes. Allow the consumer to rest for 5 minutes before you take the blood pressure.
2. Select an appropriate size cuff.
3. Position the consumer. The consumer may be sitting, standing or supine.
At the rst encounter, all three positions are recommended.
4. Position the arm or leg to be used so that the extremity is at a level equal to
the heart to prevent a false reading ( the sitting position is usually the position of choice. Ensure that the consumer’s feet are rmly on the oor in this position.
5. Apply the deated blood pressure (BP) cuff: a. Upper arm: Wrap the BP cuff snugly around the bare upper arm. The
bottom of the BP cuff should be 3 to 5cm above the antecubital fossa. The centre of the bladder should be directly above the brachial artery.
b. Leg: Wrap the BP cuff around the bare thigh, with the bottom of the BP
cuff slightly above the knee. The popliteal artery below the cuff is used for BP measurement. This is usually easier if the consumer is in the prone position.
6. Establish a baseline systolic blood pressure (palpating the blood pressure),
if needed:
a. Palpate the brachial or radial artery with the finger pads of your
nondominant hand distal to the BP cuff.
b. Inflate the BP cuff, and note when the artery pulsation is no longer
palpable.
c. Release the air from the BP cuff, and wait 1 to 2 minutes.
7. Palpate the pulse distal to the BP cuff.
8. Place the bell of the stethoscope over the blood pressure site (the diaphragm
may be used if sounds are hard to hear):
a. If a Doppler ultrasonic stethoscope is to be used, apply conducting gel to
the site where the pulse was palpated.
b. Place the Doppler transducer over the site.
9. Inate the BP cuff to approximately 20mmHg above the established
baseline blood pressure or 20mmHg above where the Korotkoff sounds disappear.
10. Slowly open the valve and release the pressure at a rate of 2 to 3mmHg
per second.
11. Listen for the Korotkoff sounds: a. Onset of Korotkoff sounds correlates to systolic pressure. b. Muffling or disappearance of sounds correlates to diastolic pressures.
12. Deate the BP cuff completely.
13. Record the blood pressure reading(s). The extremity used and position of
the consumer are important data to record along with the blood pressure reading. Refer to Clinical reasoning practice tip: Accurate documentation of blood pressure.
Figure 6.19). At subsequent encounters,
A. When taking a blood pressure measurement, the nurse can manually elevate the arm to the level of the consumer’s heart.
B. The nurse can also rest a consumer’s elevated arm at the level of the heart using equipment. Note that this seated consumer has her feet resting on the oor.
FIGURE 6.19 Measuring a consumer’s
blood pressure
CHAPTER 6
E
ExaminationNNormal ndingsAAbnormal ndingsPPathophysiology
164 PHYSICAL EXAMINATION
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
UNIT 2
TABLE 6.5 Normal blood pressure range according to age and sex*
AGE (FEMALE) SYSTOLIC (MMHG) DIASTOLIC (MMHG)
1 97–103 52–56 5 103–109 66–70 10 112–118 73–76 15 120–127 78–81
18 < 120 < 80
AGE (MALE) SYSTOLIC (MMHG) DIASTOLIC (MMHG)
1 94–103 49–54 5 104–112 65–70 10 111–119 73–78 15 122–131 76–81
18 < 120 < 80
*The measurements listed for paediatric consumers are included; however blood pressure is not taken routinely for children unless they are in
ICU or in a critical condition.
TABLE 6.6 Errors in blood pressure measurement
IF READING SHOWS SUSPECT
inaccurately high blood pressure
high diastolic blood pressure
inaccurately low blood pressure
low systolic blood pressure
Blood pressure cuff is too short or too narrow (e.g. using a regular blood pressure cuff on an obese arm), or the brachial artery may be positioned below the heart. The consumer may also be stressed, be in an emotional state, or have just completed physical activity.
Unrecognised auscultatory gap (a silent interval between systolic and diastolic pressures that may occur in hypertensive consumers or because you deated the blood pressure cuff too rapidly); immediate reination of the blood pressure cuff for multiple blood pressure readings (resultant venous congestion makes the Korotkoff sounds less audible). contraction can raise the diastolic blood pressure by 10%.
Blood pressure cuff is too long or too wide; the brachial artery is above the heart.
Unrecognised auscultatory gap (a rapid deation of the cuff or immediate reination of the cuff for multiple readings can result in venous congestion, thus making the Korotkoff sounds less audible and the pressure appear lower).
If the consumer supports his or her own arm, then sustained muscular
CLINICAL REASONING
Practice tip: Accurate documentation of blood pressure
It is important to record the position of the consumer during the blood pressure measurement, as there can be differences (e.g. sitting and standing can have a marked difference), and identify postural hypotension. Document consumer position as follows:
> supine > sitting or standing
Also, record where the blood pressure was taken. Use the following abbreviations:
> RA = (right arm) LA = (left arm) > RL = (right leg) LL = (left leg)
Examples of blood pressure readings are:
> 160/122mmHg LL (supine) > 98/52mmHg RA (sitting) > 118/85mmHg LA (standing)
EXAMINATION REQUIREMENTS FOR EVERY CONSUMER 165
Copyright 2024 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. WCN 02-300
https://t.me/medicina_free
CLINICAL REASONING
Practice tip: Palpating the blood pressure
If you are unable to hear a consumer’s blood pressure and there is no electronic monitor or amplication device available, you can use the palpation method. Conduct the blood pressure measurement as described in steps 1–6 using the brachial artery. When releasing air from the cuff, note when the brachial artery is palpable. This correlates with the systolic pressure. This blood pressure is then documented as a number over palpated (e.g. 115/ palpated). It should also be noted that an audible BP reading was unable to be obtained.
TABLE 6.7 Classication of blood pressure in adults
DIAGNOSTIC CATEGORY
Optimal Normal 120–129 and/or 80–84 High-normal 130–139 and/or 85–89 Grade 1 (mild) hypertension 140–159 and/or 90–99 Grade 2 (moderate) hypertension 160–17 and/or 100–109 Grade 3 (severe) hypertension Isolated systolic hypertension
a
When a consumer’s systolic and diastolic BP levels fall into different categories, the higher diagnostic category and recommended
action/s apply.
SOURC E: NATION AL HEAR T FOUNDATI ON OF AUST RALIA . GUIDELI NE FOR THE DI AGNOSIS A ND MANAG EMENT OF H YPERT ENSION IN A DULTS – 2016. ME LBOURN E: NATION AL HEA RT FOUNDAT ION OF AUST RALI A, 2016.
Normal blood pressure varies with age. As a person ages, blood pressure
N
generally increases. at different ages. Normally, baroreceptors help a consumer to maintain normal blood pressure when changing from a supine to a sitting or a standing position. (Baroreceptors are the receptors located in the walls of most of the great arteries that sense hypotension and initiate reex vasoconstriction and tachycardia to bring the blood pressure back to normal.) Processes that increase cardiac output, such as exercise, will normally increase blood pressure. Pulse pressure is normally 30 to 40mmHg. in blood pressure measurement.
Hypertension, or high blood pressure, is usually conrmed when an adult
P
consumer has blood pressure readings remaining consistently above 120mmHg systolic and 80mmHg diastolic on two consecutive visits after an initial screening. See Table 6.7 for the classication of hypertension.
The cause of hypertension in 90% of consumers who have it is unknown. It is
P
thought that the mechanisms that maintain the therapeutic uid volume in the body (e.g. the heart, kidneys, nervous system, renin–angiotensin–aldosterone system) may be abnormal. The other 10% of the population with high blood pressure have secondary hypertension. All of the following pathophysiologies of hypertension are secondary in nature.
Arteriosclerosis reduces arterial compliance. Elastic and muscular tissues of
P
arteries are replaced with brous tissue as part of the normal ageing process, making the vessels less able to contract and relax in response to systolic and diastolic pressures. When the systolic pressure alone is elevated in the elderly, it is called isolated systolic hypertension.
Processes decreasing the size of the arterial lumen cause hypertension.
P
Hypercholesterolaemia results in deposits of plaque along the inner walls of the vessels, reducing the size of the arterial lumen and increasing blood pressure.
a
SYSTOLIC (MMHG) DIASTOLIC (MMHG)
<120
180 >140
and
and/or
and
<80
110
<90
Table 6.5 presents general ranges for normal blood pressure
Table 6.6 lists errors
CHAPTER 6
E
ExaminationNNormal ndingsAAbnormal ndingsPPathophysiology