A.L. Weiner
https://t.me/med1917
active agent out of the device is likely to be related to osmotic pressure or even gravitational influence as opposed to hydraulic or mechanical forces. In that case, the device chamber volume will reach some level of equilibrium with the external tissue environment. Again, as in the above case for an active pump system, the chamber content will contain a residual lower concentration of the active agent that should be purged prior to a refill.
Using a passive system makes the ability to define the refill interval more chal­lenging. In an active system, either through a known rate of fluid delivery or feed­back measure of the chamber fluid volume, a calculated refill interval can be fairly straightforward. However in a passive system which might maintain a constant fluid volume with decreasing drug concentration, the required feedback information could require more complex sensor approaches. Alternatively, well-powered clini­cal studies of patient efficacy could provide the information needed although this may not be the most timely or efficient process.
13.2.5.3
The above discussion does not take into consideration the potential of a device to be refilled with a solid dose form of the drug, such as reintroduction of a powder, tablet, cylinder, or fiber into an already implanted device. As one might expect this to be a more invasive approach overall, it would seem to be less attractive as a design comparatively to an external refillable injection system like a pellet gun or cartridge injector (Dinius and Huizenga 1984). Nonetheless, in consideration of such a method, the solid material would still require an environment where fluid can dissolve it. If the refilled solid is not directly open to the tissue on the output side, but rather is housed in a fully enclosed chamber or container, the consider­ation of occupied solid volume vs. residual tissue fluid volume in that chamber would need to be considered.
Solid Refill
13.2.6 Contamination Potential
Multiple refilling of an implanted reservoir offers potential to introduce infectious and/or noninfectious contaminants, particularly if the injection port is located under­neath the tissue (Renard et al. 2001). However, as long as sterile techniques are applied in the refill operation, risk for introducing infection should be low. In a study of 890 refill procedures in 25 patients with implanted intrathecal infusion pumps, cultures of samples taken from extracted residual drug in the reservoir at the last pump refill was negative for either aerobic or anaerobic bacteria (Dario et al.
2005). In the eye, the ability to flush surfaces with antiseptic such as povidone-
iodine and to have the patient apply a brief course of antibiotic prior to injection can help reduce or eliminate any chance of introducing endophthalmitis.
313
13 Refillable Devices for Therapy of Ophthalmic Diseases
https://t.me/med1917
13.3 Historical Influences
13.3.1 Infusion Pumps
Concepts for engineering of refillable infusion devices intended for ocular use have evolved from longstanding research and commercial development of systemic infu­sion pumps. From about 1980 onwards extensive research was put to developing fully implantable and refillable pumps which could achieve long-term systemic delivery of drugs for several key indications including therapy of cancers (Buchwald et al. 1980; Cohen et al. 1980, 1983a, b; Phillips et al. 1982), modulation of pain (Muller et al. 1984; Levy 1997) and control of diabetes (Selam et al. 1982; Prestele et al. 1983). Devices like the SynchroMed® (Medtronic, Inc.) and other similar designs allowed for percutaneous or abdominal implantation with accessible ports for refill or direct infusion via cannula connection to the pump (Fig. 13.3). Such pumps were made to contain large volumes (e.g., 10–40 mL) and deliver therapeutic levels for minimum periods of 2–3 weeks prior to refill, depending on rate needed and concentration of drug used. The driving force in some of the designs relies on gas pressure to drive fluid from the reservoir into the tubing which is then subject to peri­staltic pumping. The fairly large size of such devices prompted others to begin looking at approaches to reduce pump sizes through engineering of different infusion mecha­nisms. An example of one such method was described by Roorda (2001) who describe a rotating arm which applies compressive force to the dispensing path.
13.3.2 Glaucoma Drainage Devices
The foundation for experimental placement of many current ophthalmic drug delivery pump designs stems from studies examining drainage devices to reduce intraocular pressure. The first successful device to gain acceptance for this purpose was the
Fig. 13.3 SynchroMed implantable infusion pump (Medtronic, Inc.)
314
A.L. Weiner
https://t.me/med1917
design introduced by Molteno (1969) in which a plate portion of the device that was implanted in the anterior subconjunctival space maintained the patency of the filtra­tion reservoir, thus overcoming the issue of flow restriction due to subconjunctival fibrosis. Since that time there have been a variety of styles (including well-known commercial styles as the Baerveldt or Ahmed drainage valves) which are mainly implanted in the posterior subconjunctival or sub-Tenon’s space (Lim et al. 1998). Important knowledge gained from these historical studies includes tissue response to various biomaterials as well as function and placement of valved drain systems. For example, early setons, tubes and preliminary designs did not address control mechanisms to prevent development of hypotony, an important factor to consider when the device bridges multiple compartments in the eye. In line with this, glau­coma valve and shunt studies contributed to an understanding of design options to achieve optimum valve cracking pressures to assure continued flow without impact­ing backflow or drop in the internal intraocular pressure (Setabutr et
al. 2006).
13.3.3 Pioneering of Refill Procedure in the Eye
As of this writing, it is important to note that there is yet to be any commercialized refillable ophthalmic device. While the current technology for ophthalmic refillable systems is actually in very early stages, particularly relative to initial human clinical evaluations, this is not a result of such concepts being new. More than 30 years ago, Refojo, Liu and colleagues (Refojo et al. 1978; Liu et al. 1979, 1983; Refojo and Liu 1981) described episcleral implanted refillable silicone devices for treatment of intraocular malignancies using 1,2-bis [2-chloroethyl]-1-nitrosourea (BCNU). A silicone balloon was constructed by cementing two silicone sheets around a cannula that formed a reservoir accessible by the cannula (Fig. 13.4). Following episcleral implantation the device was refilled through the cannula with drug dissolved in oil or ethanol, which diffused rapidly from the device leaving drug depot in the reser­voir. Even though the device was elemental in its construction, it demonstrated in principle that episcleral implantation of a device having an external refill port could function as a potential mode of delivery. Surprisingly, despite these seminal reports, the potential of ophthalmic refillable systems went unrealized for many years and further interest in possible designs did not surface until the 1990s. This is perhaps related to a paucity of drugs deemed viable at the time for such an approach, the lack of knowledge concerning validation of drug efficacy by intraocular administration and a diversion of attention to the learning curve associated with developing more conventional intravitreal implant approaches eventually leading to products such as the Vitrasert® (Bausch and Lomb), Retisert® (Bausch and Lomb), and Ozurdex® (Allergan) devices. It can be noted that as ancillary to the development of the Ozurdex (Posurdex) system, variations of that implant approach were also reported using a drug core filled into a hollow impermeable cylinder with one or more orifices for drug release (Wong et al. 2001). Within the context of that variation, provision for refillability was discussed as an option, although specific enabling features were not described.
315
13 Refillable Devices for Therapy of Ophthalmic Diseases
https://t.me/med1917
Fig. 13.4 Refillable silicone
episcleral implanted device used for injection of carmustine solutions (from Liu et
al. 1979). Copyright 1979 Association for Research in Vision and Ophthalmology. Reproduced with permission of INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE in the format Other Book via Copyright Clearance Center
13.4 Ophthalmic Refillable Devices
13.4.1 Invasiveness and Refilling Frequency
To be successful for ophthalmic purposes, surgical implantation followed by subsequent repetitive refill procedures should accommodate both the patient’s and physician’s practices with respect to management of the specific disease. The factors include (a) the accepted re-visit interval of the patient to the ophthalmologist’s office, (b) the patient’s visual status, and (c) the speed of the disease progression. In the case of glaucoma treatment, it is common practice that patient follow-up visits are typically no less than every 3 months and often are at 4–6-month intervals. These match reasonable periods for a refill procedure. However, glaucoma patients whose intraocular pressure is easily controlled with drop medication and whose vision has not yet been dramatically impacted by the disease are unlikely to be willing candi­dates for surgical implantation of a refilling device. Rather, patients at later disease
316
A.L. Weiner
https://t.me/med1917
stages or those having difficulty with eyedrop compliance have shown willingness to undergo various surgical steps including implantation of drainage valves or trab­eculectomy to arrest further progression. Moreover, similar to medically controlled and compliant glaucoma patients, in a slow progressing disease state such as the dry form of age-related macular degeneration (AMD), where vision may be reasonably good, the patient will be more risk adverse and thus resistant to surgical procedures which have a high level of complexity or invasiveness. Furthermore, their desire for retreatment will gravitate toward at least a 6–12-month interval. In more rapidly progressing disease, such as the wet form of AMD, historical experience has shown that patients have been willing to accept treatments as frequent as every 6 weeks, as has been the case for anti-VEGF therapies, although this is not the most desired practice either from the patient or physician perspective.
As a general rule, a device should offer advantages that are commensurate with the complexity or invasiveness of the procedure to implant it. Key advantages to the patient can include potential for long refill intervals, automated dosing, precise con­trol of symptoms or disease progression via feedback sensors or mechanisms, and a high degree of safety or comfort. Key advantages should be offered to the physician as well. This would include worthwhile reimbursement for the procedure, low sur­gical risks, ease of technical operation of the device (i.e., remote charging, simple refill process, data uploads or downloads, custom settings for dosing, etc.) and flex­ibility of the design of the device to accommodate more than one therapeutic medi­cation, whether refilled or dosed serially or concomitantly.
13.4.2 Intravitreal Delivery Through the Pars Plana
Along with the imminent commercialization of fully implanted intravitreal devices for long term for delivery of therapeutic agents, Weiner et al. (1995) proposed an alternative intravitreal delivery method in which a small refillable tack-shaped drug delivery device could be anchored across the sclera, having the delivery chamber in vitreous and an injectable refill port in the proximal end cap which was accessible under the conjunctiva (Fig. which single or dual reservoir chambers could be designed. In the dual reservoirs, a conduit between them (shown as element number 64 in Fig. 13.5) could optionally contain a one-way valve or diaphragm for flow control.
The above initial concept for pars plana anchored refillable systems with an accessible subconjunctival injection port was broadened further by Varner, De Juan, and colleagues a number of years later (Varner et al. 2002, 2004). In the earlier patent by these authors, a refillable reservoir was designed that had expansion capa­bility upon filling, thus allowing for large loading capacity (Fig. 13.6). The latter reference describes a modification of the shape of the device to that of a coil which could deliver drug through several mechanisms including a hollow lumen of the coil to accommodate liquid refill though the proximal end. A nonrefill-coated style of this coil device termed I-Vation™ is currently being evaluated clinically for
13.5). Several styles of the device were presented in
317
13 Refillable Devices for Therapy of Ophthalmic Diseases
https://t.me/med1917
Fig. 13.5 Refillable pars
plana implanted tack device with refillable injection port under the conjunctiva. Reprinted from Weiner et
1995)
(
al.
long-term delivery of steroid. The refillable tack and coil pars plana designs have been shown comparatively to a similar design approach but using a fully erodible format (Weiner 2007).
In general, the concept of a refillable device bridging through multiple tissues but containing an accessible injection port region under a layer of tissue has been extended to include other areas of the body. Ashton et al. (1998) and Watson et al. (2005) utilized similar techniques to describe a refillable system to reach inner portions of the brain with a compartmentalized device implanted under the scalp, through the skull and having a delivery tube extending distally into the target tissue (Fig. 13.7). The device further contained a semi-permeable membrane to control the rate of drug flow.
As opposed to having the entire refillable device located in the pars plana, Avery and Luttrull (1998) developed a design in which the majority of the device was located more posteriorly in the episcleral or sub-Tenon’s space, but from which a cannula would be directed from the device, penetrate through the pars plana, and terminate within the vitreous (Fig. 13.8). While the refill reservoir was more poste­riorly located, the designed injection port region (element 122 in Fig. 13.8) was angled and encompassed a broad area to facilitate easier access with a needle. As part of additional embodiments, this design further incorporated either valve- or baffle-type elements to reduce backflow, prevent flow out of the device during refill, or to allow for dosing by means of applying external pressure on the reservoir.
In a more concerted effort to bring the initial Avery concept to commercial utility, researchers at the University of Southern California and the California Institute of Technology have developed further engineering advancements of this style of refillable
318
A.L. Weiner
https://t.me/med1917
Fig. 13.6 Refillable pars plana implanted device with expandable balloon type reservoir chamber
within the vitreous. Reprinted from Varner et al. (2002)
device to automate its operation (Li et al. 2008; Lo et al. 2009; Saati et al. 2009; Pang et al. 2010; Avery et al. 2010). A primary added feature includes a microelec­tromechanical (MEMS) controlled electrolysis chamber which, when remotely acti­vated, expands from the gas pressure, forcing drug-containing fluid out of a second adjacent reservoir and through the cannula which terminates either in the vitreous or anterior chamber (Fig. 13.9a, b). The enhancements also accounted for a hardened baseplate underneath the refill port to prevent a reinjection needle from penetrating through to the electronic componentry (Fig. 13.10) or via a stop built onto the shaft of the needle itself (Meng et al. 2009). Recent results with prototype devices con­taining glaucoma agents have shown controllability of anterior chamber dosing from the picoliter/minute to microliter/minute rates and IOP lowering efficacy in dogs comparable to controls (Avery 2010).
319
13 Refillable Devices for Therapy of Ophthalmic Diseases
https://t.me/med1917
Fig. 13.7 Refillable ported controlled release device design showing extended delivery tube (element
530) to reach distal tissue target. The injection reservoir contains a stop plate (element 539) to prevent needle penetration. The stop plate further contains holes (element 538) to allow fluid diffusion. A rate-limiting permeable membrane (element 550) prevents injection of small foreign particles into the tissue. Reprinted from Watson (2005)
13.4.3 Episcleral Implantation for Trans-Scleral Delivery
Although injections into the anterior subconjunctival space have been a longstanding practice dating back to the 1950s, particularly for injections of antibiotics or steroids, the placement of drugs or devices in the posterior sub-Tenon’s space did not gain favor until the beginning of the new millennium following new investigations utilizing techniques to place a depot of the anti-angiogenic agent anecortave acetate above the macula (Slakter et al. 2002; Dahlin et al. 2003). Concurrent with the develop­ment of an injection cannula for placement of such suspensions, a solid silicone based
320
A.L. Weiner
https://t.me/med1917
Fig. 13.8 Episcleral implanted device with drug reservoir connected to a cannula to deliver drugs
through the pars plana to the vitreous. The injection port region above the drug reservoir has an angled and large surface area to facilitate needle placement for reinjections. Reprinted from Avery and Luttrull (1998)
device was also devised as a means to put a more controlled rate form of the drug near the macula (Yaacobi 2002; Yaacobi et al. 2003). This original device had a length that followed along the border of the lateral rectus muscle and allowed place­ment of a drug load above the macular region with a distal end that was accessible anteriorly near the limbus to allow for easy retrieval. Continued modifications of that device were subsequently designed (Fig. 13.11) to allow for refill from an anterior port position connected to pathways that would communicate fluid posteriorly (Yaacobi 2006a). Furthermore, recognizing the possible need to more broadly dis­tribute drug throughout the eye posterior, as might be required in disease states such as dry AMD, a circumferential modification of the design was proposed by
Fig. 13.9 (a) Schematic of an episcleral implanted MEMS-controlled refillable drug delivery
https://t.me/med1917
pump. Design elements include an electrolysis chamber (element 140) which is shown full expanded from water hydrolysis and gas pressure, remotely controlled MEMS chip to generate current for the water hydrolysis (shown at the base of the reservoir), adjacent fluid chamber con­taining the drug (element 130), and cannula from the drug chamber (element 120) which contains a check valve (element 200). Reprinted from Pang et placement of the MEMS-controlled refillable device. Positioning shows location between the rectus muscles, a low profile to avoid irritation (<2 Reproduced from Lo et al. (2009). Copyright 2009 with permission from Springer
Fig. 13.10 Design features to assure proper placement of a 30-gauge reinjection needle in a refillable
port. Diagram (a) and sham device (b) show a hard polymer baseplate to prevent the needle from penetrating components underneath the reservoir and a visible refill ring demarcating the port position. Reproduced from Lo et al. (2009). Copyright 2009 with permission from Springer
al. (2010). (b) Illustration of profile and ocular
mm) and approximate cannula insertion point.