DICOM PS3.16 2020a - Content Mapping Resource |
Page 381 |
Row 1 |
For an index type of measurement, the concept name of this row 1 will still be the original measurement concept |
|
|
name; it is row 2 that gives the indication that row 1 is actually an index type of measurement. When this happens, |
|
|
the measurement value of row 1 should be a value after being indexed and the measurement unit of row 1 should |
|
|
be an index type of unit. |
|
|
For example, to insert a "Stroke Volume Index" measurement to this SR object, the concept name of row 1 will be |
|
|
"StrokeVolume",itsnumericalvaluewillbethecalculationresultof"StrokeVolume/BSA"anditsunitsare"ml/cm2". |
|
Row 2 |
When this row is available, the row 1 is an index calculation of the object. |
|
TID 5225 Cardiac Ultrasound Fetal Characteristics
Contains a list of Fetus Specific characteristics.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
|
Table TID 5225. Cardiac Ultrasound Fetal Characteristics |
|
NL Rel with Parent VT |
Concept Name |
VM |
Req |
Condition |
Value Set |
||
|
|
|
|
|
|
Type |
|
Constraint |
1 |
|
|
CONTAINEREV (125015, DCM, "Fetus |
1 |
M |
|
|
|
|
|
|
|
Characteristics") |
|
|
|
|
2 |
> |
HAS OBS |
INCLUDE |
DTID 1008 “Subject |
1 |
MC |
IF this Template is |
|
|
|
CONTEXT |
|
Context, Fetus” |
|
|
invokedmorethan |
|
|
|
|
|
|
|
|
once to describe |
|
|
|
|
|
|
|
|
more than one |
|
|
|
|
|
|
|
|
fetus. |
|
3 |
> |
CONTAINS |
NUM |
EV (18185-9, LN, |
1 |
U |
|
UNITS = DCID |
|
|
|
|
"Gestational Age") |
|
|
|
7456 “Units of |
|
|
|
|
|
|
|
|
Measure for |
|
|
|
|
|
|
|
|
Age” |
4 |
> |
CONTAINS |
DATE |
EV (11778-8, LN, "EDD") |
1 |
U |
|
|
5 |
> |
CONTAINS |
NUM |
EV (8867-4, LN, "Heart |
1 |
U |
|
|
|
|
|
|
Rate") |
|
|
|
|
TID 5226 Cardiac Ultrasound Summary Section
Comments and observations of the procedure of immediate clinical interest.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
|
Table TID 5226. Cardiac Ultrasound Summary Section |
|
NL |
Rel with |
VT |
Concept Name |
VM |
Req |
Condition |
Value Set Constraint |
|
|
Parent |
|
|
|
Type |
|
|
1 |
|
|
CONTAINEREV (121111, DCM, |
1 |
M |
|
|
|
|
|
|
|
"Summary") |
|
|
|
|
2 |
> |
CONTAINS CODE |
EV (121071, DCM, |
1-n |
MC |
IF row 3 doesBCID 12248 “Cardiac |
||
|
|
|
|
"Finding") |
|
|
not exist |
Ultrasound Summary |
|
|
|
|
|
|
|
|
Codes” |
3 |
> |
CONTAINS TEXT |
EV (121071, DCM, |
1-n |
MC |
IF row 2 does |
|
|
|
|
|
|
"Finding") |
|
|
not exist |
|
- Standard -

Page 382 |
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
|||||
|
NL Rel with |
VT |
Concept Name |
VM |
Req |
Condition |
Value Set Constraint |
|
|
|
Parent |
|
|
|
Type |
|
|
4 |
> |
CONTAINS |
CODE |
EV (387713003, SCT, |
1-n |
U |
|
BCID 12247 “Pediatric, |
|
|
|
|
"Surgical Procedure") |
|
|
|
Fetal and Congenital |
|
|
|
|
|
|
|
|
Cardiac Surgical |
|
|
|
|
|
|
|
|
Interventions” |
TID 5227 Cardiac Ultrasound Fetal Summary Section |
|
|
|
|||||
Comments and observations of the procedure of immediate clinical interest. |
|
|
|
|||||
Type: |
|
Extensible |
|
|
|
|
||
Order: |
|
Significant |
|
|
|
|
||
Root: |
|
No |
|
|
|
|
|
|
|
|
Table TID 5227. Cardiac Ultrasound Fetal Summary Section |
||||||
|
NL |
Rel with |
VT |
Concept Name |
VM |
Req |
Condition |
ValueSetConstraint |
|
|
Parent |
|
|
|
Type |
|
|
1 |
|
|
CONTAINERDT (125008, DCM, |
1 |
M |
|
|
|
|
|
|
|
"Fetus Summary") |
|
|
|
|
2 |
> |
HAS OBS |
INCLUDE DTID 1008 “Subject |
1 |
MC IF this Template is |
|||
|
|
CONTEXT |
|
Context, Fetus” |
|
invoked more than |
||
|
|
|
|
|
|
once to describe |
||
|
|
|
|
|
|
more than one fetus |
||
3 |
> |
CONTAINS |
CODE |
EV (121071, DCM, |
1-n |
MC IF row 4 does not BCID 12249 “Cardiac |
||
|
|
|
|
"Finding") |
|
exist |
Ultrasound Fetal |
|
|
|
|
|
|
|
|
|
Summary Codes” |
4 |
> |
CONTAINS |
TEXT |
EV (121071, DCM, |
1-n |
MC IF row 3 does not |
||
|
|
|
|
"Finding") |
|
exist |
|
|
5 |
> |
CONTAINS |
CODE |
EV (387713003, SCT, |
1-n |
U |
|
BCID 12247 |
|
|
|
|
"Surgical Procedure") |
|
|
|
“Pediatric, Fetal and |
|
|
|
|
|
|
|
|
Congenital Cardiac |
|
|
|
|
|
|
|
|
SurgicalInterventions” |
TID 5228 Cardiac Ultrasound Fetal Measurement Section |
|
|
||||||
Type: |
|
Extensible |
|
|
|
|
||
Order: |
|
Significant |
|
|
|
|
||
Root: |
|
No |
|
|
|
|
|
|
|
|
Table TID 5228. Cardiac Ultrasound Fetal Measurement Section |
||||||
|
NL |
Rel with |
VT |
Concept Name |
VM Req Condition Value Set Constraint |
|||
|
|
Parent |
|
|
Type |
|
|
|
1 |
|
CONTAINEREV (125016, DCM, "Fetal |
1 |
M |
|
|
||
|
|
|
|
Measurements") |
|
|
|
|
2 |
> HAS OBS INCLUDE |
DTID 1008 “Subject |
1 |
MCIF this |
|
|||
|
CONTEXT |
|
Context, Fetus” |
|
Templateis |
|
||
|
|
|
|
|
|
invoked |
|
|
|
|
|
|
|
|
more than |
|
once to describe more than one fetus.
- Standard -

|
|
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
Page 383 |
||
|
NL Rel with |
VT |
Concept Name |
VM |
Req Condition |
Value Set Constraint |
||
|
|
Parent |
|
|
|
Type |
|
|
3 |
> |
CONTAINSINCLUDE |
DTID 300 “Measurement” |
1-n |
U |
$Measurement = DCID 12279 |
||
|
|
|
|
|
|
|
“CardiacUltrasoundFetalGeneral |
|
|
|
|
|
|
|
|
Measurements” |
|
4 |
> |
CONTAINSINCLUDE |
DTID 5222 “Pediatric, Fetal 1-n |
U |
$SectionSubject = EV (4432005, |
|||
|
|
|
|
and Congenital Cardiac |
|
|
SCT, "Ductus arteriosus") |
|
|
|
|
|
Ultrasound Section” |
|
|
$MeasType = DCID 12218 |
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
“Echocardiography Congenital” |
|
5 |
> |
CONTAINSINCLUDE |
DTID 5222 “Pediatric, Fetal 1-n |
U |
$SectionSubject = EV |
|||
|
|
|
|
and Congenital Cardiac |
|
|
(367624001, SCT, "Ductus |
|
|
|
|
|
Ultrasound Section” |
|
|
venosus") |
|
|
|
|
|
|
|
|
$MeasType = DCID 12218 |
|
|
|
|
|
|
|
|
“Echocardiography Congenital” |
|
6 |
> |
CONTAINSINCLUDE |
DTID 5222 “Pediatric, Fetal 1-n |
U |
$SectionSubject=EV(50536004, |
|||
|
|
|
|
and Congenital Cardiac |
|
|
SCT, "Umbilical artery") |
|
|
|
|
|
Ultrasound Section” |
|
|
$MeasType = DCID 12218 |
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
“Echocardiography Congenital” |
|
7 |
> |
CONTAINSINCLUDE |
DTID 5222 “Pediatric, Fetal 1-n |
U |
$SectionSubject = EV |
|||
|
|
|
|
and Congenital Cardiac |
|
|
(367567000, SCT, "Umbilical |
|
|
|
|
|
Ultrasound Section” |
|
|
vein") |
|
|
|
|
|
|
|
|
$MeasType = DCID 12218 |
|
|
|
|
|
|
|
|
“Echocardiography Congenital” |
|
8 |
> |
CONTAINSINCLUDE |
DTID 5222 “Pediatric, Fetal 1-n |
U |
$SectionSubject=EV(17232002, |
|||
|
|
|
|
and Congenital Cardiac |
|
|
SCT, "Middle cerebral artery") |
|
|
|
|
|
Ultrasound Section” |
|
|
$MeasType = DCID 12218 |
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
“Echocardiography Congenital” |
Simplified Adult Echocardiography Templates
The templates that comprise the Simplified Adult Echocardiography Report are interconnected as in Figure A-10b.
TID 5300 |
|
|
TID 1001 |
Simplified Echo |
|
|
Observation |
Procedure Report |
|
|
Context |
|
|
|
TID 1204 |
|
|
|
Language of |
|
|
|
Content Item and |
|
|
|
Descendants |
|
|
|
TID 3602 |
|
|
|
Cardiovascular |
|
|
|
Patient |
|
|
|
|
|
|
|
Characteristics |
|
|
|
TID 5301 |
|
|
|
Pre-Coordinated |
|
|
|
Echo Measurement |
|
|
|
TID 5302 |
|
|
|
Post-Coordinated |
|
|
|
Echo Measurement |
|
|
|
TID 5303 |
|
|
|
Ad hoc Measurement |
|
|
|
TID 5204 |
|
|
|
Wall Motion Analysis |
Figure A-10b. Simplified Adult Echocardiography Template Structure
- Standard -

Page 384 |
DICOM PS3.16 2020a - Content Mapping Resource |
TID 5300 Simplified Echo Procedure Report
This template forms the top of a content tree that allows an ultrasound device to describe the results of an adult echocardiography imaging procedure.
The template is instantiated at the root node. It can also be included in other templates that need to incorporate echocardiography findings into another report as quoted evidence.
This template does not include an Image Library. Image Content Items in the Echo Measurement templates (for example to indicate Source of Measurement) shall be included with by-value relationships, not with by-reference relationships.
Measurements in this template (except for the Wall Motion Analysis) are collected into one of three containers, each with a specific sub-template and constraints appropriate to the purpose of the container.
•Pre-coordinated Measurements
•Are fully standardized measurements (many taken from the ASE practice guidelines).
•Each has a single pre-coordinated standard code that fully captures the semantics of the measurement.
•The only modifiers permitted are to indicate coordinates where the measurement was taken, provide a brief display label, and indicate which of a set of repeated measurements is the preferred value. Other modifiers are not permitted.
•Post-coordinated Measurements
•Are non-standardized measurements that are performed with enough regularity to merit the control and configuration to capture thefullsemanticsofthemeasurement.Forexamplethesemeasurementsmayincludethoseconfiguredonthecartbythevendor or user site. Some of these may be variants of the Pre-coordinated Measurements.
•A set of mandatory and conditional modifiers with controlled vocabularies capture the essential semantics in a uniform way.
•A single pre-coordinated code is also provided so that when the same type of measurement is encountered in the future, it is not necessarytoparseandevaluatethefullconstellationofmodifervalues.Sincethismeasurementhasnotbeenfullystandardized, the pre-coordinated code may use a private coding scheme (e.g., from the vendor or user site).
•Adhoc Measurements
•Are non-standardized measurements that do not merit the effort to track or configure all the details necessary to populate the set of modifiers required for a post-coordinated measurement.
•The measurement code describes the elementary property measured.
•Modifiers provide a brief display label and indicate coordinates where the measurement was taken. Other modifiers are not per- mitted.
For an example of this encoding and a discussion of the benefits and use cases, see Annex CCCC Populating The Simplified Echo Procedure Report Template (Informative) in PS3.17.
Type: |
Non-Extensible |
Order: |
Significant |
Root: |
Yes |
|
Table TID 5300. Simplified Echo Procedure Report |
|
NL RelwithParent VT |
Concept Name |
VM |
ReqConditionValue Set Constraint |
|
|
|
|
Type |
1 |
CONTAINEREV (125200, DCM, "Adult |
1 |
M |
|
|
|
Echocardiography Procedure |
|
|
|
|
Report") |
|
|
2 |
> HASCONCEPTINCLUDE |
DTID 1204 “Language of |
1 |
U |
|
MOD |
Content Item and |
|
|
|
|
Descendants” |
|
|
- Standard -

|
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 385 |
|||
|
NL RelwithParent VT |
Concept Name |
VM |
ReqConditionValue Set Constraint |
|||
|
|
|
|
|
|
Type |
|
3 |
> |
HAS OBS |
INCLUDE |
DTID 1001 “Observation |
1 |
M |
|
|
|
CONTEXT |
|
Context” |
|
|
|
4 |
> |
CONTAINS |
CONTAINERDT (55111-9, LN, "Current |
1 |
U |
|
|
|
|
|
|
Procedure Descriptions") |
|
|
|
5 |
>> |
CONTAINS |
CODE |
DT (125203, DCM, |
1-n |
M |
BCID12001“Ultrasound |
|
|
|
|
"Acquisition Protocol") |
|
|
Protocol Types” |
6 |
> |
CONTAINS |
CONTAINEREV(18785-6,LN,"Indications |
1 |
U |
|
|
|
|
|
|
for Procedure") |
|
|
|
7 |
>> |
CONTAINS |
CODE |
EV (121071, DCM, "Finding") 1-n |
U |
DCID 12246 “Cardiac |
|
|
|
|
|
|
|
|
UltrasoundIndicationfor |
|
|
|
|
|
|
|
Study” |
8 |
>> |
CONTAINS |
TEXT |
EV (121071, DCM, "Finding") |
1 |
U |
|
9 |
> |
CONTAINS |
INCLUDE |
DTID 3602 “Cardiovascular |
1 |
U |
|
|
|
|
|
Patient Characteristics” |
|
|
|
10 |
> |
CONTAINS |
CONTAINEREV (125301, DCM, |
1 |
M |
|
|
|
|
|
|
"Pre-coordinated |
|
|
|
|
|
|
|
Measurements") |
|
|
|
11 |
>> |
CONTAINS |
INCLUDE |
DTID 5301 “Pre-coordinated |
1-n |
M |
$Measurement = DCID |
|
|
|
|
Echo Measurement” |
|
|
12300 “Core Echo |
|
|
|
|
|
|
|
Measurements” |
|
|
|
|
|
|
|
$Preferred = DCID |
|
|
|
|
|
|
|
12301 “Measurement |
|
|
|
|
|
|
|
Selection Reasons” |
12 |
> |
CONTAINS |
CONTAINEREV (125302, DCM, |
1 |
M |
|
|
|
|
|
|
"Post-coordinated |
|
|
|
|
|
|
|
Measurements") |
|
|
|
13 |
>> |
CONTAINS |
INCLUDE |
DTID 5302 “Post-coordinated 1-n |
U |
$Preferred = DCID |
|
|
|
|
|
Echo Measurement” |
|
|
12301 “Measurement |
|
|
|
|
|
|
|
Selection Reasons” |
14 |
> |
CONTAINS |
CONTAINEREV (125303, DCM, "Adhoc |
1 |
M |
|
|
|
|
|
|
Measurements") |
|
|
|
15 |
>> |
CONTAINS |
INCLUDE |
DTID 5303 “Adhoc |
1-n |
U |
$Property=DCID12304 |
|
|
|
|
Measurement” |
|
|
“Echo Measured |
|
|
|
|
|
|
|
Properties” |
16 |
> |
CONTAINS |
INCLUDE |
DTID 5204 “Wall Motion |
1-n |
U |
$Procedure = DT |
|
|
|
|
Analysis” |
|
|
(35757004, SCT, |
|
|
|
|
|
|
|
"Echocardiography for |
|
|
|
|
|
|
|
Determining Ventricular |
|
|
|
|
|
|
|
Contraction") |
17 |
> |
CONTAINS |
CONTAINEREV (125310, DCM, "Staged |
1 |
U |
|
|
|
|
|
|
Measurements") |
|
|
|
18 |
>> |
HAS ACQ |
CODE |
EV (18139-6, LN, "Stage") |
1 |
M |
BCID 3207 “Stress Test |
|
|
CONTEXT |
|
|
|
|
Procedure Phases” |
19 |
>> |
CONTAINS |
CONTAINEREV (125301, DCM, |
1 |
M |
|
|
|
|
|
|
"Pre-coordinated |
|
|
|
|
|
|
|
Measurements") |
|
|
|
- Standard -

Page 386 |
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
||||
|
NL RelwithParent VT |
Concept Name |
VM |
ReqConditionValue Set Constraint |
|||
|
|
|
|
|
|
Type |
|
20 |
>>> |
CONTAINS |
INCLUDE |
DTID 5301 “Pre-coordinated |
1-n |
U |
$Measurement = DCID |
|
|
|
|
Echo Measurement” |
|
|
12300 “Core Echo |
|
|
|
|
|
|
|
Measurements” |
|
|
|
|
|
|
|
$Preferred = DCID |
|
|
|
|
|
|
|
12301 “Measurement |
|
|
|
|
|
|
|
Selection Reasons” |
21 |
>> |
CONTAINS |
CONTAINEREV (125302, DCM, |
1 |
M |
|
|
|
|
|
|
"Post-coordinated |
|
|
|
|
|
|
|
Measurements") |
|
|
|
22 |
>>> |
CONTAINS |
INCLUDE |
DTID 5302 “Post-coordinated 1-n |
U |
$Preferred = DCID |
|
|
|
|
|
Echo Measurement” |
|
|
12301 “Measurement |
|
|
|
|
|
|
|
Selection Reasons” |
23 |
>> |
CONTAINS |
CONTAINEREV (125303, DCM, "Adhoc |
1 |
M |
|
|
|
|
|
|
Measurements") |
|
|
|
24 |
>>> |
CONTAINS |
INCLUDE |
DTID 5303 “Adhoc |
1-n |
U |
$Property=DCID12304 |
|
|
|
|
Measurement” |
|
|
“Echo Measured |
|
|
|
|
|
|
|
Properties” |
Content Item Descriptions |
|
|
|
|
|
||
Row 8 |
Atextstringcontainingoneormoresentencesdescribingoneormoreindications,possiblywithadditionalcomments |
||||||
|
|
from the physician or tech. |
|
|
|
|
Row 11 These are measurements from a standardized list of pre-coordinated codes. See CID 12300 “Core Echo Measurements”. Measurements which do not correspond to the full semantics of one of the pre-coordinated codes in CID 12300 can likely be encoded in Row 13 instead.
Multiple instances of the same measurement code may be present in the container. Each instance represents a different sample or derivation.
This template makes no requirement that any or all samples be sent. For example, a mean value of all the samples ofagivenmeasurementcouldbesentwithoutsendingalloranyofthesamplesfromwhichthemeanwascalculated. Device configuration and/or operator interactions determine what measurements are sent.
Row 13 These are measurements that can be encoded using a standardized structure of post-coordinated codes. Measurements which correspond to the full semantics of one of the pre-coordinated codes in CID 12300 “Core Echo Measurements” should be encoded in Row 11 instead.
$Measurement shall be provided, but is not constrained to a CID.
Multiple instances of the same measurement code may be present in the container. Each instance represents a different sample or derivation.
This template makes no requirement that any or all samples be sent. For example, a mean value of all the samples ofagivenmeasurementcouldbesentwithoutsendingalloranyofthesamplesfromwhichthemeanwascalculated. Device configuration and/or operator interactions determine what measurements are sent.
Row 15 These are adhoc measurements encoded with minimal semantics.
Row 13 can be used to encode measurements with more complete semantics.
$Units shall be provided, but is not constrained to a CID.
Device configuration and/or operator interactions determine what measurements are sent.
Rows 17-24
When present, these rows contain measurements and associate them with a specific stage of a staged procedure.
- Standard -

DICOM PS3.16 2020a - Content Mapping Resource |
Page 387 |
TID 5301 Pre-coordinated Echo Measurement
This template codes numeric echo measurements where most of the details about the nature of the measurement have been pre- coordinated in the measurement code. In contrast, see TID 5302 “Post-coordinated Echo Measurement”.
The pre-coordinated measurement code is provided when this Template is included from a parent Template.
|
|
|
|
Table TID 5301. Parameters |
|
|||
|
|
Parameter Name |
|
|
Parameter Usage |
|
||
$Measurement |
|
Coded term or Context Group for Concept Name of measurement |
||||||
$Preferred |
|
|
Flag the preferred value by indicating the reason it was selected as preferred. |
|||||
Type: |
|
Non-Extensible |
|
|
|
|
||
Order: |
|
Significant |
|
|
|
|
||
Root: |
|
No |
|
|
|
|
|
|
|
|
|
Table TID 5301. Pre-coordinated Echo Measurement |
|
||||
|
NL |
Rel with |
VT |
Concept Name |
VM |
Req |
Condition |
Value Set Constraint |
|
|
Parent |
|
|
|
Type |
|
|
1 |
|
|
NUM |
$Measurement |
1 |
M |
|
|
2 |
> |
HAS |
CODE |
EV (121404, DCM, |
1 |
MC IFF this |
$Preferred = MemberOf |
|
|
|
PROPERTIES |
"Selection Status") |
|
|
measurement has{DCID 12301 |
||
|
|
|
|
|
|
|
been selected as “Measurement Selection |
|
|
|
|
|
|
|
|
thesinglepreferredReasons”} |
|
|
|
|
|
|
|
|
value for the |
|
|
|
|
|
|
|
|
measured concept. |
|
3 |
> |
HAS |
CODE |
EV (121401, DCM, |
1 |
MC IFF this |
EV (373098007, SCT, |
|
|
|
CONCEPT |
|
"Derivation") |
|
|
measurementisnot"Mean") |
|
|
|
MOD |
|
|
|
|
a sample. |
|
4 |
> |
|
INCLUDEDTID 320 “Image or |
1-n |
U |
|
$Purpose = EV (121112, |
|
|
|
|
|
Spatial Coordinates” |
|
|
|
DCM, "Source of |
|
|
|
|
|
|
|
|
measurement") |
5 |
> |
|
INCLUDEDTID 321 “Waveform or 1-n |
U |
|
$Purpose = EV (121112, |
||
|
|
|
|
Temporal Coordinates” |
|
|
|
DCM, "Source of |
|
|
|
|
|
|
|
|
measurement") |
6 |
> |
HAS |
TEXT |
EV (125309, DCM, |
1 |
U |
|
|
|
|
PROPERTIES |
"Short Label") |
|
|
|
|
Content Item Descriptions
Row 2 The reason that this value was selected as the preferred value for the measured concept.
The parent template may allow TID 5301 “Pre-coordinated Echo Measurement” to be included multiple times with the same Measurement Concept Name, for example to allow multiple samples of the measurement.
A given Measurement Concept Name might appear only once in the instance, in which case this this row may or may notbepresent.AgivenMeasurementConceptNamemayappearmultipletimes,howeverthisrowshallnotbepresent for more than one value of the given Measurement Concept Name. E.g. multiple measurements of (11706-9, LN, "Aortic Valve Peak Systolic Flow") may be present, but only one may be selected as preferred.
- Standard -

Page 388 |
DICOM PS3.16 2020a - Content Mapping Resource |
Row 3 The method used to derive this measurement value from multiple samples of the Measurement Concept Name.
If Row 3 is not present, then this measurement value is simply a single sample of the Measurement Concept Name.
Note
A measurement value that is a mean value of other measurements and was also selected as the preferred value because it is the mean will have both Row 2 and Row 3 present.
Row 6 This may be used to label the measurement value when space is limited on the screen or report page. E.g. a Short Label of "LVIDD" might be provided for a measurement of the left ventricle internal diameter at end diastole.
Note
Short Labels are not standardized and may omit details of the measurement, thus it is not recommended to use them for purposes such as matching.
TID 5302 Post-coordinated Echo Measurement
This template codes numeric echo measurements where most of the details about the nature of the measurement have been post- coordinated in modifiers and acquisition context. In contrast, see TID 5301 “Pre-coordinated Echo Measurement” .
This template is intended to be used for User-defined and Vendor-defined Echo Measurements.
Several modifier rows are conditional and are omitted when the modifier concept is not significant for the measurement encoded in the item. When these modifiers are included by the sender, it indicates that the modifier concept is significant and receivers will gen- erally treat the measurements differently than similar measurements sent that omit that modifier.
Note
The codes in the CIDs referenced below were sufficient to accurately encode all the best practice echo measurements re- commendedbytheASE.If,however,anewcodeisneededtorecordaspecificUser-definedorVendor-definedmeasurement, most of the CIDs are extensible. It is not unreasonable to expect that measurements might be made at other Finding Sites than those listed in CID 12305 “Basic Echo Anatomic Sites” , or using Measurement Methods beyond those listed in CID 12227 “Echocardiography Measurement Method” .
Theconceptmodifiersinthetemplatebelowweresufficienttoaccuratelyencodeallthebestpracticeechomeasurementsrecommended by the ASE. Although TID 5302 “Post-coordinated Echo Measurement” is extensible and adding new modifiers is not prohibited, the meaning and significance of such new modifiers will generally not be understood by receiving systems, delaying or preventing import of such measurements. Further, adding modifiers that replicate the meaning of an existing modifier is prohibited.
If such measurements cannot be encoded with the following structure, an implementation may choose to code the measurement in TID 5303 “Adhoc Measurement” , or to use TID 5200 “Echocardiography Procedure Report” instead of TID 5300 “Simplified Echo Procedure Report” .
|
|
|
Table TID 5302. Parameters |
|
|
|
|
Parameter Name |
|
|
Parameter Usage |
|
|
$Measurement |
|
Coded term or Context Group for Concept Name of measurement |
||||
$Preferred |
|
|
Flag the preferred value by indicating the reason it was selected as preferred. |
|||
Type: |
|
Extensible |
|
|
|
|
Order: |
|
Significant |
|
|
|
|
Root: |
|
No |
|
|
|
|
|
Table TID 5302. Post-coordinated Echo Measurement |
|
||||
NLRel with Parent |
VT |
Concept Name |
VM Req |
Condition |
Value Set Constraint |
|
|
|
|
|
Type |
|
|
1 |
NUM |
$Measurement |
1 M |
|
|
- Standard -

|
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 389 |
||||
|
NLRel with Parent VT |
Concept Name |
VM Req |
Condition |
Value Set Constraint |
|||
|
|
|
|
|
|
Type |
|
|
2 |
> |
HAS |
CODE |
EV (121050, DCM, |
1-n |
U |
|
|
|
|
PROPERTIES |
|
"Equivalent Meaning of |
|
|
|
|
|
|
|
|
Concept Name") |
|
|
|
|
3 |
> |
HAS |
CODE |
EV (121404, DCM, |
1 |
MC IFF this measurement$Preferred=MemberOf |
||
|
|
PROPERTIES |
|
"Selection Status") |
|
|
has been selected as {DCID 12301 |
|
|
|
|
|
|
|
|
the single preferred |
“Measurement |
|
|
|
|
|
|
|
value for the measuredSelection Reasons”} |
|
|
|
|
|
|
|
|
concept. |
|
4 |
> |
HAS CONCEPTCODE |
EV (121401, DCM, |
1 MC IFF this measurementEV (373098007, SCT, |
||||
|
|
MOD |
|
"Derivation") |
|
|
is not a sample. |
"Mean")) |
5 |
> |
|
INCLUDEDTID 320 “Image or |
1-n |
U |
|
$Purpose=EV(121112, |
|
|
|
|
|
Spatial Coordinates” |
|
|
|
DCM, "Source of |
|
|
|
|
|
|
|
|
measurement") |
6 |
> |
|
INCLUDEDTID 321 “Waveform or 1-n |
U |
|
$Purpose = EV |
||
|
|
|
|
Temporal Coordinates” |
|
|
|
(121112,DCM,"Source |
|
|
|
|
|
|
|
|
of measurement") |
7 |
> |
HAS CONCEPTCODE |
EV (125306, DCM, |
1 |
M |
|
DCID 12303 “Echo |
|
|
|
MOD |
|
"Measurement Type") |
|
|
|
Measurement Types” |
8 |
> |
HAS CONCEPTCODE |
EV (363698007, SCT, |
1 |
M |
|
DCID 12305 “Basic |
|
|
|
MOD |
|
"Finding Site") |
|
|
|
Echo Anatomic Sites” |
9 |
> |
HAS CONCEPTCODE |
EV (125305, DCM, |
1 |
M |
|
DCID 12302 “Echo |
|
|
|
MOD |
|
"Finding Observation |
|
|
|
Finding Observation |
|
|
|
|
Type") |
|
|
|
Types” |
10 |
> |
HAS CONCEPTCODE |
EV (125307, DCM, |
1 |
M |
|
DCID 12304 “Echo |
|
|
|
MOD |
|
"Measured Property") |
|
|
|
Measured Properties” |
11 |
> |
HAS CONCEPTCODE |
EV (260674002, SCT, |
1 MC IFF Row 9 is |
DCID 12306 “Echo |
|||
|
|
MOD |
|
"Flow Direction") |
|
|
(44324008, SCT, |
Flow Directions” |
|
|
|
|
|
|
|
"Hemodynamic |
|
|
|
|
|
|
|
|
Measurements") and |
|
|
|
|
|
|
|
|
the Flow Direction is |
|
|
|
|
|
|
|
|
significant for this |
|
|
|
|
|
|
|
|
measurement. |
|
12 |
> |
HAS CONCEPTCODE |
EV (370129005, SCT, |
1 MC IFF the Measurement DCID 12227 |
||||
|
|
MOD |
|
"Measurement Method") |
|
|
Methodissignificantfor“Echocardiography |
|
|
|
|
|
|
|
|
this measurement. |
Measurement Method” |
13 |
> |
HAS ACQ |
CODE |
EV (399264008, SCT, |
1 MC IFF the Image Mode isDCID 12224 |
|||
|
|
CONTEXT |
|
"Image Mode") |
|
|
significant for this |
“Ultrasound Image |
|
|
|
|
|
|
|
measurement. |
Modes” |
14 |
> |
HAS ACQ |
CODE |
EV (111031, DCM, |
1 MC IFF the Image View isDCID 12226 |
|||
|
|
CONTEXT |
|
"Image View") |
|
|
significant for this |
“Echocardiography |
|
|
|
|
|
|
|
measurement. |
Image View” |
15 |
> |
HAS CONCEPTCODE |
EV (272518008, SCT, |
1 MC IFF the Cardiac CycleDCID 12307 “Cardiac |
||||
|
|
MOD |
|
"Cardiac Cycle Point") |
|
|
Point is significant for Phases and Time |
|
|
|
|
|
|
|
|
this measurement. |
Points” |
16 |
> |
HAS CONCEPTCODE |
EV (272517003, SCT, |
1 MC IFF the Respiratory |
DCID 12234 |
|||
|
|
MOD |
|
"Respiratory Cycle |
|
|
CyclePointissignificant“Respiration State” |
|
|
|
|
|
Point") |
|
|
for this measurement. |
|
- Standard -

Page 390 |
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
||||
|
NLRel with Parent VT |
Concept Name |
VM Req |
Condition |
Value Set Constraint |
||
|
|
|
|
|
Type |
|
|
17 |
> |
HAS CONCEPTCODE |
EV (125308, DCM, |
1 MC IFF the value of Row 7 |
|||
|
|
MOD |
|
"Measurement Divisor") |
|
is (125313, DCM, |
|
|
|
|
|
|
|
"Indexed") or |
|
|
|
|
|
|
|
(118586006, SCT, |
|
|
|
|
|
|
|
"Ratio") or (125314, |
|
|
|
|
|
|
|
DCM, "Fractional |
|
|
|
|
|
|
|
Change") |
|
18 |
> |
HAS |
TEXT |
EV (125309, DCM, |
1 U |
|
|
|
|
PROPERTIES |
|
"Short Label") |
|
|
|
Content Item Descriptions
Row 1 A fully pre-coordinated code that incorporates all the semantics of Rows 7-17 for this measurement.
Thecodeisintendedtoallowparserstorecognizepost-coordinatedmeasurementsthathavebeenpreviouslyencountered, thus facilitating incorporation of the measurement into databases, report templates, registries, etc. Typically these codes will be from a vendor or site specific coding scheme, e.g., 99ACME. Sending the same code consistently in different reports will depend on the recording system maintaining a stable list of these pre-coordinated codes. Such a list might be configured or internally generated and managed.
This shall be populated by the recording system. If the recording system does not have a method to ensure that all occurrences of the same post-coordinated measurement use the same code, it shall use the code (125304, DCM, "Untrackable Measurement").
Note
1.Two measurements with the same pre-coordinated code have, by definition, the same semantics (except for "Untrackable Measurements")
2.Two measurements with the same constellation of modifier values have the same semantics but may have different pre-coordinated codes because they
•come from carts of different vendors who don't share the same code table
•come from carts of the same vendor, but the carts don't share the same code table
•come from the same cart, but it's code table has been modified
•come from the same cart, but it does not maintain a code table
3.Two measurements with the same constellation of modifier values and different pre-coordinated codes have the same semantics and the receiver is entitled to treat them as the same (with respect to the scope of those modifiers)
4.Recommended units for various Measured Properties (Row 10) can be found in the Units column of CID 12304 “Echo Measured Properties” .
5.WhentheMeasurementType(Row7)is(125313,DCM,"Indexed"),(118586006,SCT,"Ratio")or(125314, DCM, "Fractional Change") , the Units for Row 1 corresponds to the fully calculated $Measurement, incorporatingboththenumerator(Row10)andthedenominator(Row17).E.g.ameasureofLeftVentricular Outflow Tract Diameter / BSA would have units of (cm/m2, UCUM, "cm/m2") in Row 1, (125313, DCM, "Indexed") in Row 7, (81827009, SCT, "Diameter") in Row 10, and (8277-6, LN, "Body Surface Area") in Row 17.
Row 2 One or more additional fully pre-coordinated codes which are semantically equivalent to the code in Row 1.
This may be used to communicate known mappings, such as to national registry codes or other vendors' codes.
- Standard -