Ординатура / Офтальмология / Английские материалы / Textbook of Visual Science and Clinical Optometry_Bhattacharya_2009
.pdf
C H A P T E R
16 Ophthalmic
Instrumentation
INTRODUCTION
This chapter highlights the basic principles, examination techniques of common instruments used in ophthalmology. The instruments are:
•Slit-lamp biomicroscope and vital stains
•Tonometry
•Gonioscopy
•Indirect biomicroscopy
•Geneva lens measure
•Keratometer
•Lensometer
•Direct ophthalmoscope.
SLIT-LAMP BIOMICROSCOPE
It offers a magnified, stereoscopic, noninvasive and detailed view of the anterior segment of the eye. However, in conjunction with some accessory optical lenses, it can be used for detailed examination of the angle of the anterior chamber (gonioscopy), measurement of intraocular pressure (applanation tonometry) and minute examination of the retina (biomicroscopic indirect ophthalmoscopy). Different types of lasers (YAG, Frequency– doubled YAG, Argon, Diode, etc.) are also delivered through slitlamp for therapeutic purpose. Vital staining of cornea and conjunctiva is also observed through slit-lamp biomicroscope.
Slit-lamp examination is done in a room semidarkened or darkened, free from dust, humidity and heat. The patient should
232 Textbook of Visual Science and Clinical Optometry
be comfortably seated in a revolving stool with his chin and forehead pressed against chinrest and head rest respectively. Mastering the art of slit-lamp examination is a prerequisite for a good ophthalmologist and an efficient optometrist.
SALIENT TECHNICAL INFORMATIONS
•Nomenclature–Since the microscope has two eyepieces, it is binocular and offers the observer a binocular, stereoscopic and magnified view of the eye. Further, the light beam can be made very narrow producing a slit of light. Hence, it is called slit-lamp
biomicroscope (Fig. 16-1).
•Magnification–Magnification of image is obtained by multiplying the power of the oculars with the power of the objective lens. Thus, if the oculars are of 10X strength and the objective lens is of 2.5X, the resultant magnification will be 25X. Majority of slit-lamp biomicroscope are usually available with magnifications between 8X and 40X with intermediate magnifications of 12X, 16X and 25X. Magnification is changed by either changing the objective lens or by replacing the oculars, i.e. eyepieces. Objective lens is changed by either turning a dial on the side of the observation system or flipping a lever located below the observation system.
•Focussing–Focussing is controlled by a joystick. It is designed to move the slit-lamp biomicroscope laterally, axially and vertically. Vertical movement is controlled by rotating the joystick.
•Intensity of light–The illumination is controlled either by a rheostat or by rotating a knob. The height, diameter of the light beam can be altered by another knob. The width of the light beam is also controlled by another knob, located in the lower end of the illumination system.
•Click-stop position–A “click-stop” position indicates that the illumination system and the observation system are coaxial.
•The illumination system or light source can be moved away from the observation system, to change the angle of the incident light beam which is essential for various types of illumination techniques, discussed later in this chapter.
Ophthalmic Instrumentation 233
•A frosted ground glass (if present) can be positioned in the path of the incident light by flipping a lever in the diffuse illumination technique.
•Halogen light is used in most of the examinations. Coloured filters (cobalt blue, red free, i.e. green) can be used by sliding a knob for procedures discussed later.
Fig. 16-1: Slit-lamp AIA–11 5 step, inset showing observation system of Slitlamp AIA–11 2 step (photo courtesy Appasamy Associates, India)
DIFFERENT TECHNIQUES OF ILLUMINATION
Diffuse Illumination (wide beam)
•The diffuse illumination is provided by placing a frosted or ground glass filter in front of the focussed light beam of slitlamp biomicroscope (Fig. 16-2).
•Usually, in diffuse illumination the incident light is projected intotheeyeobliquely(45°)andviewedunderlowmagnification.
•The light will be defocussed and will provide even illumination over the entire field of view of opaque tissues of the anterior segment (sclera, iris, eyelid margins, bulbar conjunctiva, etc).
234 Textbook of Visual Science and Clinical Optometry
Fig. 16-2: Slit-lamp biomicroscopy—Diffuse illumination
Direct Illumination
This technique refers to viewing of structures within the focussed light beam. The type of direct illumination varies according to height, width of the light beam, angle of the incident light and angle between incident light beam and observation system.
i. Optical section
•The beam is very narrow and slit.
•The light beam is projected obliquely. The angle between oculars and illumination source is 30 to 60°. The more is the angle, the more wide is the optical section (parallelepiped).
•It gives a cross section view of different layers of the cornea (Fig. 16-3).
•In dilated pupil, it is possible to see the lens and anterior third of the vitreous in optical section with light beam being brightest in an absolutely dark room (Fig. 16-3).
•It is used to locate the site of a corneal lesion, scar, foreign body and depth of anterior chamber (Van Herick-Shaffer technique, see later in this chapter).
Ophthalmic Instrumentation 235
Fig. 16-3: Slit-lamp biomicroscopy—Optical section of ocular structures. Inset showing different layers of the cornea (schematic). 1 = Tear film–bright, 2 = Epithelium–dark, 3 = Bowman’s membrane–bright, 4 = Stroma–wide zone,
5 = Descemet’s membrane–bright, 6 = Endothelium–dark, 7 = Anterior chamber, 8 = Lens and 9 = Vitreous humour
ii. Conical/pinpoint illumination
•The light beam is very narrow and short and the light beam is focussed into the aqueous against dark pupillary background to enhance the contrast. The room should be absolutely dark.
•It is used to detect “aqueous flare” and “cells” in the anterior chamber in iritis.
•The optical section may be similarly used to detect flare and cells in anterior chamber. Aqueous flare appears as yellowish particles and cells in anterior chamber appear as whitish reflections.
•Cells in the tear meniscus may be similarly observed against background of iris by pushing the lower lid up.
iii. Specular Reflection
•The angle of the illumination source must be set to equal the angle of the oculars (Fig. 16-4).
•It is used to visualise corneal endothelium.
•The illumination beam is parallelepiped, i.e. a wider optical section (corneal width).
236 Textbook of Visual Science and Clinical Optometry
•First start with lower magnification and focus the corneal surface. Now move the oculars 20º to 30° away from the illumination source. Then, move the illumination source 20° to 30° on the opposite direction until a bright mirror-like reflection is seen through one ocular, i.e. eyepiece. Specular image is seen at every interface between structures of different refractive index. Hence, there should be three reflexes (Fig. 16-4);
Specular reflection
bright reflection |
dim reflection |
blurred image |
from tearfilm |
from endothelium |
of light source |
•Focus finely on the intermediate dim reflection to observe endothelial mosaic. Endothelium is best seen using only one ocular.
•Now the magnification may be increased to view the endothelial cells.
Fig. 16-4: Slit-lamp technique—Specular reflection, inset shows specular reflections withtheendotheliuminfocus(schematic). 1=Tearfilm–brightreflection,2=Endothelial mosaic–dim reflection and 3 = Light source–blurred image
iv. Tangential or oblique illumination
•In this technique, the angle between the observation system and illumination source is set at 90° (Fig. 16-5).
Ophthalmic Instrumentation 237
•It produces long shadows for any elevation particularly on the iris surface. Even depressions are better appreciated.
Fig. 16-5: Slit-lamp technique—Tangential illumination
Indirect Illumination
In this technique the area of interest is viewed indirectly by an illumination from reflected light of the direct beam, i.e. structures
not within the focussed light beam are under observation.
i. Proximal illumination
•A moderately wide light beam is focussed on the areas adjacent to the lesion of interest (Fig. 16-6).
•The lesion is observed with scattered light against dark background. This results in a higher contrast of the lesion against a dark pupil.
ii. Sclerotic scatter
•A parallelepiped illumination beam is focussed at the temporal limbus.
•Illumination beam is set at an angle between 45° and 60° with the observation system, i.e. the microscope is focussed centrally on the cornea.
238 Textbook of Visual Science and Clinical Optometry
Fig. 16-6: Slit-lamp technique—Proximal illumination
•Theslit-lamplightbeamundergoestotalinternalreflectionwith- in the corneal substance and reaches opposite limbus (Fig. 16-7).
•This produces a bright circumcorneal glow with the cornea being dark.
•A scar or lesion within the cornea will be visible by scattering of light against dark papillary background.
Fig. 16-7: Slit-lamp technique—Sclerotic scatter
Ophthalmic Instrumentation 239
Retroillumination
•In this technique, the incident light beam is reflected from the iris, anterior lens surface or retina to illuminate more anteriorly located structures of interest.
i. Direct retroillumination
•This refers to viewing the structure against an illuminated background.
•In direct retroillumination corneal opacities appear black against an illuminated background. Posterior capsular opacities or lental opacities appear dark against red glow of retina.
•The light beam should be directed from 45° angle and focussed behind the area of interest (Fig.16-8).
•In direct retroillumination objects that normally appear bright will appear dark.
Fig.16-8: Slit-lamp technique—Direct retroillumination
ii. Indirect retroillumination
•This refers to viewing the structures of interest against a dark background.
•The light beam is positioned in such way that the dark background is behind the structure of interest (Fig. 16-9).
240 Textbook of Visual Science and Clinical Optometry
•The structure should not be within the pathway of the reflected light.
Fig.16-9: Slit-lamp technique—Indirect retroillumination
• = Corneal opacity
iii. Transillumination
•In transillumination the iris tissue is tested for passage of light through a defect within it.
•The slit-lamp light source and the microscope is positioned coaxially, i.e. click-stop position (Fig. 16-10) and focussed on the iris surface.
•Use a circular light beam equal to pupillary size and project through the pupil.
•Use lower magnification.
USE OF FILTERS IN SLIT-LAMP BIOMICROSCOPE
a.Cobalt blue filter: It is used in applanation tonometry, assessment of rigid contact lens fitting and fluorescein staining of cornea. It should be kept in mind that significant amount of light has been filtered out by the cobalt blue filter. Hence, brighter illumination should be employed while using cobalt blue filter.
