DICOM PS3.16 2020a - Content Mapping Resource Page 431
Table TID 10003B. Irradiation Event X-Ray Source Data
NL Rel with |
VT |
Concept Name |
VM |
Req |
Condition |
Value Set Constraint |
Parent |
|
|
|
Type |
|
|
1 |
NUM |
EV (113738, DCM, "Dose |
1 |
MC IF TID 10001 Row 2UNITS = EV (Gy, |
||
|
|
(RP) ") |
|
|
= (113704, DCM, |
UCUM, "Gy") |
"Projection X-Ray") AND any of the valuesofTID(10001) Row 18 are not (113858, DCM, "MPPS Content")
2 |
TEXT |
EV (113780, DCM, |
1 |
|
|
"Reference Point |
|
|
|
Definition") |
|
3 |
CODE |
EV (113780, DCM, |
1 |
|
|
"Reference Point |
|
|
|
Definition") |
|
4 |
NUM |
EV (111631, DCM, |
1 |
|
|
"AverageGlandularDose") |
|
5 |
CODE |
EV (113732, DCM, "Fluoro 1 |
|
|
|
Mode") |
|
6 |
NUM |
EV (113791, DCM, "Pulse |
1 |
|
|
Rate") |
|
7 |
NUM |
EV (113768, DCM, |
1 |
|
|
"Number of Pulses") |
|
8 > HAS |
CODE |
EV (121401, DCM, |
1 |
CONCEPT |
"Derivation") |
|
|
MOD |
|
|
|
MC IF Row 1 is present and Row 3 is not present
MC IF Row 1 is presentDCID 10025 “Radiation and Row 2 is not Dose Reference Points” present
MC IFF TID 10001 RowUNITS = EV (mGy, 2 = (71651007, SCT,UCUM, "mGy") "Mammography")
UC IFF TID 10003 RowDCID 10004 “Fluoro
7value=(44491008,Modes”
SCT, "Fluoroscopy")
MC IFF Row 5 value = UNITS = EV ({pulse}/s, (113631, DCM, UCUM, "pulse/s") "Pulsed")
MC IFF Row 5 is not UNITS = EV (1, UCUM, present or Row 5 is"no units")
present and equals (113631, DCM, "Pulsed")
MC IFFcountofpulsesinEV (414135002, SCT,
Row 7 is estimated "Estimated")
9 |
NUM |
EV (113793, DCM, "Pulse 1-n |
U |
UNITS = EV (ms, |
|
|
|
Width") |
|
|
UCUM, "ms") |
10 |
NUM |
EV (113742, DCM, |
1 |
U |
UNITS = EV (s, UCUM, |
|
|
"Irradiation Duration") |
|
|
"s") |
11 |
NUM |
EV (113733, DCM, "KVP") 1-n |
M |
UNITS=EV(kV,UCUM, |
|
|
|
|
|
|
"kV") |
12 |
NUM |
EV (113734, DCM, "X-Ray1-n |
MC IF Row 15 is not |
UNITS = EV (mA, |
|
|
|
Tube Current") |
|
present |
UCUM, "mA") |
13 |
NUM |
EV (113767, DCM, |
1 |
U |
UNITS = EV (mA, |
|
|
"Average X-Ray Tube |
|
|
UCUM, "mA") |
|
|
Current") |
|
|
|
14 |
NUM |
EV (113824, DCM, |
1 |
MC IF Row 15 is not |
UNITS = EV (ms, |
|
|
"Exposure Time") |
|
present |
UCUM, "ms") |
15 |
NUM |
EV (113736, DCM, |
1-n |
MC IF Row 12 or 14 is UNITS = EV (uA.s, |
|
|
|
"Exposure") |
|
not present |
UCUM, "uA.s") |
16 |
NUM |
EV (113766, DCM, "Focal |
1 |
U |
UNITS = EV (mm, |
|
|
Spot Size") |
|
|
UCUM, "mm") |
- Standard -

Page 432 |
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
||||
|
NL Rel with |
VT |
Concept Name |
VM |
Req |
Condition |
Value Set Constraint |
|
|
|
Parent |
|
|
|
Type |
|
|
17 |
|
|
CODE |
EV (111632, DCM, "Anode 1 |
U |
|
DCID 10016 “Anode |
|
|
|
|
|
Target Material") |
|
|
|
Target Material” |
18 |
|
|
CONTAINEREV (113771, DCM, "X-Ray1-n |
U |
|
|
||
|
|
|
|
Filters") |
|
|
|
|
19 |
> |
CONTAINSCODE |
EV (113772, DCM, "X-Ray 1 |
U |
|
DCID 10007 “X-Ray |
||
|
|
|
|
Filter Type") |
|
|
|
Filter Types” |
20 |
> |
CONTAINSCODE |
EV (113757, DCM, "X-Ray 1 |
U |
|
DCID 10006 “X-Ray |
||
|
|
|
|
Filter Material") |
|
|
|
Filter Materials” |
21 |
> |
CONTAINSNUM |
EV (113758, DCM, "X-Ray 1 |
U |
|
UNITS = EV (mm, |
||
|
|
|
|
FilterThicknessMinimum") |
|
|
|
UCUM, "mm") |
22 |
> |
CONTAINSNUM |
EV (113773, DCM, "X-Ray 1 |
U |
|
UNITS = EV (mm, |
||
|
|
|
|
FilterThicknessMaximum") |
|
|
|
UCUM, "mm") |
23 |
|
|
NUM |
EV (113790, DCM, |
1 |
U |
|
UNITS = EV (m2, |
|
|
|
|
"Collimated Field Area") |
|
|
|
UCUM, "m2") |
24 |
|
|
NUM |
EV (113788, DCM, |
1 |
U |
|
UNITS = EV (mm, |
|
|
|
|
"Collimated Field Height") |
|
|
|
UCUM, "mm") |
25 |
|
|
NUM |
EV (113789, DCM, |
1 |
U |
|
UNITS = EV (mm, |
|
|
|
|
"Collimated Field Width") |
|
|
|
UCUM, "mm") |
26 |
|
|
CODE |
EV (111635, DCM, "X-Ray1-n |
U |
|
DCID 10017 “X-Ray |
|
|
|
|
|
Grid") |
|
|
|
Grid” |
27 |
|
|
INCLUDE |
DTID 1021 “Device |
1 |
MC Required if the |
$DeviceProcedureRole |
|
|
|
|
|
Participant” |
|
|
irradiating device is= EV (113859, DCM, |
|
|
|
|
|
|
|
|
not the recording |
"Irradiating Device") |
|
|
|
|
|
|
|
device. |
|
Content Item Descriptions |
|
|
|
|
|
|||
Row 1 |
Dose applied by this irradiation event, relative to defined reference point. |
|
||||||
Row 7 |
If a precise count of pulses is not available, an estimated number shall be provided, and the Row 8 Concept |
|||||||
|
|
Modifier shall indicate "Estimated" |
|
|
|
|
||
Row 9 |
Pulse width as measured/recorded by the system, either as a single total value, or as multiple values. If |
|||||||
|
|
multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". |
||||||
Row 11 |
KVP value as measured/recorded by system, either as a single mean value, or as multiple values. If multiple |
|||||||
|
|
values are provided, their number shall match the value in Row 7 "Number of Pulses". |
||||||
Row 12 |
Tube current as measured/recorded by system, either as a single mean value, or as multiple values. If |
|||||||
|
|
multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". |
||||||
Row 14 |
Exposure time as measured/recorded by the system. |
|
|
|||||
Row 15 |
Exposure as measured/recorded by system, either as a single total value, or as multiple values. If multiple |
|||||||
|
|
values are provided, their number shall match the value in Row 7 "Number of Pulses". The Exposure will be |
||||||
|
|
affected by the shape of the pulse and other factors, and may not be a simple multiplication of tube current |
||||||
|
|
and exposure time. |
|
|
|
|
||
Row 18 |
If one or more Filter(s) were applied during this irradiation event |
|
||||||
Row 23 |
Collimated area at the receptor plane. |
|
|
|
|
|||
Row 27 |
The device that produced the irradiation in this Irradiation Event. I.e., the X-Ray source. |
This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10001 “Projection X-Ray Radiation Dose” Row 5, which in turn may be absent if identical to the content in the Enhanced General Equipment Module.
- Standard -

DICOM PS3.16 2020a - Content Mapping Resource |
Page 433 |
TID 10003C Irradiation Event X-Ray Mechanical Data
This Template contains data that is expected to be available to the gantry or mechanical component of the equipment.
Type: |
Extensible |
Order: |
Non-Significant |
Root: |
No |
Table TID 10003C. Irradiation Event X-Ray Mechanical Data
NL Rel |
VT |
Concept Name |
VM |
Req Condition |
Value Set Constraint |
with |
|
|
|
Type |
|
Parent |
|
|
|
|
|
1 |
CODE |
EV (113956, DCM, "CR/DR |
1 |
U |
DCID 10031 “CR/DR |
|
|
Mechanical Configuration") |
|
|
Mechanical Configuration” |
2 |
NUM |
EV (112011, DCM, "Positioner |
1 |
UC XOR Row 6 UNITS = EV (deg, UCUM, |
|
|
|
Primary Angle") |
|
|
"deg") |
3 |
NUM |
EV (112012, DCM, "Positioner |
1 |
UC XOR Row 6 UNITS = EV (deg, UCUM, |
|
|
|
Secondary Angle") |
|
|
"deg") |
4 |
NUM |
EV (113739, DCM, "Positioner |
1 |
UC IFF TID |
UNITS = EV (deg, UCUM, |
|
|
Primary End Angle") |
|
10003 Row 7"deg") |
|
|
|
|
|
value = |
|
|
|
|
|
(113613, |
|
|
|
|
|
DCM, |
|
|
|
|
|
"Rotational |
|
|
|
|
|
Acquisition") |
|
5 |
NUM |
EV (113740, DCM, "Positioner |
1 |
UC IFF TID |
UNITS = EV (deg, UCUM, |
|
|
Secondary End Angle") |
|
10003 Row 7"deg") |
|
|
|
|
|
value = |
|
|
|
|
|
(113613, |
|
|
|
|
|
DCM, |
|
|
|
|
|
"Rotational |
|
|
|
|
|
Acquisition") |
|
6 |
NUM |
EV (113770, DCM, "Column |
1 |
UC XOR Rows 2,UNITS = EV (deg, UCUM, |
|
|
|
Angulation") |
|
3 |
"deg") |
7 |
NUM |
EV(113754,DCM,"TableHeadTilt |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Angle") |
|
|
"deg") |
8 |
NUM |
EV (113755, DCM, "Table |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Horizontal Rotation Angle") |
|
|
"deg") |
9 |
NUM |
EV (113756, DCM, "Table Cradle |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Tilt Angle") |
|
|
"deg") |
10 |
NUM |
EV (111633, DCM, "Compression |
1 |
U |
UNITS = EV (mm, UCUM, |
|
|
Thickness") |
|
|
"mm") |
10a |
NUM |
EV (111647, DCM, "Compression |
1 |
U |
UNITS = EV (N, UCUM, |
|
|
Force") |
|
|
"Newton") |
10b |
NUM |
EV (111648, DCM, "Compression |
1 |
U |
UNITS = EV (kPa, UCUM, |
|
|
Pressure") |
|
|
"kilopascal") |
10c |
NUM |
EV (111649, DCM, "Compression |
1 |
U |
UNITS = EV (mm2, UCUM, |
|
|
Contact Area") |
|
|
"mm2") |
11 |
NUM |
DCID 10008 “Dose Related |
1-n |
U |
UNITS = EV (mm, UCUM, |
|
|
Distance Measurements” |
|
|
"mm") |
12 |
NUM |
EV (128757, DCM, "Positioner |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Isocenter Primary Angle") |
|
|
"deg") |
- Standard -

Page 434 |
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
||
NL Rel |
VT |
Concept Name |
VM |
Req Condition |
Value Set Constraint |
with |
|
|
|
Type |
|
Parent |
|
|
|
|
|
13 |
NUM |
EV (128758, DCM, "Positioner |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Isocenter Secondary Angle") |
|
|
"deg") |
14 |
NUM |
EV (128759, DCM, "Positioner |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
IsocenterDetectorRotationAngle") |
|
|
"deg") |
15 |
NUM |
EV (128760, DCM, "Positioner |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Isocenter Primary End Angle") |
|
|
"deg") |
16 |
NUM |
EV (128761, DCM, "Positioner |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Isocenter Secondary End Angle") |
|
|
"deg") |
17 |
NUM |
EV (128762, DCM, "Positioner |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Isocenter Detector Rotation End |
|
|
"deg") |
|
|
Angle") |
|
|
|
18 |
NUM |
EV(128763,DCM,"TableHeadTilt |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
End Angle") |
|
|
"deg") |
19 |
NUM |
EV (128764, DCM, "Table |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Horizontal Rotation End Angle") |
|
|
"deg") |
20 |
NUM |
EV (128765, DCM, "Table Cradle |
1 |
U |
UNITS = EV (deg, UCUM, |
|
|
Tilt End Angle") |
|
|
"deg") |
Content Item Descriptions |
|
|
|
|
|
Row 2 |
Angle in patient's "equatorial" plane (LAO to RAO). For dynamically changing angle during the event, |
||||
|
the start value shall be provided. Equivalent to (0018,1510) in an image instance. |
||||
Row 3 |
Angle in patient's "sagittal" plane (CRAN to CAUD). For dynamically changing angle during the event, |
||||
|
the start value shall be provided. Equivalent to (0018,1511) in an image instance. |
||||
Row 4 |
In case of motion during irradiation event, Positioner Primary ending angle |
||||
Row 5 |
In case of motion during irradiation event., Positioner Secondary ending angle |
||||
Row 6 |
Column device Angle in equipment based coordinates |
|
|||
Rows 12 to 20 |
Refer to the definitions of the X-Ray Isocenter Reference System (“X-Ray Isocenter Reference System |
||||
|
Macro” in PS3.3). |
|
|
|
TID 10004 Accumulated Fluoroscopy and Acquisition Projection X-Ray Dose
This general Template provides detailed information on projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).
Type: |
Extensible |
Order: |
Non-Significant |
Root: |
No |
Table TID 10004. Accumulated Fluoroscopy and Acquisition Projection X-Ray Dose
NL Rel |
VT |
Concept Name |
VM |
Req |
Condition |
ValueSetConstraint |
with |
|
|
|
Type |
|
|
Parent |
|
|
|
|
|
|
1 |
NUM |
EV (113726, DCM, |
1 |
MC IFF TID 10003 Row 7 value = |
UNITS = EV (Gy.m2, |
|
|
|
"Fluoro Dose Area |
|
|
(44491008, SCT, "Fluoroscopy") forUCUM, "Gy.m2") |
|
|
|
Product Total") |
|
|
at least one irradiation event |
|
- Standard -

|
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 435 |
|||
NL Rel |
VT |
Concept Name |
VM |
Req |
Condition |
ValueSetConstraint |
with |
|
|
|
Type |
|
|
Parent |
|
|
|
|
|
|
2 |
NUM |
EV (113728, DCM, |
1 |
MC IFF TID 10003 Row 7 value = |
UNITS = EV (Gy, |
|
|
|
"Fluoro Dose (RP) Total") |
|
(44491008, SCT, "Fluoroscopy") forUCUM, "Gy") |
||
|
|
|
|
|
atleastoneirradiationeventANDany |
|
|
|
|
|
|
of the values of TID (10001) Row 18 |
|
|
|
|
|
|
are not (113858, DCM, "MPPS |
|
|
|
|
|
|
Content"). |
|
3 |
NUM |
EV (113730, DCM, "Total 1 |
MC IFF TID 10003 Row 7 value = |
UNITS = EV (s, |
||
|
|
Fluoro Time") |
|
|
(44491008, SCT, "Fluoroscopy") forUCUM, "s") |
|
|
|
|
|
|
at least one irradiation event. |
|
4 |
NUM |
EV (113727, DCM, |
1 |
MC IF any of the values of TID (10001) UNITS = EV (Gy.m2, |
||
|
|
"Acquisition Dose Area |
|
|
Row 18 are not (113858, DCM, |
UCUM, "Gy.m2") |
|
|
Product Total") |
|
|
"MPPS Content"). |
|
5 |
NUM |
EV (113729, DCM, |
1 |
MC IF any of the values of TID (10001) UNITS = EV (Gy, |
||
|
|
"Acquisition Dose (RP) |
|
|
Row 18 are not (113858, DCM, |
UCUM, "Gy") |
|
|
Total") |
|
|
"MPPS Content"). |
|
6 |
NUM |
EV (113855, DCM, "Total 1 |
MC IF any of the values of TID (10001) UNITS = EV (s, |
|||
|
|
Acquisition Time") |
|
|
Row 18 are not (113858, DCM, |
UCUM, "s") |
|
|
|
|
|
"MPPS Content"). |
|
Content Item Descriptions |
|
|
|
|
||
Rows 1-3 |
|
Fluoroscopic component only |
|
|
|
|
Row 3 |
|
Total clock time of Fluoroscopy accumulated over the defined scope of accumulation (i.e., the sum of |
||||
|
|
the Irradiation Duration values for accumulated fluoroscopy irradiation events) |
||||
Rows 4-6 |
|
Acquisition component only |
|
|
|
|
Row 6 |
|
Total clock time of acquisitions accumulated over the defined scope of accumulation (i.e., the sum of |
||||
|
|
the Irradiation Duration values for accumulated acquisition irradiation events) |
TID 10005 Accumulated Mammography X-Ray Dose
This modality specific Template provides detailed information on breast imaging projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
Table TID 10005. Accumulated Mammography X-Ray Dose
|
NL Rel with Parent VT |
Concept Name |
VM |
Req TypeConditionValue Set Constraint |
|
1 |
NUM |
EV (111637, DCM, |
1-2 |
M |
UNITS = EV (mGy, |
|
|
"Accumulated Average |
|
|
UCUM, "mGy") |
|
|
Glandular Dose") |
|
|
|
2 |
> HAS CONCEPTCODE |
EV (272741003, SCT, |
1 |
M |
DCID 6022 “Side” |
|
MOD |
"Laterality") |
|
|
|
TID 10006 Accumulated Cassette-based Projection Radiography Dose
This Template provides information on Projection Radiography dose values accumulated on Cassette-based systems over one or more irradiation events (typically a study or a performed procedure step) from the same equipment.
Type: |
Extensible |
Order: |
Non-Significant |
- Standard -

Page 436 |
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
||||
Root: |
|
No |
|
|
|
|
|
|
Table TID 10006. Accumulated Cassette-Based Projection Radiography Dose |
||||||
NL Rel with VT |
Concept Name |
VMReq Type |
Condition |
Value Set Constraint |
|||
|
Parent |
|
|
|
|
|
|
1 |
CODE |
EV (113947, DCM, |
1 |
MC |
IFTID10001Row8isabsentDCID10030“Detector |
||
|
|
"Detector Type") |
|
|
orvalueis(373066001,SCT,Types” |
||
|
|
|
|
|
"Yes") |
|
|
2 |
NUM |
EV (113731, DCM, "Total 1 |
MC |
IFTID10001Row8isabsentUNITS = EV (1, |
|||
|
|
Number of Radiographic |
|
|
orvalueis(373066001,SCT,UCUM, "no units") |
||
|
|
Frames") |
|
|
"Yes") |
|
|
Content Item Descriptions |
|
|
|
|
|
|
|
Row 2 |
The number of radiographic frames recorded by the X-Ray detector or the number of exposures recorded |
||||||
|
by the X-Ray source, whichever is known to be greater. |
|
|
TID 10007 Accumulated Total Projection Radiography Dose
This Template provides information on total Projection Radiography dose values accumulated on Integrated or combined fluoro- scopy/acquisition systems over one or more irradiation events (typically a study or a performed procedure step) from the same equipment.
Type: |
Extensible |
Order: |
Non-Significant |
Root: |
No |
|
Table TID 10007. Accumulated Total Projection Radiography Dose |
NL Rel |
VT |
Concept Name |
VM |
Req |
Condition |
ValueSetConstraint |
with |
|
|
|
Type |
|
|
Parent |
|
|
|
|
|
|
1 |
NUM |
EV (113722, DCM, "Dose |
1 |
M |
|
UNITS = EV (Gy.m2, |
|
|
Area Product Total") |
|
|
|
UCUM, "Gy.m2") |
2 |
NUM |
EV (113725, DCM, "Dose |
1 |
MC IF TID 10001 Row 4 = (113958, |
UNITS = EV (Gy, |
|
|
|
(RP) Total") |
|
|
DCM, "Integrated Projection |
UCUM, "Gy") |
|
|
|
|
|
RadiographySystem")oranyofthe |
|
|
|
|
|
|
values of TID (10001) Row 18 are |
|
|
|
|
|
|
not (113858, DCM, "MPPS |
|
|
|
|
|
|
Content"). |
|
3 |
NUM |
EV (113737, DCM, |
1 |
U |
|
UNITS = EV (mm, |
|
|
"Distance Source to |
|
|
|
UCUM, "mm") |
|
|
Reference Point") |
|
|
|
|
4 |
NUM |
EV (113731, DCM, "Total |
1 |
U |
|
UNITS = EV (1, |
|
|
Number of Radiographic |
|
|
|
UCUM, "no units") |
|
|
Frames") |
|
|
|
|
5 |
CODE |
EV (113780, DCM, |
1 |
MC IF any of (113725, DCM, "Dose |
DCID 10025 |
|
|
|
"Reference Point |
|
|
(RP) Total"), (113728, DCM, |
“Radiation Dose |
|
|
Definition") |
|
|
"Fluoro Dose (RP) Total") or |
Reference Points” |
|
|
|
|
|
(113729, DCM, "Acquisition Dose |
|
|
|
|
|
|
(RP) Total") are present, and Row |
|
|
|
|
|
|
6 is not present. |
|
- Standard -

|
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 437 |
|||||
NL Rel |
VT |
Concept Name |
VM |
|
Req |
Condition |
ValueSetConstraint |
|
with |
|
|
|
|
Type |
|
|
|
Parent |
|
|
|
|
|
|
|
|
6 |
TEXT |
EV (113780, DCM, |
1 |
|
MC IF any of (113725, DCM, "Dose |
|
||
|
|
"Reference Point |
|
|
|
|
(RP) Total"), (113728, DCM, |
|
|
|
Definition") |
|
|
|
|
"Fluoro Dose (RP) Total") or |
|
|
|
|
|
|
|
|
(113729, DCM, "Acquisition Dose |
|
|
|
|
|
|
|
|
(RP) Total") are present, and Row |
|
|
|
|
|
|
|
|
5 is not present. |
|
Content Item Descriptions |
|
|
|
|
|
|
||
Row 1 |
|
Accumulated Dose Area Product |
|
|
||||
Row 2 |
|
Accumulated dose relative to reference point. |
|
|||||
Row 3 |
|
A single value for Radiography systems calculating reference point dose based on fixed distance. |
||||||
Row 5 |
|
A coded definition of the Reference Point (RP) used for RP-related dose values. |
||||||
Row 6 |
|
A text definition of the Reference Point (RP) used for RP-related dose values. |
||||||
CT Radiation Dose SR IOD Templates |
|
|
||||||
The Templates that comprise the CT Radiation Dose SR are interconnected as in Figure A-15. |
|
|||||||
|
|
|
TID 10011 |
|
TID 1002 |
|
|
|
|
|
|
CT Radiation Dose |
|
|
Observer |
|
|
|
|
|
|
|
Context |
|
|
|
|
|
|
|
|
|
|
|
TID 10012
CT Accumulated
Dose Data
TID 10013
CT Irradiation
Event Data
Figure A-15. CT Radiation Dose SR IOD Template Structure
TID 10011 CT Radiation Dose
ThisTemplatedefinesacontainer(theroot)withsubsidiaryContentItems,eachofwhichcorrespondstoasingleCTX-Rayirradiation event entry. There is a defined recording observer (the system or person responsible for recording the log, generally the system). Accumulated values shall be kept for a whole Study or at least a part of a Study, if the Study is divided in the workflow of the examin- ation, or a performed procedure step. Multiple CT Radiation Dose objects may be created for one Study.
Type: |
Extensible |
Order: |
Significant |
Root: |
Yes |
|
Table TID 10011. CT Radiation Dose |
|
NL Rel with Parent VT |
Concept Name |
VM |
Req ConditionValue Set Constraint |
||
|
|
|
|
|
Type |
|
1 |
|
CONTAINEREV (113701, DCM, "X-Ray |
1 |
M |
Root node |
|
|
|
|
Radiation Dose Report") |
|
|
|
1b |
> |
HAS CONCEPTINCLUDE |
DTID 1204 “Language of |
1 |
U |
|
|
|
MOD |
Content Item and |
|
|
|
|
|
|
Descendants” |
|
|
|
2 |
> |
HAS CONCEPTCODE |
EV (121058, DCM, |
1 |
M |
EV (77477000, SCT, |
|
|
MOD |
"Procedure reported") |
|
|
"ComputedTomography |
|
|
|
|
|
|
X-Ray") |
- Standard -

Page 438 |
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
||||
|
NL Rel with Parent VT |
Concept Name |
VM |
Req ConditionValue Set Constraint |
|||
|
|
|
|
|
|
Type |
|
3 |
>> |
HAS CONCEPTCODE |
EV (363703001, SCT, "Has |
1 |
M |
DCID 3629 “Procedure |
|
|
|
MOD |
|
Intent") |
|
|
Intent” |
4 |
> |
|
INCLUDE |
DTID 1002 “Observer |
1-n |
M |
|
|
|
|
|
Context” |
|
|
|
5 |
> |
HAS OBS |
DATETIME EV (113809, DCM, "Start of |
1 |
M |
|
|
|
|
CONTEXT |
|
X-Ray Irradiation") |
|
|
|
6 |
> |
HAS OBS |
DATETIME EV (113810, DCM, "End of |
1 |
M |
|
|
|
|
CONTEXT |
|
X-Ray Irradiation") |
|
|
|
7 |
> |
HAS OBS |
CODE |
EV(113705,DCM,"Scopeof |
1 |
M |
DCID 10000 “Scope of |
|
|
CONTEXT |
|
Accumulation") |
|
|
Accumulation” |
8 |
>> |
HAS |
UIDREF |
DCID 10001 “UID Types” |
1 |
M |
|
|
|
PROPERTIES |
|
|
|
|
|
9 |
> |
CONTAINS |
INCLUDE |
DTID 10012 “CT |
1 |
M |
|
|
|
|
|
Accumulated Dose Data” |
|
|
|
10 |
> |
CONTAINS |
INCLUDE |
DTID 10013 “CT Irradiation |
1-n |
M |
|
|
|
|
|
Event Data” |
|
|
|
11 |
> |
CONTAINS |
TEXT |
EV (121106, DCM, |
1 |
U |
|
|
|
|
|
"Comment") |
|
|
|
12 |
> |
CONTAINS |
CODE |
EV (113854, DCM, "Source |
1-n |
M |
DCID 10021 “Source of |
|
|
|
|
of Dose Information") |
|
|
CT Dose Information” |
13 |
> |
CONTAINS |
INCLUDE |
DTID 1020 “Person |
1 |
U |
$PersonProcedureRole |
|
|
|
|
Participant” |
|
|
= EV (113850, DCM, |
|
|
|
|
|
|
|
"IrradiationAuthorizing") |
Content Item Descriptions |
|
|
|
|
|
||
Row 4 |
The observer context may include both a Person Observer identification, as well as the identity of the |
||||||
|
|
equipment providing the values for the irradiation event (Device Observer identification), if not inherited. |
|||||
Row 5 |
Start, DateTime of the first CT Irradiation Event of the accumulation |
|
|||||
Row 6 |
End, DateTime of the last CT Irradiation Event of the accumulation |
|
|||||
Row 12 |
The primary source of information from which this dose object was constructed. |
||||||
Row 13 |
Thephysicianresponsiblefordeterminingthattheirradiatingprocedurewasappropriatefortheindications. |
||||||
|
|
ThevaluemaycomefromRequestingPhysician(0032,1032),RequestingPhysicianIdentificationSequence |
|||||
|
|
(0032,1031) or somewhere else based on hospital policies. |
|
|
TID 10012 CT Accumulated Dose Data
This general Template provides detailed information on CT X-Ray dose value accumulations over several irradiation events from the same equipment and over the scope of accumulation specified for the report (typically a Study or a Performed Procedure Step).
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
|
Table TID 10012. CT Accumulated Dose Data |
NL Rel with |
VT |
Concept Name |
VM Req |
Condition Value Set Constraint |
Parent |
|
|
Type |
|
1 |
CONTAINEREV (113811, DCM, "CT |
1 M |
|
AccumulatedDoseData")
- Standard -

|
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 439 |
||||
|
NL |
Rel with |
VT |
Concept Name |
VM Req |
Condition |
Value Set Constraint |
|
|
|
Parent |
|
|
|
Type |
|
|
2 |
> |
CONTAINS |
NUM |
EV(113812,DCM,"Total 1 |
M |
|
UNITS = EV ({events}, |
|
|
|
|
|
Number of Irradiation |
|
|
|
UCUM, "events") |
|
|
|
|
Events") |
|
|
|
|
3 |
> |
CONTAINS |
NUM |
EV (113813, DCM, "CT |
1 |
M |
|
UNITS = EV (mGy.cm, |
|
|
|
|
Dose Length Product |
|
|
|
UCUM, "mGy.cm") |
|
|
|
|
Total") |
|
|
|
|
4 |
> |
CONTAINS |
NUM |
EV (113814, DCM, "CT |
1 |
U |
|
UNITS = EV (mSv, |
|
|
|
|
Effective Dose Total") |
|
|
|
UCUM, "mSv") |
5 |
>> |
HAS |
TEXT |
EV (121406, DCM, |
1 |
MC XOR row 6 |
|
|
|
|
PROPERTIES |
"Reference Authority") |
|
|
|
|
|
6 |
>> |
HAS |
CODE |
EV (121406, DCM, |
1 |
MC XOR row 5 |
DCID 10015 “CT Dose |
|
|
|
PROPERTIES |
"Reference Authority") |
|
|
|
Reference Authorities” |
|
7 |
>> |
HAS |
CODE |
EV (370129005, SCT, |
1 |
M |
|
DCID 10011 “Effective |
|
|
CONCEPT |
|
"Measurement Method") |
|
|
|
Dose Evaluation |
|
|
MOD |
|
|
|
|
|
Method” |
8 |
>> |
HAS |
TEXT |
EV (113815, DCM, |
1 |
MC IF the value of row 7 |
|
|
|
|
PROPERTIES |
"Patient Model") |
|
|
equals(113800,DCM, |
||
|
|
|
|
|
|
|
"DLP to E conversion |
|
|
|
|
|
|
|
|
via MC computation") |
|
|
|
|
|
|
|
|
or equals (113801, |
|
|
|
|
|
|
|
|
DCM, "CTDIfreeair to |
|
|
|
|
|
|
|
|
E conversion via MC |
|
|
|
|
|
|
|
|
computation") |
|
9 |
>> |
HAS |
CONTAINEREV (113816, DCM, |
1 |
MC IF the value of row 7 |
|
||
|
|
PROPERTIES |
"ConditionEffectiveDose |
|
equals(113802,DCM, |
|||
|
|
|
|
measured") |
|
|
"DLP to E conversion |
|
|
|
|
|
|
|
|
via measurement") or |
|
|
|
|
|
|
|
|
equals(113803,DCM, |
|
|
|
|
|
|
|
|
"CTDIfreeair to E |
|
|
|
|
|
|
|
|
conversion via |
|
|
|
|
|
|
|
|
measurement") |
|
10 |
>>> CONTAINS |
TEXT |
EV (113817, DCM, |
1 |
M |
|
|
|
|
|
|
|
"Effective Dose Phantom |
|
|
|
|
|
|
|
|
Type") |
|
|
|
|
11 |
>>> CONTAINS |
TEXT |
EV (113818, DCM, |
1 |
M |
|
|
|
|
|
|
|
"Dosimeter Type") |
|
|
|
|
12 |
> |
CONTAINS |
TEXT |
EV (121106, DCM, |
1 |
U |
|
|
|
|
|
|
"Comment") |
|
|
|
|
13 |
> |
CONTAINS |
INCLUDE |
DTID 1021 “Device |
1 |
MC Required if the |
$DeviceProcedureRole |
|
|
|
|
|
Participant” |
|
|
irradiating device is |
= EV (113859, DCM, |
|
|
|
|
|
|
|
not the recording |
"Irradiating Device") |
|
|
|
|
|
|
|
device and the dose |
|
|
|
|
|
|
|
|
was accumulated on |
|
|
|
|
|
|
|
|
a single device. |
|
Content Item Descriptions
- Standard -

Page 440 |
DICOM PS3.16 2020a - Content Mapping Resource |
Rows 5, 6 |
Total Number of CT irradiation events. |
|
A CT irradiation event is one continuous irradiation procedure and is defined through consistent acquisition |
|
parameters. |
|
In the case of dose modulation the calculations are based on the effective parameters (e.g., the effective mA |
|
recorded in the Mean X-Ray Tube Current), and these acquisition parameters are consistent. |
Rows 5, 6 |
The Dose Length Product (DLP) is calculated for every irradiation event. The Dose Length Product Total is the |
|
sum of the DLP values. The calculation is based on the CTDIvol result of each irradiation event. |
Rows 5, 6 |
Effective dose (E, in units of mSv) evaluated as a total over the scope is defined in Row 6 of Template TID |
|
10011 “CT Radiation Dose”. |
|
Effective dose is defined by the reference in Rows 5 or 6 of this Template. |
|
It may be calculated from a product of DLP and an 'Effective Dose Conversion Factor' (E/DLP). Or it may be |
|
calculatedfromaproductoftheMeanCTDIfreeairandtheratioE/CTDIfreeair.TheratiosE/DLPorE/CTDIfreeair |
|
may be evaluated either from computer simulations applying Monte Carlo (MC) sampling techniques or from |
|
dosimetric measurements in an anthropomorphic phantom, e.g., the Alderson-Rando phantom.. The specific |
|
method used is identified in Rows 7 through 11. |
Row 5 - 6 |
ReferenceofthebasepublicationdefiningtheEffectiveDose,eitherasacodedvalue,oratextualbibliographic |
|
reference. ICRP Publications shall be referenced using their assigned coded values. |
Row 7 |
Description of the method used for Effective Dose evaluations. |
Row 8 |
Description of the reference-patient mathematical or computational model used when Effective Dose is derived |
|
viaMonteCarlosimulationsofradiationtransportinsuchmodels.Examplesofpublicationsthatspecifyparticular |
|
referencepatientmodelsareNUREG/CR-1159,ORNL/NUREG/TM-367(1980);NRPB-R186(1985);GSF-Bericht |
|
S-885 (1986); Fill et al., Health Physics Vol. 86 (3): 253-272 (2004). |
Row 9 |
Description of the condition Effective Dose measured |
Row 10 |
Type of Effective Dose phantom used, e.g., Alderson-Rando |
Row 11 |
Type of dosimeter used, e.g., TLD (Thermo Luminescence Dosimeter) |
Row 13 |
ThedevicethatproducedtheirradiationaccumulatedinthisTemplate.I.e.,theCTScanner.Thisisnotrequired |
|
to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying |
|
Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10011 “CT Radiation Dose” Row |
|
4, which in turn may be absent if identical to the content in the Enhanced General Equipment Module, or if more |
|
than one device produced the accumulated irradiation. |
TID 10013 CT Irradiation Event Data
This Template conveys the dose and equipment parameters of a single irradiation event.
ACTirradiationeventistheloadingofX-Rayequipmentcausedbyasinglecontinuousactuationoftheequipment'sirradiationswitch, from the start of the loading time of the first pulse until the loading time trailing edge of the final pulse. Any on-off switching of the ra- diation source during the event shall not be treated as separate events; rather the event includes the time between start and stop of radiation as triggered by the user, e.g., a single sequence of scanning comprised of multiple slices acquired with successive tube rotations and table increments shall be treated as a single irradiation event. Depending on the examination workflow and the anatom- icaltargetregiontheCTirradiationeventdatamaysplitintomultipleinstancesofthisTemplateforbetterdoseestimation.Theirradiation event is the "smallest" information entity to be recorded in the realm of Radiation Dose reporting. Individual Irradiation Events are describedbyasetofaccompanyingphysicalparametersthataresufficienttounderstandthe"quality"ofirradiationthatisbeingapplied. This set of parameters may be different for the various types of equipment that are able to create irradiation events.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |