Ординатура / Офтальмология / Английские материалы / Practical Ophthalmology A Manual for the Beginning Ophthalmology Residents 4th edition_Wilson_1996
.pdf46 Chapter 4: Visual Acuity Examination
Table 4.2 Visual Impairment and Estimates of Visual Disability
Visual |
Visual |
Comment |
Impairment |
Disability |
|
20/12 to 20/25 |
Normal vision at |
Healthy young adults average |
|
normal reading |
better than 20/20. |
|
distance |
|
20/30 to 20/70 |
Nearly normal |
Usually causes no serious |
|
vision; normal |
problems, but vision should be |
|
reading perfor- |
checked for potential improve- |
|
mance using |
ment or possible early disease. |
|
shorter working |
Most states will issue a driving |
|
distance |
license to individuals with this |
|
|
level of vision in at least one eye. |
20/80 to 20/160 |
Moderate low |
Strong reading glasses or vision |
|
vision; (near) nor- |
magnifiers usually provide ade- |
|
mal performance |
quate reading ability; this level is |
|
with magnifiers |
usually insufficient for a driving |
|
|
license. |
20/200 to 20/400 Severe low vision: legal blindness by US definition
Gross orientation and mobility generally adequate, but difficulty with traffic signs, bus numbers, etc. Reading requires high-power magnifiers; reading speed is reduced, even with reading aids.
20/500 to 20/1000 Profound visual |
Limited spot reading with visual |
impairment |
aids. |
CF 8 ft to 4 ft |
Unreliable vision |
Increasing problems with visual |
|
|
orientation and mobility. Long |
|
|
cane is useful to explore environ- |
|
|
ment. Talking devices and vision |
|
|
substitutes are useful. |
Less than CF |
Nearly total |
Vision unreliable, except under |
4 ft |
blindness |
ideal circumstances; must rely on |
|
|
nonvisual devices. |
NLP |
Total blindness |
No light perception; must rely |
|
|
on talking devices and vision |
|
|
substitutes. |
Reading Distance:
Reading Aids
>33 cm: Regular bifocals
(up to 3D)
33-20 cm:
Stronger bifocals (3-5 D)
Low-power magnifiers (5D)
16-10 cm:
Half-eye glasses (6-10 D),with prisms for binocularity
Stronger magnifiers (>8D)
8-5 cm (cannot be binocular): Highpower reading lenses (12-20 D) High-power magni-
fiers (> 16 D)
4-2 cm (cannot be binocular): Highpower reading lenses (24-28 D) High-power magni-
fiers (>28 D) Video magnifier Talking devices and
vision substitutes
•V'-itfi!:-:t'"-: Other lests oi Sensory Visual Function 47
when a patient is 8 to 10 years of age, recent evidence indicates that some improvement in vision can be achieved until even 12 or 13 years of age. Therefore, a trial of amblyopia therapy should be attempted in an older child after a thorough discussion with the parents and the
child of the risks and benefits of such treatment. - • • • , • ; • > •
Other Tests of Sensory Visual Function
Sensory visual functions include the ability to use both eyes together, referred to as fusion or single binocular vision. Single binocular vision y» ^ ;v':*i-1 results from simultaneous stimulation of corresponding retinal elements that have the same visual direction. For example, an object to a person's left may stimulate a spot on the temporal retina of the right eye and a corresponding spot on the nasal retina of the left eye. The brain then perceives the object being viewed as a single image in space.
Tests for binocular vision necessarily rely on both eyes being stimulated at the same time. These tests are discussed in detail in most orthoptic or ocular motility textbooks.
Stereopsis, considered a type of fusion, is the perception of depth or three-dimensionality when noncorresponding retinal elements are stimulated at the same time. To achieve stereopsis, both eyes must be used simultaneously. The tests used for stereopsis measurements consist of either polarized images or random-dot stereograms that may be nonpolarized or polarized. Figure 4.5 illustrates the Titmus (fly) test, which is probably the most common test for near stereopsis. The procedure for the Titmus stereoacuity test is summarized in Clinical Protocol 4.7. Random-dot stereogram tests also are used to test near
Figure 4.5 The Titmus stereopsis test. (A) Light-polarizing eyeglasses and targets. (B) The patient views the target through polarizing filters and reports perception of depth.
'48 |
|
Chapter 4: Visual Acuity Examination |
|
|
|
|
stereoacuity. The BVAT system and vectographic polarized projection |
|
|
|
slides can be used for distance stereopsis testing. Stereoscopic acuity is |
|
|
|
expressed in seconds of arc, but there are no standards for stereoacuity |
|
|
|
tests that are comparable the way that Snellen acuity charts are. |
|
|
|
Entopticphenomenon is the visualization of images of one's internal reti- |
' |
> |
|
nal blood vessels and corpuscles after a light has been applied. Although |
' |
• |
' |
- • die presence of these images implies that the retina is functioning, it is |
' |
•. |
|
not a measure of acuity and is rarely used to characterize vision. "Eliciting |
• '.' |
'• |
entoptic phenomenon is particularly helpful in determining a reasonable |
|
|
|
|
level of function behind very opaque media. |
Contrast Sensitivity, Glare, and Color Vision Tests
Contrast sensitivity refers to the ability to discern relative darkness and brightness and the ability to see details, edges, or borders of images. Contrast sensitivity can be impaired even in the presence of excellent Snellen acuity. Alterations in contrast sensitivity imply abnormalities in the anterior visual receptive systems, from the tear film to the optic nerve. Specific patterns of contrast sensitivity function alterations are discussed in more advanced textbooks.
In the simplest contrast sensitivity tests, patients are shown printed charts with contrasting lines, referred to as gratings, which are presented in varying orientations. The difference in the intensity between the background of the chart and the printed lines is gradually decreased, and the patient is asked to identify the direction of the lines. The endpoint is reached when the patient can no longer correctly identify either the presence of any lines or the direction of their orientation. More technical testing methods involve presenting grating patterns or letters on an oscilloscope screen. The endpoints and reporting methods are similar in the two techniques.
Glare occurs when light from a single bright source scatters across the visual field, dazzling sight and reducing the quality of the visual image. The perception of troublesome glare, which causes distorted vision and, in some cases, mild pain, is also a symptom of potential cataract. As with contrast sensitivity testing, glare testing (by exposing the patient to bright light under controlled circumstances) can reveal the presence of cataract or other opacity
The most commonly recognized color vision abnormalities are the sex-linked congenital red-green deficiencies or confusions, but, many other color vision anomalies exist. Optic nerve or retinal disease may result in color vision defects that may be acquired or asymmetric. Most
•' patients with inherited color vision defects see red as less bright than
6'* |
nonaffected individuals and, according to a standard established by test- |
Pitfalls and Pointers |
49 |
ing nonaffected individuals, they fail to identify color mixtures that include red. Although not disabling, color vision anomalies can impair performance in some careers or activities, although often, as in some computer graphics applications, colors can be altered to accommodate those who have difficulty discriminating certain shades.
Evaluation of color vision is often performed with a book that displays circles in multicolored patterns, called pseudoisochromatic color plates. Patients with normal color vision easily detect specific numbers and figures composed of and embedded in the dot pattern, but patients with impaired color vision do not detect the same numbers. Various combinations of colors are used to identify the nature of the color vision deficit.
Another test of color vision, the 15-hue test (Farnsworth-Munsell D-15 test), consists of 15 pastel-colored chips, which the patient must arrange in a related color sequence. The sequence is obvious to patients with normal color vision, but patients with color deficits arrange the chips differently.
Principles and performance of contrast sensitivity, glare, and color vision testing are covered more thoroughly in other, more detailed textbooks and in manufacturer's instructions that accompany testing materials.
Pitfalls and Pointers
Be aware of the variables possible in visual acuity measurement. Particularly note that all lenses, projectors, and charts are clean.
Avoid glare on the viewing chart or screen.
Avoid glare in patient's eyes from overhead lights or outside windows.
Learn appropriate ways of interacting with patients who have lowvision or are blind. Do not shout at the patient; a visual impairment does not imply an auditory one. Warn the patient of your movement, particularly if there is severe visual limitation. Offer an arm to the patient, but do not attempt to grab his or her hand or arm.
Ensure that you are using the appropriate tests for the patient's abilities.
Refrain from using the term blindness when counseling parents of children with severe visual impairment.
50 Chapter 4: Visual Acuity Examination
Suggested Resources
Amblyopia [Preferred Practice Pattern]. San Francisco: American Academy of Ophthalmology; 1992.
Frisen L. Clinical Tests of Vision. New York: Raven Press; 1990.
Optics, Refraction, and Contact Lenses. Basic and Clinical Science
Course, Section 3. San Francisco: American Academy of
Ophthalmology; updated annually.
Pediatric Ophthalmology and Strabismus. Basic and Clinical Science Course, Section 6. San Francisco: American Academy of - Ophthalmology; updated annually.
Rehabilitation: The Management of Adult Patients With Low Vision
, [Preferred Practice Pattern]. San Francisco: American Academy of Ophthalmology; 1994.
Testing Distance Visual Acuity
1. Ask the patient to stand or sit at a designated testing distance (20 feet
'.; |
from a well-illuminated wall chart is ideal). If a projected chart is used, |
' |
distance may vary; the projected optotype size must be focused and |
: adjusted to be equivalent to the corresponding Snellen acuity for the distance used. Most clinics are set up to test vision with projected charts at predetermined distances.
2.Occlude the left eye. Ask the patient or check for yourself to be sure that the occluder is not touching or pressing against the eye. Observe the patient during the test to make sure there is no conscious or inadvertent peeking.
3.Ask the patient to say aloud each letter or number, or name the picture
object on the lines of successively smaller optotypes, from left to right or, alternatively, as you point to each character in any order, until the patient correctly identifies only half the optotypes on a line. •;,•.
4.Note the corresponding acuity measurement shown at that line of the chart. Record the acuity value for each eye separately, with correction and without correction, as illustrated below. If the patient misses half or
fewer than half the letters on the smallest readable line, record how i& many letters were missed; for example, 20/40"2. If acuity is less than
20/20, recheck with a 2.4 mm pinhole (see Clinical Protocol 4.2).
5.Repeat steps 1 through 4 for the left eye, with die right eye covered.
6.Retest acuity with the patient using both eyes simultaneously and record acuity OU (see example below).
DV sc |
OD |
20/200 |
OU |
20/80 |
OS |
20/100 |
|||
1) V cc |
OD |
20/20 |
OU |
20/20 |
OS |
20/25 |
7. Record the power of die corrective lenses worn for the distance acuity determination (see Clinical Protocol 5.1, "Performing Manual Lensometry").
Clinical Protocol 4.2
. 1 ' •
•i •
Testing Pinhole Visual Acuity
1.Position the patient and occlude die eye not being tested, as done for the distance acuity test.
2. Ask the patient to hold the pinhole occluder in front of the eye that is to be tested. The patient's habitual correction should be worn for the test.
3.Instruct the patient to look at the distance chart through the single pinhole or through any one of the multiple pinholes.
4.Instruct the patient to use small hand or eye movements to align the pinhole to resolve the sharpest image on the chart.
5.Ask the patient to begin to read the line with the smallest letters that are legible as determined on the previous vision test without the use of the pinhole.
6.Record the Snellen acuity obtained and precede or follow it with the abbreviation PH.
51
J t
Clinical Protocol 4.3
Testing Near Visual Acuity
1.With the patient wearing the habitual corrective lens for near and the near card evenly illuminated, instruct the patient to hold the test card at the distance specified on the card.
2. |
Ask the patient to occlude the left eye. |
-r |
3. |
Ask the patient to say each letter or read each word on the line of small- |
|
|
est characters that are legible on the card. |
|
4. Record the acuity value for each eye separately in the patient's chart
vaccording to the notation method accepted (see example below).
5.Repeat the procedure with the right eye occluded and the left eye viewing the test chart.
6.Repeat the procedure with both eyes viewing the test card.
7.Record the binocular acuity achieved (see example below).
OD |
20/200 |
|
|
AJs c o s |
20/100 |
o u |
2 0 / 8 0 |
OD |
20/20 |
|
|
X V cc OS |
20/25 |
OU |
20/20 |
Clinical Protocol 4.4
Testing for NPA and NPC
Testing for Near Point of Accommodation (NPA)
1.With the patient wearing full distance correction and with the left eve occluded, place the near testing card at a distance of 16 inches (40 cm) from the patient and ask the patient to read the 20/40 line with the unoccluded eye.
2. |
Move the test card slowly forward as you ask the patient to state when |
|
. the letters have become blurred. |
3. |
Record this distance in centimeters or inches (whichever you are using). |
52
4.Move the test card slowly away from the patient as you ask the patient to state when the letters become clear again.
5.Record this distance in centimeters or inches.
6. Repeat steps 1-5 for the other eye and record as described above.
7.The NPA is that point (in centimeters or inches) where the patient can no longer bring the image into clear focus.
Testing for Near Point of Convergence (NPC)
1.With the patient wearing appropriate correction and with neither eye occluded, hold a target, such as a pencil tip, at a distance of approximately 16 inches (40 cm) from the patient and ask the patient to fixate on it.
2.Move the object slowly forward and ask the patient to tell you when the object doubles.
3.Observe whether both eyes are converging.
4.Record the distance (in centimeters or inches) at which the image doubles or one eye deviates. The NPC is that point (in centimeters or inches) where a single image can no longer be maintained.
Clinical Protocol 4.5
Testing Acuity for Patients With
Low Vision
1.If the patient is unable to resolve the largest optotype on the distance acuity chart from the standard testing distance, ask the patient to stand or sit 10 feet from the well-illuminated test chart. A projected chart is less desirable to use in this situation than a printed wall chart. A lowvision test chart, if available, is preferable for these patients.
2.Occlude the eye not being tested.
3.Repeatedly halve the testing distance (up to 2.5 feet) and retest the distance visual acuity at each stage until the patient successfully identifies half the optotypes on a line.
continued
53
4.Note the corresponding acuity measurement shown at that line of the chart. Record the acuity value for each eye separately, with correction
and without correction, as would be done for standard distance acuity testing, recording the distance at which the patient successfully reads the chart as the numerator of the Snellen acuity designation (eg, 5/80).
5. If the patient is unable to resolve the largest optotypes on the chart from a distance of 3 feet, display two or more fingers of one hand and ask the patient to count the number of fingers displayed. Record the distance at which counting is done accurately; for example, CF at 2 ft.
6.If the patient cannot count fingers, move your hand horizontally or vertically before the patient at a distance of approximately 2 feet. Record the distance at which the patient reported seeing your hand movement; for example, HM at 2 ft.
7.If the patient cannot detect your hand motion, shine a penlight toward the patient's face from approximately 1 foot and turn it on and off to determine if light perception is present. If the patient cannot see the light, dim the room lights and shine a bright light toward the patient's eye again. If the patient cannot see even the brightest light, record the response as NLP (no light perception). If the patient can see the light, record the response as LP (light perception). No record of distance is required.
8. If light is perceived from straight ahead, move the light sequentially into
. each of the four quadrants of the visual field. Turn the penlight on and off in each field, and ask if the patient can see the light. If the light can be seen, colored filters can be placed in front of the light and the patient asked to identify the color of the light.
9.If the patient correctly identifies the direction from which the light is coming, record the response as LP with projection. Specify the quadrant(s) in which light projection is present. State whether or not color perception is present. Color perception is usually tested only in the straightahead position. If the patient is unable to properly identify any direction but is able to discern light in the straight-ahead position, record the response as LP without projection.
10.Repeat steps 1-9 tor the fellow eve, as appropriate.
i |
|
|
Clinical Protocol 4.6 |
|
|
||
|
|
Testing an Infant's Visual Ability |
|
: |
• |
|
|
i |
• |
|
|
i
*1. Seat the infant on an adult's lap, so that the infant is comfortable.
2.Select a small toy or other attention-attracting object that stimulates sight only; do not use a sound-producing object. Hold the object about
1to 2 feet from the infant's face and move it horizontally to either side.
3.Watch die infant's eyes for fixation and following movements. Infants will respond with better following movements for objects moved from temporal to nasal field; this preference decreases after an infant is approximately 6 months old.
4. If no visual response is noted, repeat the test in a darkened room using a small, bright penlight and watch the infant's fixation and following movements to the penlight. If the light is too bright, the infant will
. ••" avoid it and look away.
'":
• • • •
Clinical Protocol 4.7
i- |
Testing for Stereoacuity |
|
|
1. |
Check the polarization of the glasses being used to ensure that each eye |
- ? |
sees a different image; the left eye should see the L on the lower left of |
- • |
|
'•• |
the fly, and the right eye should see only the R. |
2.Place the polarizing glasses on the patient. If glasses are usually worn, place the polarizing glasses over them.
•
J.Hold the fly image facing the patient approximately 16 inches (40 cm) away, with the surface of the page parallel to the surface of the glasses.
4.Ask the patient to touch the wings or to pick up the flyReassure young children who may respond witli surprise or fright if the image appears too real to them. If the image is perceived as having height, the patient will attempt to touch the wings as if they were above the page surface when instructed by the examiner.
5.If the fly test is positive, show the patient the three rows of animal figures. Ask which figure in each row is coming forward, is above the page, or can be picked up. You can vary these directions as you become more familiar with the test and with judging patients' responses.
continued
55
