- •Preface to the fourth edition
- •Preface to the first edition
- •Applied anatomy
- •Applied physiology
- •Physical properties of materials
- •Manufacture of lenses
- •References
- •Slit lamp
- •Keratometers and autokeratometers
- •Corneal topographers
- •Anterior segment photography
- •Specialist instruments for higher magnification
- •Other instruments
- •References
- •Further reading
- •Legal implications
- •Record cards
- •Clinical grading
- •Computerization of patient records
- •References
- •Further reading
- •Hygienic procedures to avoid cross-infection
- •Solutions and drugs
- •Decontamination and disinfection of trial lenses
- •In case of accident
- •Other procedures
- •Insertion and removal by the practitioner
- •References
- •Further reading
- •Discussion with the patient
- •Indications and contraindications
- •Advantages and disadvantages of lens types
- •Visual considerations
- •External eye examination
- •Patient suitability for lens types
- •References
- •The tear film
- •Dry eyes
- •Assessment of tears
- •Contact lens signs
- •Treatment and management
- •Contact lens management
- •References
- •Rigid gas-permeable lenses
- •Polymethyl methacrylate
- •Soft lenses
- •Silicone hydrogels
- •Biocompatible lenses
- •Silicone lenses
- •References
- •Basic principles of rigid lens design
- •Forces controlling design
- •Concept of edge lift
- •Tear layer thickness
- •Lid attachment lenses
- •Interpalpebral lenses
- •References
- •Introduction
- •Current bicurve, tricurve and multicurve designs
- •Current aspheric lenses
- •Reverse geometry lenses
- •References
- •Introduction
- •Back optic zone radius (BOZR)
- •Total diameter (TD)
- •Back optic zone diameter (BOZD)
- •Peripheral curves
- •Lens design by corneal topographers
- •Recommended reading
- •Use of fluorescein
- •Examination techniques
- •Fitting
- •Correct fitting
- •Flat fitting
- •Steep fitting
- •Astigmatic fitting
- •Peripheral fitting
- •References
- •Advantages and disadvantages of aspherics
- •Aspheric designs
- •Principles of fitting
- •Fluorescein patterns compared with spherical lenses
- •References
- •International Standards
- •Examples of rigid lens types and fittings
- •Rigid lens verification
- •Tolerances
- •References
- •Historical
- •Current approach
- •Reverse geometry lenses
- •Clinical appearance of reverse geometry lenses
- •Corneal topography
- •Fitting routine
- •References
- •Further reading
- •Fitting considerations
- •Corneal diameter lenses
- •Semi-scleral lenses
- •Reference
- •Characteristics of a correct fitting
- •Characteristics of a tight fitting
- •Characteristics of a loose fitting
- •Summary of soft lens fitting characteristics
- •Lens power
- •Lens flexibility and modulus of elasticity
- •Additional visual considerations
- •Thin lenses
- •Aspheric lenses
- •Spun-cast lenses
- •Unusual lens performance
- •References
- •Frequent replacement lenses
- •Disposable lenses
- •Types of disposable lens
- •Fitting disposable lenses
- •Aftercare with disposable lenses
- •Practice management
- •Other uses for disposable lenses
- •References
- •Fitting disposable silicone hydrogels
- •Fitting custom made silicone hydrogels
- •Complex lenses
- •Dispensing silicone hydrogels
- •Aftercare
- •References
- •Further reading
- •International standards and tolerances1
- •Soft lens specification (Tables 20.1, 20.2)
- •Soft lens verification
- •References
- •Physiological requirements
- •Approaches to extended wear
- •Patient selection
- •Soft lens fitting and problems
- •Rigid gas-permeable fitting and problems
- •Other lenses for extended wear
- •Long-term consequences of extended wear
- •References
- •Residual and induced astigmatism
- •Patient selection
- •Lens designs
- •Methods of stabilization
- •Fitting back surface torics
- •Fitting bitorics
- •Compromise back surface torics
- •Fitting front surface torics
- •Fitting toric peripheries
- •Computers in toric lens fitting
- •References
- •Patient selection
- •Stabilization
- •Lens designs
- •Fitting
- •Fitting examples
- •References
- •Patient selection
- •Monovision
- •Presbyopic lens designs
- •Fitting rigid multifocals and bifocals
- •Fitting soft bifocals
- •References
- •Lens identification
- •Tinted, cosmetic and prosthetic lenses
- •Fenestration
- •Overseas prescriptions
- •Contact lenses and sport
- •References
- •Components of solutions
- •Solution for soft lenses
- •Disinfection
- •Solutions for rigid gas-permeable lenses
- •Compliance and product misuse
- •References
- •Lens collection
- •Insertion and removal
- •Suggested wearing schedules
- •General patient advice
- •First aftercare visit
- •Visual problems
- •Wearing problems
- •Aftercare at yearly intervals or longer
- •References
- •Emergencies and infections
- •Grief cases (drop-outs)
- •Side effects of systemic drugs
- •Lens ageing
- •References
- •Refitting PMMA wearers
- •Prescribing spectacles for contact lens wearers
- •Rigid lens modification
- •Management
- •Instrumentation
- •Non-therapeutic fitting
- •Refractive applications
- •Therapeutic applications
- •References
- •High myopia and hypermetropia
- •Keratoconus
- •Aphakia
- •Corneal grafts (keratoplasty)
- •Corneal irregularity
- •Albinos
- •Combination lenses
- •Silicone rubber lenses
- •Bandage lenses
- •Additional therapeutic uses
- •References
- •Appendix 1
- •Journals
- •Teaching resources
- •Professional
- •General interest
- •Technology
- •Investigative techniques
- •Ophthalmology
- •Glossary
- •Index
Toric rigid lenses 22 Chapter 
As induced astigmatism is neutralized, the final powers are:
BVP along flatter meridian: = −2.00 D
BVP along steeper meridian: = −2.00+ (−2.33) = −4.33
Equivalent to: −2.00/−2.33 along flatter meridian
22.6 Fitting bitorics
A bitoric is required when a back surface toric has created sufficient induced astigmatism (usually over 0.75 D) to warrant correction with a front surface cylinder.
•If corneal and ocular astigmatism are equal, the induced astigmatism created may require correction.
•If the ocular astigmatism is less than the corneal astigmatism, lenticular astigmatism is revealed by a rigid lens and requires neutralizing together with the induced astigmatism. In this case, a large front surface cylinder is needed.
Example: |
|
|
|
|
|
Ocular refraction: |
−2.00/−3.00 × 180 |
||||
Keratometry |
8.00 mm along 180, 7.40 mm along 90 |
||||
Corneal astigmatism |
3.00 D |
|
|
||
Residual astigmatism |
0.00 D |
|
|
||
If we choose BOZRs of: |
r1 = 7.50 mm, r2 = 8.00 mm |
||||
For: |
n = 1.336, n’ = 1.480 |
||||
Induced astigmatism = |
−144 |
− |
−144 |
||
|
8.00 |
|
|||
|
7.5 |
|
|
||
= (−19.20) − (−18.00) = −1.20 D
This degree of induced astigmatism requires correction. The astigmatism induced by the back surface in air:
−480 − −480 = (−64.00) − (−60.00) = −4.00 D 7.50 8.00
BVP along flatter meridian |
= −2.00 D |
BVP along steeper meridian |
= (−2.00) + (−4.00) − (−1.20) |
|
= −4.80 D |
Equivalent to −2.00/−2.80 along flatter meridian
22.7 Compromise back surface torics
If the ocular astigmatism is greater than the corneal astigmatism, a compromise fitting option is worth considering to avoid the use of a front surface cylinder. In these cases, some of the induced astigmatism will be neutralized by the residual astigmatism and the remaining induced astigmatism can be further eliminated by reducing the difference between the flatter and steeper radii.
259
Section four Complex lenses
•Convert the spectacle prescription to an ocular prescription.
•Calculate the resultant induced astigmatism when the back surface toric lens is designed close to alignment.
•Calculate the residual astigmatism or use rule-of-thumb.
•Calculate the difference needed in BOZRs to create induced astigmatism that is equal or close to the residual astigmatism.
•Redesign the lens so that the flatter BOZR aligns with the flattest corneal meridian and the steeper BOZR satisfies the above calculation.
Example: |
|
Ocular refraction: |
−4.00/−5.00 × 180 |
Keratometry |
8.00 mm along 180, 7.20 mm along 90 |
Corneal astigmatism |
4.00 D |
Residual astigmatism |
(−5.00) − (−4.00) = −1.00 D |
RULE OF THUMB
Choosing the lens radii to be approximately 75% of the corneal toric difference often avoids the need for a bitoric lens.
If the BOZRs were chosen to align r1 = 7.20 r2 = 8.00 For n = 1.336 n’ = 1.480
Induced astigmatism = −7.20144 − −8.00144
= (−20.00) − (−18.00) = −2.00 D
Front surface cylinder to compensate = (−2.00) − (−1.00) = −1.00 D However, by flattening the steeper BOZR to 7.40 mm
Induced astigmatism = −7.40144 − −8.00144
= (−19.45) − (−18.00) = −1.45 D Uncorrected astigmatism = (−1.45) − (−1.00) = −0.45 D
The patient would be expected to tolerate this small amount of astigmatism, therefore a bitoric is not needed with a back surface toric of 8.00 × 7.40 mm.
22.8 Fitting front surface torics
Method 1
•Use a large spherical diagnostic lens with power as near as possible to the final Rx and over-refract.
•Order a lens based on this power, usually with 1.5 .
•Always have the prism base marked so that the lens orientation can be measured on the eye.
260
Toric rigid lenses 22 Chapter 
•Over-refract and record the cylinder and axis.
•Note the orientation of the prism base.
•If the lens is unstable, consider a truncation of 0.50 mm, increasing the prism, or a larger TD.
•Order the front cylinder, taking into account any rotation of the lens.
•The cylinder can be worked on the initial lens if no other changes are needed.
Method 2
•Over-refract with a spherical diagnostic lens and order the final prescription lens directly with the front surface cylinder.
•This procedure is quicker, but an estimate of the final lens rotation is required (usually 5–15° nasal).
Example: |
|
Spectacle Rx: |
−2.75/−1.50 × 80 |
‘K’ readings: |
7.55 along 80, 7.45 along 170 |
Lens order: |
C3 7.55:7.50/9.50 AEL 0.12 Dk 60 |
|
−2.75/−1.00 × 75 1.5 base down (along 270) |
|
Single truncation: 0.50 mm along 180 |
|
Mark dots at prism base and apex |
PRACTICAL ADVICE
•If only one eye requires a toric, there may be problems with tolerance.
•Use truncation if there is a problem with binocular vision because of prism in only one eye.
22.9Fitting toric peripheries
•Fitted when the cornea is more astigmatic in the periphery and a small lens design fails to work.
•Normally used when the corneal astigmatism is 2.00–3.00 D.
•With a spherical diagnostic set, the fitting should be assessed for central alignment and correct peripheral clearance in the flatter meridian.
•The radius for the steeper peripheral meridian can be determined either from the ‘K’ readings and looking up axial edge lift tables for each meridian, or from the fluorescein pattern. A minimum toric difference of 0.60 mm is usually required.
•The secondary curves are approximately 0.80–1.20 mm flatter.
•All toric periphery lenses with a spherical BOZR give an oval optic with the smallest diameter along the flattest meridian.
•If there are manufacturing problems with the theoretically designed periphery, an additional spherical edge curve can be used.
261
Section four Complex lenses
•A toric periphery diagnostic set can be used for fitting.
•The prescription lens should give good acuity since the liquid lens neutralizes corneal astigmatism.
•A toric periphery is less expensive than a fully toroidal back surface design but represents a compromise.
Example: |
|
|
‘K’ 7.90 mm × 7.50 mm |
||
C3 7.90 : 7.00 |
8.90 |
: 8.20 11.30 : 9.00 |
|
8.30 |
10.70 |
C3 7.90 : 7.00 |
8.90 |
: 8.20 10.75 : 9.00 |
|
8.30 |
|
22.10 Computers in toric lens fitting
Computer programmes are available which help in the design and ordering of bitoric rigid gas-permeable contact lenses. The required data are keratometry readings, spectacle Rx, BVD and, possibly, the selected BOZR, BOZD and TD. Most programmes provide the final lens specification along the flatter and steeper meridians as well as the ocular astigmatism, corneal astigmatism, residual astigmatism and induced astigmatism. From this information it becomes clear whether a front surface cylinder is necessary. Some programmes suggest the compromise fit required to avoid a front surface cylinder.
Most corneal topographers have a contact lens module which can create lens designs for back surface torics, bitorics or front surface torics. It is then possible for the practitioner to set the level of corneal toricity at which the fitting module will begin to suggest a toric lens. The programmes will simulate the expected fluorescein picture so changes to the BOZR can be made before lens ordering. However, the effect of lid tone, position and other ocular surface factors will affect the performance of the lens on the eye. The simulated fluorescein pattern is therefore only a starting point and assessment needs to be carried out in the usual way to determine the optimum fitting.
References
1.Douthwaite WA. Contact Lens Optics and Lens Design. 3rd ed. Oxford: Butterworth-Heinemann; 2006.
2.Christie C, Keirl A, editors. The correction of astigmatism with rigid gas permeable contact lenses. In: Clinical Optics and Refraction. Elsevier; 2007. p. 225–40.
262
