Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Офтальмология / Английские материалы / Handbook of Optical Coherence Tomography_Bouma, Tearney_2002

.pdf
Скачиваний:
0
Добавлен:
28.03.2026
Размер:
37.86 Mб
Скачать

Reference Optical Delay Scanning

111

then given by Eqs. (9) and (10). The major drawback of this OPL is the nonlinearity of the scan. The scan velocity changes through the course of the scan, so the interferogram center frequency and bandwidth also change. This ODL configuration was demonstrated by rotating a 1.27 cm cube at 22 revolutions per second (rev/s), for a scan repetition rate of 88 Hz. Through the course of the 1.5 mm scan, the center frequency varied from 0 to 1000 kHz, corresponding to a varying scan velocity of 0–650 mm/s. The duty cycle was almost 50%, but if double-sided scanning were used the duty cycle would be almost 100%.

Single Pass, Two Internal Reflections

Figure 9 illustrates a configuration in which the reference beam is retroreflected inside the cube. Mirrors are deposited on portions of each face of the cube as illustrated in the inset of Fig. 10. This configuration was used to demonstrate the highest speed scanner yet implemented for OLCR. A 5 5 2 mm3 cube was turned at 7117 rev/s by an air turbine, resulting in a scan repetition rate of 28.5 kHz [13]. A pathlength variation of 2 mm was swept at 176 m/s. The scanning duty cycle was approximately 32%. The scan linearity and dispersion effects were not reported but should be comparable to those of the configuration described in the following subsection. The function describing pathlength scan as a function of the angular position of the cube given in the next section should be valid for this configuration if divided by a factor of 2. This ODL required custom deposition of mirrors on portions of each cube face and extremely precise mounting and alignment of the cube on the turbine shaft. At this high repetition rate, mechanically resonant instabilities resulted in cyclical variation in recoupling efficiency over the time scale of several scans. The OLCR system using this ultrahigh speed scanning ODL achieved a reflectivity sensitivity of only 36 dB, which is not sufficient for imaging in biological samples.

Double Pass, Four Internal Reflections

The configuration illustrated in Fig. 10 is similar to that of Fig. 9 except that it double-passes the reference beam through the cube, doubling the pathlength scan

Figure 9 Rotating cube configuration with single pass, two internal reflections.

112

Rollins and Izatt

Figure 10 Optical pathlength configuration with double pass, four internal reflections. L2, lens; P, right-angle prism; M, mirror; EGM, reflecting gold layers; RC, rotating cube. (#Copyright 1998 IEEE.)

[14,15]. This configuration was demonstrated using a 12 mm 12 mm 4 mm cube rotated at 310 rev/s, resulting in a scan repetition rate of 1240 Hz. A pathlength of 11 mm was scanned at 42 m/s. The resultant duty cycle is approximately 32%. In most biological samples, only about 3 mm of pathlength scanning is useful, however, so this configuration would yield an effective duty cycle of about 9%. A smaller cube could be used to improve this. The (free space) pathlength variation as a function of angular position of the cube is given by

dð Þ ¼ L(2n 1 þ

 

ð

Þ

!

 

1

 

 

cos sin 1 sin =n

 

o

þ sin h1 þ tan 2

2 tan sin 1ðsin =nÞ

 

 

 

 

 

where L is the length of the side of the cube and n is the refractive index of the cube material. This function is obviously not linear, and a scan speed variation of17% was measured. When the cube is rotated at a fixed angular frequency !, then ¼ !t, and the scan velocity is s ¼ @d=@t. The interferogram center frequency and bandwidth are then given by Eqs. (9) and (10). The autocorrelation width measured using this ODL was reported to be only 1% broader than the theoretical width (the source coherence length), so dispersion mismatch was apparently not significant. The type of glass that the cube was composed of was not specified. The optical sources used for this measurement had a center wavelength of 1310 nm and a bandwidth of 45 nm. This ODL required custom deposition of mirrors on portions of each cube face.

Reference Optical Delay Scanning

113

Double Pass for Sample and Reference Light, four Internal Reflections for Each path

The configuration illustrated in Fig. 11 is identical to the configuration in Fig. 10, but in this case both the reference and sample light beams are passed through the rotating cube [16]. In this way, the effective pathlength scan is doubled, and the scan nonlinearity is largely canceled. To optimize the nonlinearity cancellation, the incidence angle of the sample beam with respect to the reference beam must be carefully selected. The (free space) pathlength variation as a function of the angular position of the cube is given by

Dð Þ ¼ dð Þ dð 0 Þ

where dð Þ is defined above and 0 is the angle of incidence of the sample beam with respect to the reference beam, as illustrated in Fig. 11. This configuration was demonstrated by rotating a 60 mm cube at 76.25 rev/s, resulting in a scan repetition rate of 305 Hz. This ODL was designed for a long scan range of greater than 100 mm for OLCR measurements. The scan velocity was 95 m/s and the duty cycle was about 32%. The scan speed variation was only 0:14%. In the reported demonstration, the OLCR using this ODL measured reflectivity with a sensitivity of 65 dB. This ODL requires the same type of custom cube as the previous configurations and obviously requires extremely careful alignment of both beams passing through the cube.

Figure 11 Double pass for sample and reference light, four internal reflections for each path. L1, L2, L3, lenses; Pc, polarization controllers; P1, P2, right-angle prisms; M1, M2, mirrors, M3, M4, reflecting gold layers; Rc, rotating cube.

114

Rollins and Izatt

4.3.2Scanning Mirror

Another ODL based on the transformation of angular motion into pathlength variation is illustrated in Fig. 12 [17]. As with the rotating cubes and the galvan- ometer-mounted retroreflector, the motivation for using a rotational scanning element is speed. In addition to an angular scanning mirror, three additional mirrors and a lens are used. An examination of Figs 12a and 12b will instruct the reader as to the operation of this configuration, particularly the need to place the scanning mirror and mirror 2 one focal length from the lens. This configuration reflects the beam four times off the scanning mirror, amplifying the scan length and velocity by 4. It should be noted that 0 pathlength variation is effected by scanning the beam in a line across mirror 2. All of the pathlength variation arises from the beam being offset from the pivot of the scanning mirror so that there is an axial shift imposed on the beam in addition to the angular scan. In other words, if the beam were reflected from the pivot of the scanning mirror, then no pathlength variation would result from scanning the mirror. The pathlength variation d as a function of scan angle (referenced from the plane parallel to the lens) is given by d ¼ 4x tan 4x for small scan angles, where x is the distance that the beam is offset from the pivot of the scanning mirror. The scan velocity s is given by s 4x =t. For a constant angular velocity !, s 4x!. The interferogram center frequency and bandwidth will then be given by Eqs. (9) and (10). The pathlength variation achievable for a given configuration of this design will be limited by the need to keep the scanning mirror within the depth of focus of the lens and by the size of the lens and mirror 2. Optical power loss will be determined by the quality of the mirrors and the alignment. There should be no significant scan-dependent polarization or dispersion variation.

This ODL was demonstrated using a 1.2 kHz resonant scanner and 50 mm focal length lens. Pathlength was scanned by 3 mm at 1.2 kHz and a peak scan velocity of 11.3 m/s. Most likely the full 3 mm was not usable for imaging, because the motion of a resonant scanner is sinusoidal. The nonlinear pathlength variation with time would warp the image to an unacceptable degree near the extrema of the scan. A reasonable duty cycle for single-sided scanning using a resonant ODL is 33% (66% for double-sided scanning). In order to maintain a narrow detection bandwidth over the course of the scan (and prevent SNR degradation), the signal was bandpass filtered by a voltage-controlled oscillator that was controlled by the reso-

Figure 12 Schematic illustrations of an ODL based on the transformation of an angular mirror scan to pathlength variation. (# Copyright 1998, Taylor & Francis, Ltd.)

Reference Optical Delay Scanning

115

nant scan frequency control. This scheme does not compensate for image warping, however.

Figures 13 and 14 illustrate other ODLs based on pathlength variation due to rotation of reflective elements [18,19]. These designs have not been used for OCT and will not be discussed in detail here.

4.4FIBER STRETCHING

Another method for scanning optical delay is stretching a long optical fiber [20,21,22]. This has been implemented in OCT by coiling a length of optical fiber on a piezoelectric plate or cylinder. With many fiber windings, the small expansion of the PZT actuator can create several millimeters of delay. For the case illustrated in Fig. 15, the pathlength change as a result of a change in the cylinder radius r is given by d ¼ 2 m r, where m is the number of windings. The scan velocity is given by s ¼ 2 m @ =@t. The interferogram center frequency and bandwidth will then be given by Eqs. (9) and (10). This type of ODL has been demonstrated for OCT at scan rates of up to 1200 scans/s (600 Hz triangle drive waveform and doublesided scanning). This type of ODL resulted in the first demonstration of real-time imaging using OCT at 1 frame per second (fps) [20] and 4 fps [21]. The demonstrated duty cycle was 75% [22]. The fiber and air lengths in the reference and sample arms can easily be matched, so dispersion mismatch is not an issue. Optical power loss has not been reported but should be equivalent to linear translation ODLs. The major disadvantage of fiber-stretching ODLs is that there are polarization effects. The static effects of winding the fiber on the actuator can be compensated for by winding the matching length of fiber in the sample arm on a matching actuator. This second actuator can also be driven by the inverse of the reference delay waveform, generating a differential delay scan range of two times the range of a single scanner. As the fiber is stretched, dynamic stretch-induced birefringence will change the SOP of the light propagating through the fiber throughout the course of the scan stroke. This

Figure 13 Schematic of an ODL based on a rotating roof prism.

116

Rollins and Izatt

Figure 14 Schematic of an ODL based on a rotating pair of parallel mirrors. (Reprinted with permission from Elsevier Science.)

Figure 15 Schematic of an ODL based on optical fiber stretching. This design uses a cylindrical PZT element to uniformly stretch the fiber wound around the barrel of the cylinder.

Reference Optical Delay Scanning

117

effect can be compensated for by introducing 45 Faraday rotators (FRs) in the air gaps of the reference and sample arms so that on a double pass through the FRs, the SOP is conjugated and the return propagation through the fiber reverses the SOP change [23]. Alternatively, polarization-maintaining fiber can be used to minimize these effects. In addition, the temperature of the actuator should be carefully controlled in order to prevent temperature-dependent drift of the delay and SOP.

4.5FOURIER DOMAIN RAPID SCANNING OPTICAL DELAY LINE

The Fourier domain rapid scanning optical delay line (RSOD) was developed for femtosecond pulse measurement [24–26] and has recently been applied to OCT [1,27,28]. The RSOD is based on Fourier transform pulse-shaping techniques, which have been demonstrated to also be capable of shaping the temporal properties of broadband incoherent light [29]. This delay line achieves scans of several millimeters at repetition rates of several kilohertz and also allows separation of group and phase delay, which provides an additional degree of control over the center frequency and bandwidth of the OCT signal.

The RSOD consists of a grating–lens pair in a folded, double-passed Fourier domain pulse-shaping configuration (Fig. 16). A flat mirror serves as a spatial phase filter, which imposes a linear phase ramp on the optical frequency spectrum. The delay line is based on the well-known property of the Fourier transform that a phase ramp in the frequency domain corresponds to a group delay in the time domain:

x

t

 

t

Þ

I

X

! e j!t0

ð

11

Þ

ð

 

0

!

ð

Þ

 

The angle of the incident light on the grating can be selected such that the center wavelength 0 of the diffracted beam is normal to the grating and the entire grating is in the focal plane of the lens. This is done to prevent introduction of group velocity dispersion (GVD), which varies throughout the course of the scan. If the distance from the grating to the lens is not one focal length, then GVD is introduced. Therefore, if the grating is at an angle not normal to the optical axis of the lens, then as light is laterally displaced on the grating by the scanning mirror it is also displaced from the focal plane of the lens, introducing GVD [30].

The mirror pivot can be offset from the center wavelength by an arbitrary distance x by a simple lateral translation of the scanning mirror. The phase shift ð Þ as a function of wavelength for a given mirror tilt angle can be written as (refer to Fig. 16)

 

8 x

 

8 lf ð 0Þ

 

þ p cos 0

ð Þ ¼

 

where lf is the focal length of the lens, p is the pitch of the grating, and 0 is the firstorder diffraction angle of the center wavelength from the grating (measured from the grating normal). Recall that in the configuration shown in Fig. 16, 0 ¼ 0, i.e., the grating is in the focal plane of the lens. In this case cos 0 evaluates to unity, so this factor will not be propagated further in the subsequent analysis.

The function ð Þ was derived by using the grating equation and assuming that the small-angle approximation sin holds for diffraction angles between different wavelengths off the grating and for small deflection angles effected by the scan-

118

Rollins and Izatt

(a)

(b)

Figure 16 Schematics of the Fourier domain optical delay line. (a) View from above. The incident, collimated broadband light is diffracted from the grating and spectrally dispersed. The lens parallelizes the dispersed spectrum while focusing it to a line on the scanning mirror. The scanning mirror imposes a linear phase ramp on the spectrum and redirects the light back through the lens, which recollimates the beam and reconverges the spectrum onto the grating. The beam then diffracts in a reverse manner from the grating and propagates collimated and undispersed toward the double-pass mirror collinear with the incident beam. The double-pass mirror turns the light back through an identical path. Double passing is illustrated clearly in (b). (Courtesy of OSA.)

ning mirror [25]. The distance y that a wavelength component traverses as a function of scanning mirror tilt angle is calculated (taking into account the fact that the beam traverses the path y four times) and multiplied by 2= to convert from displacement to phase shift. This phase shift can also be written as a function of angular optical frequency !:

Reference Optical Delay Scanning

119

!

4 x!

 

8 lf ð! !0Þ

ð

12

Þ

 

c

p!

ð Þ ¼

 

 

where !0 is the center angular optical frequency. From Eq. (12) and the definition of phase delay, t ¼ ð!0Þ=!0, the phase delay is given by t ¼ 4 x=c. This corresponds to the free-space phase pathlength difference l (referenced from a zero scan angle),

l ¼ 4 x

ð13Þ

From the definition of group delay, tg ¼ @ð!Þ=@!j!¼!0 , the group delay is given by

tg ¼

4 x

 

4 lf 0

 

c

 

 

cp

 

 

 

 

which corresponds to the free-space group pathlength difference lg,

 

lg ¼ 4 x

4 lf

0

 

ð14Þ

 

 

 

 

p

 

 

 

This expression for group pathlength difference can alternatively be derived by evaluating the change in pathlength experienced by the center wavelength as a result of an arbitrary mirror tilt angle. It can be seen that the group pathlength difference is equal to the phase pathlength difference plus an additional term that is a function of the properties of the lens, grating, and light source. Both the phase pathlength difference and the group pathlength difference are proportional to the tilt angle of the scanning mirror, so an angular scan of the mirror effects a scan of both phase and group pathlength differences. The second term in the expression for group pathlength difference is dominant in typical configurations; thus the group pathlength difference is much larger than the phase pathlength difference for a given angle . This feature is key to the value of this delay line. The mirror need be moved only a very small amount to generate a large scan of the group pathlength difference. It is also possible to adjust the offset x for a desired center frequency without significantly affecting the scan length of the group pathlength difference. If the center wavelength is located at the mirror pivot (i.e., x ¼ 0), then the phase pathlength difference term l vanishes and there is no phase pathlength scanning. This allows for the application of an external phase-modulating element that is truly independent of the group delay scanning if such a configuration is desired.

Differentiating Eq. (13) with respect to time and using Eq. (6), the center frequency of the interferogram measured using the Fourier domain scanning delay line is given as

f

 

¼

4x

@ðtÞ

ð

15

Þ

0

0 @t

 

 

Similarly, differentiating Eq. (14) with respect to time and using Eq. (8), the bandwidth of the interferogram measured using the Fourier domain scanning delay line is

f

 

2

2x

2lf 0

 

@ðtÞ

 

16

 

 

 

 

 

 

 

 

¼

02

p

 

 

ð

Þ

 

 

@t

 

The design equations for the high speed OCT system using this delay line are Eqs. (14)–(16). Equation (14) is used to specify the grating pitch, the lens focal length, and the maximum angular excursion of the scanning mirror for a given

120 Rollins and Izatt

source center wavelength and desired group delay scan. Equations (15) and (16) are used to calculate the center frequency and bandwidth, respectively, of the OCT signal for a given mirror offset x and mirror scan rate @ðtÞ=@t. If the mirror is tilted at a fixed angular frequency !, then ðtÞ ¼ !t and @ðtÞ=@t ¼ !. Estimation of the signal bandwidth f from Eq. (16) can be simplified by recognizing that for a typical

configuration, 2x 2lf 0=p.

A resonant scanning mirror tilts as a sinusoidal function of time,

ðtÞ ¼ b sinð2 fm

where b is the maximum excursion of the tilt angle and fm is the resonance frequency of the scanning mirror. Thus, from Eq. (15) the center frequency becomes

 

 

 

4x

 

 

 

 

 

 

 

 

 

f0ðtÞ ¼

 

 

 

 

b2 fm cosð2 fm

 

 

ð17Þ

0

 

 

 

From Eq. (16) the interferogram bandwidth becomes

 

 

 

 

 

 

2

 

 

2lf 0

 

 

 

 

 

f

ð

Þ ¼

 

 

 

 

 

 

ð

Þ

ð

18

Þ

 

02

p

t

 

 

 

 

 

 

2x

 

 

 

b2 fm cos 2 fmt

 

 

 

This sinusoidal variation of center frequency and bandwidth with time is the major disadvantage of using a resonant scanner in the delay line. The advantage is that resonant scanners can operate at much higher repetition rates than galvanometermounted scanners. Using a resonant scanner, a reasonable compromise between duty cycle and nonlinearity is to record the interferometric signal only during the middle two-thirds of the forward scan, resulting in an overall duty cycle of 33%. Double-sided scanning increases the duty cycle to 67%. A linear scanner can achieve a duty cycle of nearly 100% (at the expense of scan speed). Unless a tracking bandpass filter is used, as discussed in Section 5.2.3, the use of a resonant scanner with a reduced duty cycle requires the detection bandwidth to be increased compared to an equivalent linear scanner in order to accommodate the varying center frequency. This corresponds to degradation of the SNR. Thus, a resonant scanner is advantageous only in cases where the needed scan rate cannot be achieved with a linear scanner.

As discussed above, if the distance from the grating to the lens is not one focal length, then GVD is introduced. This requires careful alignment to prevent broadening of the cross-correlation due to dispersion mismatch. This property of the RSOD also provides a convenient means to compensate for dispersion mismatch due to dissimilar components in the reference and sample arms. For example, an endoscopic probe in the sample arm may not include an air gap to match the air gap in the RSOD. This dispersion mismatch requires compensation, which can be introduced by the RSOD [28,30,31].

This ODL has few disadvantages. Besides complexity, the major disadvantage is optical power loss. In a double-passed configuration, the light is diffracted into the first order four times. A grating blazed at the appropriate angle can increase the efficiency of diffraction into the first order, but the aggregate loss will be significantly greater than most of the other ODLs reviewed here. In the typical OCT configuration using a Michelson interferometer, the reference ODL must be attenuated to