DICOM PS3.16 2020a - Content Mapping Resource​

Page 331​

NL​ Rel with Parent​ VT​

Concept Name​

VM​

Req​ Condition​Value Set Constraint​

 

 

 

 

Type​

 

4​ >​ INFERRED​

INCLUDE​ DTID 4127 “Colon CAD Single​ 1-n​

U​

 

FROM​

 

Image Finding”​

 

 

 

Content Item Descriptions​

 

 

 

 

CAD Processing and​

This code value is used to express if and why the Colon CAD Findings Summary sub-tree is empty.​

Findings Summary​

The Summary of Detections and Summary of Analyses sub-trees of the Document Root node contain​

 

detail about which (if any) algorithms succeeded or failed.​

 

If the code value indicates that there were no findings, then the code value can be used to determine​ whether colon CAD processing occurred successfully, without parsing the Summary of Detections and​ Summary of Analyses sub-trees.​

TID 4122 CAD Common Image Properties Entry​

Each instance of the CAD Common Image Properties Entry Template contains selected attributes for a set of parallel contiguous​ equally spaced slices (with identical properties) from which CAD findings are derived.​

Type:​

Extensible​

Order:​

Significant​

Root:​

No​

Table TID 4122. CAD Common Image Properties Entry​

 

NL​

Rel with​

VT​

Concept Name​

VM​ Req​ Condition​

Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

 

1​

 

 

CONTAINER​EV (112224, DCM,​

1​

M​

 

 

 

 

 

 

"Image Set Properties")​

 

 

 

 

2​

>​

CONTAINS​UIDREF​

EV (112227, DCM,​

1​

M​

 

 

 

 

 

 

"Frame of Reference​

 

 

 

 

 

 

 

 

UID")​

 

 

 

 

3​

>​

CONTAINS​UIDREF​

EV (110180, DCM,​

1​

M​

 

 

 

 

 

 

"Study Instance UID")​

 

 

 

 

4​

>​

CONTAINS​DATE​

EV (111060, DCM,​

1​

M​

Shall be taken from Study Date​

 

 

 

 

"Study Date")​

 

 

(0008,0020)intheImageInstances.​

5​

>​

CONTAINS​TIME​

EV (111061, DCM,​

1​

M​

Shall be taken from Study Time​

 

 

 

 

"Study Time")​

 

 

(0008,0030)intheImageInstances.​

6​

>​

CONTAINS​CODE​

EV (121139, DCM,​

1​

M​

Shall be taken from Modality​

 

 

 

 

"Modality")​

 

 

(0008,0060)intheImageInstances.​

7​

>​

CONTAINS​NUM​

EV (111026, DCM,​

1​

M​

Shall be taken from value 2 of Pixel​

 

 

 

 

"Horizontal Pixel​

 

 

Spacing (0028,0030) in the Image​

 

 

 

 

Spacing")​

 

 

Instances.​

 

 

 

 

 

 

 

UNITS = EV (mm/{pixel}, UCUM,​

 

 

 

 

 

 

 

"millimeters per pixel")​

8​

>​

CONTAINS​NUM​

EV (111066, DCM,​

1​

M​

Shall be taken from value 1 of Pixel​

 

 

 

 

"Vertical Pixel Spacing")​

 

 

Spacing (0028,0030) in the Image​

 

 

 

 

 

 

 

Instances.​

UNITS = EV (mm/{pixel}, UCUM,​ "millimeters per pixel")​

- Standard -​

Page 332​

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

 

 

NL​

Rel with​

VT​

Concept Name​

VM​ Req​ Condition​

Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

 

9​

>​

CONTAINS​NUM​

EV(112225,DCM,"Slice​ 1​

M​

Shall be taken from Slice Thickness​

 

 

 

 

Thickness")​

 

 

(0018,0050)intheImageInstances.​

 

 

 

 

 

 

 

UNITS = EV (mm, UCUM,​

 

 

 

 

 

 

 

"millimeter")​

10​

>​

CONTAINS​NUM​

EV (112226, DCM,​

1​

M​

Shall be computed from the Image​

 

 

 

 

"Spacingbetweenslices")​

 

Position (Patient) (0020,0032)​

 

 

 

 

 

 

 

projected onto the normal to the​

 

 

 

 

 

 

 

Image Orientation (Patient)​

 

 

 

 

 

 

 

(0020,0037); may or may not be the​

 

 

 

 

 

 

 

sameastheSpacingBetweenSlices​

 

 

 

 

 

 

 

(0018,0088) if present.​

 

 

 

 

 

 

 

UNITS = EV (mm, UCUM,​

 

 

 

 

 

 

 

"millimeter")​

11​

>​

CONTAINS​CODE​

EV (112228, DCM,​

1​

MC​Required if​Shall be derived from Patient​

 

 

 

 

"Recumbent Patient​

 

Patient​

Position (0018,5100) in the Image​

 

 

 

 

Position with respect to​

 

Position​

Instances.​

 

 

 

 

gravity")​

 

(0018,5100)​

 

DCID 6206 “Recumbent Patient​ is present in​

the image​ Orientation for Colon”​ instances​

and has a​ value.​

TID 4125 Colon CAD Composite Feature​

This Template collects a composite feature for a lesion, non-lesion object, or correlation of related objects (see TID 4121 “Colon CAD​ Findings Summary”). The details of the composition are expressed in the Colon CAD Composite Feature Body (see TID 4126 “Colon​ CAD Composite Feature Body”). The data from which the details are inferred, are expressed in the Composite Features (see TID​ 4125 “Colon CAD Composite Feature”) and/or Single Image Findings (see TID 4127 “Colon CAD Single Image Finding”), of which​ there may be several.​

Type:​

Non-Extensible​

Order:​

Significant​

Root:​

No​

Table TID 4125. Colon CAD Composite Feature​

 

NL​ Rel with Parent​ VT​

Concept Name​

VM​

Req​

Condition​ Value Set Constraint​

 

 

 

 

 

Type​

 

 

1​

 

CODE​

EV (111015, DCM,​

1​

M​

DCID 6201

“Colon​

 

 

 

"Composite Feature")​

 

 

Finding or Feature”​

2​

>​

HAS CONCEPT​CODE​

EV (112023, DCM,​

1​

U​

DCID 6202

“Colon​

 

 

MOD​

"Composite Feature​

 

 

Finding or Feature​

 

 

 

Modifier")​

 

 

Modifier”​

 

3​

>​

HAS CONCEPT​CODE​

EV (111056, DCM,​

1​

M​

DCID 6034

“Intended​

 

 

MOD​

"Rendering Intent")​

 

 

Use of CAD Output”​

- Standard -​

DICOM PS3.16 2020a - Content Mapping Resource​

Page 333​

NL​ Rel with Parent​ VT​

Concept Name​

VM​ Req​ Condition​

Value Set Constraint​

 

 

Type​

 

4​ >>​ HAS​

NUM​ EV (111071, DCM, "CAD​ 1​

PROPERTIES​

Operating Point")​

UC​ IFF value of row 3​UNITS = DT ({1:n},​ is (111151, DCM,​UCUM, "range: 1:n"),​ "Presentation​ where n is the​ Optional") and row​maximum specified in​ 1ofTID4023“CAD​Row 1 of TID 4023​ OperatingPoints”is​“CADOperatingPoints”​

present for the​

forthefeatureidentified​

feature identified in​in row 1. Value is​

row 1.​

restricted to being an​

 

integer.​

5​

>​

HAS OBS​

INCLUDE​DTID 4108 “Tracking​

1​

U​

 

 

CONTEXT​

Identifier”​

 

 

6​

>​

HAS OBS​

INCLUDE​DTID 4022 “CAD​

1​

MC​ Shall be present​

 

 

CONTEXT​

Observation Context”​

 

IFF this feature is​

 

 

 

 

 

duplicated from a​

 

 

 

 

 

differentreportthan​

 

 

 

 

 

its parent.​

7​

>​

HAS OBS​

INCLUDE​DTID 4019 “Algorithm​

1​

M​

 

 

CONTEXT​

Identification”​

 

 

8​

>​

HAS​

INCLUDE​DTID 4126 “Colon CAD​

1​

M​

 

 

PROPERTIES​

Composite Feature Body”​

 

 

9​

>​

INFERRED​

INCLUDE​DTID 4125 “Colon CAD​

1-n​

U​

 

 

FROM​

Composite Feature”​

 

 

10​

>​

INFERRED​

INCLUDE​DTID 4127 “Colon CAD​

1-n​

U​

 

 

FROM​

Single Image Finding”​

 

 

Content Item Descriptions​

Rendering Intent​

This Content Item constrains the SCP receiving the Colon CAD SR IOD in its use of the contents of this​

 

TemplateanditsTargetContentItems.ColonCADdevicesmayopttousedatamarked"NotforPresentation"​

 

or"PresentationOptional"asinputtosubsequentcolonCADprocessingsteps.RefertoSectionO.2“Structured​

 

Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4.​

TID 4126 Colon CAD Composite Feature Body​

The details of a composite feature are expressed in this Template. It is applied to Colon CAD Composite Feature (TID 4125 “Colon​ CAD Composite Feature”).​

Type:​

Non-Extensible​

Order:​

Significant​

Root:​

No​

Table TID 4126. Colon CAD Composite Feature Body​

NL​

Rel with​

VT​

Concept Name​

VM​

Req​

Condition​

Value Set Constraint​

 

Parent​

 

 

 

Type​

 

 

1​

 

CODE​

EV (111016, DCM,​

1​

M​

 

DCID 6035 “Composite​

 

 

 

"Composite type")​

 

 

 

Feature Relations”​

2​

 

CODE​

EV (111057, DCM,​

1​

M​

 

DCID 6036 “Scope of​

 

 

 

"Scope of Feature")​

 

 

 

Feature”​

3​

 

NUM​

EV (111011, DCM,​

1​

U​

 

UNITS = EV (%, UCUM,​

 

 

 

"Certainty of feature")​

 

 

 

"Percent")​

Value = 0 - 100​

- Standard -​

Page 334​

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

 

NL​ Rel with​

VT​

Concept Name​

VM​

Req​

Condition​

Value Set Constraint​

 

 

Parent​

 

 

 

Type​

 

 

4​

 

 

INCLUDE​DTID 4129 “Colon CAD​

1​

U​

 

 

 

 

 

 

Geometry”​

 

 

 

 

5​

 

 

INCLUDE​DTID 4128 “Colon CAD​

1​

U​

 

 

 

 

 

 

Descriptors”​

 

 

 

 

6​

 

 

NUM​

DCID 6207 “Colon​

1-n​

UC​ MaybepresentIFF​

 

 

 

 

Quantitative Temporal​

 

 

the value of row 1​

 

 

 

 

 

Difference Type”​

 

 

is (111153, DCM,​

 

 

 

 

 

 

 

 

"Target content​

 

 

 

 

 

 

 

 

items are related​

 

 

 

 

 

 

 

 

temporally")​

 

7​

>​

R-INFERRED​NUM​

 

2​

U​

 

The referenced numeric​

 

 

FROM​

 

 

 

 

 

values shall have the same​

 

 

 

 

 

 

 

 

Concept Name. Their​

 

 

 

 

 

 

 

 

UNITSshallbethesameas​

 

 

 

 

 

 

 

 

row 6​

8​

 

 

CODE​

EV (111049, DCM,​

1-n​

UC​ May be present​

DCID 6134 “Chest​

 

 

 

 

"Qualitative Difference")​

 

 

only if the value of​Qualitative Temporal​

 

 

 

 

 

 

 

row 1 is (111153,​ Difference Type”​

 

 

 

 

 

 

 

DCM, "Target​

 

 

 

 

 

 

 

 

content items are​

 

 

 

 

 

 

 

 

relatedtemporally")​

9​

>​

HAS​

TEXT​

EV (111021, DCM,​

1​

U​

 

 

 

 

PROPERTIES​

"DescriptionofChange")​

 

 

 

 

10​

>​ R-INFERRED​CODE​

 

2​

M​

 

The referenced Content​

 

 

FROM​

 

 

 

 

 

Items shall have the same​

 

 

 

 

 

 

 

 

Concept Name and their​

 

 

 

 

 

 

 

 

code values shall be from​

 

 

 

 

 

 

 

 

the same context group.​

Content Item Descriptions​

 

 

 

 

 

Certainty of Feature​

The CAD device's certainty that the feature analyzed and classified by the CODE, as specified in the​

 

 

 

Composite Feature parent Template is, in fact, that type of feature.​

 

Row 6​

 

Values ≤ 0 are allowed. The two referenced numeric values are Target Content Items of the first​

 

 

 

generation Composite Feature or Single Image Finding children of this composite feature. Given the​

 

 

 

equation, A - B, the value representing A shall be referenced first.​

 

Qualitative Difference​

The two referenced code values are Target Content Items of the first generation Composite Feature​

 

 

 

or Single Image Finding children of this composite feature.​

 

TID 4127 Colon CAD Single Image Finding​

This Template describes a single image finding for a lesion or other object. The details of the finding are expressed in this Template​ and/or more specific Templates.​

Type:​

Non-Extensible​

Order:​

Significant​

Root:​

No​

- Standard -​

DICOM PS3.16 2020a - Content Mapping Resource​ Page 335​

Table TID 4127. Colon CAD Single Image Finding​

 

NL​Rel with Parent​ VT​

Concept Name​

VM​ Req​

Condition​

Value Set Constraint​

 

 

 

 

 

 

Type​

 

1​

 

 

CODE​

EV (111059, DCM,​

1​

M​

 

DCID 6201 “Colon​

 

 

 

 

"Single Image​

 

 

 

Finding or Feature”​

 

 

 

 

Finding")​

 

 

 

 

2​

>​

HAS CONCEPT​CODE​

EV (112024, DCM,​

1​

U​

 

DCID 6202 “Colon​

 

 

MOD​

 

"Single Image Finding​

 

 

 

Finding or Feature​

 

 

 

 

Modifier")​

 

 

 

Modifier”​

3​

>​

HAS CONCEPT​CODE​

EV (111056, DCM,​

1​

M​

 

DCID 6034 “Intended​

 

 

MOD​

 

"Rendering Intent")​

 

 

 

Use of CAD Output”​

4​

>>​

HAS​

NUM​

EV (111071, DCM,​

1​

UC​ IFF value of row 3 is​

UNITS = DT ({1:n},​

 

 

PROPERTIES​

 

"CAD Operating​

 

 

(111151, DCM,​

UCUM, "range: 1:n"),​

 

 

 

 

Point")​

 

 

"Presentation Optional")​wherenisthemaximum​

 

 

 

 

 

 

 

and row 1 of TID 4023​ specifiedinRow1ofTID​

 

 

 

 

 

 

 

“CAD Operating Points”​4023 “CAD Operating​

 

 

 

 

 

 

 

is present for the finding​Points” for the finding​

 

 

 

 

 

 

 

identified in row 1​

identifiedinrow1.Value​

 

 

 

 

 

 

 

 

is restricted to being an​

 

 

 

 

 

 

 

 

integer.​

5​

>​

HAS OBS​

INCLUDE​DTID 4108 “Tracking​

1​

U​

 

 

 

 

CONTEXT​

 

Identifier”​

 

 

 

 

6​

>​

HAS OBS​

INCLUDE​DTID 4022 “CAD​

1​ MC​ Shall be present IFF this​

 

 

CONTEXT​

 

Observation Context”​

 

 

findingisduplicatedfrom​

 

 

 

 

 

 

 

a different report than its​

 

 

 

 

 

 

 

parent.​

 

7​

>​

HAS OBS​

INCLUDE​DTID 4019 “Algorithm​ 1​

M​

 

 

 

 

CONTEXT​

 

Identification”​

 

 

 

 

8​

>​

HAS​

NUM​

EV (111012, DCM,​

1​

U​

 

UNITS = EV (%, UCUM,​

 

 

PROPERTIES​

 

"Certainty of Finding")​

 

 

 

"Percent")​

 

 

 

 

 

 

 

 

Value = 0 - 100​

9​

>​

HAS​

TEXT​

EV (111058, DCM,​

1​ MC​ Shall be present IFF​

 

 

 

PROPERTIES​

 

"Selected Region​

 

 

value of row 1 is​

 

 

 

 

 

Description")​

 

 

(111099,DCM,"Selected​

 

 

 

 

 

 

 

region")​

 

10​

>​

HAS​

INCLUDE​DTID 4129 “Colon​

1​ MC​ Shall be present unless​

 

 

PROPERTIES​

 

CAD Geometry”​

 

 

value of row 1 is​

 

 

 

 

 

 

 

 

(111101, DCM, "Image​

 

 

 

 

 

 

 

 

quality")​

 

11​

>​

HAS​

INCLUDE​DTID 4128 “Colon​

1​

U​

 

 

 

 

PROPERTIES​

 

CAD Descriptors”​

 

 

 

 

12​

>​

INFERRED​

IMAGE​

 

1​ MC​ Shall be present IFF​

 

 

 

FROM​

 

 

 

 

value of row 1 is​

 

 

 

 

 

 

 

 

(111101, DCM, "Image​

 

 

 

 

 

 

 

 

quality") and row 13 is​

 

 

 

 

 

 

 

 

not present​

 

13​

>​

INFERRED​

SCOORD​EV (111030, DCM,​

1-n​

MC​ Shall be present IFF​

 

 

 

FROM​

 

"Image Region")​

 

 

value of row 1 is​

 

 

 

 

 

 

 

 

(111101, DCM, "Image​

 

 

 

 

 

 

 

 

quality") and row 12 is​

 

 

 

 

 

 

 

 

not present​

 

- Standard -​

Page 336​

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

 

NL​Rel with Parent​ VT​

Concept Name​

VM​ Req​

Condition​

Value Set Constraint​

 

 

 

 

 

Type​

 

14​

>>​

SELECTED​

IMAGE​

 

1​ M​

 

All the row 13 Content​

 

 

FROM​

 

 

 

 

Items in a single​

 

 

 

 

 

 

 

invocation of this​

 

 

 

 

 

 

 

Templateshallreference​

 

 

 

 

 

 

 

the same IMAGE​

15​

>​

HAS​

INCLUDE​DTID 4014 “CAD​

1​ MC​ Shall be present IFF​

$QualityFinding = DCID​

 

 

PROPERTIES​

Image Quality”​

 

value of row 1 is​

6135 “Image Quality​

 

 

 

 

 

 

(111101, DCM, "Image​Finding”​

 

 

 

 

 

 

quality")​

$QualityStandard =​

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DCID 6208 “Colon​

 

 

 

 

 

 

 

Types of Quality Control​

 

 

 

 

 

 

 

Standard”​

Content Item Descriptions​

 

 

 

 

Rendering Intent​

This Content Item constrains the SCP receiving the Colon CAD SR IOD in its use of the contents of this​

 

 

 

Template and its Target Content Items. Colon CAD devices may opt to use data marked "Not for​

 

 

 

Presentation" or "Presentation Optional" as input to subsequent colon CAD processing steps. Refer to​

 

 

 

Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4.​

CAD Operating Point​

AdditionalinformationtousewhenRenderingIntentis"PresentationalOptional".ACADOperatingPoint​

 

 

 

of zero is not sent, and is encoded as a Rendering Intent of "Presentation Required". See Section O.2​

 

 

 

“Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 and Section E.4 “CAD​

 

 

 

Operating Point” in PS3.17.​

 

 

 

Certainty of Finding​

The certainty of the CAD device that the finding detected and classified by the Single Image Finding​

 

 

 

CODE specified is in fact that type of finding.​

 

 

TID 4128 Colon CAD Descriptors​

This Template provides qualitative detail for a Single Image Finding or Composite Feature. It is applied to Colon CAD Composite​ Feature (TID 4125 “Colon CAD Composite Feature”) and Colon CAD Single Image Finding (TID 4127 “Colon CAD Single Image​ Finding”).​

Type:​

Non-Extensible​

Order:​

Significant​

Root:​

No​

Table TID 4128. Colon CAD Descriptors​

NL​

Rel with​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

Parent​

 

 

 

Type​

 

1​

 

CODE​

EV (116676008, SCT,​

1-n​

U​

DCID6209“ColonMorphology​

 

 

 

"Associated Morphology")​

 

 

Descriptor”​

2​

 

CODE​

EV (363698007, SCT,​

1​

U​

DCID 6210 “Location in​

 

 

 

"Finding Site")​

 

 

Intestinal Tract”​

3​

 

CODE​

EV (111014, DCM,​

1​

U​

DCID6205“ClockfaceLocation​

 

 

 

"Clockface or region")​

 

 

for Colon”​

- Standard -​

 

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 337​

NL​ Rel with​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

Parent​

 

 

 

Type​

 

4​

CONTAINS​

INCLUDE​ DTID 300 “Measurement”​

1-n​

U​

$Measurement = DCID 6212​

 

 

 

 

 

 

“Calculated Value for Colon​

 

 

 

 

 

 

Findings”​

 

 

 

 

 

 

$Derivation = DCID 6140​

 

 

 

 

 

 

“Calculation Methods”​

 

 

 

 

 

 

$DerivationParameter = EV​

 

 

 

 

 

 

(112032, DCM, "Threshold​

 

 

 

 

 

 

Attenuation Coefficient")​

 

 

 

 

 

 

$DerivationParameterUnits =​

 

 

 

 

 

 

EV ([hnsf'U], UCUM,​

 

 

 

 

 

 

"Hounsfield unit")​

5​

 

INCLUDE​ DTID 1400 “Linear​

1-n​

U​

 

 

 

 

Measurement”​

 

 

 

6​

 

INCLUDE​ DTID 1401 “Area​

1-n​

U​

 

 

 

 

Measurement”​

 

 

 

7​

 

INCLUDE​ DTID 1402 “Volume​

1-n​

U​

 

 

 

 

Measurement”​

 

 

 

8​

 

INCLUDE​ DTID 1406 “Three​

1-n​

U​

 

 

 

 

Dimensional Linear​

 

 

 

 

 

 

Measurement”​

 

 

 

9​

 

NUM​

DCID 6141 “Attenuation​

1-n​

U​

UNITS = EV ([hnsf'U], UCUM,​

 

 

 

Coefficient Measurements”​

 

 

"Hounsfield unit")​

10​ >​

HAS​

CODE​

EV(112009,DCM,"Typeof​

1​

U​

DCID 6211 “Colon CAD​

 

PROPERTIES​

Content")​

 

 

Material Description”​

Content Item Descriptions​

 

 

 

 

Row 3​

12 o'clock position is the anterior direction of the patient regardless of the positioning with respect to gravity;​

 

clockwise is from the point of view of an observer located closer to the anus than the finding being observed.​

TID 4129 Colon CAD Geometry​

All geometry Template invocations require specification of either the location of the center of the object, the outline, or both. Geometry​ isapropertyofsingleimagefindings(seeTID4127“ColonCADSingleImageFinding”)andcompositefeatures(seeTID4125“Colon​ CAD Composite Feature”).​

Type:​

Non-Extensible​

Order:​

Significant​

Root:​

No​

 

Table TID 4129. Colon CAD Geometry​

NL​

Rel with​

VT​

Concept Name​

VM​

Req​

Condition​

Value Set​

 

Parent​

 

 

 

Type​

 

Constraint​

1​

 

SCOORD​

EV (111010, DCM,​

1​

MC​

At least one of rows 1,​GRAPHIC TYPE =​

 

 

 

"Center")​

 

 

3, 4, 6 or 10 shall be​ {POINT}​

 

 

 

 

 

 

present.​

 

2​ >​

SELECTED​IMAGE​

 

1​

M​

 

 

 

FROM​

 

 

 

 

 

 

- Standard -​

Page 338​

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

 

NL​

Rel with​

VT​

Concept Name​

VM​

Req​

Condition​

Value Set​

 

 

Parent​

 

 

 

Type​

 

Constraint​

3​

 

 

SCOORD3D​EV (111010, DCM,​

1​

MC​

At least one of rows 1,​GRAPHIC TYPE =​

 

 

 

 

"Center")​

 

 

3, 4, 6 or 10 shall be​

{POINT}​

 

 

 

 

 

 

 

present.​

 

4​

 

 

SCOORD​

EV (111041, DCM,​

1​

MC​

At least one of rows 1,​

 

 

 

 

"Outline")​

 

 

3, 4, 6 or 10 shall be​

 

 

 

 

 

 

 

 

present.​

 

5​

>​

SELECTED​IMAGE​

 

1​

M​

 

 

 

 

FROM​

 

 

 

 

 

 

6​

 

 

SCOORD3D​EV (111041, DCM,​

1​

MC​

At least one of rows 1,​

 

 

 

 

"Outline")​

 

 

3, 4, 6 or 10 shall be​

 

 

 

 

 

 

 

 

present.​

 

7​

 

 

SCOORD​

DCID 6166 “CAD​

1-n​

U​

 

 

 

 

 

 

Geometry Secondary​

 

 

 

 

 

 

 

 

Graphical​

 

 

 

 

 

 

 

 

Representation”​

 

 

 

 

8​

>​

SELECTED​IMAGE​

 

1​

M​

 

 

 

 

FROM​

 

 

 

 

 

 

9​

 

 

SCOORD3D​DCID 6166 “CAD​

1-n​

U​

 

 

 

 

 

 

Geometry Secondary​

 

 

 

 

 

 

 

 

Graphical​

 

 

 

 

 

 

 

 

Representation”​

 

 

 

 

10​

 

 

IMAGE​

EV (112229, DCM,​

1​

MC​

At least one of rows 1,​Referenced image​

 

 

 

 

"Identifying Segment")​

 

 

3, 4, 6 or 10 shall be​

shall be a​

 

 

 

 

 

 

 

present.​

Segmentation and​

 

 

 

 

 

 

 

 

the Content Item​

 

 

 

 

 

 

 

 

shall include​

Referenced​ Segment Number​ (0062,000B)​

Breast Imaging Report Templates ​

The Templates that comprise the Breast Imaging Report are interconnected as in Figure A-10.​

- Standard -​

DICOM PS3.16 2020a - Content Mapping Resource​

Page 339​

TID 4200

 

TID 1001

 

 

 

 

 

 

 

 

 

 

 

Breast Imaging

 

 

Observation Context

 

 

 

 

 

 

 

 

 

 

 

Report

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TID 4202

 

TID 1002

 

 

 

 

 

 

 

 

 

Breast Imaging

 

 

Observer Context

 

 

 

 

 

 

 

 

 

Report Narrative

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TID 1204

 

TID 350

 

 

 

 

 

 

 

 

 

Language of

 

References to

 

 

 

 

 

 

 

 

 

Content Item and

 

Supporting Evidence

 

 

 

 

 

 

 

 

 

Descendants

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TID 4208

 

TID 4201

 

 

 

 

 

 

 

 

 

Breast Imaging

 

Breast Imaging

 

 

 

 

 

 

 

 

 

Report

 

 

Procedure Reported

 

 

 

 

 

 

 

 

 

Supplementary Data

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TID 4203

 

 

TID 4207

 

 

 

 

 

 

 

 

 

Breast Imaging

 

 

Breast Imaging

 

 

 

 

 

 

 

 

 

Assessment

 

 

 

Pathology Results

 

 

 

 

 

 

 

 

 

TID 4205

 

 

TID 4201

 

 

 

 

 

 

 

 

 

Breast Composition

 

 

Breast Imaging

 

 

 

 

 

 

 

 

 

Section

 

 

 

Procedure Reported

 

 

 

 

 

 

 

 

 

TID 4204

 

 

 

 

TID 4201

 

 

 

 

 

 

 

Breast Imaging

 

 

 

 

Breast Imaging

 

 

 

 

 

 

 

Report

 

 

 

 

 

 

 

 

 

 

 

 

 

Intervention Section

 

 

 

 

Procedure Reported

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TID 4206

 

 

TID 1002

 

TID 1009

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Breast Imaging

 

 

 

 

Subject Context,

 

 

 

 

 

 

 

Report

 

 

 

Observer Context

 

Specimen

 

 

 

 

 

 

 

Finding Section

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TID 350

 

TID 4207

 

 

 

 

 

 

 

 

 

 

 

 

 

References to

 

Pathology

 

 

 

 

 

 

 

 

 

 

 

 

 

Supporting Evidence

 

Results

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TID 4201

 

TID 350

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Breast Imaging

 

References to

 

 

 

 

 

 

 

 

 

 

 

 

 

Procedure Reported

 

Supporting Evidence

 

TID 1402

Volume

Measurement

TID 1401

Area

Measurement

TID 1400

Linear

Measurement

TID 4203

TID 4207

Breast Imaging

Pathology Results

Assessment

Figure A-10. Breast Imaging Report Template Structure​

TID 4200 Breast Imaging Report​

This Template forms the content tree that allows a Breast Imaging Report device to describe the results of a radiologist's diagnostic​ interpretation of Breast Imaging (e.g., X-Ray mammography or breast ultrasound) evidence. This Template, together with its subor-​ dinate Templates, describes the results for presentation to clinicians, or for consumption by Breast Imaging Report devices for sub-​ sequent Breast Imaging Reports.​

This Template shall be instantiated at the Root node only.​

See Figure Q.1-1 “Top Level of Breast Imaging Report Content Tree” in PS3.17.​

Type:​

Non-Extensible​

Order:​

Significant​

Root:​

Yes​

 

Table TID 4200. Breast Imaging Report​

NL​

Rel with Parent​

VT​

Concept Name​

VM​

Req​ Condition​Value Set​

 

 

 

 

 

Type​

Constraint​

1​

 

CONTAINER​ EV (111400, DCM, "Breast​

1​

M​

Root node​

 

 

 

Imaging Report")​

 

 

 

2​ >​

HAS CONCEPT​ INCLUDE​

DTID 1204 “Language of Content​

1​

M​

 

 

MOD​

 

Item and Descendants”​

 

 

 

- Standard -​

Page 340​

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

 

 

NL​

Rel with Parent​

VT​

Concept Name​

VM​

Req​ Condition​Value Set​

 

 

 

 

 

 

Type​

Constraint​

2a​

>​

HAS OBS​

INCLUDE​

TID 1001 “Observation Context”​

1​

U​

 

 

 

CONTEXT​

 

 

 

 

 

2b​

>​

CONTAINS​

INCLUDE​

DTID 4209 “Breast Patient​

1​

U​

 

 

 

 

 

Characteristics”​

 

 

 

3​

>​

CONTAINS​

INCLUDE​

DTID 4202 “Breast Imaging​

1​

M​

 

 

 

 

 

Report Narrative”​

 

 

 

4​

>​

CONTAINS​

INCLUDE​

DTID 4208 “Breast Imaging​

1​

U​

 

 

 

 

 

Report Supplementary Data”​

 

 

 

TID 4201 Breast Imaging Procedure Reported​

A procedure that is reported in a Breast Imaging Report is expressed in this Template. The results of more than one procedure may​ be included in a single report instance (see TID 4208 “Breast Imaging Report Supplementary Data”).​

See Figure Q.1-2 “Breast Imaging Procedure Reported Content Tree” in PS3.17.​

Type:​

Non-Extensible​

Order:​

Significant​

Root:​

No​

Table TID 4201. Breast Imaging Procedure Reported​

 

NL​

Rel with Parent​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

 

 

 

 

 

Type​

 

 

1​

 

 

CODE​

EV (121058, DCM,​

1​

M​

BCID 6050 “Breast​

 

 

 

 

"Procedure reported")​

 

 

Procedure Reported”​

2​

>​

HAS CONCEPT​ CODE​

EV (111464, DCM,​

1-n​

U​

DCID

6058 “Procedure​

 

 

MOD​

 

"Procedure Modifier")​

 

 

Modifiers for Breast”​

3​

>​

HAS CONCEPT​ CODE​

EV (272741003, SCT,​

1​

M​

DCID

6022 “Side”​

 

 

MOD​

 

"Laterality")​

 

 

 

 

4​

>​

HAS​

CODE​

EV (111401, DCM,​

1​

U​

DCID

6051 “Breast​

 

 

PROPERTIES​

 

"Reason for procedure")​

 

 

Procedure Reason”​

5​

>>​

HAS CONCEPT​ CODE​

EV (118578006, SCT,​

1​

U​

DCID

12102 “Temporal​

 

 

MOD​

 

"Relative time")​

 

 

Periods Relating to​

 

 

 

 

 

 

 

Procedure or Therapy”​

6​

>>​

HAS CONCEPT​ CODE​

EV (111402, DCM,​

1-n​

UC​ IFF row 4​DCID

6055 “Breast Clinical​

 

 

MOD​

 

"Clinical Finding")​

 

value is​

Finding or Indicated​

 

 

 

 

 

 

"Clinical​

Problem”​

 

 

 

 

 

 

Finding"​

Breast Clinical Finding or​

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Indicated Problem​

7​

>>>​

HAS​

CODE​

EV (272741003, SCT,​

1​

U​

DCID

6022 “Side”​

 

 

PROPERTIES​

 

"Laterality")​

 

 

 

 

8​

>​

HAS​

DATE​

EV (111060, DCM, "Study​

1​

U​

 

 

 

 

PROPERTIES​

 

Date")​

 

 

 

 

Content Item Descriptions​

 

 

 

 

 

 

Row 5 "Relative time"​

This Content Item indicates whether the value of "Reason for procedure" (row 4) is modified​

 

 

 

with "pre-" or "follow-up".​

 

 

 

 

- Standard -​