Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_3592_Библиотеки_им_академика_М_И_Перельмана

.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
89 Мб
Скачать
g
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
Figure 11.5. Full system workow diagram.
The IVUS images required processing to enhance the appearance of the vessel lumen and to remove noise and labels. As a rst step, a region of interest (ROI) was dened to reduce the size of the image to only the IVUS output. To remove the region dened by the catheter silhouette, appearing as a constant circle in the centre of the IVUS images, each image was transformed rst to polar coordinates. A mask of minimum intensity (usually zero) pixels [21], of a width dened by the known catheter diameter, was then applied to the upper part of each polar image to remove this catheter artefact. The resulting image was then transferred back to Cartesian coordinates. To remove the grid markings in the IVUS images, a 6 × 10 neighbourhood median lter was applied at each point (known in advance).
IVUS images tend to be distorted by both speckle noise and modality-specic artefacts. To reduce speckle noise in the images, a global speckle reducing bilateral lter (SRBF) [4] was used. This incorporates local noise statistics to provide a full automated high-performance image despeckling and originates from classical bilateral ltering that takes into account both space- and intensity-oriented similarity in a specied region around the pixel. The lter combines both domain ltering, using pixel weights that decay as the distance to the central pixel rise, and range ltering, where pixel weights decrease when a dissimilarity-in-intensity measure increases. SRBF has been shown to demonstrate better results in ultrasound images, maintaining edge details, over other lters [36]. The lter is shown in the equations below:
+
in
WstWstgst
(, ) (, )(, )
dr
+
in
WstWst
(, ) (, )
dr
(11.6)
fij
(, )
∑∑
=−+=−
snintn
=
11
∑∑
=−+=−
snintn
11
is every pixel value of the input image and the ltered pixel is
ij(, )
size of the window lter was w = 7 (an odd number for symmetry) and the span was
22
−+−
is jt() ()
Wst e(, )
Ws t e(, ) .
r
=
d
∥−∥
=
2
σ
2
d
gi j gs t(, ) (, )
2
σ
2
r
2
11-13
fij(, )
(11.7)
(11.8)
. The
1
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
n = 3(
=+wn2
index for range parameters.
). d is an index corresponding to domain parameters while r is an
and
2
are parameters (variances) that regulate the
σ
r
2
σ
d
weight behaviour in the domain and range elds [42]. The choice of these values depends on the statistical features of noise; high values lead to edges loss while low values create a weak lter that is unable to eliminate noise. In our implementation,
22
σσ==4
dr
. Finally, to increase the contrast of the IVUS images, contrast
stretching (histogram dynamic full range expansion) [32] was applied.
A number of IVUS segmentation methods have already been proposed and these can be divided into four main strategies: (1) edge detection techniques, (2) active contour methods, (3) statistical approaches and (4) multiscale analysis. Due to the absence of peripheral tissues and blood ow in our phantom and the structural homogeneity of the phantom vascular wall, a straightforward global thresholding strategy was implemented. Otsus method [28] was chosen to distinguish between vessel wall and noise components. The method assumes that the pixel set of every image can be divided into two discrete classes: objects and background. The classication is performed through an optimisation problem which aims to detect an optimal threshold in the image histogram.
Finally, dilation and erosion were applied to the binary images. This was followed by a connect-component labelling (region labelling) whereby a graph-based process assigned a specic label to a set of pixels within a clearly bounded region. A two pass process was applied. In the rst step, initial labels were assigned to the objects and label correspondences were recorded. In the second step, the algorithm substituted each previously specied label for the smallest of the corresponding class. For each region, a centre of mass was calculated and these values were used to estimate the main ring of image energy; this allowed for the safe application of vessel-specic geometrical constraints to further remove non-object (non-lumen) regions. First, regions of size below a certain threshold were removed. Geometric constraints on the size of the average human aorta were then used to remove regions that were too far away from the geometric centroid of all the regions.
It should be noted that the appearance of the IVUS images collected from the phantom currently does not resemble those from patients; however, in vivo data do exhibit the same kind of speckle noise that we encountered. For that reason, while the image processing steps presented here are not directly transferrable to in vivo sequences, many of the steps can be used.
11.3.2.2 Landmark detection
The detection of landmarks is a key step in the proposed IVUS-based navigation platform. In our platform, we cater for both anatomical and morphological landmarks. Morphological landmarks are structures which describe a specic shape, are present in all shapes of the same class, and are easily detected. Anatomical landmarks along the aorta may be as suggested by the ACCF/AHA (American College of Cardiology Foundation/American Heart Association) Task Force [17]; here the authors identied nine well-established anatomical landmarks to effectively describe the anatomy of a normal aorta. These include: the aortic sinuses of Valsalva, the sinotubular junction, the mid ascending aorta, the proximal aortic
11-14
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
arch, the mid aortic arch, the proximal descending thoracic arch, the mid descending aorta, the aorta at diaphragm and the abdominal aorta at the coeliac axis origin.
As the landmarks used in the proposed framework must be effective descriptors of the vascular structures, some of these anatomical landmarks are unsuitable. For example, the mid aortic arch only refers to a length of the descending aorta, effectively a tube. For this reason, we only consider branches, being present in each artery, as well as morphological variations in the aorta, dilations and stenoses, representing the two main pathologies of the aorta, aneurysms and atherosclerotic lesions.
To detect these landmarks, two ellipses were tted to the inner wall edge and outer boundary of the vessel in the processed IVUS frames. The ellipses were tted using the direct least squares tting approach by Fitzgibbon et al [13]. The method searches for an optimally tted curve to a known set of scattered 2D data by minimising the sum value of the calculated residuals of the given points to the curve. A Cartesian quadratic representation of an ellipse was used:
2
+++++=ax bxycy dxeyf 0.
1
11211
1
(11.9)
In addition to tting the ellipses, the main ellipse parameters were calculated: the centre of each ellipse, semi-axis lengths and ellipse orientation. Subsequently, these parameters form the basis of the description of vessel morphology.
The absolute lengths of the major and minor axes of the ellipses dened the area covered by the ellipses and indicated the lumen area. Their relative lengths (calculating the eccentricity, dened by the ratio of major to minor lengths) describe the shape of the lumen. Finally, the absolute difference between the corresponding axes of the inner and outer ellipses described wall thickness.
A combination of these descriptors can be used to indicate the presence of landmarks along the aorta. It is challenging to identify certain landmarks in a single IVUS image in isolation but examining these parameters along the vessel axial dimension, i.e. along a series of IVUS images in a pullback sequence, it is possible to interpret the parameter uctuations.
An aneurysm would be characterised by an increase and then a decrease in the inner or outer ellipse axes lengths. An atheromatic stenosis will be a decrease and then an increase. The presence of a branch would be indicated by a localised increase in the eccentricity of the inner ellipse.
11.3.2.3 3D IVUS–CT fusion
After detection of landmarks in the IVUS sequence, the landmarks must be matched to the equivalent landmarks detected in 3D. For this, we used a preoperatively obtained 3D surface mesh of the vessel. This was derived from a preoperative rotational-CT scan of the phantom obtained using a GE Innova 4100 interventional imaging suite. CT visible markers were attached to the phantom for registration purposes. ITK-SNAP [48] was used to segment the vessel from the images and a 3D surface mesh was produced.
11-15
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
To dene the mesh skeleton (vessel centrelines), a mesh thinning algorithm was applied to the 3D surface mesh. Binvox [25, 27], followed by Thinvox [26, 29], was used for this purpose. Binvox rst converts the 3D surface mesh to a voxel model while Thinvox is an iterative procedure that detects and removes boundary vessels until a single voxel curve is left. By taking the centroid of each of the voxels, the mesh skeleton was produced.
To nd branch-points along the skeleton, all voxels along the initial voxel curve were examined. Each voxel could have up to 26 neighbouring voxels; however, as we are starting with a voxel curve, no voxel should have all 26 neighbours. Any voxel with a neighbouring voxel will either (1) share a common side (four common vertices), (2) share a common edge (two common vertices), or (3) share a single common vertex. After examining all neighbours of each voxel, it was determined that voxels with at least three neighbouring voxels were branch-points, i.e. were voxel junctions, and voxels with only one neighbouring voxel were end-points.
Subsequently, after all end-points and branch-points were identied along the vessel centreline, points along the centreline must be grouped, corresponding to each of the junctions. K-means clustering was used, with k set to the ideal number of junctions in the shape. After the lengths of centre points have been clustered together, the central axis of the aorta was determined using a Dijkstra-based directed path search algorithm. This sorted points within each junction and identied erroneous branch- and end-points. The total nal output path was the centreline of the vessel, which was smoothed using a spline.
The nal step was registration to the IVUS pullback sequence. For this, the start and end of the IVUS image sequence were registered to the end-points. Branches identied in the pullback were associated with the branch-points along the centre­line. Images in between were linearly interpolated between these xed points. As we currently do not have information on the orientation of the tip of the catheter, the IVUS images were oriented perpendicularly to the central axis.
11.3.2.4 Results
The image processing steps transforming each initial IVUS image to the nal centralised binary frame are presented in gure 11.6.
Figure 11.7 depicts the main branch detection procedure. The three peaks of the inner ellipse eccentricity index plot along the vessel correspond to the three main branches of the aortic arch. The detection of other lesions is based on tracking other features of the tted ellipses.
The calculated skeleton of the 3D CT model is presented in gure 11.8 along with the discretised vascular volume as well as the result of the IVUS–CT fusion procedure.
In gure 11.9 we present the results of our system in terms of navigation precision. This diagram provides a comprehensive comparison of the catheter tip points tracked by the EM system (red: ground truth) and the tip points calculated by our method (green). The point cloud of the 3D model (yellow points) as well as its smoothed central axis (blue curve) are also depicted. The absolute differences
11-16
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
Figure 11.6. The processing stages transforming each raw image to the nally centralised frame.
Figure 11.7. Branch detection and localisation based on eccentricity index of the inner tted ellipse. Left:
topEI plot along the vascular structure; bottomthe three main branches of our aorta phantom. Right: the three branch-corresponding binary IVUS frames before the centralisation step.
between the tip point coordinates provided by the EM system and those estimated via our system are also given in the form of three supplementary boxplots.
The current implementation of the proposed programme is in MATLAB and can be seen in gure 11.10.
11.4 The future of intravascular imaging for navigation
There still remain a number of challenges to be addressed in the use of intravascular imaging for navigation and these remain the research priority areas for any potential
11-17
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
Figure 11.8. From left to right: Voxelised CT-derived vessel model, central axis with end- and branch-points and IVUS–CT data fusion.
Figure 11.9. Precision results and absolute error in space. Yellow points: 3D model volume; red points: catheter tip tracked with EM system (ground truth); blue points: smoothed calculated central axis; green points: catheter tip according to the proposed method.
framework to enter clinical usage. The rst is down to the limitations in vessel coverage using an imaging modality acquired at the tip of a catheter. The quality of the acquired image-sets may be degraded by unpredictable abrupt movements of the catheter tip (jumps) during the pullback scan due to localised bending, torsional stiffness and lumen irregularities. These events may create extended lling decits of the vascular wall surface produced by the frame-sequence 3D repositioning. Furthermore, the relative position of the catheter and the wall at a cross section
11-18
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
Figure 11.10. The graphical user interface for the proposed framework.
may cause an incomplete capture of the wall inner circumference in cases of large distances between the lumen and the frame centres; this is, in particular, possible in the aorta. This decit may also extend several centimetres along the vessel, requiring rescanning. Scans conducted by using an automatic pullback device are more prone to these kinds of errors, while the manual steering of the catheter allows for better results. This way, the operator adapts the catheter conguration to the geometrical shape of the vascular structure by applying appropriate torque in a real-time model. However, multiple decit regions may be present in the endovascular images due to imaging noise and artefacts and these may cause a loss of potentially valuable information that interferes with the different processing steps. As a solution we propose the implementation of a mechanism that detects the position and extent of the decit and suggests a corrective rescanning of the corresponding vascular segment. This mechanism could be based on detecting discontinuities of the wall representation in a longitudinal mode, or tracking gaps in the wall region in polar coordinates.
For the use of a static 3D scan, such as a preoperative or intra-operative 3D model, co-registration is required. For example, in our framework, the detection of useful anchoring points that would serve as a reference pattern for this task is necessary on both IVUS and CT datasets, and consequently they must be matched. However, the lack of information on the exact diameter of the CT-derived model at every discrete cross section hinders the use of any point corresponding to a stenosis or an aneurysm limiting our landmark set to that containing only the branch- and end-points of the 3D model axis. One way to expand the usable landmark set could be the application of appropriate mesh-skeletonisation algorithms that would give feedback on the
11-19
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
number of cycles (cropped voxels) needed to produce an ideal one-voxel central line. Furthermore, the use of a more sophisticated method of IVUS image segmentation could further enlarge the set of landmarks via embedding tissue characterisation features and, thus, allowing for the discretisation among different lesions of the same type and the creation of a plaque map. These improvements would increase navigation precision and get us one step further towards the fully automated detection of all landmarks. Our proposed system currently relies on a friendly graphical user interface which allows for the manual detection and classication of each landmark based on the uctuations of several tted-ellipses parameters along an IVUS image stack. The development of novel training algorithm strategies, which may also encompass statistical modelling data as well as anatomical and epidemiological information, should further facilitate this process.
The current commercial electromagnetic tracking systems provide excellent feedback on the position as well as the spatial orientation of the catheter tip (5 or 6 DOF sensors), as utilised in our rst proposed system. However, the difculties in incorporating EM sensing during arterial procedures are clear. While our second proposed framework provides independence from EM trackers, the system can only partially compensate for the loss of the 3D orientation of the catheter tip. Development of a strategy relying solely on IVUS image data that would resolve this limitation and lead to better precision results is thus an important research direction. The preliminary technique assumes a perfectly cylindrical model for the vessel under investigation, with a constant wall thickness. By using measured wall thickness, aided by an extension to in vivo data, it may be possible to determine the true angular orientation of the IVUS scanning plane. This would also provide clinicians with information about the current state of the catheter in vivo, without resorting to the use of uoroscopy for conrmation.
The proposed methods also cannot depict the changes to the entire vessel structure caused by patient motion or the introduction of stiff endovascular devices during a procedure. With the rst proposed method, it is possible to perform an extra IVUS pullback after deformation to reconstruct the new shape; however, the second method is only able to identify the location of an endovascular imaging device relative to a known map of the vasculature. The use of biomechanical modelling has a role here in the simulation of the shape of the vessels after deformation, in particular, with the use of limited intravascular imaging data to act as constraints. The current limitation in its use is, as reviewed earlier in this chapter, the computation time required.
In summary, while the benets of using a safer and established intravascular imaging modality for navigation purposes are clear, there are still a number of challenging problems that require research to bring its application to clinical use.
11.5 Conclusion
We have presented here two intravascular navigation frameworks, both reliant on IVUS imaging. The rst requires EM tracking for the reconstruction of the vasculature in situ, a technique able to compensate for any motion between the
11-20
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
preoperative scan and the endovascular procedure. The second is based on image processing of intra-operative IVUS data and the use of a preoperative model, without the need for any external tracking equipment. This system aims to facilitate endovascular procedures and especially in providing the operator with the oppor­tunity to tag a specic point of interest along a vessel and re-visit it in less time and with better precision.
While there remain issues to be addressed for the proposed frameworks to be accepted clinically, the use of endovascular imaging methods for navigation is a promising approach that limits the use of x-ray uoroscopy and the accompanying contrast agents required to delineate the arteries.
Acknowledgements
Partial support of this research was provided by the FP7-EU Project Smart Catheterization (SCATh).
References
[1] Abi-Jaoudeh N, Glossop N, Dake M, Pritchard W F, Chiesa A, Dreher M R, Tang T,
Karanian J W and Wood B J 2010 Electromagnetic navigation for thoracic aortic stent-graft deployment: a pilot study in swine J. Vasc. Interv. Radiol.
[2] Alberti M, Balocco S, Gatta C, Ciompi F, Pujol O, Silva J, Carrillo X and Radeva P 2012
Automatic bifurcation detection in coronary IVUS sequences IEEE Trans. Biomed. Eng.
[3] Baert S A M, Viergever M A and Niessen W J 2003 Guide-wire tracking during
endovascular interventions IEEE Trans. Med. Imaging
[4] Balocco S, Gatta C, Pujol O, Mauri J and Radeva P 2010 SRBF: speckle reducing bilateral
ltering Ultrasound Med. Biol.
[5] Besl P J and McKay N D 1992 A method for registration of 3-D shapes IEEE Trans. Pattern
Anal. Mach. Intell.
[6] Brost A, Wimmer A, Liao R, Hornegger J and Strobel N 2010 Catheter tracking: lter-based
versus learning-based Pattern Recognition (Lecture Notes in Computer Science vol 6376)
(Berlin: Springer) sect 30, pp [7] Cascade. http://www.cascade-fp7.eu/ [8] Ciompi F, Balocco S, Caus C, Mauri J and Radeva P 2013 Stent shape estimation through a
comprehensive interpretation of intravascular ultrasound images Medical Image Computing
and Computer-Assisted InterventionMICCAI 2013 (Lecture Notes in Computer Science vol
8150) (Berlin: Springer) sect 43, pp
[9] Cochennec F, Riga C, Hamady M, Cheshire N and Bicknell C 2013 Improved catheter
navigation with 3D electromagnetic guidance J. Endovasc. Ther.
[10] Davison A J, Reid I D, Molton N D and Stasse O 2007 Monoslam: real-time single camera
slam IEEE Trans. Pattern Anal. Mach. Intell.
[11] Dore A, Smoljkic G, van der Poorten E, Sette M, Sloten J V and Yang G Z 2012 Catheter
navigation based on probabilistic fusion of electromagnetic tracking and physically-based
simulation IEEE/RSJ Int. Conf. on Intelligent Robots and Systems (IROS) (Piscataway, NJ:
IEEE) pp
[12] Evans J L et al 1996 Accurate three-dimensional reconstruction of intravascular ultrasound
data: spatially correct three-dimensional reconstructions Circulation
3806–11
14 239–56
36 1353–63
293–302
345–52
29 1052–67
21 888–95
59 1022–31
22 965–72
20 39–47
93 567–76
11-21
Vascular and Intravascular Imaging Trends, Analysis, and Challenges, Volume 1
https://t.me/medicina_free
[13] Fitzgibbon A, Pilu M and Fisher R B 1999 Direct least square tting of ellipses IEEE Trans.
Pattern Anal. Mach. Intell.
[14] Froggatt M E, Klein J W, Gifford D K and Kreger S T 2011 Optical position and/or shape
sensing US patent US20110109898 a1
[15] Gao B, Hu K, Guo S and Nan X 2013 Mechanical analysis and haptic simulation of the
catheter and vessel model for the MIS VR operation training system 2013 IEEE Int. Conf. on
Mechatronics and Automation (ICMA) (Piscataway, NJ: IEEE) pp
[16] Goldenberg I and Matetzky S 2005 Nephropathy induced by contrast media: pathogenesis,
risk factors and preventive strategies Can. Med. Assoc. J.
[17] Hiratzka L F et al 2010 ACCF/AHA/AATS/ACR/ASA/SCA/SCAI/SIR/STS/SVM guide-
lines for the diagnosis and management of patients with thoracic aortic disease: a report of
the American college of cardiology foundation/American heart association task force on
practice guidelines, American association for thoracic surgery, American college of radiol-
ogy, American stroke association, society of cardiovascular anesthesiologists, society for
cardiovascular angiography and interventions, society of interventional radiology, society of
thoracic surgeons, and society for vascular medicine Circulation
[18] Kandarpa K 2013 Magnetic resonance imagingguided endovascular interventionsare we
there yet? J. Vasc. Interv. Radiol.
[19] Karlas A and Lee S L 2015 Towards an IVUS-driven system for endovascular navigation
IEEE Int. Symp. on Biomedical Imaging (Piscataway, NJ: IEEE)
ISBI.2015.7164119
[20] Langelaar M and Keulen F V 2004 Modeling of a shape memory alloy active catheter
Structures, Structural Dynamics, and Materials and Co-located Conf. (Reston, VA:
American Institute of Aeronautics and Astronautics)
[21] Lazrag H, Aloui K and Naceur M 2013 Automatic segmentation of lumen in intravascular
ultrasound images using fuzzy clustering and active contours Proc. Eng. Technol. 1 58–63
[22] Lenoir J, Cotin S, Duriez C and Neumann P 2006 Interactive physically-based simulation of
catheter and guidewire Comput. Graph.
[23] Liu H, Fu Y L, Zhou Y Y, Li H X, Liang Z G and Wang S G 2010 An in vitro investigation
of image-guided steerable catheter navigation Proc. Inst. Mech. Eng. H
[24] Manstad-Hulaas F, Tangen G A, Dahl T, Hernes T A N and Aadahl P 2012 Three-
dimensional electromagnetic navigation versus uoroscopy for endovascular aneurysm
repair: a prospective feasibility study in patients J. Endovasc. Ther.
[25] Min P B 3D mesh voxeliser http://www.cs.princeton.edu/min/binvox/ (Accessed: 8 September
2014)
[26] Min P B 3D voxel model thinning http://www.cs.princeton.edu/min/thinvox/ (Accessed: 8
September 2014)
[27] Nooruddin F S and Turk G 2003 Simplication and repair of polygonal models using
volumetric techniques IEEE Trans. Vis. Comput. Graph.
[28] Otsu N 1979 A threshold selection method from gray-level histograms IEEE Trans. Syst.
Man Cybern.
[29] Palágyi K and Kuba A 1999 Directional 3D thinning using 8 subiterations Proc. of the 8th
Int. Conf. on Discrete Geometry for Computer Imagery (Lecture Notes in Computer Science
vol 1568) (Berlin: Springer) sect 25, pp
[30] Rotger D, Radeva P and Bruining N 2010 Automatic detection of bioabsorbable coronary stents
in IVUS images using a cascade of classiers IEEE Trans. Inf. Technol. Biomed.
9 62–6
21 476–80
1372–7
172 1461–71
121 e266–369
24 891–3
doi:https://doi.org/10.1109/
doi:https://doi.org/10.2514/6.2004-1653
30 416–22
224 945–54
19 70–8
9 191–205
325–36
14 535–7
11-22