Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:
Ординатура / Хирургия / Библиотека им академика М.И. Перельмана / Книга_541_Библиотеки_им_академика_М_И_Перельмана.pdf
Скачиваний:
0
Добавлен:
29.08.2026
Размер:
86 Мб
Скачать
5 Robotic Gastric Neurostimulator Placement
39
Fig. 5.9 The needle is then tunneled through the muscularis layer of the stomach, taking care to avoid the larger vessels on the greater cur­vature of the stomach. The lead must be placed precisely so as not to enter the gastric lumen or be visible under the serosa
Fig. 5.10 The needle direction is perpendicular to the major axis of the stomach and traverses the muscularis layer
Fig. 5.11 After a minimum of 1cm of gastric wall tunneling, the nee­dle may be resurfaced and exit the serosa. It is then grasped again and pulled through until the bare metal of the neurostimulator lead enters the tunnel and the provided silicone ange is ush with the stomach wall
Fig. 5.12 The second neurostimulator lead is similarly placed in the stomach wall, 1cm away from the rst lead in parallel fashion
40
Fig. 5.13 Neurostimulator lead placement checks are performed before the leads are secured to the stomach wall. The robotic instru­ments can be used to stabilize the leads in position while the placement checks are performed
W. C. Sherrill III and M. M. Awad
Fig. 5.14 The rst placement check is performed endoscopically. The upper endoscope is used to ensure no metal from the neurostimulator leads are visible within the gastric lumen. The robotic TilePro feature can be used to simultaneously view the robotic and endoscopic elds during this step
Fig. 5.15 The other end of the leads are gently held outside of the 12mm assistant port to prevent them from being inadvertently pulled into the abdomen
5 Robotic Gastric Neurostimulator Placement
41
Fig. 5.16 The second lead placement check is then performed by con­rming an appropriate current through the circuit. First, the two neuro­stimulator leads are inserted into the gastric neurostimulator in the provided slots. Either lead can be placed in either slot
Fig. 5.17 Leads are then secured in place using the provided torque wrench. All four screws are tightened through the holes in the plastic generator housing. The torque wrench prevents overtightening of the screws
42
W. C. Sherrill III and M. M. Awad
Fig. 5.18 The leads are gently tugged to conrm they are securely anchored to the generator
Fig. 5.19 The paddle of the programming device is then placed into a transparent, sterile cover and introduced into the surgical eld. The paddle of the programmer is placed in contact with the neurostimulator generator. (The entire programmer can also be sterilely prepped into the eld and controlled by the bedside assistant)
5 Robotic Gastric Neurostimulator Placement
43
Fig. 5.20 Slide the “P” button on the side of the programmer to turn it on
Fig. 5.21 Tap the “OK” icon to begin the diagnostic check
44
W. C. Sherrill III and M. M. Awad
Fig. 5.22 The programmer will take a few moments to communicate with the neurostimulator generator
Fig. 5.23 Impedance through the circuit is checked. Leads are prop­erly placed when the impedance is between 200 and 800 ohms
Fig. 5.24 Once these checks are successful, the neurostimulator leads can be secured to the gastric wall. The leads are then unscrewed from the stimulator using the torque wrench
5 Robotic Gastric Neurostimulator Placement
45
Fig. 5.25 Next, the silicone anges of the two leads are secured to the gastric wall. This is performed by placing interrupted 2-0 Ethibond™ sutures between a serosal bite of the stomach wall and through the pro­vided holes in the anges. Care must be taken to avoid damage to the tunneled leads when placing these sutures
Fig. 5.26 Shown is the cranial stitch being placed for the proximal gastric lead
Fig. 5.27 Completed image showing all securing stitches placed through the anges
Fig. 5.28 The leads are further secured at their distal ends where they exit the stomach wall. The provided silicone discs are used for this purpose
46
W. C. Sherrill III and M. M. Awad
Fig. 5.29 The gastric neurostimulator ski needle is passed through the discs as shown, threading the Prolene suture portion of the lead through the appropriate holes
Fig. 5.30 The disc is then secured to the anterior stomach wall. Additional 2-0 Ethibond suture is placed through the provided side holes of the disc and then to the stomach wall using a serosal bite
Fig. 5.31 Completed xation of the disc to the anterior gastric wall is shown
Fig. 5.32 The process is then repeated with the other neurostimulator lead using another silicone disc. The disc is secured away from the rst one
5 Robotic Gastric Neurostimulator Placement
47
Fig. 5.33 A robotic clip applier is then used to further secure the leads where the Prolene suture portion exits the silicone discs
Fig. 5.34 Two clips are placed on each Prolene suture, close to where it exits the silicone disc
Fig. 5.35 The two ski needles are then cut off the suture and removed from the abdomen. Completed placement of the two gastric neurostim­ulator leads is shown. This concludes the robotic portion of the procedure
Fig. 5.36 The robot is undocked from the patient, pneumoperitoneum is released, and the 12mm assistant trocar is carefully removed leaving the ends of the leads outside of the patient’s body
48
W. C. Sherrill III and M. M. Awad
Fig. 5.37 The generator pocket is then prepared. The 12mm trocar site incision is extended medially to 4cm in length
Fig. 5.38 Monopolar electrosurgical energy must not be used at any point in the remaining portion of the procedure. Current may travel through the leads causing stomach wall thermal injury or damaging the leads themselves. Hemostasis may be achieved with simple direct pres­sure or a bipolar forceps device