Добавил:
kiopkiopkiop18@yandex.ru t.me/Prokururor I Вовсе не секретарь, но почту проверяю Опубликованный материал нарушает ваши авторские права? Сообщите нам.
Вуз: Предмет: Файл:

Ординатура / Офтальмология / Английские материалы / Elevation Based Corneal Tomography 2nd_Belin, Khachikian, Ambrósio_2011

.pdf
Скачиваний:
0
Добавлен:
28.03.2026
Размер:
18.35 Mб
Скачать

40

ELEVATION BASED CORNEAL TOMOGRAPHY

While not an elevation map, the corneal thickness (pachymetric map) represents the spatial difference between the anterior and posterior corneal surfaces and as such is totally dependent on accurate elevation data. In addition to identifying thin corneas, the overall pachymetric distribution may be another indicator of pathology. Normal corneas are typically thinnest in the central region and thicken in the periphery. Displacement of the thinnest region is often seen in keratoconus and may at times predate changes on either the anterior or posterior surfaces (FIGURE 15 – SAMPLE Isolated Pach Displacement).

Figure 15. The 4-view composite map (Oculus Pentacam) above shows an asymptomatic patient presenting for refractive surgical evaluation. Although there are no obvious abnormalities in the curvature or elevation the large displacement of the thinnest point from the corneal apex may suggest early ectatic disease.

DISPLACED APEX SYNDROME

Previously we discussed some of the limitations of trying to use curvature to depict shape. Early studies in patients seeking refractive surgery reported an incidence of “form fruste” keratoconus or “keratoconus suspect” as high as 17% of apparent normal individuals.34 Certain investigators initially pointed out that this high false-positive rate was related to the limitations of sagittal or axial-based curvature reconstructions and Placido-derived topography

CHAPTER 3. UNDERSTANDING ELEVATION BASED TOPOGRAPHY: How Elevation Data is Displayed

41

systems.12,35 Curvature maps on asymmetric corneas are prone to pattern errors due to the difference between the curvature map’s reference axis, the line of sight, and the corneal apex.4,5 Many of these so-called keratoconus patients have what is now recognized as a displaced corneal apex (commonly inferior).1 These patients demonstrate an elevated I-S ratio, inferior corneal axial power > 1.5 D steeper than the comparable superior corneal region. However, they have no other clinical or topographic (elevation) aspects of keratoconus. These patients have a more normal topography pattern when imaged on an elevation based topography system and commonly do not meet the keratoconus criteria of some of the newer keratoconus detection subprograms (FIGURE 16 - SAMPLE Displaced Apex) (FIGURE 17 –

SAMPLE Displaced Apex SUPERIOR).

Figure 16. A four-image composite map (Oculus Pentacam) of a normal astigmatic cornea. The asymmetric bowtie pattern seen in the curvature map is created when the reference axis and the corneal apex do not coincide. The anterior elevation map shows that the apex of the cornea is displaced inferiorly but the elevation and pachymetry are both normal.

42

ELEVATION BASED CORNEAL TOMOGRAPHY

 

 

 

 

 

 

 

 

Figure 17. A four-image composite map of a normal astigmatic cornea (Oculus Pentacam). The asymmetric bowtie pattern seen in the curvature map is created when the reference axis and the corneal apex do not coincide. The anterior elevation map shows that the apex of the cornea is displaced superiorly but the elevation is normal.

The classic asymmetric inferior bowtie pattern can be produced by a completely normal astigmatic eye if the curvature’s reference axis does not go through the corneal apex

(SEE FIGURE 2). (In actuality what more typically occurs is that the patient does not look through the center of their cornea, the so called positive angle kappa). Patients with a displaced apex syndrome typically have normal pachymetry, orthogonal astigmatism, stable refractions, and BSCVA of 20/20 or better.1 Many patients in the literature who have been described as having early keratoconus based solely on curvature maps (and reported to have excellent results from refractive surgery) have instead what is more likely a “displaced apex syndrome” and would probably not meet the criteria for keratoconus on elevation topography.36-38

CHAPTER 3. UNDERSTANDING ELEVATION BASED TOPOGRAPHY: How Elevation Data is Displayed

43

CONE LOCATION

Similar to the above discussion, sagittal or axial curvature maps are poor indicators of the location of the cone in keratoconus and commonly exaggerate its peripheral appearance. Both anterior elevation maps, posterior elevation maps and pachymetric maps more accurately locate the true cone position (FIGURE 18 – SAMPLE Faulty Location).

It should be understood the limitations on axial or sagittal curvature are the same limitations whether the maps are Placido generated or elevation generated. The limitations are not with the machine or the technology, but are innate limitations in that type of curvature measurement. The recent increase in diagnosing Pellucid Marginal Degeneration is, at least in part, due to a reliance on trying to use a curvature map to depict shape.

Figure 18. A four-image composite map of a patient with keratoconus (Oculus Pentacam). The curvature map (upper right) does not accurately reflect the location of the pathology and suggests a superior cone. The posterior elevation and pachymetry maps are able to accurately localize the cone inferiorly.

44

ELEVATION BASED CORNEAL TOMOGRAPHY

SUMMARY

Elevation based topography offers important advances over Placido based devices. The ability to image the posterior cornea and to produce an accurate pachymetric map is in itself significant. Elevation maps are also more accurate in determining the cone morphology and in separating the false positive keratoconus suspect often due to a displaced corneal apex.

REFERENCES

1.Belin MW, Khachikian SS. New devices and clinical implications for measuring corneal thickness. Clin Experiment Ophthalmol. 2006;34:729-31.

2.Miller D, Greiner JV: Corneal measurements and tests. In Albert DM, Jakobiec FA (eds): Principles and Practice of Ophthalmology. Philadelphia: W.B. Saunders Company, 1994, p 7.

3.Dabezies OH, Holladay JT: Measurement of corneal curvature: keratometer (ophthalmometer). In Kastle PR (ed): Contact Lenses: The CLAO Guide to Basic Science and Clinical Practice, vol. 1. Dubuque: Kendall/Hunt Publishing Company, 1995, pp 253-289.

4.Arffa RC, Klyce SD, Busin M: Keratometry in refractive surgery. J Refract Surg 2:6, 1986

5.Rubin ML: Optics for Clinicians. Gainesville: Triad Publishing Company, 1993.

6.Brody J, Waller S, Wagoner M: Corneal topography: history, technique and clinical uses. International Ophthalmology Clinics 34: 197-207, 1994.

7.Levine JR: The true inventors of the keratoscope and photokeratoscope. Br J Hist Sci 2:324-341, 1965.

8.Brody J, Waller S, Wagoner M: Corneal topography: history, technique and clinical uses. International Ophthalmology Clinics 34: 197-207, 1994.

9.Maquire LJ: Keratometry, photokeratoscopy and computer-assisted topographic analysis. In Krachmer JH, Mannis MJ, Holland EJ (eds): Cornea - Fundamentals of Cornea and External Disease. St. Louis: Mosby, 1997, pp 223-235.

10.Wilson SE, Klyce SD: Advances in the analysis of corneal topography. Surv Ophthalmol 35: 269-277, 1991.

11.Klyce SD: Computer-assisted corneal topography. High-resolution graphic presentation and analysis of keratoscopy. Invest Ophthalmol Vis Sci 25:1426-1435, 1984.

12.Committee on Ophthalmic Procedures Assessment Cornea Panel, Cohen EJ(Chair): Corneal Topography. Ophthalmology 1999;106:1628-1638.

13.Belin MW, Zloty P: Accuracy of the PAR Corneal Topography System with spatial misalignment. CLAO J. 1993;19:64-68.

14.Mandell RB. The enigma of the corneal contour. CLAO J.1992;18:267-273.

15.Arffa RC, Warnicki JW, Rehkopf PG: Corneal topography using rasterstereography. Refract Corneal Surg. 1989;5:414-417.

16.Belin MW, Litoff D, Strods SJ, et al: The PAR Technology Corneal Topography System. Refrac Corneal Surg. 1992;8:88-96.

17.Schultze RL. Accuracy of corneal elevation with four corneal topography systems. J Refract Surg. 1998;14:100- 4.

18.Cairns G, Ormonde SE, Gray T, Hadden OB, Morris T, Ring P, McGhee CN. Assessing the accuracy of Orbscan II post-LASIK: apparent keratectasia is paradoxically associated with anterior chamber depth reduction in successful procedures. Clin Experiment Ophthalmol. 2005;33:147-52

19.Cairns G, McGhee CN. Orbscan computerized topography: attributes, applications, and limitations. J Cataract Refract Surg. 2005;31:205-20.

CHAPTER 3. UNDERSTANDING ELEVATION BASED TOPOGRAPHY: How Elevation Data is Displayed

45

20.Hashemi H, Mehravaran S. Corneal changes after laser refractive surgery for myopia: comparison of Orbscan II and Pentacam findings. J Cataract Refract Surg. 2007;33:841-7.

21.Prisant O, Calderon N, Chastang P, et al. Reliability of pachymetric measurements using Orbscan after excimer refractive surgery. Ophthalmology 2003;110:511-5.

22.Kamiya K, Oshika T, Amano S, et al. Influence of excimer laser photorefractive keratectomy on the posterior corneal surface. J Cataract Refract Surg 2000;26:867-71.

23.Naroo SA, Charman WN. Changes in posterior corneal curvature after photorefractive keratectomy. J Cataract Refract Surg 2000;26:872-8.

24.Seitz B, Torres F, Langenbucher A, et al. Posterior corneal curvature changes after myopic laser in situ keratomileusis. Ophthalmology 2001;108:666-72.

25.Wang Z, Chen J, Yang B. Posterior corneal surface topographic changes after laser in situ keratomileusis are related to residual corneal bed thickness. Ophthalmology 1999;106:406-9.

26.Baek T, Lee K, Kagaya F, et al. Factors affecting the forward shift of posterior corneal surface after laser in situ keratomileusis. Ophthalmology 2001;108:317-20.

27.Ciolino JB, Belin MW. Changes in the posterior cornea after laser in situ keratomileusis and photorefractive keratectomy. J Cataract Refract Surg 2006;32:1426-31.

28.Buehl W, Stojanac D, Sacu S, et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol 2006;141:7-12.

29.Lackner B, Schmidinger G, Pieh S, et al. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci 2005;82:892-9.

30.Lackner B, Schmidinger G, Skorpik C. Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan. Optom Vis Sci 2005;82:858-61.

31.O’Donnell C, Maldonado-Codina C. Agreement and Repeatability of Central Thickness Measurement in Normal Corneas Using Ultrasound Pachymetry and the OCULUS Pentacam. Cornea 2005;24:920-4.

32.Ucakhan OO, Ozkan M, Kanpolat A. Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry. J Cataract Refract Surg 2006;32:970-7.

33.Ciolino JB, Khachikian SS, Cortese MJ, Belin MW. Long-term stability of the posterior cornea after laser in situ keratomileusis. J Cataract Refract Surg. 2007;33:1366-70

34.Wilson SE, Klyce SD: Screening for corneal topographic abnormalities before refractive surgery. Ophthalmology 1994;101:147-52.

35.McGhee CNJ, Weed KH: Computerized videokeratography in clinical practice. In McGhee CNJ, Taylor HR, Gartry DS, et al (eds): Excimer Lasers in Ophthalmology: Principles and Practice. London: Martin Dunitz Ltc., 1997.

36.Bilgihan K, Ozdek SC, Konuk O, Akata F, Hasanreisoglu B: Results of photorefractive keratectomy in keratoconus suspects at 4 years. J Refract Surg. 2000;16:438-43.

37.Sun R, Gimbel HV, Kaye GB: Photorefractive keratectomy in keratoconus suspects. J Cataract Refract Surg. 1999;25:1461-6.

38.Kremer I, Shochot Y, Kaplan A, Blumenthal M: Three year results of photoastigmatic refractive keratectomy for mild and atypical keratoconus. J Cataract Refract Surg. 1998;24:1581-8.

46

ELEVATION BASED CORNEAL TOMOGRAPHY

Chapter

4

The Importance of

Understanding the

Reference Surface

Michael W. Belin, MD, FACS

Stephen S. Khachikian, MD

Renato Ambrósio Jr., MD, PhD

One limitation of curvature is that the same shape can have different curvatures depending on the axis or orientation.1,2 Typically, the clinician views elevation data not in its raw form (actual elevation data) but compared to some reference shape to allow the clinician to quantitatively examine the maps for clinically significant changes. The maps display how the actual corneal elevation data deviates when compared from a known shape. The choice of the reference shape is determined by the clinical situation. A properly chosen reference surface will magnify the differences, highlight the “abnormal” areas and allow the clinician a qualitative map which will emphasize clinically significant areas. The reason for viewing elevation data in this format is that the actual raw elevation data lacks qualitative patterns that would allow the clinician to easily separate normal from abnormal corneas.3 In other words, raw elevation data for normal eyes looks surprisingly similar to the raw elevation data in abnormal eyes (e.g. Keratoconus) (FIGURE 1).

Figure 1. Raw elevation data from the PAR CTS (PAR Technology). The raw elevation data displays the elevation values without comparing them to a reference surface. The reference surface serves to highlight or magnify the surface changes. Without this, the raw elevation data from normal and pathologic eyes looks remarkably similar.

48

ELEVATION BASED CORNEAL TOMOGRAPHY

This is not an uncommon approach. When one wants to highlight an abnormality, you typically attempt to remove the “background noise.” In the case of elevation data, the “background” noise is any shape that will help demonstrate the clinically significant abnormalities.

For refractive surgery screening and for most clinical situations using a best-fit-sphere gives the most useful qualitative map (i.e. easiest to read and understand). Fitting a best-fit- sphere to the central 8.0 mm zone appears best. Since the normal eye is an aspherical prolate surface the central 8.0 mm zone yields a reference surface that allows for subtle identification of both ectatic disorders and astigmatism. Larger zones would typically yield a flatter BFS and smaller zones a steeper BFS. Because the maps we call “elevation” are actually subtraction maps

Effects of BFS Diameter on the Appearance of the Elevation Map

 

 

 

 

 

 

 

 

Diameter = 9.0 mm

Diameter = 7.0 mm

 

 

 

 

 

 

 

 

 

 

 

Diameter = 11.94 mm

Figure 2. Three anterior elevation maps showing the effect of varying the area used to compute the BFS. The upper left sample uses an area of 7.0 mm, the upper right 9.0 mm and the bottom 11.94 mm. As the area gets larger it incorporates more peripheral (flatter portions of the cornea) data. A flatter reference surface will make the “normal” prolate cornea appear to have a central “island.” This is a good example of the importance in keeping the BFS computation area consistent.

CHAPTER 4. THE IMPORTANCE OF UNDERSTANDING THE REFERENCE SURFACE

49

(showing how the raw elevation data appears compared to the reference surface) the shape of the reference surface will greatly affect the appearance of the map. Since the normal eye is steeper centrally and flattens in the periphery a flatter reference surface (i.e. a very large optical zone) will accentuate the central steep zone and the eye will appear to have an “island” (FIGURE 2). Similarly, a smaller optical zone will be steeper and may mask subtle cones.4

The Pentacam has two settings (AUTO & MAN) that determine how the machine selects the area used for the BFS. An understanding of these settings is mandatory to utilize the full capability of the system. The AUTO (Automatic) setting looks for the largest circle around the apex that can be drawn without any extrapolated data. The machine then selects the area set at 90% of this size. This insures that only valid data is utilized. The advantage of the AUTO system is that only valid data points are used. The disadvantage is that the area used to compute the BFS is variable in size. This makes comparisons difficult and the development of normal values impossible. If the image was of very high quality (no extrapolated data > 10.0 mm) it was possible to have the area used to define the BFS greater than > 9.0 mm. As the zone gets larger and incorporates more peripheral data (peripheral cornea is flatter) the Best-Fit shape would be flatter. When the cornea is then compared to a flatter reference surface, you can see positive islands of elevation that would normally be considered abnormal when compared to a reference surface computed form a smaller optical zone (e.g. 8.0 mm). Similarly, if a smaller optical zone was chosen the BFS would be steeper since it does not utilize the flatter periphery for its computation. A steeper Best-Fit shape will hide or mask conical areas. In the example of FIGURES 3A & 3B, the posterior elevation map appears to have a significant central island, but a closer inspection reveals a large area (diameter 9.00 mm) was used to compute the BFS.

When the map is recomputed with a smaller diameter (7.5 mm) the prominent anterior island disappears. This shows the importance of standardizing the area used to compute the BFS for both qualitative and quantitative analysis.

Because of the variable nature of the BFS and the need for the clinician to have normal values to screen patients, we no longer recommend the routine use of the AUTO setting.

The MAN (manual) setting is user selectable. The user may choose any size optical zone and the system will utilize all the data within that user defined optical zone to compute the Best-Fit-Shape. The value of this setting is that it allows a consistent area to be specified (see discussion in the following pages). The limitation is that the system will accept all the data points within that area whether the data is actual or extrapolated. The incorporation of extrapolated data, if excessive (poor quality scan), can lead to erroneous results. To allow for the generation of normal values and to allow for both patient comparisons over time and to compare different patients it is necessary to utilize the same BFS constructs. It is our recommendation that setting the system to MAN with a BFS diameter fixed at 8.0 mm gives the best results for the following reasons:

Соседние файлы в папке Английские материалы