Ординатура / Офтальмология / Английские материалы / Visual Prosthesis and Ophthalmic Devices New Hope in Sight_Rizzo, Tombran-Tink, Barnstable_2007
.pdf
Glaucoma Drainage Devices |
185 |
glaucoma surgery, recently published studies that address current use of these agents, and new methods of wound modulation refer to Yoon and Singh, 2004 [83]).
ANIMAL MODELS
Increased utilization of animal models of glaucoma will become crucial for novel GDDs to be more critically evaluated in comparison with existing implants. For a comprehensive review of spontaneously occurring and experimental animal models of glaucoma refer to Gelatt et al. (4,86). Such animal models are useful when studying, not only the effects of new GDD designs, but also the effects of new drainage sites for the diversion of aqueous humor (69,87) and antimetabolites to reduce the occurrence of postoperative fibrosis.
FLOW CONTROL: NEW VALVE DESIGNS
The Ahmed Glaucoma Valve (New World Medical, Inc., Rancho Cucamonga, CA) was invented by Mateen Ahmed, and was approved by the US Food and Drug Administration in November 1993 (88). The Ahmed valve design offers a unidirectional, pressure-sensitive valve created of a silicone, elastomer membrane that resides in a trapezoidal chamber. The membranes of this valve have an opening IOP of 8 mmHg and a Venturi effect is created to assist with the flow of aqueous through the chamber (88). The Ahmed valve design has been reported to reduce the complications associated with overfiltration and subsequent hypotony in the immediate postoperative period compared with other drainage devices, although the long-term results are similar to other GDDs (89). As such, new Ahmed valve designs have subsequently been developed and studied, and different designs of Ahmed GDDs are available varying in size, shape, number of explant plates, and materials used (88).
Future studies continuing to investigate novel designs for GDD valve mechanisms, building on the significant information gained from current valve mechanisms may help to prevent acute and chronic hypotony associated with GDDs.
TISSUE BIOCOMPATIBILITY: ALTERNATIVE MATERIALS
Inflammation associated with the biomaterials used in the construction of GDDs, and their explant plates in particular may contrbute to bleb failure secondary to fibrosis (59). Consequently, choosing a biomaterial with limited potential to cause inflammation may improve the success rate of GDDs (59).
Experimental GDDs made of new polymers or polymer modifications to help reduce the risks of chronic inflammation and subconjunctival fibrosis have been described. The alternative materials used in these experimental studies with potential application in GDDs used the management of glaucoma in humans, include: GDDs constructed of expanded polytetrafluoroethylene (90); hydroxylapatite reservoir tube implants (91); and the potential to use the same material used to manufacture Acrosof, a new acrylic (polymethylmethacrylate), foldable intraocular lens implant, for producing explant plates of GDDs (59). In addition, successful development of a new group of bioinert materials synthesized from monomers based on phosphorylcholine appears promising regarding improving the biocompatibility of medical devices, such as GDDs (92). One study reported significantly less protein (fibrinogen and fibrin) and cellular (macrophages and scleral fibroblasts) adhesion to phosphorylcholine-coated polymethylmethacrylate discs compared with levels of adhesion to materials used in contemporary GDDs
186 Cullen
including uncoated phosphorylcholine-coated polymethylmethacrylate, PTFE, polypropylene, and silicone (92).
NEW DRAINAGE SITES: EPITHELIUM-LINED EXTRAORBITAL SITES
Postoperative fibrosis at the explant site may be prevented with GDDs, which divert aqueous humor to extraorbital sites with epithelial lining. Recipient sites in the periorbital region including the lacrimal drainage apparatus and paranasal sinuses have been implanted with GDDs in patients with keratoprosthesis and severe glaucoma (71,72). Based on studies conducted not only in humans (71,72), but also in dogs (69,70), additional studies investigating these and similar innovative surgical procedures for shunting aqueous humor to these epithelium-lined sites in different forms of glaucoma are warranted. In particular, because the explant plate placed subconjunctivally is the main site of the fibrosis, designing shunts for extraorbital diversion without explant plates, as is performed in dogs (69,70), may be beneficial.
CONCLUSIONS
In order to effectively and critically evaluate such modifications in GDDs, future research should be directed at prospective studies using animal models of glaucoma with systematic investigation of variables including biocompatible materials, novel drainage sites, and current and newer antimetabolites, and standard guidelines regarding success with shortand long-term follow-up.
REFERENCES
1.Weinreb RN, Khaw PT. Primary open-angle glaucoma. Lancet 2004;363:1711–1720.
2.Quigley HA. Number of people with glaucoma worldwide. Br J Ophthalmol 1996;80:389–393.
3.Shiose Y. Intraocular pressure: new perspectives. Surv Ophthalmol 1990;34:413–435.
4.Gelatt KN, Brooks DE, Samuelson DA. Comparative glaucomatology. I: The spontaneous glaucomas. J Glaucoma 1998;7:187–201.
5.Bellezza AJ, Rintalan CJ, Thompson HW, Downs JC, Hart RT, Burgoyne CF. Deformation of the lamina cribrosa and anterior scleral canal wall in early experimental glaucoma. Invest Ophthalmol Vis Sci 2003;44:623–637.
6.Stefansson E, Pedersen DB, Jensen PK, et al. Optic nerve oxygenation. Prog Retin Eye Res 2005;24:307–332.
7.Dreyer EB, Zurakowski D, Schumer RA, Podos SM, Lipton SA. Elevated glutamate levels in the vitreous body of humans and monkeys with glaucoma. Arch Ophthalmol 1996;114: 299–305.
8.Yan X, Tezel G, Wax MB, Edward DP. Matrix metalloproteinases and tumor necrosis factor alpha in glaucomatous optic nerve head. Arch Ophthalmol 2000;118:666–673.
9.Sacca S. Nitric oxide as a mediator of glaucoma pathogenesis. Med Sci Monit 2002;8: LE33–LE34.
10.Tezel G, Edward DP, Wax MB. Serum autoantibodies to optic nerve head glycosaminoglycans in patients with glaucoma. Arch Ophthalmol 1999;117:917–924.
11.Yucel YH, Zhang Q, Weinreb RN, Kaufman PL, Gupta N. Effects of retinal ganglion cell loss on magno-, parvo-, koniocellular pathways in the lateral geniculate nucleus and visual cortex in glaucoma. Prog Retin Eye Res 2003;22:465–481.
12.Tielsch JM, Sommer A, Witt K, Katz J, Royall RM. Blindness and visual impairment in an American urban population. The Baltimore Eye Survey Arch Ophthalmol 1990;108:286–290.
Glaucoma Drainage Devices |
187 |
13.Lim KS, Allan BDS, Lloyd AW, Muir A, Khaw PT. Glaucoma drainage devices; past, present, and future. Br J Ophthalmol 1998;82:1083–1089.
14.Khaw PT, Shah P, Elkington AR. Glaucoma—2: treatment. Br Med J 2004;328:156–158.
15.Chen CW, Huang HT, Bair JS, Lee CC. Trabeculectomy with simultaneous topical application of mitomycin-C in refractory glaucoma. J Ocul Pharmacol 1990;6:175–182.
16.Smith MF, Sherwood MB, Doyle JW, Khaw PT. Results of intraoperative 5-fluorouracil supplementation on trabeculectomy for open-angle glaucoma. Am J Ophthalmol 1992;114: 737–741.
17.Assaad MH, Baerveldt G, Rockwood EJ. Glaucoma drainage devices: pros and cons. Curr Opin Ophthalmol 1999;10:147–153.
18.Hong CH, Arosemena A, Zurakowski D, Ayyala RS. Glaucoma drainage devices: a systematic literature review and current controversies. Surv Ophthalmol 2005;50:48–60.
19.Rodrigues AM, Corpa MV, Mello PA, de Moura CR. Results of the Susanna implant in patients with refractory primary congenital glaucoma. J AAPOS 2004;8:576–579.
20.Muckley ED, Lehrer RA. Late-onset blebitis/endophthalmitis: incidence and outcomes with mitomycin C. Optom Vis Sci 2004;81:499–504.
21.Rollett M, Moreau M. Traitement de hypopyon par le drainage capillary de la chambre anterieure. Rev Gen Ophtalmol 1906;25:481–489.
22.Rollett M, Moreau M. Le drainage au crin de la chambre anterieure contre l’hypertonie et la douleur. Rev Gen Ophtalmol 1907;26:289–292.
23.Zorab A. The reduction of tension in chronic glaucoma. Ophthalmoscope 1912;10:258–261.
24.Stefansson J. An operation for glaucoma. Am J Ophthalmol 1925;8:681–693.
25.Muldoon WE, Ripple PH, Eilder HC. Platinum implant in glaucoma surgery. Arch Ophthalmol 1951;45:666–672.
26.Row H. Operation to control glaucoma:preliminary report. Arch Ophthalmol 1934;12: 325–329.
27.Trancosco MU. Use of tantalum implants for inducing a premanent hypotony in rabbits’ eyes. Am J Ophthalmol 1949;32:499–508.
28.Gibson G. Transcleral lacrimal canaliculus transplants. Trans Am Ophthalmol Soc 1942;40:499–515.
29.Bick MW. Use of tantalum for ocular drainage. Arch Ophthalmol 1949;42:373–388.
30.Bock RH. Subconjunctival drainage of the anterior chamber by a glass seton. Am J Ophthalmol 1950;33:929.
31.Losche W. Vorschlage zur Verbesserung der Zyklodialyse. Klin Monatsbl Augenheilkd 1952;121:715–716.
32.Bietti GB. The present state of the use of plastics in eye surgery. Acta Ophthalmol (Copenh) 1955;33:337–370.
33.La Rocca V. Int. Cong. Ophthalmol (Brussels) 1958.
34.Lee PF, Wong WT. Aqueous-venous shunt for glaucoma: report on 15 cases. Ann Ophthalmol 1974;6:1083–1088.
35.Mascati NT. A new surgical approach for the control of a class of glaucomas. Int Surg 1967;47:10–15.
36.Molteno AC. New implant for drainage in glaucoma. Clinical trial Br J Ophthalmol 1969;53:606–615.
37.Molteno AC. New implant for drainage in glaucoma. Animal trial Br J Ophthalmol 1969;53:161–168.
38.Molteno AC, Straughan JL, Ancker E. Long tube implants in the management of glaucoma. S Afr Med J 1976;50:1062–1066.
39.Krupin T, Podos SM, Becker B, Newkirk JB. Valve implants in filtering surgery. Am J Ophthalmol 1976;81:232–235.
188 |
Cullen |
40.Ayyala RS, Zurakowski D, Smith JA, et al. A clinical study of the Ahmed glaucoma valve implant in advanced glaucoma. Ophthalmology 1998;105:1968–1976.
41.Molteno AC. The optimal design of drainage implants for glaucoma. Trans Ophthalmol Soc NZ 1981;33:39–41.
42.Britt MT, LaBree LD, Lloyd MA, et al. Randomized clinical trial of the 350-mm2 versus the 500-mm2 Baerveldt implant: longer term results: is bigger better? Ophthalmology 1999;106:2312–2318.
43.Smith MF, Sherwood MB, McGorray SP. Comparison of the double-plate Molteno drainage implant with the Schocket procedure. Arch Ophthalmol 1992;110:1246–1250.
44.Freedman J. Clinical experience with the Molteno dual-chamber single-plate implant. Ophthalmic Surg 1992;23:238–241.
45.Baerveldt G, Chou JS, Longren B. Comparison of the Baerveldt glaucoma implant with bioseal to the Baerveldt implant in rabbits. Invest Ophthalmol Vis Sci 1997;38:52.
46.Price FW Jr., Whitson WE. Polypropylene ligatures as a means of controlling intraocular pressure with Molteno implants. Ophthalmic Surg 1989;20:781–783.
47.Egbert PR, Lieberman MF. Internal suture occlusion of the Molteno glaucoma implant for the prevention of postoperative hypotony. Ophthalmic Surg 1989;20:53–56.
48.Latina MA. Single stage Molteno implant with combination internal occlusion and external ligature. Ophthalmic Surg 1990;21:444–446.
49.Schocket SS, Lakhanpal V, Richards RD. Anterior chamber tube shunt to an encircling band in the treatment of neovascular glaucoma. Ophthalmology 1982;89:1188–1194.
50.White TC. Clinical-Results of Glaucoma Surgery Using the White Glaucoma Pump Shunt. Ann Ophthalmol 1992;24:365–373.
51.White TC. A new implantable ocular pressure relief device. A preliminary rep Glaucoma 1985;7:289–294.
52.Hitchings RA, Joseph NH, Sherwood MB, Lattimer J, Miller M. Use of one-piece valved tube and variable surface area explant for glaucoma drainage surgery. Ophthalmology 1987; 94:1079–1084.
53.Joseph NH, Sherwood MB, Trantas G, Hitchings RA, Lattimer L. A one-piece drainage system for glaucoma surgery. Trans Ophthalmol Soc UK 1986;105(Pt 6):657–664.
54.Susanna R Jr. Modifications of the Molteno implant and implant procedure. Ophthalmic Surg 1991;22:611–613.
55.Nyska A, Glovinsky Y, Belkin M, Epstein Y. Biocompatibility of the Ex-PRESS miniature glaucoma drainage implant. J Glaucoma 2003;12:275–280.
56.Kaplan-Messas A, Traverso CE, Glovinski Y, et al. The ExPRESS miniature glaucoma implant: Intermediate results of a prospective multi-center study. Invest Ophthalmol Vis Sci 2001;42:S552.
57.Wamsley S, Moster MR, Rai S, Alvim HS, Fontanarosa J. Results of the use of the ExPRESS miniature glaucoma implant in technically challenging, advanced glaucoma cases: a clinical pilot study. Am J Ophthalmol 2004;138:1049–1051.
58.Traverso CE, De Feo F, Messas-Kaplan A, et al. Long term effect on IOP of a stainless steel glaucoma drainage implant (Ex-PRESS) in combined surgery with phacoemulsification. Br J Ophthalmol 2005;89:425–429.
59.Ayyala RS, Michelini-Norris B, Flores A, Haller E, Margo CE. Comparison of different biomaterials for glaucoma drainage devices: part 2. Arch Ophthalmol 2000;118: 1081–1084.
60.Bentley E, Nasisse MP, Glover T, Nelms S. Implantation of filtering devices in dogs with glaucoma: preliminary results in 13 eyes. Vet Comp Ophthalmol 1996;6:243–246.
61.Glover TL, Nasisse MP, Davidson MG. Effects of topically applied mitomycin-C on intraocular pressure, facility of outflow, and fibrosis after glaucoma filtration surgery in clinically normal dogs. Am J Vet Res 1995;56:936–940.
Glaucoma Drainage Devices |
189 |
62.Tahery MM, Lee DA. Review: pharmacologic control of wound healing in glaucoma filtration surgery. J Ocul Pharmacol 1989;5:155–179.
63.Pasquale LR, Dorman-Pease ME, Lutty GA, Quigley HA, Jampel HD. Immunolocalization of TGF-beta 1, TGF-beta 2, and TGF-beta 3 in the anterior segment of the human eye. Invest Ophthalmol Vis Sci 1993;34:23–30.
64.Kostick AM, Grahn BH, Murphy PH, Romanchuk KG. Comparison of aqueous humor TGF-B-2 from cataract and glaucoma patients. Invest Ophthalmol Vis Sci 1996;37:717.
65.Kostick AM, Grahn BH, Murphy PH, Romanchuk KG. Analysis of Growth-Factors in Aqueous-Humor from Glaucoma and Cataract Patients. Invest Ophthalmol Vis Sci 1995;36:S726.
66.Tripathi RC, Li J, Chan WF, Tripathi BJ. Aqueous humor in glaucomatous eyes contains an increased level of TGF-beta 2. Exp Eye Res 1994;59:723–727.
67.Grahn BH, Wilcock B, Hayes MA. Demonstration of mediators of fibrous and fibrovascular reactions in the canine, feline and equine eye. Vet Pathol 1992;25:216.
68.Cullen CL, Allen AL, Grahn BH. Anterior chamber to frontal sinus shunt for the diversion of aqueous humor: a pilot study in four normal dogs. Vet Ophthalmol 1998;1:31–39.
69.Cullen CL. Cullen frontal sinus valved glaucoma shunt: preliminary findings in dogs with primary glaucoma. Vet Ophthalmol 2004;7:311–318.
70.Cullen CL, Corcoran KA, Bartoe J, et al. Preliminary findings from the multicenter clinical study group evaluating the Cullen canine frontal sinus valved glaucoma shunt. Vet Ophthalmol 2003;6:356.
71.Dohlman CH, Grosskreutz CL, Dudenhoefer EJ, Rubin PA. Can a glaucoma shunt tube be safely extended to the lacrimal sac or the ethmoid sinus in keratoprosthesis patients? Digital J of Ophthalmol 2002;9:http://www.djo.harvard.edu/site.php?url/physicians/oa/361& PHPSESSIDba3e304919c94c495374a52f318ef18d.
72.Rubin PA, Chang E, Bernardino CR, Hatton MP, Dohlman CH. Oculoplastic technique of connecting a glaucoma valve shunt to extraorbital locations in cases of severe glaucoma. Ophthal Plast Reconstr Surg 2004;20:362–367.
73.Marcet MM, Woog JJ, Bellows AR, Mandeville JT, Maltzman JS, Khan J. Orbital complications after aqueous drainage device procedures. Ophthal Plast Reconstr Surg 2005;21: 67–69.
74.Joos KM, Lavina AM, Tawansy KA, Agarwal A. Posterior repositioning of glaucoma implants for anterior segment complications. Ophthalmology 2001;108:279–284.
75.Gedde SJ, Scott IU, Tabandeh H, et al. Late endophthalmitis associated with glaucoma drainage implants. Ophthalmology 2001;108:1323–1327.
76.Greenfield DS, Tello C, Budenz DL, Liebmann JM, Ritch R. Aqueous misdirection after glaucoma drainage device implantation. Ophthalmology 1999;106:1035–1040.
77.Tuli SS, WuDunn D, Ciulla TA, Cantor LB. Delayed suprachoroidal hemorrhage after glaucoma filtration procedures. Ophthalmology 2001;108:1808–1811.
78.Zalta AH, Sweeney CP, Zalta AK, Kaufman AH. Intracameral tissue plasminogen activator use in a large series of eyes with valved glaucoma drainage implants. Arch Ophthalmol 2002;120:1487–1493.
79.Lavina AM, Creasy JL, Tsai JC. Orbital cellulitis as a late complication of glaucoma shunt implantation. Arch Ophthalmol 2002;120:849–851.
80.Danesh-Meyer HV, Spaeth GL, Maus M. Cosmetically significant proptosis following a tube shunt procedure. Arch Ophthalmol 2002;120:846–847.
81.Oh KT, Alward WL, Kardon RH. Myositis associated with a Baerveldt glaucoma implant. Am J Ophthalmol 1999;128:375–376.
82.Nazemi PP, Chong LP, Varma R, Burnstine MA. Migration of intraocular silicone oil into the subconjunctival space and orbit through an Ahmed glaucoma valve. Am J Ophthalmol 2001;132:929–931.
190 |
Cullen |
83.Yoon PS, Singh K. Update on antifibrotic use in glaucoma surgery, including use in trabeculectomy and glaucoma drainage implants and combined cataract and glaucoma surgery. Curr Opin Ophthalmol 2004;15:141–146.
84.Trible JR, Brown DB. Occlusive ligature and standardized fenestration of a Baerveldt tube with and without antimetabolites for early postoperative intraocular pressure control. Ophthalmology 1998;105:2243–2250.
85.Jacob JT, LaCour OJ, Burgoyne CF. Slow release of the antimetabolite 5-fluorouracil (5-FU) from modified Baerveldt glaucoma drains to prolong drain function. Biomaterials 2001;22:3329–3335.
86.Gelatt KN, Brooks DE, Samuelson DA. Comparative glaucomatology. II: The experimental glaucomas. J Glaucoma 1998;7:282–294.
87.Hakanson NW. Extraorbital diversion of aqueous humour in the treatment of glaucoma in the dog: a pilot study including two recipient sites. Vet Comp Ophthalmol 1996;6:82–88, 90.
88.Bhatia LS, Chen TC. New Ahmed valve designs. Int Ophthalmol Clin 2004;44:123–138.
89.Coleman AL, Hill R, Wilson MR, et al. Initial clinical experience with the Ahmed Glaucoma Valve implant. Am J Ophthalmol 1995;120:23–31.
90.Boswell CA, Noecker RJ, Mac M, Snyder RW, Williams SK. Evaluation of an aqueous drainage glaucoma device constructed of ePTFE. J Biomed Mater Res 1999;48:591–595.
91.Pandya AD, Rich C, Eifrig DE, Hanker J, Peiffer RL. Experimental evaluation of a hydroxylapatite reservoir tube shunt in rabbits. Ophthalmic Surg Lasers 1996;27:308–314.
92.Lim KS. Cell and protein adhesion studies in glaucoma drainage device development. The AGFID project team. Br J Ophthalmol 1999;83:1168–1171.
13
The Efficacy and Safety of Glaucoma Drainage Devices
Jenn-Chyuan Wang, MB BCH BAO, MRCSED, MMED, FRCS (OPHTH), FAMS and Paul Chew, MBBS, FRCOPHTH, FRCSED, FAMS
CONTENTS
INTRODUCTION
EFFICACY IN IOP CONTROL
SAFETY
FUTURE
REFERENCES
INTRODUCTION
The glaucoma drainage devices (GDD) are increasingly popular as the primary procedure over trabeculectomy and cyclo-destructive in the management of refractory glaucoma. Interest in this field is warranted because studies have found significant rates of desirable intraocular pressure (IOP) control as well as maintenance of visual acuity in these difficult cases. GDD have evolved from the first crude attempt, in 1907, using horsehair to drain aqueous from the anterior chamber into the subtenon/conjunctival space at the limbus (1). Since then, various materials, such as silk, gold, platinum, tantalum, glass rod, and polythene tube have been used near the limbus and have failed because of conjunctival erosion, implant migration, extrusion as well as scarring. To overcome these problems, current implants were based on the Molteno implant, which has a long tube attached to an explant placed 9–10 mm posterior to the limbus. Current GDD are made from various biomaterials, such as acrylic, silicone, and polypropylene, however, their influence on efficacy and safety has not yet been determined.
Significant progresses have been made to the design of GDD. Postoperative hypotony, flat anterior chambers, and choroidal effusion were very common observations in nonvalved implants. These complications resulted in the introduction of several modifications, such as temporary ligature of the tube. An important advancement was the addition of unidirectional valves with fixed opening IOP to the design in order to increase the resistance to aqueous outflow. Another concern was to lower IOP to a level that reduces the risk of glaucoma progression. In order to achieve this, Molteno
From: Ophthalmology Research: Visual Prosthesis and Ophthalmic Devices: New Hope in Sight
Edited by: J. Tombran-Tink, C. Barnstable, and J. F. Rizzo © Humana Press Inc., Totowa, NJ
191
192 |
|
|
Wang and Chew |
Table 1 |
|
|
|
Types of Implants |
|
|
|
|
|
|
|
Implant |
Valve |
Material |
Surface area/mm2 |
Ahmed |
Yes |
Polypropylene |
96/184/364 |
|
|
Silicone |
|
Baerveldt |
No |
Barium-impregnated |
250/350/425/500 |
|
|
silicone plate |
|
Molteno |
No |
Polypropylene |
270 |
|
|
|
|
Table 2
Indications for Glaucoma Drainage Device Implantation
Neovascular glaucoma Failed trabeculectomy
Glaucoma following Retinal detachment surgery Glaucoma following previous ocular surgery with extensive
conjunctival scarring PKP with glaucoma Congenital glaucoma Uveitic glaucoma Traumatic glaucoma Silicone oil glaucoma
Irido-corneal endothelial syndrome Sturge–Weber syndrome
Glaucoma with chronic topical medication Aphakic glaucoma
increased the surface area of his GDD by introducing the double plate implant in 1981. Surgeons are now faced with many choices of both valved and nonvalved implants (Table 1) in the management of refractory glaucoma (Table 2). The following chapter would attempt to summarize the current knowledge on the efficacy and safety of these implants.
EFFICACY IN IOP CONTROL
In the literature, GDD surgical success is usually defined as IOP less than 22 mmHg, but more than 4 mmHg and these eyes should not become totally blind or undergo phthisis. However, comparisons between studies are difficult because they differ in terms of surgical indications, surgical techniques, postoperative management, criteria for defining success as well as duration of follow-up. Furthermore, the numbers are usually small and lacked the power to elucidate risk factors for failures. The success rates were impressive in most reports considering the fact that GDD are usually implanted in refractory glaucoma where conventional surgery would otherwise have failed. They were quoted in
The Efficacy and Safety of GDD |
193 |
the region of 58–98% for the Caucasian population (2–5). Long-term follow-up of Molteno tubes over a median of 44 mo had success rates of 57% (6).
Similar short-term results were echoed in Asian patients. Aung et al. (7) reported success in 73.5% and qualified success in 12% for Molteno and Baerveldt tubes in East Asians with average follow-up of 13.41 mo. While Lai et al. (8) had 73.8% success at 12 mo for Ahmed tubes in Hong Kong Chinese patients. Success rates of 82.83% at 24 mo were seen in Indian patients (9). Complete success experienced for Baerveldt was 83.3% and 66.7% for Ahmed implants and qualified success were 4.2 and 16.7%, respectively with a mean follow-up of 22.8 mo (10). As for the intermediate term, Seah et al. (11) had 54% complete success and 22% qualified success for Baerveldt GDD. Future randomized comparative trials with long-term follow-up may shed light on the impact of race on GDD outcomes.
Plate Area
Historically, it has been assumed that the plate size determined the size of the encapsulated bleb and therefore, has a positive influence on surgical success (12). With twice the surface area, the double-plate Molteno device achieved higher success rates and lower IOP than single-plate Molteno GDD (13). Although, most surgeons considered these findings as proof of theory, Camras (14) pointed to the fact that the success rates and IOP reduction had not been twice as good. Further evidence from Smith et al. (15) and Wilson et al. (16) showed that the smaller double-plate Molteno produced similar or lower IOP when compared with the Schocket encircling device. Studies comparing similar devices, i.e., the Baerveldt 250 mm2 with 350 mm2 in Asians (11) and 350 mm2 with 500 mm2 in Western population did not find significant differences in final IOP control (17). Evidently, IOP decrease is not proportional to the plate area used as the effective surface area may be quite different from the total surface area of the GDD endplate. For example, a significant proportion of the Baerveldt endplate is tucked under both the vertical and the horizontal recti muscles and the extension of bleb under these muscles has been demonstrated echogenically (18). However, the plate-muscle adhesions and pseudocapsule here may not be as permeable to aqueous, therefore, the resultant effective surface area may be significantly less.
Molteno described the best plate-size for the development of optimal bleb surface area. He mentioned that the circumferential distance between recti muscle delimits the size and shape of the episcleral plate. Plates larger than an ocular quadrant with the involvement of recti muscles may not necessarily increase functioning encapsulation. The posterior ciliary artery limits the posterior dimension. Finally, encapsulation within the orbit as opposed to being near the conjunctiva reduces permeability to aqueous humor (19).
Second Tube Implant
In the event of suboptimal IOP control or failure of surgery, Management options include shunt revision, additional tube implant, and supplemental cyclophotocoagulation. Burgoyne et al. examined a group of patients with second tube implant and found that corneal morbidity is a common complication, whereas other rates of complications were not higher than expected. Pressure control was achieved with fewer hypotensive agents
194 |
Wang and Chew |
and stable visual acuity could be maintained (20). Shah et al. compared additional tube shunt with shunt revision and found more patients (62% vs 42%) achieved at least qualified success (no statistical significance). Moreover, the sample size was very small with overrepresentation of neovascular glaucoma in the shunt revision group (21).
Semchyshyn et al. treated a small group of tube failures with supplemental trans- scleral diode laser cyclophotocoagulation. Twenty-one eyes were treated and followed up during a mean period of 26.9 mo. The average IOP was significantly reduced from a preoperative level of 35.7 +/− 14.7 (SD) mmHg to a postoperative level of 13.6 +/− 7.1 (SD) mmHg. Therefore, in cases of tube failure, adjunctive trans-scleral diode cyclophotocoagulation treatment(s) is a viable option to lower IOP (22).
Comparison With Other Surgical Modalities
GDD do not offer any advantage in terms of improving IOP control or reducing complication rates when compared with trabeculectomy alone in uncomplicated glaucoma (23). However, for the Asian patients who have high risk of trabeculectomy failure (24), GDD may have more to offer as the primary procedure (7,10). Pars plana Baerveldt GDD surgery is superior to neodymium:YAG cyclophotocoagulation in terms of IOP control, less hypotony and with higher preservation of visual acuity. As compared with 5.6% in the Baerveldt group 26.3% of Nd:YAG failed at 6 mo. The neodymium:yttrium aluminum garnet (Nd:YAG) group lost light perception in 23.3% vs 5.6% in the Baerveldt group (25).
When Lima et al. compared endoscopic cyclophotocoagulation (ECP) with Ahmed GDD in the treatment of refractory glaucoma, efficacy was similar, but there were more complications with Ahmed GDD, specifically, choroidal detachment (Ahmed 17.64%, ECP 2.94%), shallow anterior chamber (Ahmed 17.64%, ECP 0.0%), and hyphema (Ahmed 14.7%, ECP 17.64%) (26). As such, ECP may be promising as a treatment modality for refractory glaucoma.
Augmentation With Mitomycin C
Antimitotic agents have not been used in any of the patients mainly because of the lack of evidence for their efficacy. The issue of using antimitotic agents to improve the outcome of GDD remained unresolved as the few comparative studies produced contradicting results. Kook et al. (27) reported increased success rates with Ahmed GDI without increasing complication rates whereas Costa et al. (28) did not observe any significant improvement in the IOP lowering effect. Meanwhile Irak et al. (29) reported no differences with the Baerveldt GDD in a noncomparative study. Likewise, Molteno GDD did not show significant differences in terms of survival rates, IOP control, and antiglaucoma medication use (30). Conversely, complications related to mitomycin C use, such as conjunctival melts (31), infections, and hypotony have been reported (32,33).
Combined Cataract and GDD Surgery
Combined phacoemulsification and lens implantation with GDD surgery has been met with good outcomes. Chung et al. (34) reported a series of 32 patients who underwent such surgery with both Ahmed and Baerveldt GDD, and they benefited from