PS-2020a / part16
.pdf
DICOM PS3.16 2020a - Content Mapping Resource Page 1131
Table TID 8200. Radiology Reading Task Parameters
NL |
VT |
Concept Name |
VM |
Req TypeCondition Value Set Constraint |
|
1 |
CODE |
EV (128002, DCM, "Modality to |
1 |
U |
DCID 29 “Acquisition Modality” |
|
|
Read") |
|
|
|
2 > |
CODE |
EV (128003, DCM, "Reader |
1 |
U |
DCID 7449 “Reader Specialty” |
|
|
Specialty") |
|
|
|
3 |
CODE |
EV (128004, DCM, "Modality to |
1-n |
U |
DCID 9233 “Requested Report |
|
|
Read") |
|
|
Types” |
TID 15100 Contrast Agent/Pre-Medication Protocol Context
This Template specifies medications to be administered prior to a diagnostic imaging protocol, imaging contrast agents to be used in the protocol, and/or bolus agents to be used in the protocol. Each medication or agent may be modified by a specified route of admin- istration. The top level Content Items of this Template may appear in any order in the Protocol Context Sequence, hence the order in this Template is not significant. There may be significance in the order in which the Content Items are included in the Protocol Context Sequence, e.g., the requested order in which pre-medications are to be administered.
Type: |
Extensible |
Order: |
Non-Significant |
Root: |
No |
Table TID 15100. Contrast Agent/Pre-Medication Protocol Context
|
NL |
VT |
Concept Name |
VM |
Req TypeCondition Value Set Constraint |
||
1 |
|
CODE |
EV(123011,DCM,"Contrast/Bolus |
1-n |
U |
BCID 12 |
“Radiographic |
|
|
|
Agent") |
|
|
Contrast Agent” |
|
2 |
> |
CODE |
EV (410675002, SCT, "Route of |
1 |
U |
BCID 11 |
“Route of |
|
|
|
Administration") |
|
|
Administration” |
|
3 |
|
CODE |
EV (123012, DCM, |
1-n |
U |
|
|
|
|
|
"Pre-Medication") |
|
|
|
|
4 |
> |
CODE |
EV (410675002, SCT, "Route of |
1 |
U |
BCID 11 |
“Route of |
|
|
|
Administration") |
|
|
Administration” |
TID 15101 NM/PET Protocol Context
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
Table TID 15101. NM/PET Protocol Context
NL |
VT |
Concept Name |
VM |
Req |
Condition Value Set Constraint |
|
|
|
|
Type |
|
1 |
CODE |
EV (349358000, SCT, |
1 |
M |
BCID 25 |
|
|
"Radiopharmaceutical agent") |
|
|
“Radiopharmaceuticals” |
|
|
|
|
|
BCID 4021 “PET |
|
|
|
|
|
Radiopharmaceutical” |
2 > |
CODE |
EV(89457008,SCT,"Radionuclide) |
1 |
U |
BCID 18 “Isotopes in |
|
|
|
|
|
Radiopharmaceuticals” |
BCID 4020 “PET
Radionuclide”
- Standard -

Page 1132 |
|
DICOM PS3.16 2020a - Content Mapping Resource |
|||||
|
NL |
VT |
Concept Name |
VM |
Req |
Condition Value Set Constraint |
|
|
|
|
|
|
|
Type |
|
3 |
> |
UIDREF |
EV (113503, DCM, |
1 |
U |
|
|
|
|
|
"Radiopharmaceutical |
|
|
|
|
|
|
|
Administration Event UID") |
|
|
|
|
4 |
> |
DATETIME |
EV (123003, DCM, |
1 |
U |
|
|
|
|
|
"Radiopharmaceutical Start |
|
|
|
|
|
|
|
DateTime") |
|
|
|
|
5 |
> |
DATETIME |
EV (123004, DCM, |
1 |
U |
|
|
|
|
|
"Radiopharmaceutical Stop |
|
|
|
|
|
|
|
DateTime") |
|
|
|
|
6 |
> |
NUMERIC |
EV (123005, DCM, |
1 |
U |
UNITS = DT (cm3, UCUM, |
|
|
|
|
"Radiopharmaceutical Volume") |
|
|
"cm3") |
|
7 |
> |
NUMERIC |
EV (123006, DCM, "Radionuclide |
1 |
U |
UNITS = DT (Bq, UCUM, |
|
|
|
|
Total Dose") |
|
|
|
"Bq") |
8 |
> |
NUMERIC |
EV (123007, DCM, |
1 |
U |
UNITS = DT (Bq/mol, |
|
|
|
|
"Radiopharmaceutical Specific |
|
|
UCUM, "Bq/mol") |
|
|
|
|
Activity") |
|
|
|
|
9 |
> |
CODE |
EV (410675002, SCT, "Route of |
1 |
U |
BCID 11 “Route of |
|
|
|
|
Administration") |
|
|
|
Administration” |
10 |
> |
NUMERIC |
EV (123009, DCM, "Radionuclide |
1 |
U |
UNITS = DT ({counts}/s, |
|
|
|
|
Syringe Counts") |
|
|
|
UCUM "counts/s") |
11 |
> |
NUMERIC |
EV (123010, DCM, "Radionuclide |
1 |
U |
UNITS = DT ({counts}/s, |
|
|
|
|
Residual Syringe Counts") |
|
|
UCUM "counts/s") |
|
12 |
|
NUMERIC |
EV (14749-6, LN, "Glucose") |
1 |
U |
UNITS = EV (mmol/l, |
|
|
|
|
|
|
|
|
UCUM, "mmol/l") |
13 |
> |
DATE |
EV (127857, DCM, "Glucose |
1 |
MC |
IFF Row 12 is |
|
|
|
|
Measurement Date") |
|
|
presentanddoes |
|
|
|
|
|
|
|
|
not contain |
|
|
|
|
|
|
|
Observation |
|
|
|
|
|
|
|
DateTime |
|
|
|
|
|
|
|
(0040,A032) |
14 |
> |
TIME |
EV (127858, DCM, "Glucose |
1 |
MC |
IFF Row 12 is |
|
|
|
|
Measurement Time") |
|
|
presentanddoes |
|
|
|
|
|
|
|
|
not contain |
|
|
|
|
|
|
|
Observation |
|
|
|
|
|
|
|
DateTime |
|
|
|
|
|
|
|
(0040,A032) |
Content Item Descriptions |
|
|
|
|
|
||
Row 13 |
Glucose Measurement Date |
In an earlier edition of the Standard, an incorrect DCM code was used for this |
|||||
|
|
|
|
concept, which was already assigned as (109081, DCM, "Prospective gating"). |
|||
Row 14 |
Glucose Measurement Time |
In an earlier edition of the Standard, an incorrect DCM code was used for this |
|||||
|
|
|
|
concept,whichwasalreadyassignedas(109082,DCM,"Retrospectivegating"). |
TID 15200 JJ1017 Protocol Context
This Template defines protocol context concepts to support the requirements of Japanese Guideline JJ1017. This is expected to be used with Scheduled or Performed Protocol Codes from Coding Scheme JJ1017-16M defined in Guideline JJ1017.
Type: |
Extensible |
Order: |
Significant |
Root: |
No |
- Standard -

DICOM PS3.16 2020a - Content Mapping Resource Page 1133
Table TID 15200. JJ1017 Protocol Context
NL |
VT |
Concept Name |
VM |
Req Type Condition |
Value Set Constraint |
|
1 |
CODE |
EV(123016,DCM,"Imaging |
1 |
M |
BaselinetermsfromCodingScheme |
|
|
|
Conditions") |
|
|
JJ1017-16S of JJ1017 version 3.0 |
TID 15300 RT Prescription Annotation
TheconceptsinthisTIDaretopicsofadviceorinformationprovidedbytheprescribingphysicianforplanning,preparationanddelivery of treatment for a prescription.
Type: |
Extensible |
Order: |
Non-Significant |
Root: |
No |
Table TID 15300. RT Prescription Annotation
|
NL |
VT |
Concept |
VM |
ReqConditionValue Set Constraint |
|
|
|
|
|
Name |
|
Type |
1 |
TEXT |
EV (130022, DCM, "Radiation |
|
1 |
U |
|
|
|
Characteristics Note") |
|
|
|
|
2 |
TEXT |
EV (130023, DCM, "Beam Shaping |
|
1 |
U |
|
|
|
Note") |
|
|
|
|
3 |
TEXT |
EV (130024, DCM, "Treatment Planning |
1 |
U |
|
|
|
|
Note") |
|
|
|
|
4 |
TEXT |
EV (130025, DCM, "Special Procedure |
|
1 |
U |
|
|
|
Note") |
|
|
|
|
5 |
TEXT |
EV (130026, DCM, "Patient Positioning |
|
1 |
U |
|
|
|
Note") |
|
|
|
|
6 |
TEXT |
EV(130028,DCM,"PatientSetupNote") |
1 |
U |
|
|
7 |
TEXT |
EV (130029, DCM, "Previous Treatment |
1 |
U |
|
|
|
|
Note") |
|
|
|
|
8 |
TEXT |
EV (130030, DCM, "Planning Imaging |
|
1 |
U |
|
|
|
Note") |
|
|
|
|
9 |
TEXT |
EV(130031,DCM,"DeliveryVerification |
1 |
U |
|
|
|
|
Note") |
|
|
|
|
10 |
TEXT |
EV (130032, DCM, "Simulation Note") |
|
1 |
U |
|
11 |
CODE |
DT (130033, DCM, "Radiation Therapy |
|
1-n |
U |
BCID 9525 “Radiation |
|
|
Particle") |
|
|
|
Therapy Particle” |
12 |
CODE |
DT (130037, DCM, "Ion Therapy |
|
1-n |
U |
BCID 9526 “Ion |
|
|
Particle") |
|
|
|
Therapy Particle” |
13 |
CODE |
DT (130038, DCM, "Brachytherapy |
|
1-n |
U |
BCID 9528 |
|
|
Isotope") |
|
|
|
“Brachytherapy |
|
|
|
|
|
|
Isotope” |
14 |
CODE |
DT (130040, DCM, "Teletherapy |
|
1-n |
U |
BCID 9527 |
|
|
Isotope") |
|
|
|
“Teletherapy Isotope” |
15 |
NUMERIC |
DT (130034, DCM, "RT Beam Energy") |
|
1-n |
U |
UNITS=DCID 9521 |
|
|
|
|
|
|
“Radiotherapy |
Treatment Energy
Unit”
- Standard -

Page 1134 |
DICOM PS3.16 2020a - Content Mapping Resource |
||||
|
NL |
VT |
Concept |
VM |
ReqConditionValue Set Constraint |
|
|
|
Name |
|
Type |
16 |
CODE |
DT (130035, DCM, "Patient Positioning |
1-n |
U |
BCID 9242 |
|
|
Procedure Note") |
|
|
“Radiotherapy |
|
|
|
|
|
Acquisition Workitem |
|
|
|
|
|
Definition” |
17 |
TEXT |
EV (130036, DCM, "QA Process Note") |
1 |
U |
|
18 |
TEXT |
EV (130027, DCM, "4D Radiation |
1 |
U |
|
|
|
Treatment Note") |
|
|
|
19 |
TEXT |
EV (130039, DCM, "Adaptive Radiation |
1 |
U |
|
|
|
Therapy Note") |
|
|
|
Content Item Descriptions
Rows 11, 12, 13, 14 |
The source of radiation to be used for this RT treatment. More than one source indicates that the RT |
|
treatment may use any combination for treatment. There is no defined relationship between the entries |
|
in Row 11, 12, 13, 14 and entries in the Rows 15 and 16. |
Row 15 |
Including several energies indicates that they may be used in any combination. |
Row 16 |
ThecodesidentifyproceduressupportingthepatientpositioningprocesspriortoRTtreatment.Including |
|
several procedures indicates that they may be used in any combination. |
TID 15301 RT Segment Characteristics
Type: |
Extensible |
Order: |
Non-Significant |
Root: |
No |
Table TID 15301. RT Segment Characteristics
|
NL |
VT |
Concept Name |
VM |
Req Condition Value Set Constraint |
|
|
|
|
|
|
Type |
|
1 |
NUMERIC |
EV (130082, DCM, "Relative Mass |
1 |
U |
Units = EV (ratio, UCUM, |
|
|
|
|
Density") |
|
|
"ratio") |
2 |
NUMERIC |
EV(130083,DCM,"RelativeElectron |
1 |
U |
Units = EV (ratio, UCUM, |
|
|
|
|
Density") |
|
|
"ratio") |
3 |
NUMERIC |
EV (130084, DCM, "Effective Z") |
1 |
U |
Units = EV (1, UCUM, "no |
|
|
|
|
|
|
|
units") |
4 |
NUMERIC |
EV (130085, DCM, "Effective Z per |
1 |
U |
Units = EV (/u, UCUM, "/u") |
|
|
|
|
A") |
|
|
|
5 |
NUMERIC |
EV (130086, DCM, "Relative Linear |
1 |
U |
Units = EV (ratio, UCUM, |
|
|
|
|
Stopping Power") |
|
|
"ratio") |
6 |
> NUMERIC |
EV (130087, DCM, "Reference |
1 |
M |
Units = EV (MeV, UCUM, |
|
|
|
|
Energy") |
|
|
"Megaelectronvolt") |
7 |
NUMERIC |
EV (130088, DCM, "Linear Cell Kill |
1 |
U |
Units = EV (ratio, UCUM, |
|
|
|
|
Factor") |
|
|
"ratio") |
8 |
NUMERIC |
EV (130089, DCM, "Quadratic Cell |
1 |
U |
Units = EV (ratio, UCUM, |
|
|
|
|
Kill Factor") |
|
|
"ratio") |
9 |
NUMERIC |
EV (130090, DCM, "High Dose |
1 |
U |
Units = EV (ratio, UCUM, |
|
|
|
|
Fraction Linear Cell Kill Factor") |
|
|
"ratio") |
10 |
NUMERIC |
EV (130091, DCM, "Half-time for |
1 |
U |
Units = EV (s, UCUM, |
|
|
|
|
Tissue Repair ") |
|
|
"second") |
- Standard -

|
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 1135 |
||
|
NL |
VT |
Concept Name |
VM |
Req Condition Value Set Constraint |
|
|
|
|
|
|
Type |
|
11 |
NUMERIC |
EV (130092, DCM, "High Dose |
1 |
U |
Units=EV(Gy,UCUM,"Gray") |
|
|
|
|
Fraction Transition Dose") |
|
|
|
12 |
NUMERIC |
EV(130093,DCM,"AtomicNumber") 1-n |
U |
Units = EV (1, UCUM, "no |
||
|
|
|
|
|
|
units") |
13 |
> NUMERIC |
EV (130094, DCM, "Elemental |
1 |
M |
Units = EV (ratio, UCUM, |
|
|
|
|
Composition Atomic Mass Fraction") |
|
|
"ratio") |
14 |
NUMERIC |
EV (130095, DCM, "alpha gEUD |
1 |
U |
Units = EV (ratio, UCUM, |
|
|
|
|
value") |
|
|
"ratio") |
Content Item Descriptions
Rows 12, 13 |
Thevalueof(130094,DCM,"ElementalCompositionAtomicMassFraction")annotatesthefractionalweight |
|
of the elements identified by the (130093, DCM, "Atomic Number") with respect to the total mass of the |
|
segment. The allowed value is in the range of [0, 1]. |
TID 15302 Patient Support Position Parameters
Type: |
Extensible |
Order: |
Non-Significant |
Root: |
No |
Table TID 15302. Patient Support Position Parameters
|
VT |
Concept Name |
VM |
Req Condition Value Set Constraint |
|
|
|
|
|
Type |
|
1 |
NUMERIC |
EV (126802, DCM, "IEC61217 Table Top |
1 |
U |
Units = EV (deg, UCUM, |
|
|
Continuous Pitch Angle") |
|
|
"deg") |
2 |
NUMERIC |
EV (126803, DCM, "IEC61217 Table Top |
1 |
U |
Units = EV (deg, UCUM, |
|
|
Continuous Roll Angle") |
|
|
"deg") |
3 |
NUMERIC |
EV (126801, DCM, "IEC61217 Patient |
1 |
U |
Units = EV (deg, UCUM, |
|
|
Support Continuous Yaw Angle") |
|
|
"deg") |
4 |
NUMERIC |
EV (126804, DCM, "IEC61217 Table Top |
1 |
U |
Units = EV (mm, UCUM, |
|
|
Eccentric Axis Distance") |
|
|
"mm") |
5 |
NUMERIC |
EV (126805, DCM, "IEC61217 Table Top |
1 |
U |
Units = EV (deg, UCUM, |
|
|
Continuous Eccentric Angle") |
|
|
"deg") |
6 |
NUMERIC |
EV (126806, DCM, "IEC61217 Table Top |
1 |
U |
Units = EV (mm, UCUM, |
|
|
Lateral Position") |
|
|
"mm") |
7 |
NUMERIC |
EV (126807, DCM, "IEC61217 Table Top |
1 |
U |
Units = EV (mm, UCUM, |
|
|
Longitudinal Position") |
|
|
"mm") |
8 |
NUMERIC |
EV (126808, DCM, "IEC61217 Table Top |
1 |
U |
Units = EV (mm, UCUM, |
|
|
Vertical Position") |
|
|
"mm") |
9 |
NUMERIC |
EV (126812, DCM, "Isocentric Patient |
1 |
U |
Units = EV (deg, UCUM, |
|
|
Support Continuous Pitch Angle") |
|
|
"deg") |
10 |
NUMERIC |
EV (126813, DCM, "Isocentric Patient |
1 |
U |
Units = EV (deg, UCUM, |
|
|
Support Continuous Roll Angle") |
|
|
"deg") |
11 |
NUMERIC |
EV (126814, DCM, "Isocentric Patient |
1 |
U |
Units = EV (deg, UCUM, |
|
|
Support Continuous Yaw Angle") |
|
|
"deg") |
12 |
NUMERIC |
EV (126815, DCM, "Isocentric Patient |
1 |
U |
Units = EV (mm, UCUM, |
|
|
Support Lateral Position") |
|
|
"mm") |
- Standard -

Page 1136 |
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
|
|||||
|
|
VT |
|
Concept Name |
|
VM Req Condition Value Set Constraint |
|||
|
|
|
|
|
|
|
Type |
|
|
13 |
NUMERIC |
|
EV (126816, DCM, "Isocentric Patient |
1 |
U |
|
Units = EV (mm, UCUM, |
||
|
|
|
|
Support Longitudinal Position") |
|
|
|
|
"mm") |
14 |
NUMERIC |
|
EV (126817, DCM, "Isocentric Patient |
1 |
U |
|
Units = EV (mm, UCUM, |
||
|
|
|
|
Support Vertical Position") |
|
|
|
|
"mm") |
TID 15303 Radiotherapy Treatment Scheduled Processing Parameters |
|||||||||
Type: |
|
|
|
Extensible |
|
|
|
|
|
Order: |
|
|
Non-Significant |
|
|
|
|
|
|
Root: |
|
|
|
No |
|
|
|
|
|
|
|
Table TID 15303. Radiotherapy Treatment Scheduled Processing Parameters |
|||||||
|
|
VT |
|
Concept Name |
VM |
Req TypeCondition Value Set Constraint |
|||
1 |
TEXT |
|
EV (121384, DCM, "RT Plan Label") |
1 |
U |
|
|
||
2 |
NUMERIC |
|
EV (121385, DCM, "Current Fraction |
1 |
U |
|
UNIT = EV (1, UCUM, "no |
||
|
|
|
|
Number") |
|
|
|
|
units") |
3 |
NUMERIC |
|
EV (121386, DCM, "Number of |
|
1 |
U |
|
UNIT = EV (1, UCUM, "no |
|
|
|
|
|
Fractions Planned") |
|
|
|
|
units") |
4 |
NUMERIC |
|
EV (121387, DCM, "Number of |
|
1 |
U |
|
UNIT = EV (1, UCUM, "no |
|
|
|
|
|
Fractions Completed") |
|
|
|
|
units") |
5 |
CODE |
|
EV (121388, DCM, "Checked-In |
|
1 |
U |
|
DCID 230 “Yes-No” |
|
|
|
|
|
Status") |
|
|
|
|
|
TID 15304 Radiotherapy Treatment Progress Parameters |
|
|
|||||||
Type: |
|
|
|
Extensible |
|
|
|
|
|
Order: |
|
|
Non-Significant |
|
|
|
|
|
|
Root: |
|
|
|
No |
|
|
|
|
|
|
|
|
Table TID 15304. Radiotherapy Treatment Progress Parameters |
||||||
|
VT |
|
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
||
1 |
NUMERIC |
EV (121389, DCM, "Referenced |
1 |
U |
|
UNIT = EV (1, UCUM, "no |
|||
|
|
|
Beam Number") |
|
|
|
|
units") |
|
TID 15400 Real-World Quantity Definition |
|
|
|
|
|
||||
Type: |
|
|
|
Extensible |
|
|
|
|
|
Order: |
|
|
Non-Significant |
|
|
|
|
|
|
Root: |
|
|
|
No |
|
|
|
|
|
|
|
|
|
Table TID 15400. Real-World Quantity Definition |
|
||||
|
NL |
VT |
Concept Name |
VM |
Req TypeCondition |
Value Set Constraint |
|||
1 |
|
CODE |
|
DT (246205007, SCT, |
1 |
|
M |
BCID 7180 “Abstract |
|
|
|
|
|
"Quantity") |
|
|
|
Multi-dimensional Image Model |
|
|
|
|
|
|
|
|
|
Component Semantics” |
|
2 |
|
CODE |
|
BCID 9000 “Physical Quantity |
1-n |
|
U |
|
|
|
|
|
|
Descriptors” |
|
|
|
|
|
Content Item Descriptions
- Standard -

|
|
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 1137 |
|||
Row 1 |
|
Thisrowusesaconceptnamethatspecifiesthequantifiedcharacteristic.Itisnotrequiredthat(246205007, |
|||||
|
|
SCT, "Quantity") be used if there is a reason to use a similar concept. |
|
||||
Row 2 |
|
May be concept modifiers, such as (370129005, SCT, "Measurement Method"). |
|||||
TID 15401 Real-World Quantity Definition for X-Ray Attenuation Properties |
|||||||
Type: |
|
|
Extensible |
|
|
|
|
Order: |
|
|
Non-Significant |
|
|
|
|
Root: |
|
|
No |
|
|
|
|
|
Table TID 15401. Real-World Quantity Definition for X-Ray Attenuation Properties |
||||||
NL |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
1 |
INCLUDE |
DTID 15400 “Real-World |
1 |
M |
|
|
|
|
|
|
Quantity Definition” |
|
|
|
|
2 |
NUMERIC |
EV (DCM, 130087, |
1 |
MC IF TID 15400 Row 1 |
UNITS = EV ("MeV", |
||
|
|
|
"Reference Energy") |
|
Quantity value is (130086,UCUM, |
||
|
|
|
|
|
DCM, "Relative Linear |
"Megaelectronvolt") |
|
|
|
|
|
|
Stopping Power") |
|
- Standard -

Page 1138 |
DICOM PS3.16 2020a - Content Mapping Resource |
- Standard -

DICOM PS3.16 2020a - Content Mapping Resource |
Page 1139 |
DDICOMControlledTerminologyDefinitions (Normative)
This Annex specifies the meanings of codes defined in DICOM, either explicitly or by reference to another part of DICOM or an ex- ternal reference document or standard.
The contents of this table are available in OWL format at ftp://medical.nema.org/medical/dicom/resources/ontology/dcm/dcm.owl and
in Bioportal.
Table D-1. DICOM Controlled Terminology Definitions (Coding Scheme Designator "DCM" Coding
Scheme Version "01")
Code Value |
Code Meaning |
Definition |
Notes |
AR |
Autorefraction |
Anacquisitiondevice,processormethodthatmeasures |
|
|
|
autorefraction. |
|
ARCHIVE |
Archive |
A device, process or method that stores images and |
|
|
|
other objects for a prolonged period of time. |
|
AS |
Angioscopy |
An acquisition device, process or method that recordsRetired |
|
|
|
images during angioscopy. |
|
ASMT |
Content Assessment Result |
Adevice,processormethodthatproducesassessments |
|
|
|
of the content of other instances, e.g. for quality or |
|
|
|
suitability. |
|
AU |
Audio |
An acquisition device, process or method that records |
|
|
|
audio. |
|
BDUS |
Ultrasound Bone DensitometryAn acquisition device, process or method that performs |
|
|
|
|
ultrasound bone densitometry. |
|
BI |
Biomagnetic imaging |
An acquisition device, process or method that performs |
|
|
|
biomagnetic imaging. |
|
BMD |
Bone Mineral Densitometry |
An acquisition device, process or method that performs |
|
|
|
bone mineral densitometry by X-Ray, including |
|
|
|
dual-energy X-Ray absorptiometry (DXA) and |
|
|
|
morphometric X-Ray absorptiometry (MXA). |
|
CAD |
Computer Assisted |
An image processing device, process or method that |
|
|
Detection/Diagnosis |
performs computer assisted detection or diagnosis. |
|
CAPTURE |
Image Capture |
An acquisition device, process or method that performs |
|
|
|
image capture, includes video capture. |
|
CD |
Color flow Doppler |
An acquisition device, process or method that performsRetired |
|
|
|
color flow Doppler. |
|
|
|
Replaced by (US, DCM, |
|
|
|
"Ultrasound") |
|
CF |
Cinefluorography |
An acquisition device, process or method that performsRetired |
|
|
|
cinefluorography. |
|
|
|
Replaced by (RF, DCM, |
|
|
|
"Radiofluoroscopy") |
|
COMP |
Computation Server |
A device, process or method that performs computation |
|
|
|
as a service; includes radiotherapy planning. |
|
CP |
Culposcopy |
An acquisition device, process or method that recordsRetired |
|
|
|
images during culposcopy. |
|
CR |
Computed Radiography |
An acquisition device, process or method that performs |
|
|
|
computed radiography. |
|
- Standard -

Page 1140 |
DICOM PS3.16 2020a - Content Mapping Resource |
|
|
Code Value |
Code Meaning |
Definition |
Notes |
CS |
Cystoscopy |
An acquisition device, process or method that recordsRetired |
|
|
|
images during cystoscopy. |
|
CT |
Computed Tomography |
An acquisition device, process or method that performs |
|
|
|
computed tomography. |
|
CTPROTOCOLCT Protocol |
A device, process or method that produces CT device |
|
|
|
|
acquisition protocols. |
|
DD |
Duplex Doppler |
An acquisition device, process or method that performsRetired |
|
|
|
duplex Doppler. |
|
|
|
Replaced by (US, DCM, |
|
|
|
"Ultrasound") |
|
DF |
Digital fluoroscopy |
An acquisition device, process or method that performsRetired |
|
|
|
digital fluoroscopy. |
|
|
|
Replaced by (RF, DCM, |
|
|
|
"Radiofluoroscopy") |
|
DG |
Diaphanography |
An acquisition device, process or method that performs |
|
|
|
diaphanography. |
|
DM |
Digital microscopy |
An acquisition device, process or method that performsRetired |
|
|
|
digital microscopy. |
|
DOC |
Document |
A device, process or method that produces documents. |
|
|
|
i.e., representations of documents as images, whether |
|
|
|
by scanning or other means. |
|
DOCD |
Document Digitizer EquipmentA device, process or method that digitizes hardcopy |
|
|
|
|
documents and imports them. |
|
DS |
Digital Subtraction AngiographyAn acquisition device, process or method that performsRetired |
|
|
|
|
digital subtraction angiography. |
|
|
|
Replaced by (XA, DCM, |
|
|
|
"X-Ray Angiography") |
|
DSS |
Department System SchedulerA department-based information system (for instance, |
|
|
|
|
Radiology or Laboratory) that provides functions related |
|
|
|
to the management of orders received from external |
|
|
|
systems or through the department system’s user |
|
|
|
interface. This definition matches that of the DSS/OF |
|
|
|
Actor in the IHE Scheduled Workflow (SWF) Profile. |
|
DX |
Digital Radiography |
An acquisition device, process or method that performs |
|
|
|
digital radiography. |
|
EC |
Echocardiography |
An acquisition device, process or method that performsRetired |
|
|
|
echocardiography. |
|
|
|
Replaced by (US, DCM, |
|
|
|
"Ultrasound") |
|
ECG |
Electrocardiography |
An acquisition device, process or method that performs |
|
|
|
electrocardiography. |
|
EPS |
Cardiac Electrophysiology |
An acquisition device, process or method that performs |
|
|
|
cardiac electrophysiology. |
|
ES |
Endoscopy |
An acquisition device, process or method that records |
|
|
|
images during endoscopy. |
|
F |
Female |
Female sex. |
|
FA |
Fluorescein angiography |
An acquisition device, process or method that performsRetired |
|
|
|
fluorescein angiography. |
|
|
|
Replaced by (OP, DCM, |
|
|
|
"Ophthalmic photography") |
|
FC |
Female changed to Male |
Female sex changed to Male sex. |
|
- Standard -