Ординатура / Офтальмология / Английские материалы / Clinical Ocular Pharmacology 5th edition_Bartlett, Jaanus_2008
.pdf
16
Dyes
Jerry R. Paugh
Dyes for ophthalmic care came into use in the late 1800s, following Baeyer’s synthesis of sodium fluorescein in
1871 and Ehrlich’s use of it to study aqueous dynamics. Since then, several dyes have been used for ophthalmic diagnosis, including fluorescein sodium (e.g., to examine corneal damage, tear stability, intraocular pressure [IOP], and retinal vascular characteristics), fluorexon (a larger molecular weight fluorescein to facilitate examination in hydrogel contact lens wearers), rose bengal, and (more recently) lissamine green (for conjunctival staining). In addition, other dyes such as indocyanine green and methylene blue are developing acceptance in ocular vasculature observation and intraocular surgery, respectively.
FLUORESCEIN SODIUM
Fluorescein is probably one of the most widely used dyes for ophthalmic use. Several factors contribute to its utility, including its hydrophilicity, low toxicity, and excellent fluorescent properties in the visible spectrum, even in very dilute concentration. Early ocular applications were used in detection of corneal ulcers and aqueous flow, followed shortly thereafter by retinal diagnostic application.
Pharmacology
Fluorescein sodium, 3¢,6¢-dihydroxyspiro[isobenzofuran- 1(3H),9¢-[9H]xanthen], C20H10Na2O5, CAS number 518- 47-8, is a yellow acid dye of the xanthene series. Its molecular weight is 376 Da, and its solubility in water at 15° C is 50% (i.e., it is freely soluble). It is generally formulated as its sodium salt (Figure 16-1). When exposed to light, fluorescein maximally absorbs light at approximately 493 nm and emits (fluoresces) at approximately 520 nm.
Figure 16-2 illustrates the excitation and emission spectra of dilute fluorescein in phosphate buffer.
Because fluorescein is a weak acid, depending on the pH of the solution, it can exist in various ionic states. Below pH 2, the cationic form predominates, and a weak blue-green fluorescence occurs. Between pH 2 and
4 the cations dissociate to neutral molecules. At pH 7
negative ions prevail and are associated with a brilliant yellow-green fluorescence. The pH sensitivity has been used to noninvasively examine human stromal pH.
Several factors can alter the fluorescence of fluorescein in solution: its concentration, the pH of the solution, the presence of other substances, and the intensity and wavelength of the absorbed light. In the eye the thickness of the media being measured becomes important in quantitative measurements due to fluorescein self-absorption (or “quenching”). One clinical effect can be to obscure corneal staining if the tear concentration is too great. The intensity of fluorescence increases with increasing pH, reaching a plateau at approximately pH 8. Thus at physiologic pH the fluorescence is nearly maximum. Further increases in pH above 8 reduce the intensity of fluorescence.
Clinical Uses
Fluorescein may be applied topically to the eye in the form of a solution or by fluorescein-impregnated filter paper strips (Table 16-1). It is also available in injection form for intravenous use (Table 16-2).
Fluorescein in solution is highly susceptible to bacterial contamination, especially by Pseudomonas aeruginosa, which grows easily in the presence of fluorescein. Major methods of reducing the possibility of bacterial growth include sterile formulation and air-tight seal of solutions (e.g., injection fluorescein), use of effective
COONa
NaO |
O |
O |
|
Figure 16-1 Molecular structure of fluorescein sodium.
283
284 CHAPTER 16 Dyes
Excitation spectra |
Emission spectra |
(emission wavelength |
(excitation wavelength |
set at 515 nm) |
set at 490 nm) |
RELATIVE INTENSITY
200 |
300 |
400 |
500 |
600 |
700 |
800 |
WAVELENGTH IN NANOMETERS (nm)
Figure 16-2 Excitation and emission spectra of a 0.00005% solution of sodium fluorescein in KH2PO4–K2HPO4 buffer at pH 8. (Reprinted with permission from Romanchuk KG. Fluorescein. Physiochemical factors affecting its fluorescence. Surv Ophthalmol 1982;26:269–283.)
preservatives in fluorescein solution (e.g., Fluress®), and the development of sterile fluorescein impregnated strips.
Although both injection fluorescein and that used clinically from sterile strips are used once and discarded, fluo- rescein–anesthetic combination solutions are used repeatedly on sequential patients. Thus the problem of maintaining sterility, particularly when accidental contamination from patient contact can easily occur, becomes a major issue.
Many studies have examined the issue of microbial contamination of fluorescein–anesthetic solutions designed for clinical use. Although most have directly inoculated the liquid solution, it seems more useful to consider the evidence of contamination from products retrieved from clinical use and from direct contamination
Table 16-1
Fluorescein Preparations for Topical Ocular Use
Product (Manufacturer) Composition
Fluorescein sodium solutions
Fluress (Akorn) |
0.25% with 0.4% benoxinate |
|
HCl; chlorobutanol 1%, |
|
povidone, boric acid; 5 ml |
Fluoracaine (Akorn) |
0.25% with 0.5% |
|
proparacaine HCl; |
|
thimerosal 0.01%; 5 ml |
Fluorescein sodium |
0.25% with 0.4% benoxinate |
and benoxinate |
HCl; 5 ml |
(Bausch & Lomb) |
|
Fluorescein strips
Ful-Glo (Akorn) |
0.6 and 1.0 mg; sterile |
Fluor-I-Strip |
9 mg, 0.5% chlorobutanol, |
(Bausch & Lomb) |
polysorbate 80, with buffers |
Fluor-I-Strip A.T. |
1 mg, 0.5% chlorobutanol, |
(Bausch & Lomb) |
polysorbate 80, with buffers |
|
|
Table 16-2
Fluorescein Preparations for Intravenous Use
Product (Manufacturer) |
How Supplied |
Fluorescite (Alcon Laboratories) |
10%, 5-ml ampule |
AK-Fluor (Akorn) |
10%, 5-ml ampule and |
|
5-ml vial |
AK-Fluor (Akorn) |
25%, 2-ml ampule and |
|
5-ml vial |
Fluorescite (Alcon Laboratories) |
25%, 2-ml ampule |
|
|
of the bottle tip (if attached, as for some generic products) and dropper tips (for separate droppers), as would occur clinically.
Both dropper tip or bottle tip contamination was examined using Staphylococcus and Pseudomonas species, and Fluress® was found to prevent colonization after 1 minute but the generic fluorescein–anesthetic combinations all allowed growth up to 2 hours. It was suggested that the preservative used in Fluress®, chlorobutanol, worked more rapidly than the thimerosal used in the competing formulations and also that the benoxinate and weak boric acid in Fluress® may confer additional antibacterial properties.
Bottles of Fluress® sourced from the clinic, which might be expected to demonstrate contamination, were examined and found to be largely free of either
Staphylococcus or Pseudomonas species. Although it appears that resistance to bacterial contamination is quite good for Fluress®, the potential for viral contamination appears more serious.
The resistance to adenoviruses types 8 and 19, both common causes of epidemic keratoconjunctivitis, in Fluress® was studied and survival was found for 3 to 4 weeks for types 19 and 8, respectively. Extreme care should be taken when examining suspect patients. Conversely, resistance to contamination for Fluress® from herpes simplex virus type 1 was examined and found to be quite good. Overall, it appears that Fluress®, with its unique formulation, is generally the most effective of the combination fluorescein–anesthetic solutions for clinical use but that care must be taken when using generic versions.
Topical Ocular Applications
Assessment of Ocular Surface Integrity
Instillation of the dye in the cul-de-sac allows detection of corneal and conjunctival lesions, such as abrasions, ulcers, and edema, and aids in the detection of foreign bodies.When the cobalt blue filter of the slit lamp is used to excite the dye, the epithelial defect usually appears outlined in vivid green fluorescence.The dye turns green in the tear film, in spite of being introduced as a yellow-orange liquid, due to dilution with tear fluid.
CHAPTER 16 Dyes |
285 |
Figure 16-3 Fluorescein photograph of conjunctival staining taken without barrier filter. (From Courtney RC, Lee JM. Predicting ocular intolerance of a contact lens solution by use of a filter system enhancing fluorescein staining detection. Int Contact Lens Clin 1982;9:302–310.)
Although the use of the cobalt blue excitation illumination is adequate for some observations, the addition of a yellow barrier filter over the observation system of the slit lamp, Burton lamp, or camera greatly enhances visibility of the stained areas, especially on the conjunctiva. Kodak Wratten No. 12 or No. 15 photographic filters or Tiffen No. 2 photographic filters are relatively inexpensive and serve well in this capacity (Figures 16-3 and 16-4). The yellow barrier filter must be placed over the optics of the instrument, not in the path of the blue excitation light. It is highly recommended that the yellow barrier filter be used for staining assessment and for other tests such as fluorescein breakup time.
The mechanism of fluorescein staining of ocular epithelia has been subject to some conjecture. In earlier work it was suggested that staining occurred due to accumulation in intraepithelial spaces rather than direct staining of the cells. However, it has become clear that fluorescein can directly stain diseased human corneal cells and rabbit epithelial cells. Moreover, the hyperfluorescence that probably represents micropunctate clinical staining is likely due to optimum dye concentration and fluorescence within the cell rather than simple pooling. Cellular hyperfluorescence occurred from both mechanical abrasion and chemically induced toxicity, conditions that presumably promote an intracellular concentration that allows definitive clinical visualization. An issue that has received some attention is whether repeated
Figure 16-4 Fluorescein photograph of conjunctival staining taken with a Wratten No. 12 yellow barrier filter in place. (From Courtney RC, Lee JM. Predicting ocular intolerance of a contact lens solution by use of a filter system enhancing fluorescein staining detection. Int Contact Lens Clin 1982; 9:302–310.)
instillations of fluorescein might serve as a predictive test for corneal compromise.
Sequential instillations of fluorescein (up to six times, 5 minutes apart) may have value as a mildly provocative test of corneal integrity. Although only 19% of patients showed fluorescein staining after a single instillation of fluorescein, an additional 23% exhibited staining after repeated instillations. It was also noted that the severe degrees of staining appeared to be correlated with contact lens intolerance. However, it was demonstrated that the fluorescein itself was inducing the staining, apart from physicochemical formulation properties or preservatives. Additional work related to the predictive value of sequential staining, using nonpreserved and physiologically compatible formulations, may be warranted.
Contact Lens Fitting and Management
Fluorescein staining of the tear film is a major aid in the fitting of rigid gas-permeable contact lenses.After topical application of fluorescein to the eye the tear layer becomes visible, with a characteristic pattern of green fluorescence. Observation of the fluorescein-stained tear film with an ultraviolet light or the cobalt blue filter of the slit lamp allows determination of the fit of the lens.
However, it should be understood that many recently developed contact lens materials contain polymers that block the transmission of light in the ultraviolet region.
Therefore when an ultraviolet light source, such as a
286 CHAPTER 16 Dyes
Figure 16-5 Contact lens fluorescein pattern in eye with keratoconus. Central dark area reflects the absence of fluorescein, indicating central contact lens bearing (touch). There is also bearing in the intermediate area, surrounded by peripheral clearance indicated by the pooling of fluorescein. (Courtesy A. Christopher Snyder, O.D.)
Burton lamp, is used, the fluorescein behind the lens may not be visible. Visualizing the fluorescein necessitates changing to a blue light source in the visible region.This may be accomplished by fitting the Burton lamp or other ultraviolet light with a white light source covered with a deep blue excitation filter, such as a Kodak Wratten No. 47, 47A, or 47B photographic filter. Areas where the lens makes corneal contact show minimal fluorescence or absence of the fluorescein dye (Figure 16-5).
In addition to its usefulness during contact lens fitting procedures, fluorescein is essential for assessing the integrity of the cornea in contact lens wearers. Common contact lens complications that stain with sodium fluorescein include those thought to be due to mechanical etiologies (e.g., foreign body abrasions, 3- to 9-o’clock staining, edge desiccation, vascularized limbal keratitis, superior epithelial arcuate lesions, conjunctival flaps, etc.) and those related to other causes such as solution incompatibilities (e.g., superficial punctate keratopathy) and lack of lens tear exchange (e.g., inferior “smile face” or arcuate staining).
The practitioner should be cautious in interpreting apparent fluorescein staining in a contact lens wearer, as areas of indentation, which do not represent cellular damage, also demonstrate increased fluorescence. Indentation may result from the accumulation of bubbles (known as dimple veiling) or from compression by a lens edge or poorly finished junction, which often results in an arcuate pattern of fluorescein pooling.
Lacrimal System Evaluation
Topical ocular fluorescein can be used to evaluate two key aspects of the lacrimal system.These are the stability of the precorneal tear film and the patency of the lacrimal drainage system.
Tear breakup time (TBUT) is used clinically as a diagnostic aid in dry eye syndromes and for testing the efficacy of therapeutic approaches. Assessment of TBUT, typically defined as the interval between the last complete blink and the development of the first randomly distributed dark spot in the tear film, is commonly used to estimate tear film stability. Fluorescein can be instilled into the eye with either a pipette or wetted fluorescein strip and observed with cobalt blue excitation and with or without a yellow barrier filter for observation.* Unfortunately, there is still no global standard as to how TBUT should be determined and no consensus as to appropriate cut-off values.
Historically, TBUTs of less than 10 seconds were thought to indicate an unstable tear film. However,TBUTs in normal asymptomatic Hong Kong and Singapore Chinese were found ranging from approximately 8 seconds to 6 seconds, which suggests that a cut-off value for instability less than perhaps 7 seconds may be sensible.
Fluorescein is also useful clinically in evaluating epiphora. Fluorescein testing for lacrimal obstruction usually involves instilling the dye into the conjunctival cul-de-sac and then observing for the presence of fluorescein in the nose. Appearance of the dye in the nose or posterior oropharynx indicates that the lacrimal drainage system of that eye is functional. Generally, a 2% fluorescein solution is used, and this test can be used in conjunction with other procedures for diagnosis of lacrimal obstruction (see Chapter 24).
Applanation Tonometry
The use of topical fluorescein is an important component in the measurement of IOP with the Goldmann applanation tonometer. The dye permits visualization of the applanated area, which is 3.06 mm2 for accurate IOP measurement.
Measurement of IOP with the Goldmann applanation tonometer requires the meniscus of tear fluid surrounding the flattened corneal surface to be sufficiently stained with fluorescein so that the apex of the wedge-shaped meniscus is visible. If the fluid apex is not visible, IOP will be underestimated due to inadequate applanation (Figure 16-6).
The mire visualization problem is likely the reason why IOP measurement without fluorescein has been discredited. For example, it was found that readings without fluorescein were lower by an average of 7.01 mm Hg (Table 16-3). The mean reading with Fluress was 18.03 mm Hg compared with 11.02 mm Hg with the anesthetic Ophthetic in the absence of fluorescein. By performing a regression analysis, it was further suggested that the difference in IOP readings with and without fluorescein becomes even greater as the pressure rises.
*NB:The author recommends the yellow filter for use in TBUT because it seems to provide more definitive TBUT end points.
2.25 sq. mm |
3.06 sq. mm |
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Figure 16-6 Cornea partially flattened by applanation tonometer. The apices of the fluorescein-stained wedges above and below the flattened area are too dilute to be visible.The 3.06-mm2 end point of applanation appears to have been reached but in reality consists of a smaller flattened area. (Modified and reprinted with permission from Moses RA. Fluorescein in applanation tonometry. Am J Ophthalmol 1960;49:1149–1155.)
Intravenous Applications
The introduction of fluorescein angiography in the early 1960s provided a useful method for studying various parameters of ocular function. Intravenous fluorescein is used extensively to delineate vascular abnormalities of the fundus and occasionally to evaluate anterior segment blood and aqueous flow.
Fluorescein Angiography
In the bloodstream fluorescein is excited by a wavelength of 465 nm and emits a wavelength of 525 nm. Circulating fluorescein binds to albumin and red blood cells. It is also metabolized to a weakly fluorescent conjugate, fluorescein monoglucuronide, which exhibits less plasma protein binding than fluorescein. The amount of binding
CHAPTER 16 Dyes |
287 |
can affect the penetration of fluorescein through blood–ocular barriers. After injection of 5 ml of a 10% or 3 ml of a 25% fluorescein solution in the antecubital vein, the dye usually appears in the central retinal artery in 10 to 15 seconds. Both circulation time and integrity of the retina and choroid may be examined. Injection of the 25% concentration is well visualized and may cause fewer side effects.
Fluorescein angiography shows retinal blood vessels in high contrast. Nonvascularized pigmented retinal and subretinal lesions appear as dark areas against the green fluorescing background. The abnormal fluorescence of various retinal and choroidal lesions is explained by several mechanisms, including (1) some abnormalities in the retina, allowing for greater visibility of choroidal fluorescence; (2) neovascularization, producing enhanced fluorescence due to new vascular channels; and (3) pathologic processes resulting in enhanced capillary permeability, allowing for leakage of fluorescein into the lesion. These types of abnormalities may often be differentiated by time at onset of fluorescence.
Choroidal fluorescence appears early and usually precedes the arterial phase by about 1 second. Depending on the origin of the new vessels, neovascular fluorescence coincides with the arteriolar or venous phase of fluorescence. Enhanced capillary permeability (leakage) delays fluorescence, followed by a slow increase in fluorescence as the dye recirculates and stains the affected tissues.
Fluorescein angiography has proven helpful in the diagnosis of a variety of pathologic conditions of the fundus. Various macular lesions, central serous choroidopathy, diabetic retinopathy, and disciform macular degeneration show typical fluorescein patterns. Moreover, tumors such as malignant melanoma, those arising from metastasis, and hemangiomas of the choroid demonstrate fluorescence. Fluorescein angiography does not illustrate well the choroidal vasculature, which can be complemented by indocyanine green angiography (see below).
Chapter 31 discusses the clinical procedure and interpretation of fluorescein angiography.
Iris Angiography. Intravenous injection of fluorescein can be useful for visualization of iris tumors and vessel abnormalities such as rubeosis irides. After injection into
Table 16-3
Results of Intraocular Pressure Readings for Ophthetic and Fluress in 100 Eyes
|
Mean Tonometric |
|
|
|
|
|
Readings (mm Hg) |
Standard Deviation |
Standard Error |
R |
Regression |
|
|
|
|
|
|
Fluress |
18.03 |
4.27 |
0.427 |
0.552 |
y = 0.45 + 0.59x |
Ophthetic |
11.02 |
4.53 |
0.453 |
|
|
|
|
|
|
|
|
Reprinted with permission from Bright DC, Potter JW, Allen DC, et al. Goldmann applanation tonometry without fluorescein. Am J Optom Physiol Optics 1981;58:1120–1126.
288 CHAPTER 16 Dyes
the antecubital vein, the dye first appears in the radial vessels of the iris, which are demonstrated as linear spokes with slow leakage. The amount of iris pigmentation and the pattern of its distribution affect the amount of detail observed in a normal iris angiogram. Blue irides generally show the vessels in greater detail than do brown irides. An adapter mounted in front of a fundus camera lens has rendered possible more complete visualization of the vascular structure in heavily pigmented irides.
Aqueous Flow. Changes in the concentration of fluorescein in the anterior chamber after intravenous injection were measured as early as 1950. Using a slit lamp or objective fluorophotometer, the time course of the fluorescence in the circulating blood and the anterior chamber can be determined in humans. The rate of aqueous flow is approximately 1.5% to 2.0% of the volume of the anterior chamber per minute. Following the early work other methods were devised to measure aqueous turnover, and all have given comparable results. Anterior chamber fluorometry is also useful in monitoring inflammation after oral or injected fluorescein.
Vitreous Fluorophotometry
Vitreous fluorophotometry is a noninvasive quantitative method for measuring small amounts of fluorescein in various ocular compartments, including assessment of the blood–retinal barrier. Both slit-lamp–based and objective scanning fluorometers have been used to characterize the fluorescence of the vitreous in health and disease.
Because the normal blood–retinal barrier resists various substances, including fluorescein, the presence of fluorescein in the vitreous humor indicates a functional breakdown of this barrier. Although physiologic factors and instrument artifacts can influence vitreous fluorescence, this technique has been used to detect retinal vascular disease, especially in diabetes.The procedure has also been used to study the integrity of the blood–retinal barrier in various other diseases, including retinitis pigmentosa, optic neuritis, and essential hypertension.
Oral Fluorescein Angioscopy
Because the integrity of the normal ocular physiologic barriers to fluorescein depends less on dye administration velocity than on certain other parameters, such as retinal circulation time, fluorescein has also been administered by mouth to study posterior pole lesions. The oral procedure in adults usually involves administering 1 to 2 g of fluorescein powder or three vials of 10% injectable fluorescein mixed in a citrus drink over ice. In children, fruit juice containing 1 ml of a 10% fluorescein solution per 20 ml juice per 5 kg body weight has been used to determine macular leakage after removal of congenital cataracts. The dye begins to appear in the fundus in approximately 15 minutes, but maximal fluorescence is not obtained until 45 to 60 minutes after ingestion.
The oral route of administration yields adequate clinical angiograms in approximately 97% of cases and has the advantage that side effects are rare.
Adverse Reactions
Studies of humans undergoing fluorescein angiograms indicate an incidence range of adverse effects ranging from 1.1% to 10%. The most common mild adverse reaction is nausea, accompanied less frequently by vomiting. The nausea usually occurs 15 to 30 seconds after injection and subsides within several minutes. Moderate adverse reactions include fainting, localized reactions, and urticaria (hives), although no severe adverse reactions were reported. Interestingly, in patients with a history of adverse reaction to injected fluorescein, the incidence of adverse reactions becomes nearly 50%, suggesting that careful history and medical monitoring of these patients are imperative.A significant adverse effect that can occur with intravenous fluorescein injection includes pain at the site of injection,especially if the dye becomes extravasated. Patients should be advised that intravenous fluorescein temporarily discolors both skin and urine and can appear in breast milk for up to 76 hours after administration. Itching, discomfort, or nausea was found in 1.7% of 1,787 patients taking oral fluorescein for fundus angiography, approximately the same percentage as for injected fluorescein.
Adverse effects associated with topical fluorescein and anesthetic–fluorescein combinations are usually limited to transient irritation of the cornea or conjunctiva.
Contraindications
Because of the possibility of adverse reactions a family and personal history of allergies, and especially prior angiographic procedures, should be obtained from every patient undergoing fluorescein angiography. Appropriate emergency kits need to be available that range from a minimum of oxygen, cardiopulmonary resuscitation equipment, antihistamine, and smelling salts to a full crash cart.
Because topically administered fluorescein discolors soft contact lenses, the eye should be thoroughly irrigated with sterile saline until the tears show no discoloration or a less absorbent dye such as fluorexon (see below) should be used.
FLUOREXON
Because fluorescein sodium can penetrate into many hydrogel contact lenses, the lenses become discolored, which raises bacterial growth issues and renders the lenses cosmetically objectionable. In addition, the boundary between lens and tears becomes obscured, which precludes the use of fluorescein in soft contact lens fitting. Fluorexon, a molecule similar in fluorescent characteristics to that of fluorescein, is less readily absorbed by the soft lens material, which renders it useful in fitting and evaluating soft and hybrid design lenses.
