Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3618_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
30.08.2026
Размер:
71 Мб
Скачать
242 Quantitative postradioembolization imaging using PET/CT
homemade 2.5-mm-thick copper ring to reduce bremsstrahlung radiation emerging from a patient administered with 1.3 GBq of 90Y. Of note, typi­cal administered activities may be in the order of several GBq, therefore, possible detector saturation may not be ignored a priori.
Other degrading factors that signicantly con­tribute randomly are (1) the detection of random coincidences from two bremsstrahlung photons emitted simultaneously; (2) detection of coin­cidences from an annihilation 511 keV photon and a bremsstrahlung photon emitted simultane­ously; and (3) the presence of the naturally occur­ring isotope
176
Lu within the detection crystals (i.e., cerium-doped LYSO or cerium-doped LSO) of PET imaging systems that may generate back­ground count rates. ere is ample evidence in the literature that bremsstrahlung radiation together with the LSO background radiation greatly increases the random fraction in quantitative 90Y PET imaging.
Lutetium (Lu)-based scintillators such as LSO and LYSO are widely used in current generation PET detectors (especially in ToF scanners) due to their relatively high stopping power for 511 keV gamma rays, high light yield, and short decay time. However, 2.6% of naturally occurring Lu is the isotope
176
Lu (T
3.6 × 1010 year), a long-lived
1/2
radioactive element undergoing beta decay (maxi­mum energy 596 keV) and three major simultane­ous gamma decays at energies 88 keV (15%), 202 keV (78%), and 307 keV (94%) (Browne and Junde,
1998). While the presence of
176
Lu is generally not an issue with traditional PET radionuclides due to the high true coincidence count rate (e.g., 18F), this phenomenon is likely to introduce nonnegligible random events during 90Y PET acquisition, thereby aecting system performance.
In a PET detector, the β particles emitted from Lu-based crystals, given the short range, deliver most of their energy in the same crystal. On the other hand, γ-rays can be detected not only in the same crystal where they are generated but also in other detector elements. erefore, the background radiation generated by the radionuclide
176
Lu can contribute to the amount of random and true coin­cidences. e most likely event is the detection of coincidences originating from the detection of the β− in the crystal in which the
176
Lu decay occurred and one of the prompt γ-rays in another detector crystal (Goertzen et al., 2009). As a general rule, in
order to assess the impact of natural
176
Lu radioac­tivity on the image quality, long acquisition with no radioactivity present in the eld of view is per­formed to determine the
176
Lu background count
rate.
One last confounding factor in 90Y PET quanti­tative imaging is attributable to scatter correction. At very low counts, PET images are very noisy and the resulting scatter correction might lead to heavy under- or overestimation of the scatter contribution.
To conclude, bremsstrahlung photons and prompt gammas are likely to result in a very high random fraction in imaging 90Y on the order of 80% (Willowson et al., 2015) or even higher (Carlier et al., 2015) compared with a typical FDG scan of 30%–40%. e combination of high random frac­tion, extremely low true coincidences, and prob­lematic scatter modeling for low counting statistics results in very noisy true coincidence sinograms. In addition, a well-known problem in PET imag­ing is the introduction of a positive bias aer cor­rection for random events (Ahn and Fessler, 2004; Rahmim et al., 2005; Li and Leahy, 2006). e most common method of correcting for random coinci­dences is the real-time or oine subtraction of a delayed coincidence time window from the prompt signal. In a scenario with low true coincidences and high random fraction (as in 90Y PET imaging), negative sinogram ray-sum values can be pro­duced. ese negative sinogram values are oen truncated in commercial soware packages before iterative reconstruction (i.e., become zeroed) thus introducing a positive bias. is bias does not have a signicant impact for clinical imaging with 18F, but may become important in 90Y PET imaging. is bias was observed by several authors (Tapp et al., 2014; Carlier et al., 2015) and it is possibly responsible for hot contrast recovery obtained with
90
Y being inferior to that obtained with 18F.
It is worth mentioning that presently a number of literature studies provided ample evidence that neither detector saturation (D’Arienzo et al., 2012; Bagni et al., 2012; Carlier et al., 2013) nor intrinsic natural
176
Lu (Carlier et al., 2013) radioactivity rep­resents a major issue in 90Y PET quantication at activity concentrations commonly encountered in liver radioembolization. In particular, the presence of natural
176
Lu radioactivity produces a measurable but not limiting contribution. Carlier et al. (2013) suggested that emissions from the radionuclide
11.5 Performance characteristics of PET scanners / 11.5.6 Image reconstruction 243
176
Lu may signicantly contribute to random coin­cidences when 90Y radioactivity concentration is below 1 MBq mL−1 in the presence of high tumor to background activity concentration ratios.
An extensive study on the limitations and the accuracy achievable under conditions of low counts and high random fraction can be found in Carlier et al. (2015).
11.5.6 IMAGE RECONSTRUCTION
Iterative reconstruction has become the standard for routine clinical PET imaging. However, itera­tive algorithms are resource intensive, especially for ToF data, and OSEM algorithms are, therefore, commonly used to accelerate reconstruction. As a general rule, the image noise in the reconstructed images increases as the number of iterations pro­ceeds (Figure 11.6). On the other hand, image quality is also degraded when OSEM is used with a large number of subsets. As a consequence, there is a tradeo between the number of iterations/sub­sets and reconstructed image quality.
e impact of image reconstruction on 90Y PET quantication has been widely investigated by a number of literature studies with varying reported success. e general consensus is that the best reconstruction technique will depend on the scanner and the acquisition modality (ToF or non-ToF). Both Willowson et al. (2012) and Carlier et al. (2013) found that one iteration pro­vided the most accurate quantication on a ToF Siemens BioGraph mCT. On the other hand Bagni et al. (2012) and D’Arienzo et al. (2012) performed acquisitions on a BGO GE discovery ST scanner using two and three iterations, respectively. In another study van Elmbt et al. (2011) used three iterations (eight substeps + Gaussian lter) both on a Philips Gemini Power 16 and a Siemens Ecat Exact HR, while Lhommel et al. (2009) and Werner et al. (2010) used two iterations (33 sub­steps, ToF, RR) and eight iterations (16 substeps) on a Philips Gemini TF and a Siemens BioGraph Hi-Rez 16, respectively. A summary of the image reconstruction parameters obtained in published literature is given in Table 11.2.
1 iteration, 10 substeps2 iterations, 10 substeps 3 iterations, 10 substeps
4 iteration, 10 substeps5 iterations, 10 substeps 6 iterations, 10 substeps
Figure 11.6 Effect of number of iterations on the NEMA phantom with six llable spherical inserts. Image noise in the reconstructed images increases as the number of iterations proceeds. Images were obtained using a BGO GE Discovery ST PET scanner.
244 Quantitative postradioembolization imaging using PET/CT
e major outcome of the QUEST phantom study is that 90Y imaging capability appears to be optimal for Siemens systems using two iterations and 21 subsets with ToF and resolution recovery (RR) and an all-pass lter. For GE Healthcare systems, the use of an all-pass lter in conjunc­tion with RR and ToF (2 iterations and 24 sub­steps) provided the best quantication results. Regarding non-ToF generation scanners, mea­sures of background provided average devia­tions to within 1%, 5%, and 2% for GE Healthcare (all-pass lter, RR + ToF), Philips (4i8s ToF), and Siemens (2i21s all-pass lter, RR + ToF) ToF sys­tems, respectively.
11.6 PREPARATION OF A CALIBRATED PHANTOM
90
FOR
As previously mentioned, most qua ntitative imag­ing studies make use of phantoms and require very accurate knowledge of both the activity and the volume of liquid solution inserted into the phantoms. e general purpose of quantitative imaging studies with calibrated phantoms is (1) to assess uncertainties in the imaging process; (2) to quantify accuracy of correction factors and reconstruction algorithms; and (3) to evaluate scanner performance over time. A recent study (Sunderland et al., 2015) demonstrated that tech­nical error in phantom lling is one of the pri­mary reasons for exclusion of PET/CT scanners from clinical trials. Section 11.4 was dedicated to the importance of accurate 90Y activity measure­ments. Here, we describe further issues that need to be addressed for an optimal phantom prepara­tion dedicated to quantitative analysis with 90Y PET.
Adsorption of radionuclides on the inner walls of plastic phantoms may lead to inhomogeneous radionuclide distribution that can negatively aect quantitative imaging studies (Park et al.,
2008). erefore, the preparation of a carrier solution is recommended. e use of tap water should be avoided as minerals and other chemical impurities might stick to the phantom walls or combine with the radiopharmaceutical changing the radionuclide distribution. Carrier solutions
Y-PET STUDIES
should always be prepared using laboratory grade chemicals and puried water. erefore, it is important to ensure that a favorable chemistry is used throughout the calibration procedure in order to have a uniform and stable solution. For
90
Y PET studies, 90YCl3 in an aqueous solution of
0.1 mol dm-3 hydrochloric acid also containing inactive yttrium at a concentration of about 50 µg g-1 can be used as a carrier solution. Alternatively, diethylenetriaminepentaacetic acid (DTPA) or ethylenediaminetetraacetic acid (EDTA) at a con­centration of about 50 µgg-1 can be used to pre­vent radioactive 90Y from sticking to the phantom walls and to guarantee a homogeneous radionu­clide solution.
It is recommended that all containers be pre­lled with carrier 12 hours prior to addition of radioactive 90YCl3. is will help to “seal” the sur­face and reduce sticking or plating of activity. All containers should be emptied, dried, and the used carrier discarded before activity is added.
Activity concentration should be determined using a radionuclide dose calibrator traceable to a national standards laboratory for the geom­etry being measured, as previously discussed. Uncertainty for a typical well-calibrated eld instrument can be expected to be in the region of ~5% for low-energy gamma emitters (<100 keV) and pure beta emitters, such as 90Y. It is worth noting that if activity is determined by a national laboratory then this uncertainty can be reduced signicantly.
As a general rule, preparation of a stock solu­tion is recommended. Radioactive 90Y provided by a supplier should be diluted using the carrier solution to desired volume and concentration. Activity concentration should be determined by measuring an aliquot of the stock solution in terms of activity per unit mass (or volume). is can then be used to determine the activity of all subsequent sources produced from this stock solution.
Filling of the phantoms should be performed using a calibrated (preferably four decimal places) analytic scientic scale and with routine double or triple weighing of the sources. e overall uncer­tainty in the activity concentration determined using this method is dependent on the precision of the scale being used as well as the accuracy of the method used to determine the activity
11.8 Conclusions 245
5
±×
concentration of the solution. Radioactivity should be dispensed using calibrated pipette devices or syringes and ensuring that no air bubbles remain in the phantom. Either way, it is suggested to ush the needle to remove all activity from the syringe or the pipette. Again the uncertainty is dependent on the precision of the volume measurement as well as the activity determined.
If large background volumes are used for cali­bration purposes, the phantom can be lled with nonradioactive water to measure the llable vol­ume (and to conrm the phantom is watertight with no leaks). When lling large phantom vol­umes, a funnel should be used. When the phan­tom is nearly full, the funnel can be removed and a syringe used to complete the lling process thereby preventing the spillage of radioactive water from the background compartment.
11.7 DISCUSSION AND SUMMARY
PET imaging of 90Y microspheres is a rapidly evolv­ing eld whose features and attributes have not yet been fully addressed. At present, the general con­sensus is that accurate quantication is possible on a variety of PET scanner models, with or without ToF, although ToF is likely to improve the accu­racy at lower activity concentrations. Quantitative accuracy has been investigated by a number of authors using dierent phantoms with varying reported success. Regardless of the scanner used, partial volume eects play a major role in activity quantication showing a steady decline for spheres with diameter below 37 mm. It is expected that quantication in small hot regions may be under­estimated with all current generation scanners to a consistent degree of 15%–20% for a 37-mm-diam­eter object (Willowson et al., 2015). Higher quanti­cation accuracy can be achieved for large regions uniformly lled with 90Y. Non-ToF generation GE Healthcare and Siemens scanners are capable of recovering activity concentrations on the order of 300 kBq/mL within 2% and 9% of true values, while the deviation for ToF GE Healthcare and Siemens scanners can be expected to be in the range 1%–5% (Willowson et al., 2015). Furthermore, the recovery
of activity concentration in hot spheres is inferior to that obtained with 18F, probably due to eects of image noise on the OSEM reconstruction algo­rithms, which present an inherent nonnegativity constraint. While the intrinsic radioactivity pres­ent in lutetium-based crystals is a potential limita­tion, it has been shown to have a negligible impact at the high activity concentrations typically found in 90Y radioembolization.
Regarding the acquisition time, a 40-minute acquisition is recommended in a clinical scenario, acquired as two bed positions, 20 minutes each.
Figure 11.7 shows dierent acquisitions performed
with a GE Discovery ST BGO scanner 2 hours aer the microsphere administration (single bed acqui­sition, 20 minutes). An excellent match between microsphere accumulation and tumor areas was observed.
Last, the accuracy of quantitative 90Y PET/CT imaging is strongly related to the measurement precision of the internal pair production branch­ing ratio. At present, the β+ emission probabil­ity of 90Y is known with an uncertainty around
1.5% (
; Selwyn et al., 2007). A reduction of such uncertainty will lead to a reduc­tion of the overall uncertainty in the activity quan­tication. is is especially important in light of the current uncertainty in the clinical measure­ment of 90Y activity using a dose calibrator. In the near future, new experimental measurements of the internal pair production branching ratio are desirable in order to achieve more accurate quan­tication in postradioembolization imaging using PET/CT.
11.8 CONCLUSIONS
Currently, a number of clinics have developed their own quantitative imaging procedures using 90Y PET/ CT. Very oen there is a lack of harmonization of these procedures, most likely because quantitative imaging with 90Y PET/CT is a relatively new imaging strategy and further because imaging capabilities are strongly related to PET scanner performance. Ultimately, it is worth stressing that scanner perfor­mance is optimized for 18F imaging and dedicated imaging protocols are required for accurate 90Y PET/CT studies. Careful choice of reconstruction
246 Quantitative postradioembolization imaging using PET/CT
(a)
(b)
(c)
Figure 11.7 Clinical examples of postradioembolization imaging using 90Y-PET showing an excel- lent match between microsphere accumulation on hepatic lesions. (Left) 90Y-PET fused with CT data. (Central) CT of the liver parenchyma clearly showing tumor regions. (Right) 90Y-PET. Figure parts (a) through (c) show different axial levels of the same patient.
parameters has the potential to increase the quan­titative accuracy. However, the nal outcome will depend on the scanner and reconstruction soware.
Comprehensive guidance has yet to be pre­sented in this eld and there is no doubt that an internationally endorsed protocol on the 90Y PET/
References 247
CT quantitative imaging would lead to further advances in this area.
REFERENCES
AAPM Task Group 181. (2012). The selection, use,
calibration, and quality assurance of radio­nuclide calibrators used in nuclear medicine report of AAPM Task Group 181. College Park, MD: American Association of Physicists in Medicine.
Ahn, S., Fessler, J.A. (2004). Emission image
reconstruction for randoms-precorrected PET allowing negative sinogram values. IEEE Trans Med Imaging 23:591– 601.
Attarwala, A. et al. (2014). Quantitative and
qualitative assessment of yttrium-90 PET/CT imaging. PLoS One 9:e110401.
Bagni, O. et al. (2012). 90Y-PET for the assess-
ment of microsphere biodistribution after selective internal radiotherapy. Nucl Med Commun 33:198– 204.
Bé, M. et al. (2006). Table of radionuclides.
Monographie BIPM-5. Sèvres, Paris: Bureau International des Poids et Mesures, ISBN 92-822-2218-7.
Browne, E., Junde H. (1998). Nuclear data
sheets for A = 176. Nucl Data Sheets 1998;84:337–486.
Carlier, T. et al. (2013). Assessment of acquisition
protocols for routine imaging of Y-90 using PE T/CT. EJNMMI Res 3:11.
Carlier, T., Willowson, K., Fourkal, E., Bailey, D.,
Doss, M., Conti, M. (2015). Exploring limitations and accuracy under conditions of low counts and high random fraction. Med Phys 42:4295–4309.
Cherry, S., Sorenson, J., Phelps, M. (2012).
Physics in Nuclear Medicine. Philadelphia, PA: Elsevier/Saunders.
Conti, M. (2009). State of the art and challenges
of time-of-ight PET. Phys Med 25:1 –11.
Conti, M., Eriksson, L., Rothfuss, H., Melcher,
C. (2009). Comparison of fast scintillators with TOF PET potential. IEEE Trans Nucl Sci 56:926–933.
D’Arienzo, M. (2013). Emission of β
internal pair production in the 0
90
tion of
Zr: Historical background and current
applications. Nucl Med Imaging Atoms 1:2–12.
90
Y-PET imaging:
+
particles via
+–0+
transi-
D’Arienzo, M. et al. (2012).
90
Y PET-based dosim­etry after selective internal radiotherapy treatments. Nucl Med Commun 33:633–640.
D’Arienzo, M. et al. (2013). Absorbed dose to
lesion and clinical outcome after liver radio­embolization with 90Y microspheres: A case report of PET-based dosimetry. Ann Nucl Med 27:676–680.
Daube-Witherspoon, M.E. et al. (2002). PET
performance measures using the NEMA NU 2-2001 standard. J Nucl Med 43:1398–1409.
Dezarn, W. et al. (2011). Recommendations of
the American Association of Physicists in Medicine on dosimetry, imaging, and quality assurance procedures for
90
Y microsphere brachytherapy in the treatment of hepatic malignancies. Med Phys 38:4824–4845.
Dezarn, W., Kennedy, A. (2007a). SU-FF-T-380:
Signicant differences exist across institu­tions in 90Y activities compared to reference standard. Med Phys 34:2489.
Dezarn, W., Kennedy, A. (2007b). Resin
90
Y microsphere activity measurements for liver brachytherapy. Med Phys 34:1896 –1900.
Elschot, M. et al. (2013). Quantitative compari-
son of PET and bremsstrahlung SPECT for imaging the in vivo yttrium-90 microsphere distribution after liver radioembolization. PLoS One 8:e55742.
Fenwick, A., Baker, M., Ferreira, K., Keightley, J.
(2014). Comparison of Y-90 measurements in UK hospitals. NPL Report IR 20. United Kingdom: National Physics Laboratory.
Ferreira, K., Fenwick, A., Arinc, A., Johansson,
L. (2016). Standardisation of 90Y and deter­mination of calibration factors for 90Y microspheres (resin) for the NPL secondary ionisation chamber and a Capintec CRC-25R. Appl Radiat Isotopes 109:226–230.
Ford, K. (1955). Predicted 0
+
level of Zr90. Phys
Rev 98:1516–1517.
Fourkal, E. et al. (2013). 3D inpatient dose
reconstruction from the PET-CT imaging
90
of
Y microspheres for metastatic cancer to the liver: Feasibility study. Med Phys 40:081702.
Fowler, K. et al. (2016). PET/MRI of hepatic 90Y
microsphere deposition determines indi­vidual tumor response. Cardiovasc Intervent Radiol 39:855–864.
248 Quantitative postradioembolization imaging using PET/CT
Gadd, R. et al. (2006). Protocol for establishing
and maintaining the calibration of medical radionuclide calibrators and their quality control. Measurement Good Practice Guide No. 93. Teddington, Middlesex, UK: National Physical Laboratory.
Gates, V. et al. (2010). Internal pair production
90
of
Y permits hepatic localization of micro­spheres using routine PET: Proof of concept. J Nucl Med 52:72–76.
Goertzen, A., Stout, D., Thompson, C. (2009). A
method for measuring the energy spectrum of coincidence events in positron emission tomography. Phys Med Biol 55:535–549.
Greenberg, J.S., Deutsch, M. (1956). Positrons
from the decay of P
32
and Y90. Phys Rev
102:415 – 421.
IAEA. (1996). International basic safety standards
for protection against ionizing radiation and for the safety of radiation sources. International Atomic Energy Agency, Vienna, Austria.
Johnson, O., Johnson, R., Langer, L. (1955).
Evidence for a 0
+
rst excited state in Zr90.
Phys Rev 98:1517–1518.
Kao, Y. et al. (2013). Post-radioembolization
yttrium-90 PET/CT—Part 2: Dose-response and tumor predictive dosimetry for resin microspheres. EJNMMI Res 3 :57.
Kao, Y. et al. (2012). Yttrium-90 internal pair
production imaging using rst generation PET/CT provides high-resolution images for qualitative diagnostic purposes. Br J Radiol 85:1018–1019.
Kossert, K. et al. (2016). Comparison of 90Y
activity measurements in nuclear medicine in Germany. Appl Radiat Isot 109:247–249.
Langhoff, H., Hennies, H. (1961). Zum experimen-
tellen Nachweis von Zweiquantenzerfall beim
+–0+
0
Ubergang des Zr90. Z Phys 164:166–173.
Lewellen, T. (1998). Time-of-ight PET. Semin
Nucl Med 28:268–275.
Lhommel, R., Goffette, P., Van den Eynde, M.,
Jamar, F., Pauwels, S., Bilbao, J., Walrand, S. (2009). Yttrium-90 TOF PET scan dem­onstrates high-resolution biodistribution after liver SIRT. Eur J Nucl Med Mol Imaging 36:1696–1696.
Lhommel, R. et al. (2010). Feasibility of
90
YTOF PET-based dosimetry in liver metastasis therapy using SIR-spheres. Eur J Nucl Med Mol Imaging 37:1654–1662.
Li, Q., Leahy, R.M. (2006). Statistical model-
ing and reconstruction of randoms precor­rected PET data. IEEE Trans Med Imaging 25:1565 –1572.
Lourenço, V. et al. (2015). Primary standardization
of SIR-Spheres based on the dissolution of the 90Y-labeled resin microspheres. Appl Radiat Isot 97:170–176.
Martí-Climent, J. et al. (2014). PET optimiza-
tion for improved assessment and accurate quantication of
90
Y-microsphere biodistri­bution after radioembolization. Med Phys 41:092503.
Mo, L. et al. (2005). Development of activity
standard for 90Y microspheres. Appl Radiat Isot 63:193–199.
National Electrical Manufacturers Association.
(2007). NEMA NU 2-2007 Performance
Measurements of Positron Emission Tomographs Arlington, VA: NEMA.
Ng, S.C. et al. (2013). Patient dosimetry for
90
Y selective internal radiation treatment based on 90Y PET imaging. J Appl Clin Med Phys 14:212–221.
Nickles, R.J. et al. (2004). Assaying and PET
imaging of yttrium-90: 1>>34 ppm>0. IEEE Nucl Sci Symp Rec 6:3412–3414.
Okuda, K. et al. (1985). Natural history of hepa-
tocellular carcinoma and prognosis in relation to treatment study of 850 patients. Cancer 56:918–928.
Park, M., Mahmood, A., Zimmerman, R.,
Limpa-Amara, N., Makrigiorgos, G., Moore, S. (2008). Adsorption of metallic radionu­clides on plastic phantom walls. Med Phys 35:1606.
Pasciak, A., Bourgeois, A., Bradley, Y. (2014a).
A comparison of techniques for 90Y PET/CT image-based dosimetry following radioem­bolization with resin microspheres. Front Oncol 4:121.
Pasciak, A. et al. (2014b). Radioembolization and
the dynamic role of
90
Y PET/CT. Front Oncol
4:38.
Peller, P., Subramaniam, R., Guermazi, A. (2012).
PET-CT and PET-MRI in Oncology. Berlin: Springer.
Rahmim, A. et al. (2005). Statistical dynamic
image reconstruction in state-of-the-art high resolution PET. Phys Med Biol 50:4887– 4912.
References 249
Selwyn, R. et al. (2007). A new internal pair pro-
duction branching ratio of ment of a non-destructive assay for
90
Sr. Appl Radiat Isot 65:318–327.
90
Y: The develop-
90
Y and
Stabin, M.G., Eckerman, K.F., Ryman, J.C.,
Williams, L.E. (1994). Bremsstrahlung radia­tion dose in yttrium-90 therapy applications. J Nucl Med 35:1377–1380.
Sunderland, J., Christian, P., Kiss, T. (2015). PET/
CT scanner validation for clinical trials-reasons for failure, recipes for success: The Clinical Trials Network (CTN) experience. J Nucl Med 56(3 ):1737.
Surti, S. (2014). Update on time-of-ight PET
imaging. J Nucl Med 56:98–105.
Surti, S., Karp, J. (2016). Advances in time-of-
ight PET. Phys Med 32:12–22. doi: 10.1016/j. ejmp. 2 015 .12.0 0 7.
Tapp, K. et al. (2014). The impact of image recon-
struction bias on PET/CT 90Y dosimetry after radioembolization. J Nucl Med 55:1452–1458.
Thiam, C., Bobin, C., Lourenço, V., Chisté, V.,
Amiot, M., Mougeot, X., Lacour, D., Rigoulay, F., Ferreux, L. (2016). Investigation of the response variability of ionization chambers for the standard transfer of SIR-Spheres®. Appl Radiat Isotopes 109:231–235.
van Elmbt, L. et al. (2011). Comparison of
yttrium-90 quantitative imaging by TOF and non-TOF PET in a phantom of liver selec­tive internal radiotherapy. Phys Med Biol 56:6759–6777.
Walrand, S. et al. (2015). The impact of image
reconstruction bias on PET/CT
90
Y dosim­etry after radioembolization. J Nucl Med 56:494–495.
Wissmeyer, M. et al. (2011).
90
Y time-of-ight PET/MR on a hybrid scanner following liver radioembolisation (SIRT). Eur J Nucl Med Mol Imaging 38:1744 –1745.
Werner, M. et al. (2009). PET/CT for the assess-
ment and quantication of 90Y biodistribution after selective internal radiotherapy (SIRT) of liver metastases. Eur J Nucl Med Mol Imaging 37:407–408.
Werner, M. et al. (2010). PET/CT for the detection
and quantication of the β-emitting thera­peutic radionuclide yttrium-90 after liver SIRT [abstract]. J Nucl Med 51(Suppl 2):341.
Willowson, K. et al. (2012). Quantitative
90
Y image reconstruction in PET. Med Phys 39:7153.
Willowson, K., Tapner, M., Bailey, D. (2015). A
multicentre comparison of quantitative 90Y PET/CT for dosimetric purposes after radio­embolization with resin microspheres. Eur J Nucl Med Mol Imaging, 42:1202–1222.
Woods, M. et al. (1996). Calibration of the NPL
secondary standard radionuclide calibrator for 32P, 89Sr and
90
Y. Nucl Instrum Methods
Phys Res A 369:698–702.
Zimmerman, B., Ratel, G. (2005). Report of
the CIPM key comparison CCRI(II)-K2Y-90. Metrologia 42:06001.
12
Image-based three-dimensional dosimetry following radioembolization
ALEXANDER S. PASCIAK AND S. CHEENU KAPPADATH
12.1 Introduction 251
12.2 Fully 3D Monte Carlo transport for image-based hepatic dosimetry 252
12.3 Dose-point kernel convolution for image-based hepatic dosimetry 253
12.4 The voxel S-value MIRD approach for image-based hepatic dosimetry 254
12.1 INTRODUCTION
e general dosimetric principles of 90Y radio­embolization have been discussed in detail in several chapters in this book. For example,
Chapter 5 discusses a commonly used formula
that allows for the calculation of the absorbed dose to a volume of tissue given a uniform dis­tribution of 90Y activity. While this is useful for determining the average absorbed dose to the liver, lobe, or segment treated, it is of limited use in the estimation of biological eect. More accurate methods for determining absorbed dose have been suggested such as the partition model, discussed in Chapters 4 and 5. However, the par­tition model cannot account for inhomogeneity of microsphere distribution in tumor or normal liver. Determining the absorbed dose following
90
Y radioembolization in all areas of the tumor, uninvolved liver, and extrahepatic tissues is a critical future component of managing patient
12.5 LDM for image-based epatic dosimetry 255
12.5.1 Validation of the LDM 256
12.5.2 LDM and alternative radionuclides 258
12.6 Instances where DPK convolution and LDM may not be appropriate 259
12.7 Conclusion 260
References 260
follow-up. For example, if radiation dose and, therefore, toxicity to normal or extrahepatic tissues can be determined immediately follow­ing radioembolization, prompt administration of prophylaxis can be considered which may decrease the severity of side eects. Furthermore, undertreated areas of tumor could be identied and alternative or adjuvant therapies could be prescribed, increasing the potential ecacy of
90
Y radioembolization in some patients.
e ability to utilize posttreatment dosimetry in the aforementioned manner depends on the availability of several important pieces of data. First, one must understand the radiation biology and dose–response properties of the normal liver andthe tumor, which may vary substantially with tumor size, type, and other factors. e radiation biology of radioembolization at a macroscopic and a microscopic level has been discussed in Chapters
8 and 9, respectively. However, there are still many
unknowns yet to be addressed. Once the radiation biology is understood for a patient with a given
251